1 Introduction
PADC of CR-39 type track detectors are widely used as passive solid state nuclear detector to measure charged particles and neutrons due to their low cost, robustness, high detection efficiency, and small sensitivity to gamma- and X-rays and electromagnetic waves[1-3]. Heavy charged particles passing through the detectors leave latent tracks which are visible at modest optical magnification levels after chemical etching. Traditional manual methods to count the tracks are labor- intensive and highly operator-dependent. With the development of technology, several models of automatic counting systems have been developed, such as Radosys®,TASLIMAGE®, etc.[4-7] In these systems, the tracks are supposed to be small light spots in uniform shapes (circular or elliptical), and counted by setting a set of binary threshold, size threshold and circularity threshold[8,9]. However, the track discrimination will be affected by the track diameters due to different etching conditions (time, temperature, concentration, etc.), the artifact tracks (the scratches, bubbles, and small dots that are caused by the camera or detector), the energy of incident particles, etc.[10-13] Therefore, uncertainties of the track counting should be further studied.
In this study, based on an automatic counting system developed in our laboratory for passive radon monitors, the size distributions of both artifact and alpha tracks were systematically investigated to optimize the etching conditions, and the overlap effect of alpha tracks was simulated for different monitoring conditions (track densities) to minimize and quantify the uncertainty of the counting system.
2 Materials and Methods
2.1 The CR-39 and its exposure
The CR-39 detector material was manufactured by Fukuvi Chemical Industry Co., Ltd, Japan. Samples in 1 cm×1 cm size were cut from the material for experiments. A total of 128 sample detectors were randomly divided into two groups. Group A was used for investigating artifact tracks without any radon exposure. Group B were set in the passive integrated radon monitor (PIRM) we developed[14] to study the alpha tracks after exposed to 2.02 kBq·m−3 of 222Rn in a chamber for 168 h (about 340 kBq·m−3·h of integrated radon exposure).
2.2 Chemical etching
All of the detectors were chemically etched in 6.25 N water solution of NaOH at 90±0.1°C. In order to investigate the artifact and alpha tracks in different etching periods, the etching was performed in eight stages, with one hour for each stage. The etched samples were taken out and cleaned for track analysis.
2.3 Track analysis
Figure 1 shows the automatic track counting system we developed. It is built on an optical microscope with a ×4 objective lens and a CCD camera. A computer controlled three-dimensional mobile object stage is implemented for autofocus and the detector movements. The software in Visual C++ development environment (Fig. 1b) is responsible for the motion control, image processing, auto-focusing and track counting. In routine application, the alpha tracks can be identified automatically from the artifact tracks based on the settings of binary threshold, size threshold and circularity threshold for the images. In this study, in order to obtain more information on the size and size-distribution of artifact and alpha tracks, the settings of size threshold and circularity threshold were banned.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F001.jpg)
2.4 Overlap correction and its uncertainty
A Monte Carlo code was developed based on Matlab® to investigate track overlap for different track densities. Fig. 2 shows an example of the simulation regions and overlapped tracks. In the simulation, the code randomly placed tracks in the 1 mm ×1 mm simulation region and counted the tracks. To avoid the boundary effects, the simulation region was surrounded by a buffer region. As shown in Fig. 1(b), most of the track images in the radon system are quasi circular or circular, for the sake of simplicity, circular tracks were assumed in simulation. Track diameters were chosen from measured size distributions. When the center-to-center distance of two tracks (D) was less than the sum of their radii (R1 and R2), i.e. D≤R1 +R2, the two tracks were regarded as overlapped tracks and counted as one track[15].
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F002.jpg)
For a specific track density of m (tracks/mm2) generated by the code, the simulations were repeated for 20000 times. The relative standard uncertainty for counting due to overlap effect can be estimated by [16]:
where ni is the track density counted in each simulation (tracks/mm2); N is the repeat times of the simulations; nmean is the arithmetic mean of ni. The overlap correction factor K is defined as[17]:
2.5 Verification of overlap correction
In order to verify the simulation results, five groups of PIRM monitors, 10 pieces each, were exposed to 10 kBq·m−3 of radon concentration in the standard radon chamber at Shanghai Institute of Measurements Technology (SIMT). The five groups were exposed to integrated radon exposures of 500–2500 kBq·m−3·h. The exposed CR-39 detectors were etched in a 6.25 N water solution of NaOH at 90±0.1°C for the optimized etching period, and counted by the auto-counting system.
