logo

Variation in patient dose due to differences in calibration and dosimetry protocols

NUCLEAR CHEMISTRY, RADIOCHEMISTRY, NUCLEAR MEDICINE

Variation in patient dose due to differences in calibration and dosimetry protocols

Wazir Muhammad
Asad Ullah
Gulzar Khan
Tahir Zeb Khan
Tauseef Jamaal
Fawad Ullah
Matiullah Khan
Amjad Hussain
Nuclear Science and TechniquesVol.29, No.5Article number 66Published in print 01 May 2018Available online 30 Mar 2018
57600

For precise and accurate patient dose delivery, the dosimetry system must be calibrated properly according to the recommendations of standard dosimetry protocols such as TG-51, TRS-398, etc. However, the dosimetry protocol followed by a calibration laboratory is usually different from the protocols that are followed by different clinics, which may result in variations in the patient dose. Our prime objective in this study was to investigate the effect of the two protocols on dosimetry measurements. Dose measurements were performed for a Co-60 teletherapy unit and a high-energy Varian linear accelerator with 6 and 15 MV photon and 6, 9, 12, and 15 MeV electron beams, following the recommendations and procedures of the AAPM TG- 51 and IAEA TRS-398 dosimetry protocols. The dosimetry systems used for this study were calibrated in a Co-60 radiation beam at the Secondary Standard Dosimetry Laboratory (SSDL) PINSTECH, Pakistan, following the IAEA TRS-398 protocol. The ratio of the measured absorbed doses to water in clinical setting, Dw (TG-51/TRS-398), were 0.999 and 0.997 for 6 and 15 MV photon beams, whereas these ratios were 1.013, 1.009, 1.003, and 1.000 for 6, 9, 12, and 15 MeV electron beams, respectively. This difference in the absorbed doses to water Dw ratio may be attributed mainly due to beam quality (KQ) and ion recombination correction factor.

Radiation dosimetryDosimetry systems calibrationsPatient doseAbsorbed dose to water ratioAAPM TG- 51IAEA TRS-398

Introduction:

The quantitative determination of energy deposition in a given medium by directly or indirectly ionizing radiation is called radiation dosimetry[1, 2]. Radiation dosimetry has a vital role in the medical field because the irradiation of patients is totally based upon the values determined by it[3]. The International Atomic Energy Agency (IAEA) and several other international organizations have published their dosimetry protocols for calibration of photon and electron beams during the last several decades. With the passage of time, these protocols have been updated for improvement in the accuracy in dosimetry of high-energy photons and electrons beams [4]. In connection with these advances in dosimetry, the two latest dosimetry protocols, those of Task Group 51 (TG-51) and TRS-398, have been implemented by the American Association of Medical Physicists (AAMP) and the IAEA, respectively [5, 6]. Both protocols are based on the calibration of an ionization chamber in terms of absorbed dose to water (NDw,Q0) in a standard reference quality beam, Q0, from which the dose to water in the user radiation beam is derived [7, 8]. In these protocols, different experimental approaches have been adopted, which results in variations in the patient dose.

Huq et. al. [7] have compared these two protocols by analyzing differences in the basic input data for photon and electron beam dosimetry. Experiments were performed using 6, 18, and 25 MV clinical photon beam energies and 6, 8, 10, 12, 15, and 18 MeV electron beams energies[7]. In this study, two Farmar-type chambers and three plane-parallel chambers, calibrated by a US accredited dosimetry calibration laboratory (K & S Associates, Inc, Nashville, TN, USA), were used [7]. It should be mentioned here that, in Pakistan, there is only one secondary standard dosimetry laboratory (SSDL) which provides calibration of radiation instruments on the national level[2, 9]. The SSDL follows the IAEA TRS-398 dosimetry protocols for the calibration of radiation instruments, whereas physicists in some medical centers either follow AAPM TG-51 or IAEA TRS-398 for the radiation beam output measurement of different machines.

Calibration factors are provided under standard environmental conditions of temperature, pressure, and relative humidity according to the recommendations of followed dosimetry protocol. In hospitals, measurement conditions are normally different from the standard laboratory conditions recommended by each protocol. The difference in the environmental conditions affects the response of the ionization chamber. Therefore, correction factors must be applied for converting the cavity air mass to the reference conditions. Our main objective in the present study was to find out the effect of environmental conditions on the patient dose.

