1 Introduction
Development of silicon detectors has been receiving benefits from the rapid technological advancement of the integrated circuit (IC) since the 1970s, when the earliest attempts to introduce planar process to the fabrication of silicon detectors took place [1-4]. Techniques like oxide passivation, ion implantation and photolithography have successfully resolved the key issues encountered prior to the planar technology, granting feasibility to the fabrication of silicon strip detectors (SSD). The advantages of an SSD include precise positioning, high energy resolution, long range of linear response, rapid response time and its feasibility to be fabricated into various designated shape, facilitating its applicability in many research areas.
Since the beginning of the 21st century, many detector arrays composed of SSD like CD [5], DRAGON [6], HiRA [7], LEDA [8], MUST [9] and MUST2 [10] have been constructed to meet individual experimental objectives. In China, several institutions such as the China Institute of Atomic Energy (CIAE), the Beijing Kelixing Photoelectric Technology Co., Ltd. and the Institute of Modern Physics, Chinese Academy of Science (IMP) have begun the development of double-sided silicon strip detectors (DSSD). Prior to this project, CIAE had successfully developed the single-sided silicon strip detectors (SSSD) that have been used in many nuclear experiments [11-14] for about a decade. A more recent development by CIAE in the quadrant-silicon detectors (QSD) [15] has also been shown successful in providing a new detector technology to various nuclear experiments [16-18]. In this article, a cooperative production of DSSD by CIAE and the Beijing Kelixing Photoelectric Technology will be introduced, and performance test conducted at the University of Hong Kong will be presented.
2 Detection principle
The DSSD being developed in this work has an active area of 48 mm × 48 mm and a thickness of 300 μM. The P+ doped junction side and the N+ doped ohmic side comprise the basic framework of PN-junction. At each side, all the 48 strips are evenly distributed and mutually parallel, each with a strip width of 0.9 mm and an inter-strip distance of 0.1 mm(Fig. 1). At an adequate supply of bias voltage around 60 V, the depletion region will expand and occupy the entire Si wafer.
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F001.jpg)
When charged particle penetrates into the active region, electron-hole pairs will be produced. Due to the presence of electric field, electrons will be drifted towards the ohmic side, while the holes will be drifted towards the junction side. Electrical pulses at the electrodes are then being generated due to the induction caused by the motions of these charges. Positive signal pulses and negative signal pulses will be received by the junction side and the ohmic side, respectively, with energy loss information of the particle being encoded in the pulses. Each silicon strip could be treated as an individual detector during the signal collection.
3 Manufacturing procedure
The entire manufacturing procedure of the DSSD takes place inside a cleanroom. An N-type 4-inch silicon wafer of 300 μM in thickness and 10 kΩ in resistivity is used for fabrication. After polishing and cleaning, the Si wafer undergoes the following manufacturing procedure:
a. Oxide passivation: The Si wafer is being exposed under high-temperature environment, with O2 gas stream flowing past its surface so that an oxide layer SiO2 of 600 nm thick is formed on top of it.
b. Photolithography: With the use of photolithographic equipment and proper etching techniques, some segments of the oxide layer are removed to form pattern of inter-strips on both sides.
c. Ion implantation: Junction side is doped with boron ions, forming a P+ region; ohmic side is doped with phosphorus ions, forming an N+ region. At both sides, the thickness of doping is around 400 nm. This step creates the P+-Intrinsic-N+ (PIN) diode structure for the detector. In order to improve isolation on the ohmic side, P-stop technique has been employed to isolate the N+ strips, i.e. boron ions have been doped into the N+ inter-strip region and its surrounding to form P+ isolation implants.
d. Thermal annealing: The doped Si wafer is being exposed under dry nitrogen gas at 600 ℃ for 30 min.
e. Al metalization and patterning: A thin layer of Al is being sputtered onto both sides, and those which fall on top of the oxide layer are then being removed.
f. Alloy and encapsulation: Each individual strip is being separated and encapsulated before the detector chip is mounted onto the printed circuit board (PCB). The mounting employs ultrasonic soldering to establish permanent connection between the readout contacts on the PCB and the Al electrodes of DSSD which have been alloyed at 380 ℃.
All these procedures follow the fabrication parameters which have been optimized beforehand based on simulations or previous experience. One noteworthy issue is the tendency for N+ strips to accumulate charges, which can easily introduce electrical breakdown across inter-strip region. Hence, P-stop technique has been implemented in this work to enhance strip isolation. A schematic figure that illustrates the important structural features of DSSD is given in Fig. 2.
