logo

Application of CMOS charge-sensitive preamplifier in triple-GEM detector

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

Application of CMOS charge-sensitive preamplifier in triple-GEM detector

LAI Yong-Fang
DENG Zhi
LI Yu-Lan
LIU Yi-Nong
LI Yuan-Jing
LI Jin
Nuclear Science and TechniquesVol.17, No.5pp.289-292Published in print 20 Oct 2006
34300

Among the various micro-pattern gas detectors (MPGD) that are available, the gas electron multiplier (GEM) detector is an attractive gas detector that has been used in particle physics experiments. However, the GEM detector usually needs thousands of preamplifier units for its large number of micro-pattern readout strips or pads, which leads to considerable difficulties and complexities for front end electronics (FEE). Nowadays, by making use of complementary metal-oxide semiconductor (CMOS)-based application specific integrated circuit (ASIC), it is feasible to integrate hundreds of preamplifier units and other signal process circuits in a small-sized chip, which can be bound to the readout strips or pads of a micro-pattern particle detector (MPPD). Therefore, CMOS ASIC may provide an ideal solution to the readout problem of MPPD. In this article, a triple GEM detector is constructed and one of its readout strips is connected to a CMOS charge-sensitive preamplifier chip. The chip was exposed to an 55Fe source of 5.9 keV X-ray, and the amplitude spectrum of the chip was tested, and it was found that the energy resolution was approximately 27%, which indicates that the chip can be used in triple GEM detectors.

GEM detectorCMOS ASICCharge-sensitive preamplifierEnergy resolution
References
[1] Oed A. Nucl. Instr. and Meth., 2001, A471: 109
[2] Sauli F. Nucl. Instr. and Meth., 1997, A386: 531
[3] http://gdd.web.cern.ch/GDD/
[4] MAXWELL Commercial Finite Element Computation Package. Ansoft Co. Pittsbug, PA
[5] Charpak G, Bouclier R, Bressani T, et al. Nucl. Instr. and Meth., 1968, 62: 262
[6] LI Ben-Jun, LIU Li-Hua, XIN De-Lu.The principle and design of CMOS IC (in Chinese). Beijing: Beijing University of Posts and Telecommunication Press, 1997: 4
[7] Oed A. Nucl. Instr. and Meth., 1988, A263: 351
[8] Straver J. Nucl. Instr. and Meth., 1994, A348: 485
[9] DENG Zhi, KANG Ke-Jun, LIU Yi-Nong, et al. Tsinghua Science and Technology (in Chinese), 2005, 45(12): 1643