logo

NONLINEAR RESPONSES OF GAMMA-RAY DOSIMETERS

NONLINEAR RESPONSES OF GAMMA-RAY DOSIMETERS

Luo Daling
Yong E.C.M.
Zhang Chunxiang
Nuclear Science and TechniquesVol.05, No.01pp.58-64Published in print 01 Feb 1994
49300

Either sublinear or supralinear responses of dosimeters to γ-ray can be described by a response function derived from statistical Poisson distribution. The characteristic parameters of the function determine linearity, sublinearity and supralinearity in their responses. The experimental data of gamma dose-responses of alanine ESR dosimeters, film dosimeters, LiF(Mg, Cu, P) and LiF(Mg, Ti) thermoluminescence dosimeters are used to test the response function.

DosimetersNonlinear responsesγ-ray
REFERENCES
1 Montret-Brugerolle M. Distribution Spaitiale de L'Energie, Depose par des ions Energetigues dans les Milieux Condenses Etude Par Thermoluminescence. Audiere [France]: Universite de clermont-Ferrand II, 1980
2 Zimmerean J. J Phys, C: Sol St Phys, 1971; 4: 3277
3 Wu Dake, Shun Fuyin, Dai Hongchen. Chinese J Radiat Med Prot, 1981; 5: 13
4 Cai Gangang. J Phys, D: Appl Phys, 1989; 22: 812
5 Larsson L, Katz R. Nucl Instr Meth, 1976; 138: 631
6 Kos H J, Nink R. Nucl Instr Meth, 1980; 175: 24
7 Lakshmanan A R, Chanda B, Bhatt R C. J Phys, D:Appl Phys, 1982; 15: 1501
8 Horowitz Y S. Radiat Prot Dosim, 1990; 83: 33(1—4)
9 Mckeever S W S. Radiat Prot Dosim, 1990; 83: 33(1—4)
10 Luo Daling, Huang Naiming, Zhang Chunxiang. Nucl Tech (in Chinese), 1992; 15: 51
11 Luo Daling, Li Mianfeng, Zhang Chunxiang et al. Acta Scientiarum Naturalium Universitis Sunyatceni (in Chinese), 1988; (4): 50
12 Luo Daling, Li Mianfeng, Zhang Chunxiang et al. Acta Scientiarum Naturalium Universitis Sunyatseni (in Chinese), 1987; (3): 70