1 Introduction
For decades, gas-type detectors have been widely used to detect particles produced in experiments in nuclear and particle physics research, typical examples of which are the multiwire proportional chamber [1-3], resistive plate chamber [4-6], and drift chamber (DC) [7]. Invented by David R. Nygren, the time projection chamber (TPC) is a new type of DC with excellent imaging capability [8]. It provides good space-time resolution, high detection efficiency, and 4π solid-angle coverage. The TPC has been widely used for many years in high-energy physics experiments [9-12], and it has also gradually been applied with some relevant modifications, in low-energy nuclear physics experiments [13-15].
With recent progress in the production of radioactive beams, nuclear physics is expanding into a much larger area of the nuclear chart. New phenomena have been explored using nuclear reaction tools, such as resonant scattering, transfer reactions, Coulomb excitation, and the breakup reaction [16-18]. Because the beam intensity is usually low for radioactive projectiles, the active target TPC (AT-TPC) has been developed to take advantage of a relatively thick target and full coverage of the detector solid angle [19,20]. However, combining the target and the detector and dealing with a broad dynamic range of particle energies would require careful design and substantial new test studies. Considering the need to study the cluster structure in unstable nuclei [18], we are developing a small AT-TPC. Its initial design and performance are presented in this article.
2 Design and construction
2.1 Mechanical structure and readout GEM foils
The TPC is essentially a gas-type particle detector. When a charged particle passes through the gas-filled vessel, the deposited energy causes ionization of gas molecules along the particle track. An electric field generated by the exterior electrodes would drive the electrons (ions) to drift toward the electrodes. The drift time can be measured to determine the distance between the track-crossing position (Z position) and the arrival electrode. On the other hand, the electron arrival position on the electrode (X-Y position) may be determined by a two-dimensional readout circuit. Therefore, the TPC serves as a three-dimensional imaging system to measure the particle track. The performance of a TPC depends on the distribution of the electric field in the vessel, the properties of the gas flow, the signal amplification and readout techniques, and so on.
The present AT-TPC is a simplified two-dimensional prototype that consists of a gas vessel, a field cage, and an end-cap readout detector, together with the associated electronics. This small AT-TPC is schematically illustrated in Fig.1. The volume of the field cage is 14 cm × 14 × 14 . The cage is surrounded by four printed circuit boards (PCBs), each having 14 copper strips (potential strips) on its inside face, and is installed between the cathode plate at the top and the readout board at the bottom. The size of the drift field is 14 cm, but the readout area is only 10 cm×10 cm;, as determined by the size of the gas electron multiplier (GEM) foil [21]. The TPC based on GEM readout is a new type of tracking detector that possesses good position resolution and a fast timing response. The presently adopted GEM is 50 μM in thickness and consists of a dielectric clad on both sides of a thin layer of copper. The copper layer was photolithographically etched on both sides, and a high-density array of holes was created. When a voltage of 300 to 400 V is applied to the copper layer, the electrons generated in the field cage can be amplified by two to three orders of magnitude. In the actual application, we applied two or three layers of the GEM foil, from which a gain of 105 can be achieved for electron multiplication. The electrons from the GEM layer are collected by 64 copper readout strips (1 mm in width and spaced 0.562 mm apart) printed on a PCB. The method by which a high voltage (HV) is applied to the field cage and the GEM layers is demonstrated in Fig.2. Inside the field cage, the electric potential is stepped down along the drift direction by a series of resistors. A bias voltage is also applied to the GEM amplifiers using a custom-made circuit (Fig.2). The distance between adjacent GEM foils is 3 mm. The lower GEM foil is also 3 mm from the readout board. A guard ring made of copper strips with a width of 15 mm was designed and fixed between the field cage and the upper GEM layer (Fig.2). Its performance will be described below.
