1. Introduction
Plastic scintillators are commonly used in high-energy physics owing to their excellent optical transmission properties, simple production, low cost, and high performance in particle detections [1,2,3]. They have often served as an anticoincidence detector to provide a trigger signal [4,5,6]. However, scintillators are seldom used in position measurement owing to their poor position resolution [7]. Accordingly, a position-sensitive plastic scintillator detector is developed to realize position measurement and provide trigger information simultaneously.
A position-sensitive detector can be used for various applications. For example, combined with muon tomography, which is a technique that uses cosmic ray muons to generate three-dimensional images of volumes, the detector can be applied in the field of security imaging [8,9]. In the United States, the Decision Sciences International Corporation (DSIC) has implemented muon tracker technology in a multi-mode passive detection system (MMPDS) [10]. This port scanner located in Freeport, Bahamas can detect shield nuclear material, explosive, and contraband. A study of the image characterization metrics for MMPDS shows that the spatial resolution is centimeter magnitude order [10]. Therefore, a position-sensitive plastic scintillator may be applied in the field of homeland security owing to its large area and low cost. The spatial resolution must be of centimeter magnitude order.
In this paper, the design of grooves on the top and bottom sides of a plastic scintillator in which wavelength-shifting (WLS) fibers can be embedded is proposed. Alternate WLS fibers are gathered and routed to a multi-anode photomultiplier (PMT) for each dimension [11]. The scintillator detector is thus considered to be segmented into several small tiles in two dimensions. The design method was tested using physical simulation and an experiment.
This paper is organized as follows. In Section 2, we describe the design of the detector, including the plastic scintillator, grooves, and WLS fibers. In Section 3, we present a Monte Carlo simulation based on GEANT4 to estimate the position resolution. Section 4 describes the fabrication of the detector in detail. In Section 5, details of the calibration of multi-anode PMT are provided, and the results of cosmic test are presented.
2. Detector Design
The size of the plastic scintillator used in this design is 320×320×25 mm3, with 16 grooves milled uniformly on each side. Each groove is 9.5 mm in width and 1.1 mm in depth, as shown in Fig. 1. They are approximately 19.32 mm apart from each other, and nine WLS fibers of diameter 1 mm are embedded within each of them. Sixteen of these WLS bundles in the x-dimension correspond respectively to the 16 channels of the multi-anode PMT. Therefore, the plastic scintillator is considered to be divided into 16 tiles in the x-dimension. Furthermore, the 16 WLS fiber bundles in the y-dimension are arranged in the same way as in the x-dimension and they correspond to the y-dimension of the multi-anode PMT. Therefore, the plastic scintillator is considered to be segmented into 16×16 tiles in two dimensions [12].
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F001.jpg)
The outputs of the multi-anode PMTs are connected to the front-end electronics (FEE) board and the data acquisition (DAQ) system, with which the hitting position of an incident particle can be reconstructed in two dimensions.
3. Physical Simulation
To understand the scintillator detector with WLS fiber readout more intuitively, a model of the detector was built using GEANT4, which is the most commonly used simulation software in high-energy physics [13,14]. The scintillator detector was constructed according to the design method described above.
Consider a muon of 2 GeV passing through the detector. Its energy is deposited and converted into fluorescence in the detector. Subsequently, fluorescent photons are transmitted in the scintillator and absorbed by the WLS fiber bundles. The distribution of fluorescence transport is shown in Fig. 2.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F002.jpg)
We subsequently counted the quantities of fluorescent photons absorbed by each fiber bundle and reconstructed the center of distribution of fluorescence, which is defined as the hitting position of the muon event, by using the barycenter method. The reconstructed position is not fixed but obeys a Gaussian distribution, owing to system error and other factors. In this study, the sigma value fitting with Gaussian distribution is regarded as the position resolution. As shown in Fig. 3, the position resolution of the detector with WLS fiber readout reaches approximately 5 mm in the Monte Carlo simulation, which indicates the feasibility of the design.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F003.jpg)
4. Detector Fabrication
A piece of BC414 produced by Saint-Gobain with the dimensions of 320×320×25 mm3was milled using a programmable milling machine according to the design method described above. After the completion of milling, the scintillator was annealed at 70 °C for three hours to remove any remaining stress in order to avoid crazing and consequent performance degradation [15].
To optimize light collection [16,17], the BCF-92 was used to match the absorption spectrum of WLS fibers to the emitting spectrum of the plastic scintillator. The BCF-92 [18], also produced by Saint-Gobain, is a kind of WLS and light-transmitting fiber, with an emission peak at 492 nm. The EJ-500 [19], produced by Eljen Technology, is a clear and colorless epoxy cement and is ideal for optically bonding plastic scintillators and fibers. The H8711 (manufactured by Hamamatsu) was used to match the emission wavelength of the BCF-92. Its quantum efficiency is approximately 30% at 490 nm [20]. Moreover, the H8711 has 4×4 anodes correspondingly reading out the fluorescent signals of the 16 WLS bundles. An adapter guides the 16 WLS bundles to the corresponding 16 cathode cells of the multi-anode PMT in the x-dimension. Another adapter performs the same task in the y-dimension. By placing this arrangement into a large black box of size 440×440×50 mm3 to protect the scintillator detector from external light, the preparation was completed. The fabrication process was as follows:
1) Cut the WLS fibers to the appropriate length, and polish the end surfaces.
