X-ray fluorescence microtomography based on polycapillary-focused X-rays from laboratory source

LOW ENERGY ACCELERATOR, RAY AND APPLICATIONS

X-ray fluorescence microtomography based on polycapillary-focused X-rays from laboratory source

Binggang Feng
Fen Tao
Yiming Yang
Tao Hu
Feixiang Wang
Guohao Du
Yanling Xue
Yajun Tong
Tianxi Sun
Biao Deng
Tiqiao Xiao
Nuclear Science and TechniquesVol.29, No.6Article number 85Published in print 01 Jun 2018Available online 30 Apr 2018
61300

X-ray fluorescence microtomography (μXFCT) is a nondestructive analytical technique and has been widely used to nondestructively detect and quantify the elemental composition and distributions in samples. Usually, synchrotron radiation X-rays are used for μXFCT, due to its high flux density. In this paper, a laboratory-source-based μXFCT system was developed, in which a polycapillary lens is employed to focus the X-ray beam and improve the flux density. The maximum likelihood expectation maximization (MLEM) algorithm was used to reconstruct the computed tomography (CT) slices at a limited number of projections. The experimental results demonstrated that the developed system could reveal the elemental distribution inside the test sample, with an elemental sensitivity of 1000 ppm.

X-ray Fluorescence microtomographyPolycapillaryLaboratory X-ray source

I. Introduction

As a unique methodology in the investigation of elemental distributions, X-ray fluorescence microtomography (μXFCT) based on synchrotron radiation has been applied to numerous fields since it was first proposed in 1986, including botany, biomedicine, medicine, material science, and mineralogy[1]. Numerous SR-μXFCT systems have been established in synchrotron sources around the world, such as the European Synchrotron Radiation Facility (ESRF) in France[2], the Advanced Photon Source (APS) in America[3], Spring-8 in Japan[4], etc. To meet a user’s requirements, a SR-μXFCT system has been designed and constructed at the Shanghai Synchrotron Radiation facility (SSRF)[5-8]. However, the applications of SR-μXFCT are limited by the complex and huge experimental facilities. In recent years, the laboratory μXFCT system underwent rapid development [9-12]. The laboratory μXFCT system has many merits, such as broader applications, availability, and convenience. The distribution of iron hydroxide was obtained with fluorescent computed tomography (CT) based on focused X-rays[13]. The gold nanoparticle distribution inside different organs of mice were measured with laboratory μXFCT [14]. Flux is the critical limitation of the laboratory μXFCT system. In this paper, a polycapillary lens[15-18], which is composed of many glass micro-capillaries, is used to focus the X-rays from an X-ray tube and improve the flux density in CT scanning.

II. Set-up of the laboratory μXFCT system

Fig. 1 shows a schematic layout of the laboratory μXFCT system. The X-ray source is a micro-focus X-ray tube with a Tungsten target (Hamamatsu, 75 W). The energy of the X-ray beam is 70keV. The opening angle of the X-ray beam is 43°. The distance between the polycapillary lens and the X-ray source is 56.5 mm. We added an aperture at the exit of the X-ray source to limit the divergence of the X-ray. In addition, the aperture behind the polycapillary lens was used to collimate the X-ray beam. The polycapillary lens is fixed on a 5D adjustable stage with a lead aperture. The focal distance of the polycapillary lens is 50 mm. The photons emitted by X-ray fluorescence were collected by a silicon drift detector (SDD) (SII, Vortex-90E) perpendicular to the incident X-ray beam in order to minimize X-ray scattering. The sample is mounted on a high-precision stage which can be adjusted in four dimensions. The X-ray CCD camera is used to detect the transmission images. For a better perspective of the imaging results, the transmission CT and μXFCT images can be fused together. A picture of the μXFCT system is shown in Fig. 2.

