1 Introduction
With the development of nuclear and high-energy physics applications in industrial and medical therapy, ion therapy was proposed and investigated in the Berkley Lawrence National Laboratory by applying unique physical properties, i.e., a Bragg peak at the end of the range [1]. In addition, in 1990, the first clinical accelerator providing a proton beam was built for cancer therapy in the Loma Linda University Medical Center (LLUMC). Benefiting from the research conducted at LLUMC, several clinical proton therapy facilities [2]and centers have been organized, improving and promoting the rapid development of proton treatment studies [3, 4]. Heavy ions have identical Bragg peak-like protons; therefore, heavy ion cancer therapy studies and treatments were later started in NIRS (Japan) and GSI (Germany) [5]. There are almost 100 ion therapy centers currently in operation or under development around the world, and more than 100,000 patients have benefited from proton and heavy ion treatments [6, 7]. More than 3000 patients have been treated at the Shanghai proton and heavy ion center (SPHIC), the first heavy ion therapy center in China, which started operation and treatment in May 2015. An increasing number of ion therapy facilities require an increasing number of occupational doctors, medical physicists, clinical therapists, engineers, and nurses involved in clinical therapy, treatment planning systems (TPSs), facility operation, and maintenance. Although high-energy ions generated from an accelerator can kill a tumor, they also produce secondary radiation and activation when the ion beam hits the beam transport lines, devices, air, walls, and patients [8]. Harmful to humans, radiation protection is extremely important and mandatory, from the accelerator design to patient treatment [9, 10], and concrete shielding is the main measure used to protect the public and occupational operators from radiation [11]. Because occupational operators are close to the facility, they may receive more radiation than the public. Radiation levels can help us propose reasonable radiation protection guidelines and regulations and adopt proper methods to improve the quality of radiation protection. To minimize the absorbed radiation and evaluate the exposed radiation dose in the case of an emergency, simulation and measurement studies have been conducted, including the shielding design [2, 12-14], exposed radiation dose of clinical physicians [15, 16], calculation of the air activation caused by a proton accelerator [17], and neutrons produced from a beam [18, 19]. The major purpose of radiation protection studies is to increase the radiation safety of the accelerator and beam and reduce the personal dose of the operators [6, 20]. In this study, the secondary and activated X(γ) dose rate generated from the SPHIC accelerator under normal operation is measured, the beam loss distribution along beam lines from the secondary and activated X(γ) radiation dose rate is analyzed, and the radiation safety during operation and maintenance of the SPHIC accelerator is evaluated. The activated X(γ) radiation decay after a QA beam is applied in the treatment room was also measured, and a safety period for entering the treatment room after QA beam irradiation is recommended.
2 Accelerator in SPHIC
The SPHIC accelerator consists of seven main sections, as shown in Fig. 1: 1) ion sources (IS), which generate 8 keV/u protons and carbon ions, 2) a low-energy beam transportation (LEBT), which transports ions from the IS to the linear accelerator (Linac), 3) Linac, where the ion energy is accelerated from 8 keV/u to 7 MeV/u, 4) a medium energy beam transportation (MEBT), which transports ions from the Linac to the Synchrotron (SYNC), 5) a SYNC, which accelerates ions from 7 MeV/u to the requested energies (the maximum energy is 221 MeV/u for protons and 430 MeV/u for carbon, and the equivalent water depth at the maximum energy is 310 mm for both ions), 6) a high-energy beam transportation (HEBT), which transports high energy ions extracted from SYNC to the treatment rooms, including a common HEBT part H0 and room sections Hx from H0 to the treatment rooms (TRx), and 7) the treatment rooms (TRx), where three of the rooms have a fixed horizontal beam and one has a 45 semi-vertical fixed beam. The accelerator in SPHIC provides both proton and carbon scanning pencil beams [21], the beam performances of which are summarized in Table 1. As a medical treatment accelerator, more safety aspects are considered and integrated in this machine than in industrial and research machines. The Siemens PT onsite service team is responsible for the operation and maintenance of this facility. The SPHIC PT facility started beam commissioning in 2013, a clinical trial at the end of 2014, and official clinical treatment from May 2015.
