44500

For a passive radon detector of Solid State Nuclear Track Detector (SSNTD), the measurement results can be affected under an environment of high humidity. In order to correct this kind of measurement errors, experiments in normal radon laboratory were performed. The track films were etched in 6.25 mol/L NaOH solution at 70°C for 12 h, and rinsed with water for 6 h. The experiment results show that calibration factor of the SSNTD decreased with increasing humidity, because of increased stopping power of the moisture-absorbed film surface to α particles. Therefore, the calibration factor K should be corrected from the calculated result.

Solid state nuclear track detectorHumidityInfluence
References
[1] Urban M, Piesch E. Radiat Prot Dosim, 1981, 1: 97-109.
[2] Zhu R S. The theory and application of solid state nuclear track detector. Beijing: Science Press, 1987, 1: 1-5 (in Chinese).
[3] Li Y L, Li S Y. Radiat Prot Dosim, 1999, 5: 377-381 (in Chinese).
[4] Nan Q L, Zou S Y, Ding T X. Nucl Electron Detect Technol, 1999, 19: 72-74.
[5] Wang X G, Luo Y S, Zhang H, et al. Nucl Sci Technol, 2005, 28: 319-323.
[6] Fujimoto K. Radiat Prot Dosim, 1994, 55: 273-277.
[7] Zhou X Y. Error theory and experiment data disposal. Beijing: Beijing aviation Institute Press, 1986, 69-70 (in Chinese).
[8] Wu H S. The measuring method and application of radon. Beijing: Atomic Energy Press, 1995, 103 (in Chinese).