Fig. 1
(Color online) Schematic layout of the laboratory μXFCT system.
pic
Fig. 2
(Color online) The laboratory X-ray fluorescence microtromography system
pic

III. Polycapillary lens for μXFCT

The opening angle of the laboratory X-ray source is quite large, which limits the flux density of the emitted X-ray beam. Low flux density is a key limitation for the development of a laboratory μXFCT system. Among all the X-ray focusing strategies, the use of a polycapillary lens is a cost-effective and highly efficient candidate [19-20], which has large acceptance aperture and high transmission efficiency. Therefore, the use of a polycapillary lens to focus X-rays is feasible in a laboratory μXFCT system. For the CT imaging system, a pencil beam is usually used to achieve the projections. This means that, for the case of a focused beam, the depth of focus (DOF) of the polycapillary lens is a key parameter. If the DOF is less than the size of the sample, or the sample is not placed within the focal depth, the focused X-ray along the sample is divergent. This will result in an inaccurate reconstruction of the sample. Hence, the DOF of the polycapillary determines the maximum size of the experimental samples. The relationship between the DOF of the polycapillary lens and the sample size is shown in Fig. 3.

Fig. 3
(Color online) Relationship between the focal depth of the polycapillary lens and sample size
pic

Before experiments, the depth of focus for the polycapillary lens was tested. As shown in Fig. 1, the polycapillary lens is fixed in the center of the aperture. The distribution of the focal spot is approximately Gaussian. The full width at half maximum (FWHM) of the Gaussian distribution at the capillary focal spot is defined as the focal spot size. We adjusted the rotation angle and the pitch angle of the polycapillary lens in order to obtain the best focus spot.

The focused spot was recorded by an X-ray CCD. The optimum position of the capillary was determined by measuring the intensity profile of the focused spot in the vertical and horizontal directions. When the distance between the polycapillary lens and the X-ray CCD was set to 50 mm, as shown in Fig. 4(a), the size of each pixel was 13 μm and the FWHM of the polycapillary was 82 μm×82 μm. To measure the focal depth, the X-ray CCD was scanned along the optical axis of the incident X-ray beam in 1 mm steps. According to the experimental results, the focal spot size remains 82 μm×82 μm when the lens-detector distance ranged from 47 mm to 54 mm, as shown in Fig. 4(c). This indicated that the DOF of the polycapillary lens was 7 mm. Therefore, the spatial resolution of the μXFCT system is 82 μm, and size of the sample should be less than 7 mm.

Fig. 4
(Color online)(a) Intensity profile of the focal spot for the polycapillary lens. Fig. 4(b) Three-dimensional display of the focal spot. Fig. 4(c) Relationship between the focal spot size and the lens-detector distance in the horizontal and vertical direction.
pic

IV. Image reconstruction algorithm for μXFCT

In a typical μXFCT system, the sample is scanned point-by-point with a pencil beam of synchrotron X-rays, which requires a significant data acquisition time to collect all the projections required for 3D tomographic imaging. Reducing the projection number is a practical solution to improve the data collection efficiency. However, the limited projections always lead to fake images and increased noise during CT reconstruction. Seeking proper algorithm for image reconstruction is critical in the μXFCT system.

Among CT reconstruction algorithms, including FBP[12], algebraic reconstruction techniques (ART), ordered subset expectation maximization (OSEM), and maximum likelihood expectation maximization (MLEM), etc., the MLEM has a unique advantage for sparse projection data. Hence, the laboratory μXFCT system adopts the MLEM algorithm to reconstruct the projection data.

The principle of the MLEM algorithm is shown in the following equation[21, 22]:

Cl+1(i,j)=Cl(i,j)(m,n)K(i,j,m,n)(m,n)K(i,j,m,n)I(m,n)(i,j)K(i,j,m,n)Cl(i,j) (1)

Cl(i,j) andCl+1(i, j)are the l and l+1 iterative image values, respectively, and I(m, n) is the fluorescence intensity for the nth projection under the mth projection angle. K(i, j, m, n) is the contribution of pixel (i,j) to the fluorescent intensity I(m,n). In this case, the sample absorption of the incident and fluorescence X-rays is not taken into account. The distance from the center of each pixel (i,j) to the incident ray (m,n) is Δd, and d is the size of each pixel.

K(i,j,m,n)={1   Δdd20   Δd>d2 (2)

To test and verify the feasibility of the algorithm, numerical simulation was carried out on a phantom. The FBP, OSEM, and MLEM algorithms were used to reconstruct the CT images with different projection number. In the OSEM algorithm[23], the projection data was divided into several subsets according to the projection angles in order to improve the reconstruction speed. When the number of subsets is set to one, the OSEM algorithm and MLEM algorithm are the same.

