1 Introduction
The particular measure of the peak widths in gamma- ray spectrometry, FWHM (Full width at half maximum), indicates the resolution of the gamma spectroscopy system. The FWHM expressed in keV has a relationship with energy of incident gamma-ray.
It is important to know the FWHM of peaks as an energy function when analyzing gamma spectra, especially in peak searching and fitting[1]. Thus, an accurate FWHM calibration needs be performed as the basis of spectrum processing.
2 Physical essence of FWHM
Gamma-ray spectrum is actually the energy response of the detection system to the gamma-rays of particular energies. Peaks in the spectrum are spread over several channels that can be characterized as a Gaussian distribution, whose standard deviation can be predicted by[2]
σ is the overall uncertainty of the peak in the spectrum in energy form.
σI is the intrinsic width of nuclear energy levels, which is extremely smaller than σ, and can be ignored[2].
σP is the statistical fluctuation of the production of electron-hole pairs (n) in the detector. The number of electron-hole pairs created by a gamma-ray of energy (E) subjects to Fano distribution (F)[3]. On defining ''W'' as the average energy consumption to create a pair, the expected uncertainty of n is σn=(F×n)0.5 = [(F×E)/W)]0.5.
σC is the uncertainty in collecting the charge by the detector caused by trapping effect. Due to the complexity effect, there is no a simple way to express the uncertainty as an energy function[3] 3. However, a linear relationship of σC=cE appears to be satisfactory, but not theoretical reason[2].
σE, the contribution from electronic noise in pulse processing system, is performed as a constant to the gamma-rays with any energy.
3 FWHM calibration equations[2]
There are mainly three independent effects in influencing the resolution of gamma-ray spectrum. Using the energy form ΔE to express the FWHM of a Gaussian-shaped peak in the spectrum, we can give
ΔE≈2.355σ[3]
where p, c, and e are constant related to fluctuation, collection, and electronic noise, respectively.
The square root for both side of Eq.(2) is taken, And
Debertin and Helmer[4] introduced Eq.(4) by ignoring the factor of incomplete charge collection.
In addition, Eq.(5) with simpler expression can be found in the commercial programs of Genie 2000[5].
Gilmore conducted a series of experiments to compare the fitting FWHM calibrations with different formulae[2] (Table 1). The best fit is from the square root quadratic function, and a simple quadratic fit. The Genie 2000 fitting is worse than the linear equation.
Fitting type | Formulae | RMS differences |
---|---|---|
Linear | a0+a1E | 0.020 |
Quadratic | a0+a1E+a2E2 | 0.0065 |
Genie 2000 | a0+a1E0.5 | 0.055 |
Debertin and Helmer | (a02+a12E)0.5 | 0.026 |
Square root quadratic | (a0+a1E+a2E2) | 0.0057 |
4 Traditional FWHM calibration
Similar to energy calibration, traditional FWHM calibration method is accomplished by measuring gamma-rays of precisely known energy and estimating the relationship between the peak width and energy[1].
The three ways can provide proper gamma-ray emission source. (1) Several kinds of nuclides with single or discrete energies are chosen and mixed together; (2) One specific nuclide emitting a variety of energies is chosen, covering the entire range of the spectrum, such as 152Eu. The known nuclides in the test sample can be used to perform the calibration.
Explicit method can be used to fine calibration; but its calibration source needs be additionally measured by time consumption. Especially, when changing experiment condition, the spectroscopy should be recalibrated by employing a new measurement except applying for nuclides in any situation. In this paper, an implicit method for FWHM calibration is developed by self-information gained from the target spectrum.
5 Implicit FWHM calibration method
5.1 Method introduction
The Ge(Li) detector has high energy resolution, and makes a great number of gamma peaks with enough distance each other, separating single peaks easily.
Implicit method contains two calibration procedures. First, single peaks are examined and extracted from the spectrum. Second, the width of each singlet is calculated. Together with the peak position, the calibration curve is obtained by fitting these data.
Comparing all kinds of peak searching algorithms with implicit FWHM calibration, symmetric zero-area conversion method (SZA) has advantages in low noise sensitivity, fine multi-peaks resolving and weak peak identifying abilities[1].
5.2 Method description
The spectrum data are convoluted by the SZA method using a zero-area window. The linear base tends to zero after convolution while maximum value is at the positions of gamma peaks. The conversion process is described as Eq.(6).
where yi and
Among various forms of window functions, Gaussian second derivative has the best property6, its coefficients can be expressed as Eq.(7).
where, σw is the standard deviation from the original Gaussian distribution Gw(j).
