An active base designed in high counting rate applications for Hamamatsu R1924A photomultiplier tube

NUCLEAR PHYSICS AND INTERDISCIPLINARY RESEARCH

An active base designed in high counting rate applications for Hamamatsu R1924A photomultiplier tube

Pei-Pei Ren
Wei-Ping Lin
Roy Wada
Xing-Quan Liu1
Mei-Rong Huang
Guo-Yu Tian
Fei Luo
Qi Sun
Zhi-Qiang Chen
Guo-Qing Xiao
Rui Han
Fu-Dong Shi
Bo-Xing Gou
Nuclear Science and TechniquesVol.28, No.10Article number 145Published in print 01 Oct 2017Available online 06 Sep 2017
43200

Hamamatsu R1924A is one of the most widely used photomultiplier tubes (PMTs) in nuclear physics. Since the active base suitable for R1924A is still not available in market, an active base is designed for Hamamatsu R1924A PMT and the test results at high counting rates are presented. The active bases with two different sets of resistor chains were tested and compared by a frequency controlled green straw hat LED light. A stable signal output up to 100 kHz is achieved using frequency controlled LED pulsed light. The temperature of bases, which reflects the power consumption and is crucial for applications in vacuum, is also monitored with the same LED pulsed light. The temperature of the active base with smaller resistances reaches about twice of that of the active base with larger resistances in the resistor chain. For the applications in vacuum, the active base with resistance between the two sets of resistor chains may be preferable.

Active baseHigh counting rateHamamatsu R1924A photomultiplier tube

1 Introductioin

For the centrality CsI array, described in Ref. [1], which is going to be used in experiments at the IMP, an active bases have been studied. In the array, Hamamatsu R1924A tubes are used. R1924A is a small (25 mm in diameter) photo-multiplier tube (PMT), which is a relatively fast (1.5 ns rising time) tube with a linear-focused 10 stage dynode geometry. The borosilicate entrance window gives a spectral response of wavelength, λ, from 300 to 650 nm with the peak at 420 nm. The photo cathode is Bialkali with a 22% quantum efficiency at λ=420 nm.

R1924A is one of the most widely used PMTs in nuclear physics. It has been applied to NIMROD-ISiS [2], the CCDA project [1] in IMP, an array to classify the centrality of heavy ion collisions, the Super-TIGER project [3], the Compton gamma imager [4], the GSO detector [5], the active veto system at LBNL low-background facility [6], our previous experiment [7-9], and so on. In most applications, the PMT has been conventionally operated with passive resistor bases, which consist of resistors and capacitors, to supply the dynode bias voltages at relatively low counting rate cases, just because of their simplicity. However in most of the trigger detectors, such as a beam monitor counter or a centrality detector, the tube has to be operated in a high counting rate with large output pulses.

When the counting rate becomes high, i.e., more than 1kHz, with pulse height larger than 1V, the linear response between the light input and the output signal height does not hold and the output signals are reduced. This is called the “pulse height defect". The higher the frequency and the larger the output signal, the more the pulse height defect. The linear response is crucial for particle identification and energy measurement. The energy measurement is very important in applications not only for calorimetry [10], but also for the particle identifications, such as the ΔE-E method [2] and pulse shape discrimination [1]. In high counting rate, with large output signals, a large current is generated at the last dynode stages, which is supplied through the resistor chain and capacitors. This large current reduces the voltage between the last few dynodes and causes the pulse height defect. Field Effect Transistors (FETs) [11] are capable for fast switching operations by a gate signal with relatively small heat generation and can provide high withstand voltage between the source and drain, which leads to a large drain current (ID) when the gate is opened. Taking advantage of the FET, the output signal stability would be improved by replacing the resistors with FETs in a passive base. This stabilized base is called an “active base”. Therefore, the active base is a better choice in high counting rate applications.

This article is organized as follows: The active base designing and tests are presented in Sect. 2. The applications of the active base in high counting rate and in vacuum are discussed in Sect. 3. A brief summary is given in Sect. 4.

2 Designing and testing of active bases

To make an active base that is suitable for the R1924A tube and has a small length, four of STP2NK100Z FET, which is a n-channel MOSFET and has 1000V withstand voltage, are used in the last four stages of the voltage supply. Figure 1 shows a schematic circuit diagram of the active base designed [12]. The four 15 MΩ and one 100 MΩ resistor chain in Fig. 1 distributes the bias voltages for FETs (dynode 8 to 10) and the other dynodes 1-7 are biased by R1, R2, and R3 resistor chains, respectively. The capacitors are used in the circuit to increase the stability of PMT output [13]. Two sets of resistor chains, which differ in 10 times resistance as shown in the bottom of Fig. 1, are used for the active base. We call them “base 1”(set1) and “base 2”(set2) throughout the paper.

