I. INTRODUCTION
Tissue-equivalent proportional counters (TEPC) are widely used in radiation protection and radiation biology [1]. In neutron monitoring, real-time dosimeters are vital for radiation protection. Most of them are of the type with a thermal neutron detector inside a large polyethylene sphere [2-4]. These moderator-based instruments convert their response to dose equivalent by simple calibration in a reference neutron spectrum, which is assumed to be valid for other fields. It is obvious that such neutron dosimeters are applicable to limited situations, and may give rise to large errors sometimes [5]. Besides, such a dosimeter is usually too heavy to handle [6].
Unlike traditional dosimeters, the application of low pressure TEPCs in neutron radiation protection is based on simultaneous measurement of absorbed dose D and mean quality factor
where, y is lineal energy defined as the quotient of energy imparted ε, to the matter in a volume from a single energy deposition event by a mean chord length
Consequently, TEPCs are promising candidates for neutron radiation protection. In this work, a spherical TEPC was developed. The prototype detector was characterized by a built-in 241Am alpha source and a standard 252Cf field. The TEPC was characterized with 2.45 MeV neutrons, and its performance was satisfactory.
II. EXPERIMENTS
A. Detector configuration
Benjamin design [11, 12] is employed in the spherical TEPC to keep multiplication process spatially uniform around the anode wire. Figure 1 shows a sketch and picture of the TEPC. With an inner diameter of 44.8 mm, the TEPC is made of A-150 (Shonka [13]) plastic in wall thickness of 3.0 mm. A single tungsten wire (Φ25 μm, gold-coated) is used as the anode. An outer stainless steel shell of 3.0 mm thickness is used for air tightness. Sealed mode, instead of gas-flow system, was employed for portability consideration. Leak rate of the system was measured at <10-13 Pa m3/s, ensuring working stability of the TEPC. An 241Am alpha source is attached to the outside wall for energy calibration. Most of the alpha particles can go into the cavity through a Φ1 mm hole on the cathode wall. In this work, the TEPC was temporarily flushed with methane based tissue equivalent (MTE) gas (64.4% CH4, 32.4% CO2, 3.2% N2) of 2.12 kPa to simulate soft tissue of 1 μm size, that is, the deposited energies by a charged particle along the TEPC diameter are equivalent to the absorbed energies in soft tissue of 1 μm size. An absolute barometer was used to measure MTE gas pressure of the detector, from 0 to 100 kPa, with the precision of 0.01 kPa.
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F001.jpg)
B. Experiments in reference 252Cf field
Dose equivalent response RH is defined as the ratio of dose equivalent reading of a TEPC to the ambient dose equivalent H*(10) [14] at the same point. RH of the TEPC was measured in a standard bare 252Cf neutron radiation field. Neutron emission rate of the 252Cf source is about 4.86×106 s-1. In this measurement, the TEPC was tested with the electronic system in Fig. 2. The detector pre-amplifier (ORTEC 142PC) was set in irradiation room and the TEPC was placed 30 cm away from the 252Cf source. A 40-m signal cable connecting the pre-amplifier and amplifier brought significant noise to the measuring system. This noise was measured after the experiment and subtracted from the measured neutron pulse height distribution.
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F002.jpg)
C. Measurements in mono-energy neutron field
The TEPC was tested with 2.45 MeV neutrons generated by d(d,n)p reaction. A schematic layout of the experimental arrangement was shown in Fig. 3. Deuterium ions were accelerated to bombard deuterium target to produce 2.45 MeV neutrons. The TEPC detector was placed at 20 cm from the target. A Si(Au) surface barrier detector (SBD) was used to monitor total neutron yields by recording the protons at 135 to the beam incidence and at 1 m from the target. It is assumed that contamination of the photo events for an accelerator-based neutron source are negligible.