Materials and Methods:

In this study, dosimetry was performed for two therapeutic units, namely a Co-60 unit (THERATRON PHOENIX) and a high-energy Varian linear accelerator (CLINAC DMX 2100C) with 6 and 15 MV photon and 6, 9, 12, and 15 MeV electron beams. The dosimetry system consisted of a PTW stationary water phantom (T41014) with a cylindrical chamber (PTW-30001) connected to an electrometer (PTW UNIDOS E), which was used for the absolute dosimetry of the THERATRON PHOENIX. A Farmer-type ionization chamber (IBA-FC65-G) and a plane-parallel chamber (IBA PPC-05) were used for absolute dosimetry of the two photon and four electron beams that were generated by the CLINAC DMX 2100C, respectively. The chambers were connected to PTW UNIDOS E, attached to Blue water Phantom, and each chamber type combined with PTW UNIDOS E was calibrated in a Co-60 radiation beam with its reference point at a measuring depth of 5 cm in water at the SSDL at PINSTECH, Pakistan. The calibration was performed under the standard conditions of the IAEA TRS-398 protocol.

The clinical dosimetry was performed following the recommendations and reference conditions of the AAPM TG- 51 and IAEA TRS-398 protocols [10, 11]. The depth dose curves were obtained experimentally with the help of a Blue Phantom scanner system loaded with the software Scanditronix-Wellhöfer OmniPro-Accept for all photon and electron beams, as shown in Figs. 1a and 1b[12, 13, 14]. The relative measurements were evaluated with the software Scanditronix-Wellhöfer OmniPro-Accept. The tissue phantom ratio (TPR20, 10) and percentage depth dose (%dd(10)) were determined from the curves given in Fig. 1a, as these values are required for all of the photon beams under study, as per the requirements of the TRS-398 and TG-51 protocols. From the measured TPR20,10 and %dd(10), beam quality factors were determined. For each electron beam under study, the reference depth (zref) and R50 (i.e., "depth in water in a 10 x 10 cm2 or larger beam of electrons at an SSD of 100 cm at which the absorbed dose falls to maximum and 50% of the dose maximum, respectively" [5]) were calculated from the depth dose curves given in Fig. 1b as per the requirements of TRS-398 & TG-51. Similarly, a polarity correction factor and an ion recombination correction factor were calculated according to the guidelines provided in the TG-51 and TRS-398 protocols for each beam generated by the CLINAC DMX 2100C. Multiple measurements were performed in each session, and were used as average values to minimize the error in the experimental values. Finally, the dose calculations were performed according to the recommendation of the AAPM TG-51 and IAEA TRS-398 protocols [5, 6].

Fig. 1
(Color online) (a) 6 and 15 MV photon depth dose curves. (b) Electrons depth dose ionization curves of 6, 9, 12, and 15 MeV
pic

Results and Discussion:

Photon Dosimetry:

Table 1 shows beam quality specifiers, namely, TPR20,10 and %dd(10) X, for 6 and 15 MV photon beams which have been calculated from Fig. 1a. With these values, a beam quality correction factor, Kq, for both photon energies (6 and 15 MV) was determined against the provided values of either TPR20,10 or %dd(10) X [5, 6], summarized in Table 1. In the case of TG-51, the Kq value is smaller than that of TRS-398. The percentage difference in Kq is - 0.1% and - 0.3% for 6 and 15 MV photon beams, respectively.

Table 1:
Comparison of Kq, Kion, Kpol, and absorbed doses at Zmax (DW(Zmax)) values calculated based on the recommendations of AAPM TG-51 and IAEA TRS-398 protocols for 6, 15 MV and Co-60 photons beams
Energy AAPM TG-51 IAEA TRS-398
  KTP %dd(10)x Kq* Kion Kpol DW(Zmax) KTP TPR20,10 Kq** Kion Kpol DW(Zmax)
6 MV 1.16055 66.7 0.991 1.00734 1.00054 0.9970 1.16055 0.668 0.992 1.00714 1.00054 0.9975
15 MV 1.16055 77.63 0.970 1.01135 1.00129 0.9967 1.16055 0.762 0.973 1.01109 1.00129 0.9995
Co-60 1.16055 NA 1.0 1.0 1.0 1.6489 1.16055 NA 1.0 1.0 1.0 1.6489
Show more
* depends upon the type of ionization chamber and beam energy. For 6 MV beam, %dd(10) X =%dd(10) while for 15 MV, %dd(10) X =1.267%dd(10)-20.0 was used. Kq for the ionization chamber was determined from the tabulated values of Kq and corresponds to %dd(10) X for the ionization chamber [5]. The interpolation techniques was used while using these tabulated values.