Two phantoms, numbered in model 1 and model 2, were employed for the simulation. The bright patterns indicate the elemental distribution at a fixed concentration, and the size of each pixel is 10 μmm. A total of 30, 20, and 10 projections were collected for the each model by rotating the model from 0° to 180° in steps of 6°, 9°, and 18°, respectively. The reconstruction results of the models using the different algorithms are shown in Fig. 5. According to Fig. 5, even at the largest number of projections (30), the signal-to-noise ratio (SNR) of the reconstructed slices by MLEM algorithm is much higher than the FBP algorithm, in which obvious artifacts can be observed. When the projection number is reduced to 20, artifacts or noises arise in the reconstructed slices when using the MLEM algorithm. The slices reconstructed with the FBP algorithm become blurred, especially for model-I with a complex elemental distribution. When 10 projections were used in the reconstructed slices from MLEM, the complex elemental distribution of model-I can still be distinguished dimly, while the image reconstructed by FBP is indecipherable. Based on the simulation results shown in Fig. 5, we can conclude that MLEM is a much better algorithm than the FBP algorithm while dealing with fewer projections and at a limited projection number. MLEM is able to achieve acceptable image quality, even for a complex elemental distribution. The results could be used as reasonable guidance in related experiments.

Fig. 5.
Comparison between the reconstructed slices by FBP and MLEM algorithms, respectively.
pic

In the OSEM algorithm, adopting the subset is a useful way to increase the speed of convergence. However, when the number of subsets exceeds a critical value, the quality of the reconstructed images decreases with an increasing number of subsets. When the number of subsets exceeds the critical value, the noise will be amplified during the iterative CT reconstruction process with limited projections[24]. Considering the reconstruction efficiency, the MLEM algorithm only requires 0.635 s of computation time for 20 projections, and there is no need for acceleration. According to the concentration profiles, the MLEM algorithm can achieve a better signal-to-noise ratio (SNR) than the OSEM algorithm to some extent, as shown in Fig. 6.

Fig. 6.
(Color online) Comparisons of the profiles between the OSEM and MLEM with various iterations
pic

Numerical simulation demonstrates the validity of the MLEM algorithm and highlights its reconstruction merits for sparse projection data. The simulation results show that the MLEM algorithm can effectively suppress noise, improve SNR, and reduce distortion in the reconstructed images. The MLEM algorithm can achieve acceptable image quality with a reduced projection number, resulting in higher data collection efficiency and lower radiation dose compared to FBP.

V. Experimental results and discussion

In the laboratory μXFCT system, the maximum voltage of the X-ray tube is 150 kV. The energy of the X-ray beam was estimated to be 70 keV. The focused beam size was 82 μm×82 μm. A series of test samples were used to evaluate the capabilities of the μXFCT system. The first test sample was a thin walled borosilicate capillary with an outer diameter of 2 mm and a wall thickness of 20 μm. The capillary was filled with standard iodine solution. Iodine is a key element in biological research and plays an important role in clinical disease diagnosis. Iodine solution (1000 μg/ml, 50 ml) is a standard reference material according to the National Institute of Standards and Technology (NIST, USA). As shown in Fig. 7(a), the capillary was filled with 1000 ppm iodine solution. The sample was mounted on a high-precision stage and adjusted within the depth of focus. The sample was scanned with 40 translations per 150 μm. The exposure time per step was 6 s and a 6° angular interval was used for CT scanning. The iodine distribution reconstructed by MLEM algorithm is shown in Fig. 7(b). Fig. 7(c) shows the acquired fluorescence spectrum (sinogram data). The ordinate is the scanning angle in degrees (0–180°) and the abscissa is the number of translations. We can obtain the iodine distribution using the experiment system and achieve an elemental sensitivity of 1000 ppm for iodine. The sensitivity depends on the atomic number and the flux on the sample.