Performance | Proton | Carbon |
---|---|---|
Energy range (MeV/u) (290 Energy levelsa) | 48.08 ∼ 221.07 | 88.65 ∼ 430.12 |
Intensity (13 levels) particles/second | 5.0×107 ∼ 2.6×109 | 1.3×107 ∼ 6.5×107 |
Extraction Time (4 levels) | 2/5/8/10 second | 2/5/8/10 second |
Focus (mm) (5 levels) | 32.5 mm @E7 8.1 ∼ 13.2 @E296 | 9.8 ∼ 13.4 @E7 3.4 ∼ 9.8 @E296 |
-202104/1001-8042-32-04-009/alternativeImage/1001-8042-32-04-009-F001.jpg)
3 Detectors used in dose measurement
Two types of radiation dose measurement detectors were used in this work. A scintillation detector was used to measure the transient radiation, and a semiconductor detector was used to measure the accumulated radiation. In normal accelerator operation with the beam, there is no access to the beam transportation line, the semiconductor detectors are fixed on the measurement positions for a limited time, the accumulated dose can be read afterward, and the average radiation dose rate can be calculated from the accumulated dose and measurement time. When the beam is off, free access is provided to the beam transportation line, and the scintillation detector can be used to measure the transient activated radiation dose rate around the beam transportation line. The semi-conductor detectors we used here are “RAD-60S” and “ALOKA PDM-122-SH,” which are alarming electrical personal dosimeters (EPDs) and are used for accumulated radiation dose monitoring, the scintillation detector used here is an “Automess 6150 AD 5/H” and is used to measure only the transient radiation dose rate. All detectors are calibrated, and the details of the detector are listed in Table 2.
RAD-60S | |
---|---|
Function | Performance |
Dose measurement range | 1.0μSv ∼ 9.99 Sv |
Dose rate measurement range | 5.0μSv/h ∼ 3.0 Sv/h |
Measurement precision | ±5% and dose rate 2mSv/h |
Linearity | better than ±15% @ <3.0 Sv/h |
Energy response | better than 25% Hp(10) 60 keV ∼ 3MeV |
ALOKA PDM-122-SH | |
Dose measurement range | 1μ Sv ∼ 9999μ Sv |
Measurement precision | ±10% (1μ Sv ∼ 9999μ Sv) |
Linearity | ±10% (<0.1 Sv/h) ±20% (0.1 Sv/h ∼0.3 Sv/h |
Energy response | ±30% (50 keV ∼ 3 MeV) |
Automess 6150 AD 5/H | |
Dose rate measurement range | 1μ Sv/h ∼ 1000 mSv/h |
Linearity | <±10% |
Energy response | <±30% (45 keV ∼ 2.6 MeV) |
4 Secondary and activated X(γ) radiation measurement in accelerator area
An accelerator operation can be identified in two states: “Beam on” and “Beam off.” “Beam on” indicates the normal clinical time, during which the beam is present in the accelerator beam lines without a lengthy interruption. “Beam off” means there is no beam in the accelerator beam lines for a lengthy amount of time when the beam is stopped. Both secondary and activated X(γ) radiation occur when the beam is on. When the beam is off, the secondary radiation decays rapidly, and only the activated X(γ) radiation can be detected.
The accumulated secondary and activated X(γ) radiation levels are measured by nine semiconductor dosimeters when the beam is on. The dosimeters are fixed on the surface of the beam vacuum tube, which are marked in Fig. 1 (marker serial numbers from 1 to 64). These measurement positions were uniformly distributed along the beam line, and the device layout and distribution were also considered. The radiation measurement time was more than 24 h to cover one complete clinical day at every measurement position, and there are some repetitive measurements in some measurement positions to check the consistency of the dosimeters and the measured radiation dose. Then, average dose rate is calculated from the accumulated dose divided by the measurement time to show the average X(γ) radiation dose rate because the average dose rate is a value independent of time and can be compared directly with that of a “beam off” case. Only activated X(γ) ray radiation can be detected when the beam is off. In this case, the transient radiation dose rate is measured instead of the accumulated dose, the scintillation detector is used, and the measurement positions are same as the “beam on” case marked in Fig. 1.