Figure 1 shows a Gaussian second derivative formed SZA window function with σw=4.
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F001.jpg)
And, gamma peaks in the spectrum can be treated as Gaussian shape approximately.
where, H, p and σγ are their amplitude, position and standard deviation, respectively.
Due to the convolution nature[7], a Gaussian peak conversion operated with the window function maintains the appearance of Gaussian second derivative, as shown in Fig.2.
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F002.jpg)
Gconv(i) is the convolution of gamma peak and the original Gaussian distribution of the window which has a Gaussian form of σconv= (σ2wσ2y)0.5.
The distance between two zero-crossing points beside the conversion maximum can be obtained from the Gaussian second derivation.
The σγ of the gamma peak is deduced from intercept D and σw, subsequently, the peak FWHM is expressed by Eq.(11).
The Eq.(11) is only valid for single peaks, and the calculated FWHM is definitely a constant for every pair of D and σw. While for multi-peaks, the intercept D relates to the intervals and ratios between the component peaks, in other words, the FWHM obtained from Eq.(11) is meaningless.
Consequently, the same spectrum can be converted using several windows with different σw. When the FWHM value tends to be invariable, the corresponding peak is considered as a singlet, thus distinguishing a set of multi-peaks due to their invalid FWHM. Also, the FWHM of each single peak can be gained, implementing the FWHM calibration process.
5.3 Method modification
Applying to practical spectra, a group of simulation data set up is used to test the calibration method.
The σw of 8 window functions equal to 1 to 8 is convoluted by σγ of 4 individual Gaussian peaks equal to 1, 2, 4, and 8. The calculated and theoretical FWHM values are compared (Table 2, Fig.3).
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F003.jpg)
σγ | Theoretical FWHM | Error between calculated and theoretical FWHM values | |||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
σw=1 | σw=2 | σw=3 | σw=4 | σw=5 | σw=6 | σw=7 | σw=8 | Mean | Std | ||
1 | 2.355 | 0.321 | 0.279 | 0.204 | 0.152 | 0.116 | 0.0881 | 0.0650 | 0.0445 | 0.159 | 0.101 |
2 | 4.710 | 0.146 | 0.0542 | 0.124 | 0.142 | 0.139 | 0.127 | 0.114 | 0.0992 | 0.118 | 0.0301 |
4 | 9.419 | 0.0405 | 0.0700 | –0.00681 | 0.0527 | 0.0646 | 0.0442 | 0.00736 | –1.71e–4 | 0.0340 | 0.0299 |
8 | 18.84 | 0.00967 | 0.0258 | 0.0263 | –0.00798 | 0.0177 | –0.0202 | 0.0114 | 0.0102 | 0.00912 | 0.0161 |
To reduce the difference in Eq.(11), a modification factor is introduced to Eq.(12).
The constant (a) is determined by applying an iteration method to optimize the average error under every window function for each Gaussian peak.
Figure 4 shows the result using Eq.(12). Under every window with different σw, the mean of residual errors approaches to zero, and the fluctuation is suppressed at some levels.
In practical application, the σw values should be selected. The SZA method can separate multipeaks with small σw, but is sensitive to the noise. On the other hand, the reliable window of large σw is inadequate in multipeak processing. For σw, a value approximating to σγ suggests obtaining the optimal results. Therefore, σw can be determined within a rough range related to FWHM varying in the spectrum.
Also, this method is compared with calculation process. The average background level above and below the peak is subtracted from the estimated peak height to get the expected count. The two positions of half height on the peak are obtained by linear interpolation, and the distance is exactly the FWHM. The resultant reliability depends on the uncertainty of peak height and background level.
Quadratic formed background and noise with constant SNR are added to a Gaussian peak by σγ=4 to simulate an actual gamma peak. The 100 sets of simulation data with random background and noise are processed by the two methods (Table 3).
The FWHM calculated by Eq.(12) approximates to the theoretical value. For the direct approach, the result is relatively fluctuant due to the noise. Applied first to reduce the standard deviation, a smoothing process impacts the FWHM value to a certain extent.
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F004.jpg)
Method | Theoretical FWHM | Error of 100 experiments | |
---|---|---|---|
Mean | Std | ||
Direct-without smoothing | 9.419 | –0.0296 | 0.252 |
Direct-with smoothing | –0.0405 | 0.221 | |
This paper | 0.0167 | 0.170 |
6 Method verification
A gamma-ray spectrum from sand sample is measured by HPGe detector, and used to test the implicit FWHM calibration method (Fig.5).