The beam quality, type of chamber, and some other measurement conditions, such as cable length and position etc., may affect the polarity for a particular ionization chamber. Therefore, a polarity correction factor, Kpol, must be applied for beams of different qualities. Furthermore, due to the Lack of ion collection on the corresponding electrodes of the ion chamber, some of the ions produced may not contribute toward the actual signal. Therefore, recombination of ions (i.e., ion recombination correction factor, Kion) has to be applied. Kion depends upon the dose rate, the chamber geometry, and the applied polarizing voltage. Although the method (two-voltage method) for both protocols under study is the same, their measurement approaches are different. Similarly, the temperature and pressure correction factor, KT,P, also must be applied, as the clinical conditions are always different from the standard conditions under which the calibration of the dosimetry systems is performed. The KT,P, Kpol, and Kion were measured for both protocols according to the recommendations and procedures. These values are summarized in Table 1.

The values of KT,P for both dosimetry protocols in the case of both 6 and 15 MV-photon beams were same as the calibration for the dosimetry systems performed under the standard conditions of the IAEA TRS-398 protocol. Figs. 2a and 2b show a comparison of the Kion and Kq values for both protocols, whereas Fig. 2 (c) shows polarity correction factors for 6 and 15 MV-photon beams for both protocols. As both the protocols follow the same approach for measurement of the polarity effect, contribution of polarity correction factor in the dosimetryis similar.

Fig. 2
(a) Comparison of Kion of TG-51 and TRS-398 protocol. (b) Comparison of Kq in both protocols. (c) Calculated value of Kpol vs energy.
pic

By comparing the values of Kion and Kq, calculated based on recommendations of both protocol for corresponding beams, a measurable difference was observed for the studied photon beams (see, Fig. 2-a, -b). In the case of the TG-51 protocol, the Kion had a relatively greater value for both photon beams, as shown in Fig. 2a, while, Kq had a relatively smaller value for both photon beams, as shown in Fig.2b. The results also showed that the polarity correction factor depends upon the photon beam energy. The polarity correction factor increases with an increase in the energy of the photon beam (see Fig. 2c).

The absorbed doses (DW(Zmax)) were calculated and normalized to Zmax for the Co-60 beam and 6 and 15 MV photon beams (Table 1) by using the %depth doses (%DD) that correspond to each beam quaility. Fig. 3 shows a comparison of the DW(Zmax) ratio (i.e., DW(Zmax)TG-51/ DW(Zmax)TRS-398) of Co-60 for the 6 and 15 MV photon beams for TG-51 and TRS-398 protocols. The same values of DW(Zmax) were observed for both protocols in the case of the Co-60 tele-therapy unit, as there was no difference between the measured parameters. For the 6 and 15 MV photon beams, the percentage difference in absorbed dose was -0.05% and -0.3% respectively. The TRS-398 gives relatively higher doses as compared to that of the TG-51 protocol. Moreover, the percentage difference in the calculated dose increased with an increase in the energy.

Fig. 3
Ratio of the absorbed doses at Zmax (i.e., DW(Zmax)TG-51/ DW(Zmax)TRS-398) for Co-60, for 6 MV and 15 MV photons, by comparing the two protocols (TG-51 and TRS-398).
pic
Electron Dosimetry:

Following the recommendations and procedures adopted in both TG-51 and TRS-398 protocols, Kq, KT,P, Kpol, and Kion for 6, 9, 12, and 15 MeV-electron beams were measured. These values are summarized in Table 2. The KT,P were the same for both protocols, as explained earlier in the section on photon dosimetry. The calculated values of R50 and the corresponding values of the reference depth, Zref, for electron energies of 6, 9, 12, and 15 MeV are also summarized in Table 2. According to the TRS-398 protocol, the Kq value is directly provided against R50, whereas, for TG-51 protocol, Kq was calculated (see Table 2) from R50 for each electron energy [5]. The calculated absorbed dose normalized to Zmax for the 6, 9, 12, and 15 MeV electron beams for the TG-51 and TRS-398 protocols is also listed in Table 2.