To compare the radiation dose rate with natural environment, the radiation dose rate in an office is also measured by the detectors “RAD-60S,” “PDM-122-SH,” and “Auotmess 6150 AD 5/H.” The office is located on the second floor of the accelerator hall and is a daily working place for accelerator engineers. The radiation level in this office is labeled the “background” because it indicates the environment radiation level for the daily work under the occupational dose rate. The measured background radiation dose rate was consistent and averaged 0.11μSv/h for the three different detectors. This background radiation dose rate is also consistent with the data in the annual monitoring report of SPHIC [22], the normal official in-room dose rate in the work area and environment of the SPHIC, which is lower than the dose rate close to the accelerator facility and clinical treatment room. This background dose rate is consistently shown in the figures in this paper as a reference radiation level.
Figure 2 shows the average dose rate when the beam is on. The numbers of the x-axis indicate the labeled measurement positions in Fig. 1: measurement positions 1–7 are locate in the LEBT section, 8–18 are located in the Linac and MEBT section, 19–43 are around the synchrotron accelerator ring, and 44–64 are in the HEBT section.
-202104/1001-8042-32-04-009/alternativeImage/1001-8042-32-04-009-F002.jpg)
In the LEBT section, the particle energy is 8 keV/u from the ion source, and low-energy ions protected by the beam vacuum tube do not activate any radiation. Therefore, the radiation dose rate measured in the LEBT section is below the background level. The first peak of the radiation dose rate is located at around measurement position No. 8. Measurement position No. 7 is before the Linac, where the DC beam cuts into the macro-pulse beam, and 95% of the cut beam is dumped at around No. 7. No. 8 is at the exit of the radio frequency quadrupole accelerator (RFQ) in the Linac section. The ions are accelerated to 400 keV/u in the RFQ, and the RFQ transmission rate is only 50%. In addition, ion dumping before and in the RFQ produces the first radiation peak at measurement position No. 8. The ions are then sped up to 7 MeV/u in the Linac before being injected into the synchrotron.
The highest broad radiation peak with the measured maximum radiation occurring at a rate of 1 mSv/h from measurement positions 19–30 covers the beam injection, extraction, and synchrotron beam dumping. This broad and high peak indicates the greatest possibility of beam loss caused by these activities. From measurement positions 30–43, on the opposite half side of the synchrotron ring, the radiation dose rate is 2 orders lower than on the other half side, because in this half ring there only beam acceleration activity occurs, and less of the beam is lost. Following measurement position 44 is the HEBT section, where a mid-level radiation emission rate of around several μSv/h, and a peak at around position 48, occurs. In this section, the beam loss is caused by dumping at position 48 after a synchrotron and beam abort treatment. A valley with a radiation dose rate of less than 0.1 μSv/h from measurement positions 56–61, which are in the common HEBT line H0 between TR1 and TR4, indicates that the beam optics are well designed in this region, the beam is perfectly constrained, and beam loss rarely occurs within this range. In Fig. 2 the reference line (black dash line) is also plotted to show the background radiation dose rate and the dramatic difference in radiation level.
The secondary radiation directly from the beam loss disappears after the beam stops, and can only irradiate people in the case of an incident such as a PSS failure or emergency, because the accelerator area is normally inaccessible through PSS protection. Radiation from the activation may exist for an extremely long time after the beam stops. It can irradiate people who are working in this area, such as occupational engineers and maintenance workers. It is the main source of occupational irradiation, and this part is the most important and monitored radiation within the facility environment. In Fig. 3 both secondary and activated radiation dose rates are plotted for beam-on (blue round symbols) and beam-off (red square symbols) cases, respectively, and both dose rates are measured on the surface of a beam vacuum tube. The activation dose rate is much lower than that of the beam on, the activation dose rate is below the background level at most measurement positions along the beam line, and the activated radiation peak appears only in the region where the strongest secondary radiation occurs. This consistency reveals that the activation radiation is caused by the beam loss on the beam line components, but the activated radiation is much lower than the secondary radiation during a beam on. After the beam stops, the maximum existing activated dose rate is less than 1 μSv/h on the synchrotron ring where injection, beam dumping, and extraction activities are included, whereas the other parts have identical background dose rates as found in the office, and most are even lower.