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F003.jpg)
III. RESULTS AND DISCUSSION
A. Characterization of the TEPC with built-in 241Am alpha source
Calibration of the TEPC was implemented with an 241Am alpha source. Because of the extremely low pressure of MTE gas in the TEPC, the alpha particles lose only part of its kinetic energy in the gas cavity. A typical signal of 241Am alpha source output from pre-amplifier (142-PC) is shown in Fig. 4. Energy deposition of 5.48 MeV alpha particles of 241Am in the gas cavity was simulated by FLUKA codes [15]. In the simulation, the sampling alpha particles were set to fire into the TEPC along the cavity diameter. The result of energy deposition distribution was plotted in Fig. 5. It can be seen that an average alpha particle traversing 44.8 mm in the TEPC cavity deposits 85.11 keV. As a consequence, the lineal energy is yα=energy deposited/mean chord length=85.11 keV / (2/3)μm = 127.67 keV/μm. The calibration of lineal energy yI with channel I in the MTE-filled TEPC is thus given by,
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F004.jpg)
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F005.jpg)
where Iα is the channel number corresponding to the alpha peak of 241Am. In the calibration, counting rate of the TEPC to the built-in 241Am alpha source is about 1.2 cps. For an accurate Iα value, the data collection time should be no less than 20 min. At bias of 730 V, the lineal energy per channel using Eq. (4) was 0.152 keV/μm. With an MCA of 2048 channels, the measured lineal energy can be extended from 0.2 keV/μm to 300.0 keV/μm in a single measurement. The measured alpha peak with TEPC biased at 730 V is shown in Fig. 6.
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F006.jpg)
B. Dose equivalent response of the TEPC in 252Cf neutron field
The direct output of TEPC measurement is pulse height distribution, which is known also as energy deposition spectra. The TEPC measurement of a 252Cf source is shown in Fig. 7. After photo components [16] of 252Cf source and electronic noise subtraction, the lineal energy yi and counts ni for each channel were calculated using Eq. (4). Absorbed dose D, mean quality factor Q and dose equivalent rate were derived using Eqs. (1)–(3). Thus, measured from the 252Cf source, we had HTEPC rate = 19.2 μSv/h and H*(10) rate = 17.6 μSv/h. It should be noted that the reference value of H*(10) rate for 252Cf neutrons at the same point was 17.6 μSv/h, so that the dose response of the TEPC to 252Cf neutrons is RH=1.1, indicating a good approximation of dose equivalent rate reading of the TEPC to H*(10) rate in 252Cf reference neutron fields.
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F007.jpg)
C. Characterization of TEPC in mono-energetic neutron field
1. Microdosimetric spectra and mean lineal energy
Figure 8 shows the measured pulse height distribution of 2.45 MeV neutrons. Frequency-mean lineal energy
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F008.jpg)
where, f(y) is the probability density function of lineal energy:
d(y) is the dose distribution:
The measured and simulated mean lineal energy are given in Table 1. In general, the measurement results agree well with the simulation results, being just 8.8% less than the simulated
-201506/1001-8042-26-06-015/alternativeImage/1001-8042-26-06-015-F009.jpg)
Measured data | FLUKA | |
---|---|---|
yf (keV/μm) | 28.9 | 31.7 |
yd (keV/μm) | 50.8 | 47.5 |
2. Mean quality factor and dose equivalent
Absorbed dose D of 2.45 MeV neutrons to the TEPC was estimated at 1.9 mGy using Eq. (2). The mean quality factor was evaluated using Eq. (3) and compared with the effective quality factor [17] Qeff. The measured dose equivalent was obtained using Eq. (1), ambient dose equivalent H*(10)(10) estimated by employing fluence to H*(10)(10) conversion coefficient [18] was used for comparison. All the results are listed in Table 2. For mean quality factor, evaluated value from measurement underestimated the Qeff by 21.8%, this is mainly attributed to the approximation of LET by y. Operation of the TEPC in lower pressure with a simulated size of 0.5 μm or less may alleviate this problem. For dose equivalent values, the difference between the results was 1.66%.