Fig. 7
(Color online) The experimental results of the iodine, where (a) picture of the capillary filled with 1000 ppm iodine solution; (b) reconstructed iodine concentration; (c) the acquired fluorescence spectra.
pic

For another test sample, three iron nails were selected, each with a diameter of 1 mm. The iron nails were placed in the depth of focus of the polycapillary lens with a triangular distribution. The phantom was scanned with 80 translation steps per 150 μm at a 6° angular interval over 180°. The exposure time per step was 0.5 s. Fig. 8(b) shows the reconstruction results for the nail sample. The reconstruction results confirmed the actual distribution of the iron nails. Fig. 8(c) shows two pieces of plastic pipe with a diameter of 1.5 mm, which were also chosen as test samples. The plastic pipes were filled with 10% iodine solution and bound tightly together with glue. The phantom was scanned with 60 translation steps per 150 μm and a 6° angular interval over 180°. The exposure time per step was 2 s. As shown in Fig. 8(d), the iodine distribution in different pipes can be distinguished. The shape of the iodine distribution in the reconstructed slices is somewhat distorted due to solution overflow from the plastic pipe, which confirmed the spatial resolution of the laboratory μXFCT system.

Fig. 8
(Color online) Experimental results from iron nails and iodine. (a) Picture of three iron nails; (b) reconstruction slice for the three nails; (c) picture of two pieces of plastic pipes; (d) reconstruction slices of μXFCT for iodine concentration contained in the plastic pipes.
pic

VI. Conclusion

A laboratory μXFCT system based on X-ray focusing by a polycapillary lens was successfully developed. The simulation and experimental results demonstrated the practicality of the system. The field of view for elemental imaging is 7 mm, and the elemental sensitivity is 1000 ppm. The developed system is to be opened to users, which will expand the applications of μXFCT in material science and biomedicine.

References:
[1] R.P. Boisseau, Determination of three dimensional trace element distributions by the use of monochromatic X-ray microbeams, Massachusetts Institute of Technology, Dept. of Physics, (1986).
[2] A.S. Simionovici, B. Golosio, M.V. Chukalina, et al.,

Seven years of x-ray fluorescence computed microtomography, Optical Science and Technology

, the SPIE 49th Annual Meeting, International Society for Optics and Photonics.5535, 232-242(2004). doi: 10.1117/12.558948
Baidu ScholarGoogle Scholar
[3] N. K. Blute, D. J. Brabander, H. F. Hemond, et al.,

Arsenic Sequestration by Ferric Iron Plaque on Cattail Roots

, Environmental science & technology.38,6074-6077(2004). doi: 10.1021/es049448g
Baidu ScholarGoogle Scholar
[4] Y. Hirai, A. Yoneyama, A. Hisada, et al.,

In Vivo X‐Ray Fluorescence Microtomographic Imaging of Elements in Single‐Celled Fern Spores

,Synchrotron Radiation Instrumentation: Ninth International Conference on Synchrotron Radiation Instrumentation.879,1345-1348.(2007). doi: 10.1063/1.2436313
Baidu ScholarGoogle Scholar
[5] B. Deng, G. H. Du, G. Z. Zhou, et al.,

3D elemental sensitive imaging by full-field XFCT

, The Analyst.140, 3521-3525 (2015).doi: 10.1039/C4AN02401J
Baidu ScholarGoogle Scholar
[6] J. Qiu, B. Deng, Q. Yang, et al.,

Internal elemental imaging by scanning X-ray fluorescence microtomography at the hard X-ray microprobe beamline of the SSRF: Preliminary experimental results

, Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 269, 2662-2666 (2011). doi: 10.1016/j.nimb.2011.08.005
Baidu ScholarGoogle Scholar
[7] B. Deng, Q. Yang, G. H. Du, et al.,

The progress of X-ray fluorescence computed tomography at SSRF

, Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 305, 5-8(2013).doi: 10.1016/j.nimb.2013.04.045
Baidu ScholarGoogle Scholar
[8] P. F. Sun, B. Deng, G. H. Du, et al.,

Nondestructive rare earth element imaging of fish teeth from deep‐sea sediments

, X‐Ray Spectrometry. 44, 442-446 (2015).doi: 10.1002/xrs.2624
Baidu ScholarGoogle Scholar
[9] P. F. Sun, B. Deng, Q. Yang, et al,

An accelerated OSEM reconstruction algorithm using an accelerating factor for X-ray fluorescence tomography

, Nucl. Tech. 38, 60201-060201 (2015) (in Chinese). doi: 10.11889/j.0253-3219.2015.hjs.38.060201
Baidu ScholarGoogle Scholar
[10] H. L. Xie, B. Deng, G. H. Du, et al,.