3 Results and Discussion
3.1 Size distribution of artifact tracks
Figure 3 shows the size distributions of artifact tracks on the CR-39 detectors without any radon exposure after etching 1–8 hours. The background tracks from radon and its progeny were distinguished from artifact tracks manually. Although the number and size distribution of artifact tracks increase slightly with the etching time, most of the artifact tracks are less than 12 μm in diameter.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F003.jpg)
3.2 Size distribution of alpha tracks
Figure 4 illustrates the track images of the radon-exposed CR-39 detectors within an area of 0.5 mm ×0.5 mm after etching for 1–8 h. Some inclined tracks occur for the samples of T≤ 5 h, while most of the tracks become quasi circular or circular for the samples of T >5 h. This indicates that the tracks might be over-etched at T > 5 h[10]. Meanwhile most of the tracks enlarge with increasing etching time, hence the increase of possibility of overlapping tracks.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F004.jpg)
Figure 5 shows the size distributions of alpha tracks on the CR-39 detectors at different etching hours. They were re-plotted by subtracting size distributions of the control group from the experimental group for corresponding etching hour. Most of the tracks enlarged with etching time in different growth rates. The track sizes for 1–8 etching hours are 9.4±3.3 μm (1.6–17.3 μm), 13.2±4.9 μm (1.6–24.6 μm), 18.5±7.2 μm (1.6–36.5 μm), 24.8±9.0 μm (1.6–48.8 μm), 33.3±10.9 μm (1.6–66.2 μm), 39.5±13.8 μm (6.2–78.4 μm), 44.6±17.3 μm (6.2–90.2 μm) and 50.1±19.7 μm (6.2–96.5 μm), respectively. The standard deviation of the track sizes expanded from 3.3 μm at 1 h to 19.7 μm at 8 h. This indicates that negative effects in track counting may occur for an automatic counting system, especially in recognition of overlapped tracks.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F005.jpg)
3.3 Errors in track discrimination
Despite technique developments in automatic discrimination of tracks in the past decades[18-20], the threshold of track size was still a fundamental parameter for the automatic counting system to eliminate most of the artifact tracks. An appropriate threshold can significantly reduce uncertainty in track discrimination. In general, two types of discrimination errors are considered. Type-I is to reject small-size alpha tracks, while Type-II is to identify artifact tracks as big as alpha tracks. Having neglected the influence of geometrical parameters other than the diameter, Table 1 lists the two types of errors for different thresholds and etching time based on the experimental data. The results show that Type-II error deceases rapidly with the threshold, while Type-I error depends on the etching time obviously. At etching time T = 5–8 h, it was reasonable to set the threshold at Φ12 μm. Therefore, taking the discrimination error, deviation in track size and time cost into account, an appropriate etching time for this system is 5h.
Threshold of diameter (μm) | 1h | 2h | 3h | 4h | 5h | 6h | 7h | 8h | ||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Ia | IIb | Ia | IIb | Ia | IIb | Ia | IIb | Ia | IIb | Ia | IIb | Ia | IIb | Ia | IIb | |
6 | 45.6 | 8.3 | 16.9 | 22.4 | 9.4 | 23.7 | 3.3 | 26.4 | 0.4 | 28.9 | 0.0 | 29.0 | 0.0 | 29.2 | 0.0 | 31.2 |
12 | 97.9 | 2.1 | 70.4 | 5.6 | 32.8 | 6.1 | 14.4 | 6.5 | 3.2 | 7.3 | 2.3 | 7.6 | 1.8 | 7.8 | 1.2 | 8.1 |
18 | 99.8 | 0.0 | 92.5 | 3.2 | 63.8 | 3.8 | 34.7 | 4.0 | 14.1 | 4.4 | 10.5 | 4.2 | 8.0 | 4.0 | 6.5 | 3.9 |
3.4 Overlap correction
Based on size distribution of the tracks etched for 5 h, the overlap correction curve for the simulated track densities is plotted in Fig. 6. It shows that the overlap effect may cause nearly 10% loss in track counting when the track density is 50 tracks·mm−2, and the counting loss will exceed 70% when the track density reaches to 200 tracks·mm−2. It indicates that the overlap correction is important for an accurate counting when the track density is greater than several tens.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F006.jpg)
3.5 Experimental verification results
Figure 7 shows the relationships between track density and radon exposure before and after overlap correction. The mean track density has a non-linear response to the radon exposure, while the corrected track density has an excellent linear response to the integrated radon exposure. After overlap correction, the range of linear response for PIRM can reach to at least 2500 kBq·m−3·h with a detection efficiency of 6.6×10−3 tracks·cm−2·Bq−1·m3·h−1. The results revealed that the simulation provides a reliable overlap correction.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F007.jpg)
3.6 Uncertainty of track counting
The uncertainties of alpha track counting for an automatic system mainly result from three factors, i.e., the error of track discrimination, the uncertainty due to overlap effect and the statistical fluctuation of incident particles. The error in track discrimination is usually considered as the inherent characteristics of an automatic reading system, which is determined by the mechanical and optical properties, and the discrimination strategy as well, of the system. However, the uncertainties due to overlap effect and the statistical fluctuation of incident particles depend greatly on the track density.
