1 Introduction
Radiotherapy is a cancer treatment with the main goal of producing a high dose of radiation to the tumor and as low a dose as possible to skin and surrounding healthy organs [1]. Dosimetry is highly important in radiation treatment planning because a successful treatment requires an accurate dose to be delivered to the target volume [2].
60Co beam is a widespread radiotherapy source because of its proper energy, relatively long half-life, and high specific activity. 60Co units are still available because of their reliability compared to modern linear accelerators (LINACs). 60Co decays to 60Ni by the emission of a beta particle. The activated nickel nucleus emits two gamma photons with energies of 1.17 MeV and 1.33 MeV [3-5].
Dose calculation is practically impossible during radiotherapy. Therefore, two methods are used: experiment using a phantom and simulation with MNCP code [6-7]. Several functions are used to relate the absorbed doses to tissue and air. Percent Depth Dose (PDD) and Tissue-Air Ratio (TAR) are the main functions used for dosimetry.
Treatment techniques and calibration practices are varied in radiotherapy. There are two techniques for treating patients: SSD technique and SAD or isocentric technique. For dosimetric calculations, SSD and SAD techniques rely on PDD distributions and TARs, respectively.
60Co units have isocentric mounts. In isocentric machines, the tumor is positioned on the isocenter. Therefore, the SSD and field size at the surface are varied during therapy [7].
Several authors have studied on dose measurement and calculation for 60Co beams [8-12]. In previous studies, 60Co therapy units were simulated using EGS and MCNP codes and the PDD distributions were calculated in a water phantom. A good agreement between the results and measurements was achieved [13-16]. The PDD distributions in water phantom were evaluated and the obtained results were compared with experiment, showing a good agreement [2]. In 2011, a 60Co radiotherapy unit was simulated using GEANT4 for different beam sizes and a good agreement between the obtained results and the published data was observed [17-18]. In 2013, dosimetric consistency of a 60Co teletherapy unit was investigated for 10 years by the SSD and SAD techniques [3].
In routine radiotherapy, it is not practical to use direct method with PDDs for an isocentric unit because it is time consuming and numerous dosimetric tables are required. Another approach is to calculate the dose for each point using two TARs and the value of inverse square law and divide the results by the reference dose to obtain relative depth dose. As an alternative, the curves can be obtained for different SSDs and approximated with a single curve, called a calibration curve, which might be used for further dose calculations [7].
In this paper, curves of relative depth doses were obtained for three different SSDs using the MCNP4C Monte Carlo simulation, and a calibration curve was obtained for the first time. The values of this curve were compared to published data. Although Monte Carlo simulations and measurements were done for a 60Co source, the isocentric technique and the approach have not been simulated using MCNP4C previously. The advantages of this study are simplicity, capability to use for complex geometries, and quick response.
2 Method:
An ellipsoid phantom with radii of 25 cm, 25 cm, and 20 cm was modeled as a soft tissue using MCNP4C, as shown in Fig. 1. An isotropic point source of 60Co beam was simulated simultaneously with two different gamma energies (1.17 and 1.33 MeV) and equal possibilities using SI (Source Information) and SP (Source Probabilities) cards. The source was positioned in an infinite air sphere. The photon weight factor was 1 in all cells and zero outside the sphere. The photon transport was considered (mode p) in simulation. Three different SSDs were considered in simulation: 75 cm, 70 cm, and 65 cm. The field size of 10×10 cm2 at SAD of 80 cm (isocenter of machine) was defined using a collimator made of lead cuboid with an outer/inner side of 20cm/5 cm. The values of dose were calculated in spherical cells with a radius of 0.5 cm. The deposited energy was determined by tally F6. The number of histories was 107. The cut-off energy for photon mode was set at 0.1 MeV.
-201709/1001-8042-28-09-006/alternativeImage/1001-8042-28-09-006-F001.jpg)
Ten statistical tests were performed by the MCNP to assure the reliability of results. These tests are: (1) The mean must exhibit only random fluctuations as N increases, (2) Relative error (R) must be less than 0.1, (3) R must monotonically decrease with N, (4) R must decrease as 1/
3 Results and Discussion:
The aim of this study was to obtain curves of depth dose for different SSDs and derive a calibration curve, which can be used in isocentric technique. This was accomplished by using the MCNP4C simulation code. The problem was properly simulated because the tally passed all 10 statistical tests. The R and VOV were monotonically decreasing, the FOM was relatively constant, and the slope had a perfect value of 10. As a result, the R calculated by MCNP code, was always less than 1%.
Relative depth dose is the ratio of dose at a point on the central axis to the isocenter dose. For each SSD, the tally was calculated at distances ‒15 to 20 cm from SAD. Different curves of relative depth dose were obtained for different SSDs using MCNP4C, as shown in Fig. 2. The dose buildup region observed for each curve is defined as the region between the surface and depth of maximum dose. This happened because energetic secondary charged particles that are initially released in the patient by photon interactions have such a relatively long range where they can deposit their kinetic energy. Therefore, the relative depth dose curve increases for the first few mm, reaches a maximum at 0.5 cm, and decrease with increasing depth. Nominal value for depth of maximum dose is 0.5 cm for 60Co and this proved that the simulation was properly undertaken. The outlier of curves is accordingly related to the build up region [1, 7]. This behavior was also seen in the simulated curves.