-202104/1001-8042-32-04-009/alternativeImage/1001-8042-32-04-009-F003.jpg)
The activation dose rate is also measured at positions 1 m away from the beam line when the beam is off, where most likely operators and engineers pass through. To analyze the activated radiation space distribution and averaged/overall environmental radiation dose rate, the measured dose rates are plotted in Fig. 4, and all data are measured 1 h after the beam stops. There is a slight radiation peak matching with the dose rate peak on the beam line tube surface when the beam is on and off, indicating that the activated radiation from the beam line is the main X(γ) ray source around this location. At around the radiation peak locations, the dose rate 1 m away from the beam line is much lower than that on the beam line surface. This reveals that “increasing the distance from the source decreases the exposed radiation” under radiation safety and protection strategies.
-202104/1001-8042-32-04-009/alternativeImage/1001-8042-32-04-009-F004.jpg)
One can also note that at most measurement positions, the activation radiation dose rate on the beam line surface is approximately 20% less than dose rate of the background radiation, and less than the dose rate measured at the measurement points at 1 m away from the beam line, which is caused by the shielding effect of the beam line tube and all other devices, such as the magnets. This type of shielding effect can also be observed from the radiation dose measurements in the following treatment room.
There is only one clear activated X(γ) radiation peak on part of the synchrotron ring, arising mainly from a high-energy beam loss, the dose rate of which is much lower than that of a beam on.
5 Secondary and activated X(γ) radiation in treatment room
Radiation in the treatment room is another aspect that influences the personal dose equivalent of the occupational engineers, medical physicists, and therapists. The secondary and activated X(γ) radiation distribution in space and time and its measurement can reveal the transient and activated X(γ) radiation level in the treatment rooms.
This measurement adopts the layout for a daily beam quality check, as shown in Fig. 5. A multiwire proportional chamber (MWPC) is placed on the beam iso-center (this QA MWPC is equivalent to a water thickness of 260 μm [24, 23]. There are two carbon fiber protecting plates on the MWPC beam entry and exit window, the thickness of the beam entry protection plate was 0.65 mm, and the thickness of the beam exit window protection plate was 1.05 mm). The beam passes through the MWPC and then hits the wall along the beam direction, and consists of concrete and an iron plate [11, 12]. Fig. 5 shows the testing device layout in the treatment room, MWPC, and measurement detector positions. The light-red cross on the MWPC marks the beam iso-center, to which the MWPC is aligned, and the dark-red cross on the wall marks the downstream beam hitting position on the wall. The orange rectangles mark four X(γ) detector measurement positions when the beam is on, and the activated X(γ) radiation is also measured at these four predefined positions.
-202104/1001-8042-32-04-009/alternativeImage/1001-8042-32-04-009-F005.jpg)
The definition of the treatment room coordinates is shown by the green arrows in Fig. 5, where +x is horizontally to the right along the beam direction, +y indicates the downward direction vertically, and +z is the beam downstream direction. The beam outlet is located at the bottom left corner in Fig. 5, and the beam is emitted from the outlet, passes through the MWPC (and the iso-center) and finally hits the wall. The iso-center is defined as (0,0,0), and the beam spot hitting the wall is at (0,0,400 cm). Four dosimeters were fixed at the positions with the coordinates listed in Table 3 to measure the accumulated X(γ) dose from the ion irradiation.
point | Coordinates | Accumulated X(γ) dose |
---|---|---|
Point_1 | (20 cm, 0 cm, 400 cm) | 22.43 mSv |
Point_2 | (0 cm, 20 cm, 0 cm) | 361 μSv |
Point_3 | (50 cm, 20 cm, 0 cm) | 83 μSv |
Point_4 | (200 cm, 150 cm, 400 cm) | 6 μSv |
In addition, 1.0 × 1010 carbon ions with the maximum energy are irradiated during testing, and the accumulated X(γ) radiation doses are recorded by the dosimeters fixed on the four points defined above, as shown in Table 3. Some ions are scattered while passing through the MWPC and its protecting carbon plate; therefore, X(γ) rays are generated within this ion collision process, and most ions passing through the MWPC finally hit the wall, losing their energy and producing transient X(γ) rays. In Table 3, the accumulated dose of 22.43 mSv at Point_1 shows that it is close to a strong radiation source, i.e., the beam irradiation spot on the wall. The second radiation source is the iso-center where the beam hits the MWPC, and Point _2 is closer to this source than Point _3. The difference in dose of 361μSv at Point_2 from 83μSv at Point_3 can be attributed to their difference in distance from the iso-center. The main source of Point _4 is the beam spot point on the wall, and the shielding effect from the wall is clear. If the main radiation source for Point_4 is the beam spot on the wall, the Point _4 dose is estimated to be more than 6μSv based on the distance dependence in comparison with Point_1 because both Point_1 and Point_4 are on the wall surface. The measured dose of only 6μSv at Point_4 indicates that the shielding effect is significant for this point.
