Study of performance of small gamma camera consisting of crystal pixel array and position sensitive photomultiplier tube

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

Study of performance of small gamma camera consisting of crystal pixel array and position sensitive photomultiplier tube

ZHU Jie
LIU Shi-Tao
LEI Xiao-Wen
YAN Tian-Xin
XU Zi-Zong
WANG Zhao-Min
Nuclear Science and TechniquesVol.16, No.4pp.238-243Published in print 01 Aug 2005
23201

The performance of gamma camera with NaI(Tl) array coupled with position sensitive photomultiplier tube (PSPMT) R2486 has been studied. The pixel size of NaI(Tl) crystal is 2mm×2mm and the overall dimension of the array is 48.2mm×48.2mm×5mm. There are 484 pixels in a 22×22 matrix. Because each pixel can produce a much focused light spot and restrict the spread of photons, position resolution of the gamma camera is mainly determined by pixel size. It is shown that crystal array pixel can reduce shrinkage effect and improve intrinsic position resolution greatly via restricting the spread of photons. Experimental results demonstrate that its position resolution and linearity are much improved comparing with the gamma camera using planar crystals coupled with PSPMT.

NaI(Tl)Pixel arrayPosition sensitive photomultiplier tubePosition linearityPosition resolution
References
[1] Bird A J, Ramsden D. Nucl Instr Meth, 1990, A299: 480-483
[2] Hamamatsu Photomultiplier Tube Hamamatsu Photonics K.K. Electron Tube Center, Position sensitive photomultiplier tube with crossed wire anodes R2486 series, Specifications manual (in Japanese), 1993
[3] Clancy R L, Thompson C J, Robar J L, et al. IEEE Trans Nucl Sci, 1997, 44(3): 494-498
[4] Smith L E, He Z, Wehe D K, et al. IEEE Trans Nucl Sci, 1998, 45(3): 963-969
[5] Guru S V, He Z, Ferreria J C, et al. Nucl Instr Meth, 1994, A353: 328-333
[6] Pani R, Scafe R, Pellegrini R, et al. Nucl Instr Meth Phys Res A, 2002, 477: 72-76
[7] CN/ASD Group. PAW users guide, Program Library Q121. CERN, October 1993
[8] Pani R, Scopibaro F, Pellegrini R, et al. Phys Med, 1993, 9(2-3): 233-236
[9] Pani R, Soluri A, Pellegrini R, et al. Nucl Instr Meth, 1994, A348: 551-556
[10] Scopinaro F, Pani R, Soluri A, et al. Nucl Med Commun, 1994, 15: 905
[11] Blazek K, de Notaristerfani F, Maly P, et al. IEEE Trans Nucl Sci, 1995, NS-42(5): 1474
[12] Baccaro S, Blazek K. Nucl Instr Meth, 1995, A361: 209
[13] Barone L M, Blazek K, Bollini D, et al. Nucl Instr Meth, 1995, A 360: 302
[14] ZENG Hai-Ning.

Study of a small gamma camera based on PSPMT

, University of Science and Technology of China, Doctoral Dissertation, 1999
Baidu ScholarGoogle Scholar