Figure 1:
PMT base circuit for R1924A tube. Two different sets of resistor chains are shown under the circuit.
pic

A 3 mm green straw hat LED, which wavelength is inside the PMT’s spectral response, driven by the frequency controlled circuit shown in Fig. 2, is used to test the newly designed active bases. A fast analog signal with less than 1ns rise time is generated by a computer controlled emulator (CAEN Mod. NDT68000). After discriminating by a constant fraction discriminator (ORTEC EG&G CF8000), a 140 ns gate signal from a gate and delay generator(ORTEC GG8020) is proceeded to the green LED. The photo electrons generated by the light from the LED are collected and multiplied by the R1924A PMT inside a black box. The oscilloscope is used to read the output signal. To make a comparison, the passive base from Hamamatsu E2924 is also used as a reference. The same R1924A tube is used for all the three bases in the tests. Figure 3 shows the signal of PMT output recorded by the oscilloscope for passive base (blue line), base 1 (brown line), and base 2 (black line) at the LED pulsed light frequency of 0.1 kHz (a) and 50 kHz (b). The voltages are slightly adjusted, around -1000 V, to get the same peak value of 6 V as shown in Fig. 3(a). One can see that the passive base shows notable oscillation in the peak, while base 1 is slightly better and base 2 is the best in signal shape at 0.1 kHz LED pulsed light. In Fig. 3(b), the signal shapes are shown at 50 kHz. One can see clearly the pulse height defect for the passive base and base 1. In the following sections, more details of the pulse height defects as a function of the frequency are investigated.

Figure 2:
The frequency controlled circuit for green straw hat LED pulsed light.
pic
Figure 3:
(Color online) The signals of PMT output in the oscilloscope for the LED light frequency of 0.1 kHz (a) and 50 kHz (b). The blue, brown, and black lines correspond to passive base, base 1, and base 2, respectively.
pic

To see the stability of signal output of three bases, we change the LED pulsed light frequency from 0.1 kHz to more than 100 kHz. The output signal for each frequency of LED light is recorded by the oscilloscope. The average value from the signal peak is assigned as the amplitude, which has less than 0.5 V uncertainty. The measured amplitude as a function of LED frequency is shown in Fig. 4 for passive base (open circles), base 1 (open squares), and base 2 (open triangles). The amplitude of passive base starts to drop after a LED frequency greater than 1 kHz and decreases monotonously as the LED frequency increases. The stability of the output signal for the two active bases is significantly extended. As one can see, the amplitude stays the same up to 10 kHz for base 1 and up to 100 kHz for base 2. A slightly increasing trend is observed for base 1 after 5 kHz. This could be from the signal oscillation at the peak mentioned above, which causes some errors to determine the amplitude. There is a minimum value before the amplitude drops monotonously for both two active bases, which could be caused by the ringing in the output signal [13].

Figure 4:
(Color online) The pulse amplitude as a function of LED light frequency for passive base (open circles), base 1 (open squares), and base 2 (open triangles).
pic

Another important characteristic for PMT bases is the power consumption, which is crucial in vacuum applications. The more power is consumed, the more heat is generated by the base. One can get the relative power consumption by monitoring the temperature of the PMT bases. An apuhua TM-902C thermometer with an accuracy of 0.1∘C is used to record the temperature of the bases in the tests. The temperature sensor is attached to the last stage of FETs for the active bases and to outer cover for the passive base, respectively. The same 1 kHz frequency pulsed LED light is used for this test. To remove the slight difference of room temperature, the ratio between base temperature and room temperature is used in the measurements. Figure 5 shows the temperature ratio, T/T0 (T0 is the room temperature), as a function of monitored time for passive base (solid line), base 1 (dotted line), and base 2 (dashed line). The base 1 with FETs and with R1 = 2.7 MΩ, R2 = 680 kΩ, and R3 = 1 MΩ shows the minimum power consumption and that the passive base comes at the second place, while base 2, with 10 times less the R1, R2, and R3 resistances comparing to that of base 1, shows the maximum power consumption.

Figure 5:
(Color online) The base temperature (T) relative to the room temperature (T0) as a function of monitoring time for passive base (solid line), base 1 (dotted line), and base 2 (dashed line) with 1 kHz LED light frequency.
pic

3 Discussion

From the results of frequency tests shown in Fig. 4, we conclude that the new designed active bases (base 1 and 2) indeed extend the stability of R1924A PMT output signal to high frequencies, 10 kHz for base 1 and 100 kHz for base 2. The extension of the stability of output signal depends significantly on the resistances of R1, R2, and R3. The smaller the resistances, the more extension of stability in the ouput signal. On the other hand, the results of power consumption tests as shown in Fig. 5 reveal that the base with smaller resistance, i.e. base 2, has a much larger power consumption than that of larger resistances. This limits further reduction of the resistor values from those of base 2 to obtain the better stability for higher frequencies without extra cooling of the base.

All resistors have temperature dependent, which depends on their characteristics, and heat causes some instability, for either passive or active base application, especially for the first few hours after the voltage is turned on. Therefore, one needs to balance the stability in the high counting rate and the power consumption to get the best resistance set for R1, R2, and R3 in actual applications. For the applications in vacuum, the active base with resistance between the two sets of the resister chains, set 1 and set 2, may be preferable.

