RESEARCH ON FADING CHARACTERISTICS OF ALANINE/ESR DOSIMETERS IN HIGH DOSE RANGE

RESEARCH ON FADING CHARACTERISTICS OF ALANINE/ESR DOSIMETERS IN HIGH DOSE RANGE

Chen Lishu
H. Tachibana
Y. Haruyama
Nuclear Science and TechniquesVol.4, No.1pp.4-10Published in print 01 Feb 1993
25700

The present work has dealt extensively and systematically with the fading characteristics of polystyrene-alanine dosimeters irradiated at different dose rates and total doses under various temperatures (10—70°C) and stored a period of time in different conditions (5, 25, room temperature, 40°C). Detail investigation on this effect in practical irradiation condition is very important. Because it is difficult to correct this effect using correction coefficients so far. So the temperature coefficients and fading rates have been given along with a lot of experimental data to try to make approximation of radical formation or decay behavior.

Radiation processing dosimetryPS-alanine dosimeterESR Transfer dosimeterTemperature coefficientsFading behaviour