Table 2:
Comparison of Kq, Kion, Kpol, and absorbed doses at Zmax (DW(Zmax)) values calculated on the basis of the recommendations of the AAPM TG-51 and IAEA TRS-398 protocols for 6, 9, 12, and 15 MeV electron beams.
Energy (MeV) AAPM TG-51 IAEA TRS-398
  KTP Zref*(cm) R50**(cm) Kq Kion Kpol DW(Zmax) KTP Zmax (cm) R50***(cm) Kq Kion Kpol DW(Zmax)
6 1.17 1.19 2.15 0.94 1.023 1.013 1.026 1.17 1.08 2.15 0.92 1.023 1.013 1.013
9 1.17 1.98 3.46 0.92 1.010 1.005 1.006 1.17 1.99 3.42 0.91 1.005 1.010 0.996
12 1.17 2.85 4.91 0.91 1.013 1.008 1.006 1.17 2.44 4.83 0.91 1.013 1.008 1.003
15 1.17 3.64 6.24 0.90 1.014 1.007 0.910 1.17 2.80 6.12 0.90 1.013 1.007 0.998
Show more
* Zref for electrom beams is the depth in water in a 10 x 10 cm2 or larger beam of electrons at an SSD of 100 cm at which the absorbed dose falls to 50% of its maximum [5]. It was determined from Fig. 1b

Figs. 4a and 4b show a comparison of the Kion and Kq values for both protocols, and Fig. 4c shows Kpol for 6, 9, 12, and 15 MeV electron beams. There was no regular pattern recorded in the variation in Kion with a change in the energy of the beam. Both the dosimetry protocols followed the same irregular pattern of variation in Kion (see Fig. 4a). By inter-comparison of both protocols for the values of Kq, (i.e., by keeping the energy of the beam constant) higher values of Kq for the TG-51 protocol as compared to the TRS-398 protocol were observed for all the studied electron beams as shown in Fig. 4b. Furthermore, variation in Kq decreases as the electron beam energy increases. The percentage difference between the Kq values for 6, 9, 12, and 15 MeVelectron beams where 1.5%, 0.9%, 0.3%, and 0.0%, respectively, for both protocols. Fig. 4c shows a variation in Kpol with a change in the energy of the electron beam. Kpol decreases with an increase in the energy of the electron beam.

Fig. 4
(a) Comparison of the Kion of TG-51 and TRS-398 protocols. (b) Comparison of the beam quality factor, Kq, for both protocols. (c) Calculated Kpol vs energy.
pic

The ratio of the absorbed doses at Zmax corresponding to TG-51 and TRS-398 is illustrated in Fig. 5. The percentage differences in the measured dose for both protocols for 6, 9, 12, and 15 MeV electron beams were 1.3%, 0.9%, 0.3%, and 0.1%, respectively. In this case, the TG-51 gives relatively high doses as compared to the TRS-398 protocol, and it showed an inverse relationship with energy of the electron beams.

Fig. 5
Ratio of the absorbed doses at Zmax (i.e., DW(Zmax)TG-51/ DW(Zmax)TRS-398) for 6, 9, 12, 15 MeV electron beam by use of two protocols (TG-51 and TRS-398).
pic

Conclusion:

To conclude, small differences in the absorbed doses to water at the Zmax ratio were found for the two protocols. The TRS-398 protocol gave higher doses as compared to the TG-51 protocol. The percentage difference in the measured absorbed dose increased as the energy increased in the case of the photon beams, however, for the electron beams, TG-51 gave relatively higher doses as compered to the TRS-398 protocol. Unlike photon beams, the percentage difference in the measured absorbed dose decreased with an increase in the energy. To reduce uncertainty in the patient dose, the clinically followed dosimetry protocol and dosimetry protocol followed for the calibration of a dosimetry system should be the same.

References:
1 J. Seuntjens, W. Strydom, K. Shortt in. Radiation Oncology Physics: A Handbook for Teachers and Students, Vol. Ed. Vienna, Austria: International Atomic Energy Agency (IAEA). 2005) p. 45. (2005)
2 W. Muhammad, A. Ullah, K. Mahmood, et al.