2) Assemble nine WLS fibers as one bundle (16 bundles in all).
3) Fix all the WLS bundles into an adapter and ensure that their ends are all located in a plane; these are the readout ends of the fibers.
4) Place the WLS bundles into the corresponding grooves of the scintillator, and the other end of the fibers are covered with the reflective material.
5) Glue the WLS bundles with EJ-500 and wait for them to dry (approximately 24 h).
6) Combine the multi-anode PMT with the WLS bundles, and thus, the fabrication of the x-dimension of the detector is complete.
7) Construct the y-dimension of the detector by repeating the above steps.
A photo of the position-sensitive scintillator detector is shown in Fig. 4.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F004.jpg)
With an appropriate multi-anode PMT base design, the 16 channels of charge signals from each multi-anode PMT can be coupled to the FEE board consisting of a VA32 ASIC [21], which has 32 analog input channels for measuring the charge signals. Each channel is composed of a charge-sensitive pre-amplifier (CSA), CR-RC-shaping amplifier, and sample-hold circuit. Therefore, the PMT output charge signals are integrated by the CSA and thereafter shaped into a semi-Gaussian pulse with a peaking time of approximately 1.8 μs. Furthermore, a hold signal is used to sample all the analog channels.
Subsequently, the output current signals obtained from the VA32 chip are wired together and led to a current-to-voltage conversion circuit, digitized using a 16-bit analog-to-digital converter (ADC, AD976A [22]). When a trigger is provided by PMT dynode signals or the telescope system, the ADC digital output data is recorded by the DAQ system [23].
5. Calibration and Cosmic Test
5.1 Calibration of Multi-anode PMTs
As two multi-anode PMTs are used in the design, their gain should be calibrated first. The peak of a single photoelectron can be used as the calibration parameter because of its linearity to the gain. Figure 5 shows a block diagram of the LED calibration system.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F005.jpg)
The LED lights are driven by the pulse generator, whose output voltage can be adjusted to vary the light intensity. The photoelectron peak becomes the main part of the spectrum for each anode channel of the calibrating PMT, as shown in Fig. 6. A random trigger is used to obtain the pedestal in the test.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F006.jpg)
After the calibration, 16 spectra of each multi-anode PMT in the x- (and y-) dimension were obtained. By choosing one of the ADC peak channels as a gain reference, the others can be normalized. The relative ratios are demonstrated in Fig. 7, and they can be used as the normalization factors for the ADC counts of the corresponding anodes in position reconstruction.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F007.jpg)
In addition, the cross talk of the multi-anode PMT is less than 1%, as presented in the datasheet [20]. Therefore, it is believed that the effect of cross talk for different anodes can be ignored in the experiment.
5.2 Cosmic Test
A telescope system is required to study the performance of the detector with WLS fiber readout [24]. First, two small plastic scintillator detectors (20×20 mm2) were used as the telescope system—one each placed above and below the detector, as shown in Fig. 8.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F008.jpg)
Cosmic muon events were triggered by the small scintillator telescope. When the logic (S1*S2) equaled one, the data of the ADC counts from each of the 16 readout channels of the x- and y- dimensions were collected. First, the response efficiency of the detector was obtained by adjusting the voltage of the multi-anode PMTs. By comparing the event counts passing through the telescope and the event counts passing through the test detector, the detection efficiencies of the test detector under different voltages can be calculated. As shown in Fig. 9, the detection efficiency is over 90% when the voltage exceeds 790 V. The working voltage of the PMT in the following test is set to 800 V.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F009.jpg)
Thus, the position where the cosmic muons pass through the test detector can be reconstructed using the barycenter method. Consider the x-dimension. When a triggered incident muon (ith-event) passes through the test detector, its position can be reconstructed using the following formula:
where
Fig. 10 shows the scatter points of reconstructed cosmic events passing through the telescope and test detector. The distribution of scatter points is only a projection image of the telescope hit by cosmic muons.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F010.jpg)
To evaluate the position resolution of the detector, it is necessary to consider another telescope with better spatial resolution. In this case, Micromegas (MM) detectors with sub-millimeter position resolution [25] were used as the telescope system. During the cosmic test, two MM detectors were placed above and below the detector, as shown in Fig. 8, with MMI and MM2 replacing S1 and S2, respectively. Both the DAQ of the test detector and the MM telescope were synchronized by the trigger (MM1*MM2). Hence, the track of an incident muon (i th-event) can be identified using the two Micromegas detectors. It is regarded as the reference track (i) and is across the test detector at (
Fig. 11 shows the histograms ofΔx (left) and Δy(right).The fitted position resolution of the detector is approximately 8.6 mm.