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F001.jpg)
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F002.jpg)
To facilitate quantitative studies of the TPC performance, we define a coordinate system with the Z axis along the direction of electron drift in the cage (perpendicular to the GEM layer) and the X axis in the GEM plane but perpendicular to the readout strips (Fig.2). The Z = 0 position is fixed at the surface of the upper GEM foil.
2.2 Working gas mixture
The working gas of a typical TPC prototype should have properties such as a low dielectric strength, small diffusion coefficient, and fast drift velocity. Some gases such as Ar, CH4, CO2, and CF4 and their mixtures have been widely tested experimentally. However, for an AT-TPC, the working gas must also serve as the target material, the choice of which is very limited. In our design, which aims at detection of the cluster fragments (most likely α particles) from the decay of a resonance state in a mother nucleus, helium gas would be the best choice for the main gas component. In addition, a mixture of 4% CO2 gas was also adopted on the basis of literature reports [22] and our own tests. The latter helps to stabilize the gain of the working gas.
2.3 Field cage
The main frame of the field cage is a cube with a side length of 14 cm. The cage is fixed by PCBs having high mechanical strength. The drift field is established by an electrode board at the top of the cage and a series of potential strips located inside the upright boards (Fig. 1). The actual electric field is usually distorted near the edge of the field cage owing to the finite range of the electrode and the strips, which may degrade the time-position resolution of the TPC. To investigate possible improvement of the field uniformity, we performed simulation calculations based on the Garfield code [23]. The electrodes, strips, and resistors between the strips were specified in the code according to the parameters of the real setup (Fig. 2). Methods of tuning the field uniformity include the application of mirror strips or a guard ring [12,24]. In the present work, the latter method is employed. The ring is located 3.0 mm above the first GEM foil, as can be seen in Fig. 2. HVs of -2700 and -880 V are assumed at the cathode and the GEM foil, respectively. The guard ring is not connected to the resistor chain, and its HV is supplied independently. The electric field intensity distributions in a plane parallel to the GEM foil are plotted in Fig. 3. As an example, the Z position of this plane is chosen to be 6.5 mm. The voltage applied to the guard ring is varied and is labeled in the figure. It can be seen that the field intensity at the edge of the cage changes quite sensitively with the guard ring voltage. The most uniform field intensity can be obtained for a voltage of 900 V on the guard ring.
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F003.jpg)
3 Test results and discussion
3.1 Experimental setup
A diagram of the experimental setup and signal processing circuit is presented in Fig.4. The small AT-TPC prototype described above was installed in a sealed chamber. A gas flow composed of helium and CO2 (96:4) was supplied to the chamber at an inside gas pressure of 800 mbar. An 241Am radioactive source (α, 5.49 MeV) was placed on one side of the cage. A small home-made gas detector was placed on the opposite side of the cage to detect the α particles and to trigger the data acquisition system (DAQ). A collimation hole (2 mm in diameter and approximately 70 mm above the GEM foil) was made in front of the detector. HVs of -2700 and -880 V were applied to the cathode board and GEM foil, respectively, as in the simulation calculation.
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F004.jpg)
The PKU preamplifier is a type of charge-sensitive preamplifier designed in the subatomic particle detection laboratory at Peking University. It has been successfully used in several nuclear physics experiments [29,30]. Fig.5(a) shows the signals read out from some of the TPC strips (nos. 14, 15, and 16) and amplified by the PKU preamplifiers. The small home-made gas detector is a proportional counter. In another test experiment, two detectors of this type were located at a fixed distance, and their output was collected by VT120 preamplifiers. Fig.5(b) shows the signals produced by an α particle traversing these two detectors. The time resolution of these detector-preamplifier combinations will be investigated below.