Ion collisions with the MWPC and wall are the main radiation source. After the beam stops, the transient X(γ) radiation disappears in the accelerator area. The activated and residual radiation remains, and these residual dose rates at the pre-defined points shown in Table 3 and at the iso-center are measured after the beam stops. Their dependence on the time decay after beam stopping is shown in Table 4, with the device layout unchanged in Fig. 5. The measured dose rates are shown in Table 4, where the maximum residual dose rate is measured on the beam stopping point on the wall (beam spot) where most of the beam energy is deposited, and the iso-center is the second point where the possible ion energy is deposited. The dose rate value of Point_2 is higher than that of Point_3 because Point_2 is closer to the activated source iso-center than point _3. All measured residual radiation rates at the five points recorded in Table 4 show an exponential decay time. After 5 min, the radiation rate decayed to half of its maximum value. No additional radiation can be detected after 30 min in the three recorded positions without the beam hitting directly. For the positions where the beam hits and passes through, such as an iso-center, it is difficult to detect additional radiation after a 3-h period, but for the beam spot on the wall even after 3 h, additional radiation still exists. From these radiation measurements conducted in the treatment room, we can learn if the beam hits directly on the wall, and if there is a possibility of residual radiation for a lengthy amount of time near the hit point. All occupational staff members should avoid staying close to the beam hit point on the wall, and the amount of time before it is safe to enter the treatment room after a QA beam is applied is at least 5 min.
Time (min) | Beam spot | Iso-center | Point_2 | Point_3 | Point_4 |
---|---|---|---|---|---|
0.5 | 80 | 11 | 1.5 | 0.8 | 0.5 |
1 | 70 | 10 | 1 | 0.5 | 0.3 |
2 | 40 | 6 | 1 | 0.5 | 0.2 |
5 | 35 | 5 | 0.8 | 0.3 | 0.16 |
10 | 35 | 4 | 0.6 | 0.2 | 0.16 |
30 | 20 | 2.5 | 0.5 | 0.17 | 0.15 |
180 | 1.4 | 0.15 | 0.154 | 0.12 | 0.12 |
6 Summary
We measured the X(γ) rays from the secondary and activated radiation when the SPHIC accelerator beam was on and off. There is strong secondary X(γ) radiation when the beam is on during clinical treatment, and the maximum measured dose rate value is 1 mSv/h on some parts of the beam line. In addition, there is a structural dose rate distribution based on the beam lattice and optics. Secondary and activated X(γ) radiation mainly comes from the high-energy beam loss. The results provide a reference for the radiation level inside the accelerator hall when the beam is on. This can help us estimate the irradiated and accumulated dose in cases of an incident or emergency, evaluate any injuries, and follow up with medical treatment if needed. It is also helpful to update the radiation protection regulations and recommendations on-site.
After the beam stops, the residual radiation from the activation is low, and the distribution of activated positions is consistent with the secondary radiation and beam loss location with the beam on. The measured activated radiation data show that SPHIC is extremely safe from the viewpoint of nuclear safety and radiation protection.
The radiation measurement results of the treatment room after the QA beam showed residual radiation after the QA beam hit the wall without dumping, and the amount of time before safely entering the treatment room was 5 min after the QA beam is applied. QA beam dumping in the treatment room is used to avoid the beam hitting the wall directly, which produces strong secondary and residual radiation. Although radiation can be detected after the beam is applied in the treatment room, the area outside the treatment room is safe because of the designed shielding of the concrete wall [12, 22].