4 Summary

Hamamatsu R1924A is one of the most widely used PMTs in nuclear physics. An active base with FETs on the last four stages, which is still not available in market for Hamamatsu R1924A PMT, is designed and tested for high counting application. The active bases with two sets of the resistor chains were tested. A straw hat LED pulsed light, driven by a frequency controlled circuit, was used in the stability of output signal and power consumption tests. A stable output signal with 6 V in the amplitude up to the LED pulsed light frequency of 100 kHz is achieved for the base with smaller resistances in resistor chain. In the power consumption tests, the temperature for the active base with smaller resistances reaches about twice of that of the active base with larger resistances in resistor chain, which limits the further reduction of resistances of R1 to R3 for extending the stability of output signal to a higher counting rate. For the applications in a vacuum, the active base with resistance between the two sets of the resister chains may be preferable.

References
[1] F. Fu, W.P. Lin, X.Q. Liu, et al.,

A module test of CCDA: an array to select the centrality of collisions in heavy ion collisions

. Chin. Phys. Lett. 31, 082502 (2014). doi: 10.1088/0256-307X/31/8/082502
Baidu ScholarGoogle Scholar
[2] S. Wuenschel, K. Hagel, R. Wada et al.,

NIMROD-ISiS, a versatile tool for studying the isotopic degree of freedom in heavyion collisions

. Nuclear Instruments and Methods in Physics Research A 604, 578 (2009). doi: 10.1016/j.nima.2009.03.187
Baidu ScholarGoogle Scholar
[3] M. Sasaki, W. R. Binns, R. G. Bose et al.,

Super-TIGER 2012/2013 In-flight Instrument Performance and Preliminary Results

. International Cosmic Ray Conference, 0570 (2013). http://galprop.stanford.edu/elibrary/icrc/2013/papers/icrc2013-0570.pdf
Baidu ScholarGoogle Scholar
[4] P.R.B. Saull, L.E. Sinclair, H.C.J. Seywerd et al.,

First demonstration of a Compton gamma imager based on silicon photomultipliers

. Nuclear Instruments and Methods in Physics Research A 679, (2012) 89. doi: 10.1016/j.nima.2012.03.019
Baidu ScholarGoogle Scholar
[5] S. Yamamoto,

Optimization of the integration time of pulse shape analysis for dual-layer GSO detector with different amount of Ce

. Nuclear Instruments and Methods in Physics Research A 587, 319-323 (2008). doi: 10.1016/j.nima.2008.01.084
Baidu ScholarGoogle Scholar
[6] K.J. Thomas, E.B. Norman, A.R. Smith, et al.,

Installation of a muon veto for low background gamma spectroscopy at the LBNL low-background facility

. Nuclear Instruments and Methods in Physics Research A 724, 47-53 (2013). doi: 10.1016/j.nima.2013.05.034
Baidu ScholarGoogle Scholar
[7] M. Huang, Z. Chen, S. Kowalski, et al.,

Isobaric yield ratios and the symmetry energy in heavy-ion reactions near the Fermi energy

. Phys. Rev. C. 81, 044620 (2010). doi: 10.1103/PhysRevC.81.044620
Baidu ScholarGoogle Scholar
[8] M.R.D. Rodrigues, W. Lin, X. Liu, et al.,

Experimental reconstruction of excitation energies of primary hot isotopes in heavy ion collisions near the Fermi energy

. Phys. Rev. C 88, 034605 (2013). doi: 10.1103/PhysRevC.88.034605
Baidu ScholarGoogle Scholar
[9] R. Wada, M.R. Huang, W.P. Lin, et al.,

IMF production and symmetry energy in heavy ion collisions near Fermi energy

. Nucl. Scie. Tech. 24, 050501 (2013). doi: 10.13538/j.1001-8042/nst.2013.05.001
Baidu ScholarGoogle Scholar
[10] S.S. Gao, C.Q. Feng, D. Jiang, et al.,

Radiation tolerance studies on the VA32 ASIC for DAMPE BGO calorimeter

. Nucl. Scie. Tech. 25, 010402 (2014). doi: 10.13538/j.1001-8042/nst.25.010402
Baidu ScholarGoogle Scholar
[11]

Feild-effect transistor, wiki

, https://en.wikipedia.org/wiki/Field-effect_transistor.
Baidu ScholarGoogle Scholar
[12] S. Takeuchi and T. Nagai,

Low Power Photomultiplier Base Circuit

. IEEE Transactions on Nuclear Science 32, 78 (1985). doi: 10.1109/TNS.1985.4336794
Baidu ScholarGoogle Scholar
[13] Hamamatsu Photonics, "

Chapter 5: How to use photomultiplier tubs and peripheral circuits

." (2007). www.hamamatsu.com/resources/pdf/etd/PMT_handbook_v3aE-Chapter5.pdf
Baidu ScholarGoogle Scholar