Figure 8 shows the two uncertainties as a function of track density. The relative standard uncertainty due to overlap effect, UOL, was estimated by Eq. (1) based on Monte Carlo simulation, while the relative standard uncertainty due to statistical fluctuation was calculated by UST=(nmeanSK)−1/2,[21] where nmean is the mean track density on CR-39, in tracks·mm−2; K is the overlap correction factor; and S is the scanned area on each CR-39, in mm2. The results indicate that the uncertainty due to overlap effect increases with the track density, while the statistical fluctuation shows the opposite trend. Based on the etching conditions (6.25 N of NaOH, 90±0.1°C, 5h) in this study, it was estimated that the relative uncertainty of the track counting system would be less than 6%, at track density of 10–160 tracks·mm−2.
-201711/1001-8042-28-11-011/alternativeImage/1001-8042-28-11-011-F008.jpg)
4 Conclusion
In this study, to minimize and quantify the uncertainty of an automatic counting system, the optimal etching time and the discrimination parameters were determined through experimentally investigating the morphological characteristics of tracks on the CR-39, and the overlap correction and its uncertainty was systematically discussed. The result indicates that the overlap correction is essential for an accurate counting. By taking the overlap correction into account, the uncertainty of an automatic counting system could be controlled to be less than 6% for alpha tracks in its general application. It is expected that the results of this study would benefit the quality control for the automatic counting system.
The precise energy spectra measurement of laser-accelerated MeV/n-class high-Z ions and protons using CR-39 detectors
. Plasma Phys. Contr. F. 58, 034013 (2016). doi: 10.1088/0741-3335/58/3/034013Occupational exposure to radon in different kinds of non-uranium mines
. Radiat. Prot. Dosim. 170, 311-314 (2016). doi: 10.1093/rpd/ncw026An experimental study on optimal chemical etching condition for CR-39 used in neutron detection
. Nucl. Tech. 39, 55-60 (2016). (in Chinese) doi: 10.11889/j.0253-3219.2016.hjs.39.060402Overview of PADC nuclear track readers
. Recent trends and solutions. Radiat. Meas. 44, 821-825 (2009). doi: 10.1016/j.radmeas.2009.10.097TRIAC: A code for track measurements using image analysis tools
. Nucl. Instrum. Meth. B 244, 392-396 (2006). doi: 10.1016/j.nimb.2005.10.026A fully automated optical microscope for analysis of particle tracks in solids
. Nucl. Instrum. Meth. B 198, 98-107 (2002). doi: 10.1016/S0168-583X(02)01484-2An automatic image analysis system for the evaluation of dielectric track detectors
. Radiat. Meas. 34, 187-188 (2001). doi: 10.1016/S1350-4487(01)00148-2Automatic evaluation of solid state track detectors by artificial vision
. Comput. Electr. Eng. 39, 274-284 (2013). doi: 10.1016/j.compeleceng.2012.11.010Construction of an environmental radon monitoring system using CR-39 nuclear track detectors
. Nucl. Eng. Technol. 37, 395-400 (2005). (in Chinese)Formation and growth of tracks in nuclear track materials
. Mater. Sci. Eng. R. 46, 51-123 (2004). doi: 10.1016/j.mser.2004.07.003Effects of etching time on alpha tracks in solid state nuclear track detectors
. Sci. Total Environ. 575, 905-909 (2017). doi: 10.1016/j.scitotenv.2016.09.147Particularization of alpha contamination using CR-39 track detectors
. Pramana-J. Phys. 69, 567-574 (2007). doi: 10.1007/s12043-007-0156-8Experimental study on CR-39 response to alpha particles with different incident angles and energy
. At. Energy Sci. Technol. 42, 322-325 (2008). (in Chinese)Effects of atmospheric parameters on radon measurements using alpha-track detectors
. Rev. Sci. Instrum. 85, 022101 (2014). doi: 10.1063/1.4865155A new model to account for track overlap in CR-39 data
. Nucl. Instrum. Meth. A 681, 84-90 (2012). doi: 10.1016/j.nima.2012.04.021Assessment of the uncertainties in the Radiological Protection Institute of Ireland (RPII) radon measurements service
. J. Environ. Radioact. 99, 1578-1582 (2008). doi: 10.1016/j.jenvrad.2007.12.018A theoretical model for the overlapping effect in solid state nuclear track detectors
. Nucl. Instrum. Meth. A 652, 550-553 (2011). doi: 10.1016/j.nima.2010.12.223Measurement of solid state nuclear track parameter
. Nucl. Tech. 22, 590-593 (1999). (in Chinese)New code for digital imaging system for track measurements
. Nucl. Instrum. Meth. B 160, 550-555 (2000). doi: 10.1016/S0168-583X(99)00628-XAnalysis of 3-dimentional track parameters from 2-dimensional images of etched tracks in solid polymeric track detectors
. Nucl. Instrum. Meth. A 690, 58-67 (2012). doi: 10.1016/j.nima.2012.06.054Some problems on the measurement of 222Rn concentrations by passive cup method
. Health Phys. 69, 508-512 (1995). doi: 10.1097/00004032-199510000-00009