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F002.jpg)
4 Characterization
A schematic diagram for the readout electronics is shown in Fig. 3. Each junction strip is supplied with a negative bias from the Mesytec MHV-4 via a preamplifier of internal resistance 100 MΩ, while each ohmic strip is connected to ground through an 100 MΩ resistor. Self-developed integrated charge sensitive preamplifiers SPA02 (bandwidth of 275 MHz and signal-to-noise ratio of 5
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F003.jpg)
To investigate the performance of DSSD, alpha response is chosen to be studied by locating a triple alpha source that consists of 239Pu, 241Am and 244Cm at approximately 3.0 cm away from the detector. Throughout the test, the entire piece of DSSD as well as the preamplifiers are all placed within a vacuum chamber with an air pressure of 10-3 Pa.
4.1 Voltage amplitude, rise time and leakage current
Voltage amplitude and rise time of pre-amplified signals have been measured as functions of bias voltage using an oscilloscope. When increasing bias voltage from 0 V to 80 V, both signal amplitude and rise time turn saturated after 60 V (see Fig. 4). At full depletion, body capacitance of each strip to the opposite side is around 50 pF. In practice, a bias higher than the saturation voltage could be taken, provided that the bias does not exceed breakdown voltage. Throughout the test, a bias of 80 V has been adopted and the total leakage currents for all strips was found to be less than 720 nA, that is, each strip has less than 15 nA on average (Fig. 5).
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F004.jpg)
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F005.jpg)
4.2 Energy resolution of a single strip
A triple alpha source containing 239Pu, 241Am and 244Cm has been used. Their primary peak energies are given by 5157 keV, 5486 keV and 5805 keV, respectively. Energy calibration has been performed for each strip individually prior to analysis. Using linear calibration, E(x) = mx + c, where x is the channel number, we find that m fluctuates less than 5% in all strips, while c shifts less than 1% of the full energy range ( 10000 keV ). Energy spectra collected by the 24th junction strip and the 24th ohmic strip have been given in Fig. 6 for illustration. Limited by the energy resolution of detector, a low-energy tail is formed from the superposition of primary peak with non-primary peaks - take 239Pu for example, the major alpha decay paths include 5157 keV (71%), 5144 keV (17%) and 5106 keV (12%). At a bias of 80 V, the energy resolutions of all strips are found to fluctuate within 0.65% - 0.80% for the junction strips and 0.85% - 1.00% for the ohmic strips, while the electronic noise is found to contribute around 0.20% to the resolution [20].
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F006.jpg)
4.3 Absence of crosstalk between adjacent strips
Geometry of the strips inevitably introduces unwanted coupled capacitance. When an incoming charged particle hits a particular strip, charges are being accumulated in it, causing induced charges to arise in its adjacent strips due to the presence of inter-strip capacitance [9, 21]. Being referred as crosstalk in the literature [22, 23], this effect introduces pseudo-signal in the electronics at low energy region, even though it has no apparent effect on the energy resolution of alpha signal. The significance of crosstalk depends on inter-strip capacitance Ci and body capacitance of the strip to the opposite side Cb. Significance of crosstalk is associated with the ratio Ci / Cb, so it is desirable to have it as small as possible. Nevertheless, increasing Cb leads to weakening of signal amplitude, hence elimination of crosstalk was approached by minimizing Ci. For our detector, Cb ≈50 pF and Ci is small enough to reject crosstalk between adjacent strips, as will be discussed in the next paragraph.
Ideally during the absence of crosstalk, only the incidences that have successfully penetrated through the inter-strip SiO2 layer and injected into the active region will generate coincidence signals at adjacent strips. When plotted as a two-dimensional spectrum, these events are expected to have their energies shared by the adjacent strips, where the energy distribution is linearly related to the proximity of electron-hole pair creation point to the strip edge. This provides us a way to check the absence of crosstalk by plotting the coincidence events between the adjacent strips. As illustrated in Fig. 7, the expected energy distribution is observed for adjacent ohmic strips when particles enter from the junction side, indicating the absence of crosstalk on the ohmic side; if crosstalk had indeed occured, clusters besides from those three belts of alpha signals would have shown up. Absence of crosstalk on the junction side has also been confirmed by a similar manner. In our previous development, elimination of crosstalk could only be achieved on the junction side but not on the ohmic side [24]. In this test, such issue has been overcome as P+ isolation implants have successfully been implemented to the ohmic strips.
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F007.jpg)
4.4 Pixel
The primary objective of developing DSSD is to acquire two-dimensional spatial position information of the charged particles with the pixels resulted from the intersection of strips on both sides. In this development, 48 strips have been equipped on each side, yielding a total of 48 × 48 = 2304 pixels on the detector. The two-dimensional energy spectra of pixels have been investigated as illustrated in Fig. 8, where the clusters of events correspond to the alpha peaks in Fig. 6.
