Tiantian DAI, Tianyu MA, Hui LIU, et al. A high-resolution small animal SPECT system developed at Tsinghua. [J]. Nuclear Science and Techniques 22(6):344-348(2011)
DOI:
Tiantian DAI, Tianyu MA, Hui LIU, et al. A high-resolution small animal SPECT system developed at Tsinghua. [J]. Nuclear Science and Techniques 22(6):344-348(2011) DOI: 10.13538/j.1001-8042/nst.22.344-348.
A high-resolution small animal SPECT system developed at Tsinghua
An SPECT system dedicated to small animal imaging shall be of a millimeter spatial resolution or even better. This study was aimed at achieving 0.5-mm spatial resolution for a small animal SPECT system at low cost. It was developed from a single-head clinical SPECT scanner, with a seven-pinhole collimator and a four-degree-of- freedom motion control stage. Several key techniques were developed, including high-resolution image reconstruction algorithm, high accuracy geometrical calibration method, and optimized system matrix derivation scheme. The system matrix was derived from Monte-Carlo simulation and de-noised by fitting each point spread function to a two dimensional Gaussian function. Experiments of point source and ultra micro hot rod phantom were conducted. With a spatial resolution of 0.5–0.6 mm, this system provides a practical way for low-cost high-resolution animal imaging on a clinic SPECT system.
Chatziioannou A F. Proc Am Thorac Soc, 2005, 2: 533-536.
Meikle S R, Kench P, Kassiou M, et al. Phys Med Biol, 2005, 50: R45-R61.
Beekman F J, Vastenhouw B. Phys Med Biol, 2004, 49: 4579-4592.
Kastis G A, Barber H B, Barrett H H, et al. IEEE Trans Nucl Sci, 2000, 47: 1923-1927.
Beekman F J, van der Have F, Vastenhouw B, et al. J Nucl Med, 2005, 46: 1194-1200.
Schramm N U, Ebel G, Engeland U, et al. IEEE Trans Nucl Sci, 2003, 50: 315-320.
Shao Y, Ma T, Yao R. IEEE Nucl Sci Symp Conf Rec, 2007, 4285-4288.
DiFilippo F P. Phys Med Biol, 2008, 53: 4185-4201.
Dai Q, Qi Y. At Energ Sci Technol, 2009, 43: 1034-1038.
Rowe R K, Aarsvold J N, Barrett H H, et al. J Nucl Med, 1993, 34: 474-480.
Metzler S D, Bowsher J E, Greer K L, et al. IEEE Trans Med Imag, 2002, 21: 878-886.
Bal G, Acton P D. Phys Med Biol, 2006, 51: 4923-4950.
Refecas M, Mosler B, Dietz M, et al. IEEE Trans Nucl Sci, 2004, 51: 2597-2605.
Yao R, Ma T, Shao Y. IEEE Trans Nucl Sci, 2009, 56: 2651-2658.
Dai T, Liu Y, Ma T, et al. At Energ Sci Technol, 2011, 45: 369-373.
Liu H, Ma T, Dai T, et al. M18-224: High accuracy geometrical calibration for half-mm animal SPECT imaging. IEEE NSS MIC RTSD, Knoxville, Tennessee, Oct, 2010.
Ma T, Yao R, Shao Y, et al. IEEE Trans Med Imag, 2009, 28: 1929-1939.
Shepp L A, Vardi Y. IEEE Trans Med Imag, 1982, MI-1: 113-122.
Hudson H M, Larkin R S. IEEE Trans Med Imag, 1994, 12: 601-609.
Jan S, Santin G, Strul D, et al. Phys Med Biol, 2004, 49: 4543-4561.
Brun R, Rademakers F. Nucl Inst Meth Phys Res A, 1997, 389: 81-86. See also http://root.cern.ch/.
A new multi-parallel-beam collimators with improved sensitivity for SPECT
A new imaging mode based on X-ray CT as prior image and sparsely sampled projections for rapid clinical proton CT
Hybrid model for muon tomography and quantitative analysis of image quality
A novel 4D resolution imaging method for low and medium atomic number objects at the centimeter scale by coincidence detection technique of cosmic-ray muon and its secondary particles
Analysis of electro-optical intensity modulator for bunch arrival-time monitor at SXFEL
Related Author
No data
Related Institution
Physics School, Nanjing University
New York Institute of Technology
Suzhou Institute of Biomedical Engineering and Technology Chinese Academy of Sciences
Shanghai Key Laboratory of Cryogenics & Superconducting RF Technology
Shanghai Synchrotron Radiation Facility, Shanghai Advanced Research Institute, Chinese Academy of Sciences