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F005.jpg)
3.2 Observed tracks
The measured Z position of a track is deduced from the drift time of the ionized electrons, which is measured as the time difference between a start signal from the small α-particle detector and some associated stop signals from the readout strips (Fig.4). Distance/time calibration was realized by making a series of small collimation holes on the cage along the vertical (Z) direction and on the opposite side of the radiation source. The observed tracks with some measured timing separations therefore correspond to the hole separation distances, resulting in a distance/time calibration factor of 8.3 mm/μ s under the optimal condition. Fig.6 shows one of the measured track positions in the Z direction. It can be seen that the initial track positions are biased near the edges of the field cage (Fig.6, solid squares). This should be attributed to the field distortion in those areas, as already demonstrated by the simulation calculations. This track shift problem may be solved to a large extent by applying a voltage to the guard ring. The best result can be obtained by using a voltage of -950 V, as shown in Fig.6 by the solid circles. Fig.7 shows the charges collected [analog-to-digital converter (ADC) values] by the readout strips during the passage of an α particle in the cage. It also reflects the distortion of the initial electric field and the effect of field correction by the guard ring voltage. It is evident that a guard ring with an appropriate voltage can improve the performance of the AT-TPC. These observed timing and charge collection effects are confirmed by the corresponding simulations.
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F006.jpg)
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F007.jpg)
It is worth noting that the best guard ring voltage (-950 V) deviates slightly from that used in the simulation (-900 V). The reason is the additional resistors used at the input/output ends of the real HV circuit, which may share some HVs.
3.3 Overall performance of the small AT-TPC prototype
The performance of the present small AT-TPC prototype modified with a guard ring was studied using the experimental setup described above. The spatial resolution along the Z axis was deduced from 100 α particle tracks. Each track was fitted by a straight-line function. For the i th readout strip (i = 1–64), a residual value,
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F008.jpg)
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F009.jpg)
The uncertainty of the X position caused by the readout strips can be deduced as σx = 1.56 (mm)/sqrt(12) = 0.45 (mm) [26,27], where 1.56 mm is the spatial coverage of one strip.
To estimate the energy resolution, the energies deposited along the track were calculated by the standard LISE code [28] and were used to calibrate the ADC values strip by strip. The measured energies are presented in Fig.10 for 100 tracks. For the first strip, the extracted energy resolution (σ) is approximately 3.6 keV at an average deposited energy of approximately 18 keV. For the last strip, these values are 4.0 and 21 keV, respectively. Using the present setup of the small AT-TPC, we were able to observe real images of α + α elastic scattering events, one of which is displayed in Fig.11.
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F010.jpg)
-201807/1001-8042-29-07-006/alternativeImage/1001-8042-29-07-006-F011.jpg)
4 Conclusion
A small AT-TPC was designed and implemented. Its performance was greatly improved by inserting a guard ring just above the first GEM foil. The behavior of this guard ring was systematically studied by simulation calculations and also by experimental tests. A time resolution of less than 20 ns and a spatial resolution (standard deviation) of less than 0.2 mm along the Z direction were obtained. The uncertainty of the X position, which depends on the readout strip width, was deduced to be approximately 0.45 mm. An energy resolution (σ) of 3.6–4.0 keV can be achieved for deposited energies of 18–21 keV, respectively, as determined from each strip. Thanks to all of these good performance parameters, images of α + α elastic scattering can be clearly observed. It was demonstrated that the present technology meets the requirements for experimental investigation of cluster decay from some nuclear resonant states. The present work focused on a two-dimensional demonstration, but the AT-TPC could easily be extended to a three-dimensional detector by using just the pixel-type readout board.