-201808/1001-8042-29-08-011/alternativeImage/1001-8042-29-08-011-F011.jpg)
6. Conclusion
In this paper, the design and fabrication of a position-sensitive plastic scintillator detector with WLS fiber readout were presented. The performance of the detector was also studied. The results of the cosmic test showed that the positions of the incident particles can be reconstructed in two dimensions with a detector and the position resolution can be as high as approximately 8.6 mm, which is close to the simulation result (approximately 5 mm).
Low-cost extruded plastic scintillator
. Nucl Instrum Meth. A 466 482-491 (2001). doi: 10.1016/S0168-9002(01)00177-2Plastic scintillators with efficient neutron/gamma pulse shape discrimination
. Nucl Instrum Meth. A 668 : 88-93 (2012). doi: 10.1016/j.nima.2011.11.071The anti-coincidence detector for the GLAST large area telescope
. Astrophys. 27 339-358 (2007). doi: 10.1016/j.astropartphys.2006.12.003The AGILE anticoincidence detector
. Nucl. Instrum. Meth. A 556 228-236 (2006). doi: 10.1016/j.nima.2005.10.016Scattering of 7Be and 8B and the astrophysical S17 factor
. Phy. Rev. C 73, 025808 (2006). doi: 10.1103/PhysRevC.73.025808Analysis of neutron and photon detection position for the calibration of plastic (BC-420) and liquid (BC-501) scintillators
. Nucl Instrum Meth. A 524 92-101 (2004). doi: 10.1016/j.nima.2003.12.036Statistical reconstruction for cosmic ray muon tomography
.GEANT4 simulation of plastic scintillator strips with embedded optical fibers for a prototype of tomographic system
. Nucl Instrum Meth. A 624, 583-590 (2010). doi: 10.1016/j.nima.2010.10.012Image characterization metrics for muon tomography
.Studies of proximity focusing RICH with an aerogel radiator using flat-panel multi-anode PMTs (Hamamatsu H8500)
. Nucl Instrum Meth. A 521, 367-377 (2004). doi: 10.1016/j.nima.2003.11.384Performance of 8×8 pixel LaBr3:Ce and Gd2 SiO5:Ce scintillator arrays coupled to a 64-channel multi-anode PMT
. IEEE T Nucl. Sci., 56, No. 6, 2009. doi: 10.1109/TNS.2009.2034657GEANT4-a simulation toolkit
. Nucl Instrum Meth. A 506 250-303 (2003). doi: 10.1016/S0168-9002(03)01368-8GEANT4 simulation of plastic scintillator strips with embedded optical fibers for a ptototype of tomographic system
. Nucl Instrum Meth. A 624, 583-590 (2010). doi: 10.1016/j.nima.2010.10.012High efficiency plastic scintillator detector with wavelength-shifting fiber readout for the GLAST Large Area Telescope
. Nucl Instrum Meth. A 583, 372-381 (2007). doi: 10.1016/j.nima.2007.09.040Plastic scintillation dosimetry: Optimal selection of scintillating fibers and scintillators
. Med. Phys. 32, 2271-2278 (2005). doi: 10.1118/1.1943807A scintillating fibre detector for the Crystal Barrel experiment at ELSA
. Nucl Instrum Meth. A 538 416-424 (2005). doi: 10.1016/j.nima.2004.09.029Scintillating Fiber BCF-92
”. http://www.crystals.saint-gobain.com/uploadedFiles/SG-Crystals/Documents/SGC%20Fibers%20Brochure.pdfOPTICAL CEMENT EJ-500
”. http://www.eljentechnology.com/images/products/data_sheets/EJ-500.pdf16ch multi anode PMT assembly h8711 datasheet
”. https://www.hamamatsu.com/resources/pdf/etd/H8711_TPMH1320E.pdf33 Channel High-Dynamic Range CSA with Sample-and-Hold
”. http://www.ideas.no/products/va32hdr14-3/16-Bit, 100kSPS/200kSPS BiCMOS A/D Converters
”. http://www.analog.com/media/en/technical-documentation/data-sheets/AD976_976A.pdfDesign of the readout electronics for the BGO calorimeter of DAMPE mission
. IEEE T. Nucl. Sci. 62, No. 6, 2015. doi: 10.1109/TNS.2015.2479091Search for time-reversal violation in the β decay of polarized 8Li nuclei
. Phys.Rev.Lett, 90, 202301 (2003). doi: 10.1103/PhysRevLett.90.202301Manufacture and performance of the thermal-bonding Micromegas prototype
. J. Instrum., 9, C10028 (2014). doi: 10.1088/1748-0221/9/10/C10028