Monte-Carlo simulation to determine detector efficiency of plastic scintillating fiber

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

Monte-Carlo simulation to determine detector efficiency of plastic scintillating fiber

Mohammad Mehdi NASSERI
MA Qing-Li
YIN Ze-Jie
WU Xiao-Yi
Nuclear Science and TechniquesVol.15, No.5pp.308-311Published in print 01 Oct 2004
4400

Fundamental characteristics of the plastic scintillating fiber (PSF) as a detector for electromagnetic radiation (X & γ) are obtained by GEANT4 detector simulation tool package. The detector response to radiation with energy of 10~400 keV is found out. Energy deposition as well as detector efficiency (DE) of the PSF are studied. In order to make linear array of the PSF for imaging purpose, the optimum length of fiber is also estimated.

Plastic scintillating fiberImagingDetector's quantum efficiencyGeant 4.
References
[1] Shao H, Miller D W, Pearsall C R. Nucl Instr Meth Phys Res, 1990, A299: 528-533
[2] http://www.bicron.com/stdwaveopticalfiber.htm (2003)
[3] lmai S, Soramoto S, Mochiki K et al. Rev Sci Instrum, 1991, 62(4): 127-132
[4] Ikhlef A, Skowronek M. IEEE Trans Nucl Sci, 1994, 41(2):408-414
[5] Knoll G F. Radiation Detection and Measurement, John Wiley & Sons, 1989:202-208