Some read-out system for proportional multiwire chambers
. Nucl. Instr. Meth. 65, 217-220(1968). doi: 10.1016/0029-554X(68)90568-5Study and application of low pressure multi-wire proportional chambers
. Nucl. Instr. Meth. A 515, 718-724(2003). doi: 10.1016/j.nima.2003.08.113Study of a multi-wire proportional chamber with a cathode strip and delay-line readout
. Chin. Phys. C 33, 364-368(2009). doi: 10.1088/1674-1137/33/5/009Applications and new developments in resistive plate chambers
. IEEE Tran. Nucl. Sci. 49, 881-887(2002). doi: 10.1109/TNS.2002.1039583Beam test results of a resistive plate chamber made of Chinese bakelites
. Nucl. Instr. Meth. A 459, 513-522(2001). doi: 10.1016/S0168-9002(00)01046-9Study of spatial resolution properties of a glass RPC
. Nucl. Instr. Meth. f 663, 22-25(2012). doi: 10.1016/j.nima.2011.10.006The time projection chamber
. Physics Today 31, 46-53(1978), doi: 10.1063/1.2994775The DELPHI time projection chamber
. Nucl. Instr. Meth. A 283, 567-572(1989).doi: 10.1016/0168-9002(89)91417-4The STAR time projection chamber: a unique tool for studying high multiplicity events at RHIC
. Nucl. Instr. Meth. A 499, 659-678(2003). doi: 10.1016/S0168-9002(02)01964-2Development of a TPC detector for the ALICE experiment
. Nucl. Instr. Meth. A 419, 511-514(1998). doi: 10.1016/S0168-9002(98)00858-4Drift field improvement and test in a GEM-TPC prototype
. Chin. Phys. C 35, 56-60(2011). doi: 10.1088/1674-1137/35/1/012Nuclear Matter Distributions in 6He and 8He from Small Angle p-He Scattering in Inverse Kinematics at Intermediate Energy
. Phys. Rev. Lett. 78, 2313-2316(1997). doi: 10.1103/PhysRevLett.78.2313A time projection chamber for the three-dimensional reconstruction of two-proton radioactivity events
. Nucl. Instr. Meth. A 613, 65-78(2010). doi: 10.1016/j.nima.2009.10.140High-resolution study of Gamow-Teller transitions in the 48Ti(3He, t)48V reaction
. Phys. Rev. C 93, 064326(2016). doi: 10.1103/PhysRevC.93.064326Study of the structure of unstable nuclei through the reaction experiments
. Sci. China Phys. Mech. Astron. 54, Suppl.1 pp1-5(2011). doi: 10.1007/s11433-011-4423-9Observation of Enhanced Monopole Strength and Clustering in 12Be
. Phys. Rev. Lett. 112, 162501(2014). doi: 10.1103/PhysRevLett.112.162501Elastic scattering and breakup of 11Be on deuterons at 26.9 A MeV
. Phys. Rev. C 93, 064620(2016). doi: 10.1103/PhysRevC.94.064620Resonance state in 7 H
. Phys. Rev. Lett. 99, 062502(2007). doi: 10.1103/PhysRevLett.99.062502Energy levels of 18 F from the 14 N+ α resonant reaction
. Phys. Rev. C 90, 024302(2014). doi: 10.1103/PhysRevC.90.024302New observations with the gas electron multiplier (GEM)
. Nucl. Instr. Meth. A 396, 50-66(1997). doi: 10.1016/S0168-9002(97)00648-7Development of MAIKo: the active target with μ-PIC for RI beam experiments
. Journal of Physics: Conference Series 569, 012042(2014). doi: 10.1088/1742-6596/569/1/012042Prototype AT-TPC: Toward a new generation active target time projection chamber for radioactive beam experiments
. Nucl. Instr. Meth. A 691, 39-54(2012). doi: 10.1016/j.nima.2012.06.050Construction and calibration of the multi-neutron correlation spectrometer at Peking University
. Nucl. Instr. Meth. A 728, 47-52(2013). doi: 10.1016/j.nima.2013.06.037The simulation study and improvement of the drift field in TPC detector
,Recoil proton tagged knockout reaction for 8He
. Journal of Physics: Conference Series 436,012048 (2013). doi: 10.1088/1742-6596/436/1/012048High-lying excited states in 10Be from the 9Be(9Be,10Be)8Be reaction
, Science China PMA 60,062011(2017).doi: 10.1007/s11433-017-9023-x