-201805/1001-8042-29-05-013/alternativeImage/1001-8042-29-05-013-F008.jpg)
5 Summary
In summary, a DSSD with 48 strips on each side has been developed and performance test has been conducted at a bias voltage of 80 V. Each strip carries a leakage current of around 15 nA. When alpha particles of 5157 keV enter from the junction side, rise time of signal after preamplifiers is about 15 ns on both sides. Energy resolutions are good, with the junction sides ranging within 0.65% - 0.80% and the ohmic sides ranging within 0.85% - 1.00%. Crosstalk is found to be insignificant on both sides. The detectability of pixels has been demonstrated. Overall, the results indicate that the DSSD being developed is ready to be used in nuclear physics experiments. In the future, we plan to develop DSSD of different thicknesses as well as of different shapes to best meet the goals of various experiments. At the same time, we also wish to improve the success rate of production.
Low-noise oxide passivated p+n silicon detectors
, Nucl. Instrum. Methods 101 (1972) 43-46. doi: 10.1016/0029-554X(72)90753-7Low-noise silicon planar detectors for room temperature application
, Nucl. Instrum. Methods 104 (1972) 209-214. doi: 10.1016/0029-554X(72)90319-9Fabrication of low noise silicon radiation detectors by the planar process
, Nucl. Instrum. Methods 169 (1980) 499-502. doi: 10.1016/0029-554X(80)90948-9Improvement of detector fabrication by the planar process
, Nucl. Instrum. Meth. A 226 (1984) 89-93. doi: 10.1016/0168-9002(84)90173-6CD: A double sided silicon strip detector for radioactive nuclear beam experiments
, Nucl. Instrum. Meth. A 480 (2002) 448-455. doi: 10.1016/S0168-9002(01)00954-8The DRAGON facility for nuclear astrophysics at TRIUMF-ISAC: design, construction and operation
, Nucl. Instrum. Meth. A 498 (2003) 190-210. doi: 10.1016/S0168-9002(02)01990-3The high resolution array (HiRA) for rare isotope beam experiments
, Nucl. Instrum. Meth. A 583 (2007) 302-312. doi: 10.1016/j.nima.2007.08.248Louvainedinburgh detector array (LEDA): a silicon detector array for use with radioactive nuclear beams
, Nucl. Instrum. Meth. A 454 (2000) 350-358. doi: 10.1016/S0168-9002(00)00479-4MUST: A silicon strip detector array for radioactive beam experiments
, Nucl. Instrum. Meth. A 421 (1999) 471-491. doi: 10.1016/S0168-9002(98)01178-4MUST II: Large solid angle light charged particle telescope for inverse kinematics studies with radioactive beams
, AIP Conference Proceedings 680 (2003) 313-316. doi: 10.1063/1.1619724Experimental study of two-proton correlated emission from 29S excited states
, Phys. Rev. C 80 (2009) 014310. doi: 10.1103/PhysRevC.80.014310Investigation of two-proton emission from excited states of the odd-Z nucleus 28P by complete-kinematics measurements
, Phys. Rev. C 81 (2010) 054317. doi: 10.1103/PhysRevC.81.054317Observation of two-α emission from high-lying excited states of 18 Ne by completekinematics measurements
, Phys. Rev. C 82 (2010) 064316. doi: 10.1103/PhysRevC.82.064316Correlations of two protons emitted from excited states of 28S and 27P
, Phys. Let. B 727 (2013) 126-129. doi: 10.1016/j.physletb.2013.10.029Development of large-area quadrant silicon detector for charged particles
, Chinese Phy. C 38(12) (2014) 126001. doi: 10.1088/1674-1137/38/12/126001A detection system for charged-particle decay studies with a continuous-implantation method
, Nucl. Instrum. Meth. A 804 (2015) 1-7. doi: 10.1016/j.nima.2015.09.039β-decay study of the Tz=-2 proton-rich nucleus 20Mg
, Phys. Rev. C 95 (2017) 014314. doi: 10.1103/PhysRevC.95.014314Observation of β-delayed two-proton emission in the decay of 22 Si
, Phys. Let. B 766 (2017) 312-316. doi: 10.1016/j.physletb.2017.01.028New data acquisition system for the RIKEN radioactive isotope beam factory, Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers
, Detectors and Associated Equipment 616(1) (2010) 65-68. doi: 10.1016/j.nima.2010.02.120Characterization of an NTD Double-Sided Silicon Strip Detector Employing a Pulsed Proton Microbeam
, IEEE Trans. Nucl. Sci., 64(9) (2017) 2551-2560 doi: 10.1109/TNS.2017.2734568Crosstalk properties of 36-fold segmented symmetric hexagonal HPGe detectors
, Nucl. Instrum. Meth. A, 599(2-3) (2009) 196-280. doi: 10.1016/j.nima.2008.11.011Development and test of double-sided silicon strip detector
, Atomic Energy Science and Technology 49(2) (2015) 336-342, (in Chinese). doi: 10.7538/yzk.2015.49.02.0336