-201709/1001-8042-28-09-006/alternativeImage/1001-8042-28-09-006-F002.jpg)
The curves were approximated with a single curve, as illustrated in Fig. 3. In this approach, the outliers were not considered, as has been done in Ref [7]. Otherwise, the accuracy would be lost. The function used to fit the data, parameters, their errors, and R2 are shown in this figure. It is not scientific motivation to use the obtained function exactly because some approximations were taken into account. The parameters of the function depend on the geometry of the problem and composition of the materials used in simulation.
-201709/1001-8042-28-09-006/alternativeImage/1001-8042-28-09-006-F003.jpg)
In Fig. 4, the fitted curve was compared to the curve obtained by published data [7], differing in approximately 5% in the worst case. The differences are due to geometry and composition of the materials. The accuracy of dose calculations is highly important in radiation therapy and overall a dose error less than 5% should be delivered to patients [20]. Since the difference in accuracy was obtained less than 5% in this study, the approach can be simulated by MNCP4C.
-201709/1001-8042-28-09-006/alternativeImage/1001-8042-28-09-006-F004.jpg)
From Fig. 4, the obtained results were more accurate for distances between ‒5 cm and 10 cm from SAD and within 5% of the published curve. Moreover, the dose was always overestimated for points nearer to the surface than the reference point and slightly underestimated for points below it. This behavior was also obvious in Ref [7]. This study can be extended for more than three SSDs to cover more depths in phantom. The precision can be increased by using smaller steps of SSD.
Run time is the limitation of MCNP to calculate routine dose. In this study, required time for achieving statistically acceptable results ranged between 6 to 12 min. The required time for direct method and hand calculation depends on the calculating process. For example, the simulation can be faster by a factor of 4, approximately. If a high degree of accuracy is needed, more run time is required. In these circumstances, some techniques such as variance reduction methods are useful.
4 Conclusion:
All of the obtained results showed that the MCNP4C was successful in the calculation of the depth dose values. It is an alternative method when dose calculations are time consuming. The results serve as a starting point for better understanding the discrepancies between depth dose values of the published data and MCNP simulation. It is suggested that a similar procedure for a phantom including inhomogeneity should be repeated and the obtained results should be supported by experimental data for a satisfactory conclusion.
Calculation of analytical expressions for measured percentage depth dose data in Megavoltage photon therapy
. IRCMJ 11, 140-144 (2009).Dosimetric consistency of Co-60 teletherapy unit-a ten years study
. Int. J. Health Sci. 7, 15-21 (2013). doi: 10.12816/0006016.The NUBASE evaluation of nuclear and decay properties
. Nucl. Phys. A 729, 3-128 (2003). doi: 10.1016/S0375-9474(97)00482-X.The computation of two and three dimensional dose distributions in Co-60 teletherapy
. Brit. J. Radiol. 38, 369-377 (1965). doi: 10.1259/0007-1285-38-449-369.Automation of radiation treatment planning - VI. A general field equation to calculate percent depth dose in the irradiated volume of a cobalt 60 beam
. Brit. J. Radiol. 40, 463-74 (1967). doi: 10.1259/0007-1285-40-474-463.Analytical approach for depth dose calculations (60Co beams with fixed source-target distance)
. Radiology 115(1), 181-185 (1975). doi: 10.1148/115.1.181.Dose calculations for asymmetric fields defined by independent collimators using symmetric field data
. Brit. J. Radiol. 73, 403-9 (2000). doi: 10.1259/bjr.73.868.10844866.New formula for calculation of cobalt 60 percentage depth dose
, IJMP 2, 53-60 (2005).Monte Carlo simulation of a cobalt-60 beam
. Med. Phys. 14, 414-419 (1987). doi: 10.1118/1.596120.Monte Carlo simulation of a typical 60Co therapy source
. Med. Phys. 26, 2494-2502 (1999). doi: 10.1118/1.598770.Monte Carlo dose calculations for radiotherapy machines: Theratron 780-C teletherapy case study
. Phys. Med. Biol. 49, 807 (2004). doi: 10.1088/0031-9155/49/5/011.MCNP simulation of a Theratron 780 radiotherapy unit
. Radiat. Prot. Dosim. 116, 65 (2005). doi: 10.1093/rpd/nci125.Application of a GEANT4 Simulation to a 60Co Therapy Unit
. J. Korean Phys. Soc. 59 (1), 12-19 (2011). doi: 10.3938/jkps.59.12.Cobalt-60 gamma-ray beams
, Brit. J. Radiol. 25, 46-61 (1996).An MCNP primer
.Prescribing, recording and reporting photon beam therapy
(