Latest advances of X-ray imaging and biomedical applications beamline at SSRF

, Nucl. Tech. 26, 6-21 (2015). doi: 10.13538/j.1001-8042/nst.26.020102
Baidu ScholarGoogle Scholar
[11] H. von Busch, G. Harding, G. Martens, et al,

Investigation of externally activated x-ray fluorescence tomography for use in medical diagnostics, Medical Imaging

, International Society for Optics and Photonics. 5745,90-101(2005).doi: 10.1117/12.591663
Baidu ScholarGoogle Scholar
[12] N. Manohar, B.L. Jones, S.H. Cho,

Improving x-ray fluorescence signal for benchtop polychromatic cone-beam x-ray fluorescence computed tomography by incident x-ray spectrum optimization: A Monte Carlo study

, Medical physics.41,101906 (2014).doi: 10.1118/1.4895004
Baidu ScholarGoogle Scholar
[13] S.-K. Cheong, B.L. Jones, A.K. Siddiqi, et al,

X-ray fluorescence computed tomography (XFCT) imaging of gold nanoparticle-loaded objects using 110 kVp x-rays

, Physics in medicine and biology.55, 647(2010).doi: 10.1088/0031-9155/55/3/007
Baidu ScholarGoogle Scholar
[14] H.M. Hertz, J.C. Larsson, U. Lundstrom, et al,

Laboratory x-ray fluorescence tomography for high-resolution nanoparticle bio-imaging

, Optics letters.39, 2790-2793(2014).doi: 10.1364/OL.39.002790
Baidu ScholarGoogle Scholar
[15] M.A. Kumakhov,

Channeling of photons and new X-ray optics

, Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms. 48, 283-286(1990). doi: 10.1016/0168-583X(90)90123-C
Baidu ScholarGoogle Scholar
[16] M.A. Kumakhov. F.F. Komarov,

Multiple reflection from surface X-ray optics

, Physics Reports. 191, 289-350 (1990). doi: 10.1016/0370-1573(90)90135-O
Baidu ScholarGoogle Scholar
[17] P. Korecki, K. M. Sowa, B. R Jany, et al,

Defect-Assisted Hard X-ray Microscopy with Capillary Optics

, Phys. Rev. Lett. 116, 233902 (2016) doi: 10.1103/PhysRevLett.116.233902
Baidu ScholarGoogle Scholar
[18] K. M. Sowa, A. Last, P. Korecki,

“Grid-enhanced X-ray coded aperture microscopy with polycapillary optics”

. Sci. Rep. 7, 44944 (2017). doi: 10.1038/srep44944
Baidu ScholarGoogle Scholar
[19] B. Laforce, B. Vermeulen, J. Garrevoet, et al,

Laboratory Scale X-ray Fluorescence Tomography: Instrument Characterization and Application in Earth and Environmental Science

, Analytical chemistry.88,3386-3391 (2016).doi: 10.1021/acs.analchem.6b00137
Baidu ScholarGoogle Scholar
[20] N. Manohar, F.J. Reynoso, P. Diagaradjane, et al,

Quantitative imaging of gold nanoparticle distribution in a tumor-bearing mouse using benchtop x-ray fluorescence computed tomography

, Sci Rep. 6,22079 (2016). doi: 10.1038/srep22079
Baidu ScholarGoogle Scholar
[21] L A Shepp, Y. Vardi,

Maximum Likelihood Reconstruction for Emission Tomography

, IEEE transactions on medical imaging. 1,113-122 (1982).doi: 10.1109/TMI.1982.4307558
Baidu ScholarGoogle Scholar
[22] P J Green,

Bayesian reconstructions from emission tomography data using a modified EM algorithm

. IEEE transactions on medical imaging.9,84-93 (1990).doi: 10.1109/42.52985
Baidu ScholarGoogle Scholar
[23] Q. Yang, B. Deng, W. Lv, et al,

Fast and accurate X-ray fluorescence computed tomography imaging with the ordered-subsets expectation maximization algorithm

, Journal of Synchrotron Radiation. 19, 210-215 (2012). doi: 10.1107/S0909049511052253
Baidu ScholarGoogle Scholar
[24] H.M. Hudson, R.S. Larkin,

Accelerated Image-Reconstruction Using Ordered Subsets of Projection Data

, Ieee Transactions on Medical Imaging. 13, 601-609(1994).doi: 10.1109/42.363108
Baidu ScholarGoogle Scholar