Assessment of national dosimetry quality audits results for teletherapy machines from 1989 to 2015

. J. Appl. Clinic. Med. Phy. 17: 145-152 (2016). doi: 10.1120/jacmp.v17i2.5984
Baidu ScholarGoogle Scholar
3 W. Muhammad, H. Sang, A. Khan, et al.,

Dose non-linearity of the dosimetry system and possible monitor unit errors on medical linear accelerators used in conventional and intensity-modulated radiation therapy

. Nucl. Technol. Radiat. 27, 368-373. (2012). doi: 10.2298/NTRP1204368M
Baidu ScholarGoogle Scholar
4 P. Andreo, M. S. Huq, M. Westermark, et al.,

Protocols for the dosimetry of high-energy photon and electron beams: a comparison of the IAEA TRS-398 and previous international Codes of Practice

. Phys. Med. Biol. 47, 3033. (2002). doi: 10.1088/0031-9155/47/17/301
Baidu ScholarGoogle Scholar
5 P. R. Almond, P. J. Biggs, B. Coursey, et al.,

AAPM's TG-51 protocol for clinical reference dosimetry of high-energy photon and electron beams

. Med. Phys. 26, 1847. (1999). doi: 10.1118/1.598691
Baidu ScholarGoogle Scholar
6 P. Andreo, D. Burns, K. Hohlfeld, et al.,

Absorbed Dose Determination in External Beam Radiotherapy: An International Code of Practice for Dosimetry Based on Standards of Absorbed Dose to Water. Technical Reports Series No. 398

(2000).
Baidu ScholarGoogle Scholar
7 M. S. Huq, P. Andreo, H. Song,

Comparison of the IAEA TRS-398 and AAPM TG-51 absorbed dose to water protocols in the dosimetry of high-energy photon and electron beams

. Phys. Med. Biol. 46, 2985. (2001). doi: 10.1088/0031-9155/46/11/315
Baidu ScholarGoogle Scholar
8 S. Saju, S. Jamema, D. Kumar, et al.,

Comparison Of IREA TRS 398, TRS 381 And AAPM TG 51 protocols For High Energy Electron Beams

. J. Med. Phy. 29, 9-16. (2004).
Baidu ScholarGoogle Scholar
9 W. Muhammad, A. Ullah, A. Hussain,

EP-1502: Effects on dosimetric measurements due to difference in calibration and dosimetry protocols followed

. Radiother. Oncol. 119 (Supplement 1), S694. (2016). doi: 10.1016/S0167-8140(16)32752-9
Baidu ScholarGoogle Scholar
10 M. Affonseca, P. Andreo, M. Arib, et al. Implementation of the international code of practice on dosimetry in radiotheraphy (TRS 398): review of testing results. International atomic energy agency (IAEA).Vienna, Austria: Volum 9201050054. (2005)
11 M. S. Huq.

Everything you wanted to know about the Practical Implementation of TG-51 protocol in the clinic

; 2003. Amer Assoc Physicists Medicine Amer Inst Physics STE 1 No 1, 2 Huntington Quadrangle, Melville, NY 11747-4502 USA. p. 1378.
Baidu ScholarGoogle Scholar
12 M. Ahmad, A. Hussain, W. Muhammad, et al.,

Studying wedge factors and beam profiles for physical and enhanced dynamic wedges

. J. Med. Phys. 35, 33-41. (2010). doi: 10.4103/0971-6203.57116
Baidu ScholarGoogle Scholar
13 W. Muhammad, M. Maqbool, M. Shahid, et al.,

Accuracy checks of physical beam modifier factors algorithm used in computerized treatment planning system for a 15 MV photon beam

. Rep. Pract. Oncol. Radiot. 14, 214-220. (2009). doi: 10.1016/j.rpor.2009.12.002
Baidu ScholarGoogle Scholar
14 W. Muhammad, M. Maqbool, M. Shahid, et al.,

Assessment of computerized treatment planning system accuracy in calculating wedge factors of physical wedged fields for 6 MV photon beams

. Phys. Medica 27, 135-143. (2011). doi: 10.1016/j.ejmp.2010.06.003
Baidu ScholarGoogle Scholar