Q | H (mSv) | |
---|---|---|
Measured data | 9.7 | 18.4 |
Reference data | 12.4 | 18.1 |
Relative deviation | 21.8% | 1.66% |
IV. CONCLUSION
We have developed a spherical TEPC that used as a neutron dosimeter for radiation protection purpose. Benjamin design was employed to improve the energy resolution of the TEPC. The measured lineal energy of the TEPC can be extended from 0.2 keV/μm to 300 keV/μm in a single measurement when biased at 730 V. The dose equivalent response of the TEPC was found to be 1.1 in the 252Cf reference field. Microdosimetric spectra, frequency mean lineal energy and dose-average lineal energy of 2.45 MeV neutrons have been evaluated and showed good agreement with the FLUKA simulated results. Mean quality factor and dose equivalent H estimated from measurement agreed well with the reference data.
Preliminary results have indicated that the developed TEPC had sufficient accuracy to measure lineal energy distribution used for absorbed dose and mean quality factor estimation. The evaluated dose equivalent can be used to approximate H*(10)(10) for neutron radiation protection. Further characterization of the TEPC in low neutron fields around nuclear reactors are in progress.
The new remcounter lb6411: Measurement of neutron ambient dose equivalent H*(10)(10) according to ICRP60 with high sensitivity
. IEEE T Nucl Sci Conf R, 1996, 1: 132-134. DOI: 10.1109/NSSMIC.1996.590921Response functions of the Andersson-Braun and extended range rem counters for neutron energies from thermal to 10 Gev
. Nucl Instrum Meth A, 2002, 476: 341-346. DOI: 10.1016/S0168-9002(01)01459-0Evaluation and Monte Carlo modelling of the response function of the Leake neutron area survey instrument
. Nucl Instrum Meth A, 2004, 531: 596-606. DOI: 10.1016/j.nima.2004.05.086An accurate and portable solid state neutron rem meter
. Nucl Instrum Meth A, 2013, 719: 6-12. DOI: 10.1016/j.nima.2013.03.060The effect of ICRP(74) on the response of neutron monitors
, 1999, 421: 365-367. DOI: 10.1016/S0168-9002(98)01205-4Sensitivity and uncertainty in the measurement of H*(10)(10) in neutron fields using an REM500 and a multi-element TEPC
. Radiat Prot Dosim, 2014, 161: 205-209. DOI: 10.1093/rpd/ncu052TEPC performance as a reference standard
. Radiat Prot Dosim, 2014, 158: 246-250. DOI: 10.1093/rpd/nct221Understanding of the microdosimetric quantities obtained by a TEPC
. J Nucl Sci Technol, 2008, 45: 213-216. DOI: 10.1080/00223131.2008.10875825A high resolution spherical proportional counter
. Nucl Instrum Meth A, 1968, 59: 77-85. DOI: 10.1016/0029-554X(68)90347-9Neutron response functions and detection efficiency of a spherical proton recoil proportional counter
. Nucl Sci Tech, 2010, 21: 330-333. DOI: 10.13538/j.1001-8042/nst.21.330-333Comparison of plastics used in proportional counters for proton and heavy ion measurements
. Nucl Instrum Meth B, 2014, 333: 69-72. DOI: 10.1016/j.nimb.2014.04.016Monte Carlo design study for thick gas electron multiplier-based multi-element microdosimetric detector
. Nucl Instrum Meth A, 2014, 757: 67-74. DOI: 10.1016/j.nima.2014.04.063Intercomparison of Monte Carlo radiation transport codes to model tepc response in low-energy neutron and gamma-ray field
. Radiat Prot Dosim, 2014, 161: 257-260. DOI: 10.1093/rpd/nct260Use of low-pressure tissue equivalent proportional counters for the dosimetry of neutron beams used in bnct and bncefnt
. Med Phys, 2000, 27: 535-548. DOI: 10.1118/1.598921Effective quality factors for neutrons based on the revised ICRP/ICRU recommendations
. Radiat Prot Dosim, 2005, 115: 536-541. DOI: 10.1093/rpd/nci004