Radiological use of fast protons
. Radiology, 47(5), 487 (1946). doi: 10.1148/47.5.487Shielding calculations for a 250MeV hospital-based proton accelerator
. Nucl. Instr. and Meth. A, 374(2), 254-268 (1996). doi: 10.1016/0168-9002(96)00017-4Medical Cyclotron Used for Diagnostic or Therapy
, Strategic Study of Chinese Academy of Engineering, 12(2), 9-15, (2000)Train probe for beam diagnostics of the SC200 superconducting cyclotron
, Nucl. Sci. Tech. 31(10), 103, (2020). doi: 10.1007/s41365-020-00808-4Proton and heavy ion radiotherapy in China
, Chin. J. Radiol. Med. Prot. 36(8), 561-563, (2016). doi: 10.3760/cma.j.issn.0254-5098.2016.08.001Physical design and multi-particle simulations of a compact IH- DTL with KONUS beam dynamics for proton therapy
, Nucl. Sci. Tech. 31(7), 64 (2020). doi: 10.1007/s41365-020-00773-yRadiation shielding design of a compact single-room proton therapy based on synchrotron
. Nucl. Sci. Tech. 31(1), 1 (2020). doi: 10.1007/s41365-019-0712-1Monte Carlo method for proton and heavy ion treatment room shielding calculation
, Chin. J. Radiol. Med. Prot. 36(8), 634-638 (2016). doi: 10.3760/cma.j.issn.0254-5098.2016.08.016Study on the optimization of radiation protection for therapeutic proton synchrotron
, Chin. J. Radiol. Med. Prot. 39(9), 707-710 (2019). doi: 10.3760/cma.j.issn.0254-5098.2019.09.013Calculation of induced radioactivity of heavy concrete shielding wall in proton therapy room
. Chin. J. Radiol. Med. Prot. 40(4), 321-325 (2020). doi: 10.3760/cma.j.issn.0254-5098.2020.04.013Shielding data for a 100-250 MeV proton accelerators: attenuation of secondary radiation in think iron and concrete/iron shields
. Nucl. Instr. and Meth. B, 266(15), 3406-3416 (2008). doi: 10.1016/j.nimb.2008.05.002Assessment of occupational exposure doses to physicians in clinical practice at Shanghai Proton and heavy ion Center
. Chin. J. Radiol. Med. Prot. 38(11), 851-854 (2018). doi: 10.3760/cma.j.issn.0254-5098.2018.11.010Study on radiation field caused by proton beam loss in high intensity heavy-ion accelerator
. Atomic Energy Science and Technology, 49(Supp. Oct.), 573-577 (2015). doi: 10.7538/yzk.2015.49.S1.0573Estimating and analysis of personel doses from activated nuclides in radiotherapy place of a proton accelerator
. Chin. J. Radiol. Med. Prot. 36(8), 630-633 (2016). doi: 10.3760/cma.j.issn.0254-5098.2016.08.015Shielding data for 100-250 MeV proton accerelatos: double differential neutron distributions and attenuation in concrete
. Nucl. Instr. and Meth. B, 265(2), 581-598 (2007). doi: 10.1016/j.nimb.2007.09.046Measurements of high-energy neutrons penetrated through concrete shields using self-TOF, NE213, and activation detectors
. Nucl. Sci. Eng. 141(2), 140-153 (2002). doi: 10.13181/nse02-a2273The problems of environmental safety for application of high energy proton accelerator therapy system
. Atomic Energy Science and Technology, 38(S1), 192-196 (2004). doi: 10.7538/yzk.2004.38.S1.0192Brief introduction to the operation of Shanghai proton and heavy ion accelerator equipment
. Chin. J. Radiol. Med. Prot. 36(8), 639-640 (2016). doi: 10.3760/cma.j.issn.0254-5098.2016.08.017Annual report of SPHIC working area and environmental radiation dose rate monitoring
, 2019Collection efficiency of monitor parallel plate ionization chamber in pencil beam scanning proton therapy
, Nucl. Sci. Tech. 31(2), 13 (2020). doi: 10.1007/s41365-020-0722-zMonte Carlo simulation and verification of a scanning proton beam nozzle
. Chin. J. Radiol. Med. Prot. 39(8), 635-640 (2019). doi: 10.3760/cma.j.issn.0254-5098.2019.08.014