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F005.jpg)
The 3 windows with σw equal to 2, 3, and 4 are employed to convert the spectrum and calculate the FWHM corresponding to each peak, the three sets of FWHM deviation is applied to pick out single peaks, and their mean value is used to the fitting procedure.
HYPERMET, a mature developed γ spectrum analysis code, and Genie 2000 program are employed to validate the calculation in Table 4. The HYPERMET using a region fitting method to get accurate peak position and FWHM can be used as the standard value, while Genie 2000 applies a direct process to calculate FWHM. Together with FWHM values, the relative errors comparing to HYPERMET are calculated for the other two processes, as shown in Table 5. The relative errors within ±3% are obviously better than Genie 2000 except one point. In addition, both HYPERMET and Genie program need a rough FWHM calibration as input to obtain proper results, while the novel method is fully automatic.
Peak position | Calculated FWHM | ||||
---|---|---|---|---|---|
σw=2 | σw=3 | σw=4 | Mean | Std | |
323 | 2.729 | 2.917 | 2.833 | 2.826 | 0.0946 |
470 | 2.770 | 2.835 | 2.773 | 2.793 | 0.0368 |
631 | 2.990 | 3.111 | 3.110 | 3.070 | 0.0699 |
812 | 3.236 | 3.254 | 3.267 | 3.253 | 0.0153 |
1116 | 3.410 | 3.414 | 3.398 | 3.407 | 0.00824 |
1407 | 3.635 | 3.653 | 3.592 | 3.627 | 0.0312 |
1531 | 3.782 | 3.722 | 3.748 | 3.751 | 0.0301 |
1918 | 3.976 | 4.025 | 4.065 | 4.022 | 0.0443 |
2197 | 4.227 | 4.217 | 4.220 | 4.222 | 0.00494 |
2861 | 4.600 | 4.616 | 4.634 | 4.616 | 0.0169 |
4127 | 5.446 | 5.538 | 5.501 | 5.495 | 0.0460 |
4335 | 5.580 | 5.642 | 5.583 | 5.602 | 0.0347 |
4595 | 5.754 | 5.847 | 5.703 | 5.768 | 0.0728 |
6851 | 7.049 | 7.131 | 7.229 | 7.137 | 0.0901 |
Peak position | Calculated FWHM | ||||
---|---|---|---|---|---|
HYPERMET | Genie | Relative error / % | This paper | Relative error / % | |
323 | 2.751 | 3.024 | 9.939 | 2.826 | 2.726 |
470 | 2.905 | 2.945 | 1.371 | 2.793 | -3.855 |
631 | 3.131 | 3.210 | 2.527 | 3.070 | -1.948 |
812 | 3.190 | 3.237 | 1.463 | 3.253 | 1.975 |
1116 | 3.368 | 3.396 | 0.8266 | 3.407 | 1.158 |
1407 | 3.660 | 3.688 | 0.7560 | 3.627 | -0.9016 |
1531 | 3.685 | 3.714 | 0.7924 | 3.751 | 1.791 |
1918 | 4.025 | 4.245 | 5.461 | 4.022 | -0.07453 |
2197 | 4.190 | 4.218 | 0.6747 | 4.222 | 0.7637 |
2861 | 4.588 | 4.457 | -2.854 | 4.616 | 0.6103 |
4127 | 5.446 | 5.651 | 3.762 | 5.495 | 0.8997 |
4335 | 5.550 | 5.571 | 0.3874 | 5.602 | 0.9406 |
4595 | 5.727 | 6.261 | 9.326 | 5.768 | 0.7159 |
6851 | 7.065 | 7.137 | 1.007 | 7.137 | 1.013 |
Table 6 and Fig.6 show the compared position- FWHM data from Table 4 fitted to the 5 equations in Section 3. The square root quadratic function has the best effect, and the linear fit performs better than Genie 2000 function, this is consistent with the conclusion in Section 3.
Calibration equation | RMS differences |
---|---|
Square root quadratic | 0.0235 |
Quadratic | 0.0285 |
Linear | 0.0774 |
Debertin and Helmer | 0.162 |
Genie 2000 | 0.433 |
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F006.jpg)
The performance of above 3 typical equations is shown in Fig.7.
-201302/1001-8042-24-02-011/alternativeImage/1001-8042-24-02-011-F007.jpg)
7 Conclusions
In this paper, an implicit FWHM calibration algorithm is introduced as a complement to traditional method. The new algorithm is developed on the basis of symmetric zero-area conversion method, and uses the self-information from the object spectrum. The method is tested, modified, and compared to a direct calculation process first using simulation data. Ultimately, this method is applied to a practical gamma-ray spectrum, and the calibration result is proved to be good.