[1] HAN Yan . NDT (in Chinese), 2003, 25(9).
[2] LIU Yunzuo . Decay Scheme of Common Radioactive Nuclide (in Chinese). Beijing : Atomic Energy Press , 1982.
A test device for isotopic γ-ray imaging, which consists of an isotope γ-ray source, a CdZnTe γ-ray spectrometer and other auxiliary equipment, is studied here. Compared with the conventional X-ray, the isotope γ-ray, which is utilized in this project, has its own advantages in imaging. Furthermore, with a room-temperature high-energy-resolution CdZnTe detector and a modern imaging processing technique, this device is capable of effectively suppressing the background and gaining more information, thus it can obtain a better image than conventional X-ray devices. In the experiment of PCB imaging, all soldered points and chip components are sharply demonstrated.