logo

Noninvasive beam diagnosis based on the TM010 mode

ACCELERATOR, RAY TECHNOLOGY AND APPLICATIONS

Noninvasive beam diagnosis based on the TM010 mode

Chuang-Ye Song
Wen-Hui Huang
Nuclear Science and TechniquesVol.35, No.7Article number 127Published in print Jul 2024Available online 16 Jul 2024
63807

A resonant cavity based on the TM010 mode is an effective tool for noninvasive beam characterization. This technique has the advantages of a high signal-to-noise ratio, compact structure, and is related to multiple parameters compared with other beam monitors. In this study, high-precision measurements of the bunch charge, arrival time, bunch length, and energy parameters based on the TM010 mode are discussed. A cavity beam arrival time monitor (BAM) utilizing a phase cavity has been widely used in many facilities. Regarding bunch-length measurements, the influence of the beam energy, beam offset, and longitudinal spectrum on the TM010 mode are carefully considered to reduce errors, and the theoretical resolution of two cavities with different frequencies is analyzed. Owing to the dependence of the beam velocity of the beam loss factor, this method can also be used for the detection low beam energy using two cavities with the same frequency but different cavity lengths. A set of three cavities with different lengths and frequencies of 1.902 GHz and 11.424 GHz is presented for measuring the four aforementioned parameters.

TM010 modeNoninvasive diagnosisBeam lengthLow energy
1

Introduction

Beam diagnostics are critical for understanding the characteristics of electron beams in accelerator facilities [1, 2]. Considering the development of free-electron lasers (FELs) and ultrafast electron diffraction (UED) facilities for example, the demand for highly precise beam diagnostics is continuously increasing[3-8], especially for the ultrafast exploration in pump-probe experiments. Moreover, noninvasive devices are favorable options that can provide real-time measurements and feedback, even bunch-by-bunch. Beam monitoring based on the TM010 mode is a nondestructive method that has been utilized in many facilities owing to its many advantages, such as compactness, simple structure, and high sensitivity [9-11]. Because the TM010 mode excited by the passing beam is related to many beam parameters, including the bunch charge, arrival time, bunch energy, and length[12, 13], this method has significant potential for measuring the four aforementioned parameters with a high resolution.

Regarding a cylindrical cavity, the amplitude of the TM010 mode is proportional to the bunch charge. A resonant cavity can be employed for the bunch charge and arrival time measurements by precisely detecting the monopole mode, which is known as the cavity beam arrival time monitor (BAM) or the reference cavity of the cavity beam position monitor (CBPM)[14, 15]. This approach can achieve a high resolution and has been applied in many facilities [16-18]. The best resolution of the cavity BAM was reported ([16]) to be 13 fs at 250 pC, which was obtained at the Linac Coherent Light Source (LCLS) using a 2806 MHz cavity.

The bunch length is a key parameter of the beam longitudinal features. Many methods have been proposed to obtain the beam length [19-23]; however, only a few are nondestructive. Considering the development of FELs and UEDs, a noninvasive bunch-length monitor for short bunches is currently required, such as for those shorter than 1 ps. The TM010 mode can be used for measuring the bunch length owing to the influence of the bunch length on the amplitude of the TM010 mode [13, 24]. Compared to other techniques, this method has the potential for a high-resolution for measuring short bunches. It usually requires two TM010 modes with different frequencies to measure the amplitudes of the beam longitudinal spectrum at two frequencies. For this technique, the beam energy, transverse offset, and longitudinal distribution all have an impact on the amplitude of the TM010 mode, which may cause significant inaccuracies [25]. The errors caused by these factors will be carefully analyzed in this article.

The beam loss factor is also related to the beam energy [26, 27], indicating that the monopole mode may be used to measure the beam energy. Compared to techniques such as those using dipole magnets [28], measuring the bunch energy based on the TM010 mode is compact and noninvasive. This technique demands two cavities with the same frequencies of the TM010 modes to normalize the bunch length and charge; longer cavities can achieve a higher resolution. However, considering a limitation of this method, it can only be used for energy measurements of low-energy electron beams; therefore, it is particularly suitable for UEDs [5] and ultrafast electron microscopes (UEMs) [29].

Because the TM010 mode is related to multiple parameters, the coupling effect must be limited when measuring a certain parameter. For the bunch length or energy measurements, the two cavities should be specifically designed to normalize the other parameters. A set of three cavities with different structures and frequencies of 1.902 GHz and 11.424 GHz is proposed for the diagnostics of the four beam parameters in this study. This design is expected to provide noninvasive measurements and save considerable space in accelerators compared to other beam diagnostic devices.

The relationship between each beam parameter and the TM010 mode is theoretically analyzed in Sect. 2. Section 3 presents the simulation results of the different parameters and evaluation experiments. A design of three cavities for the diagnostics of the four beam parameters is presented in Sect. 4.

2

Theoretical basis of the TM010 mode

When a relativistic electron beam moves along a cavity, the electromagnetic field of a series of eigenmodes is excited. Applying this theory, the voltage of a particular resonant mode excited in the cavity (Vexc) is represented as follows: Vexc=klossq=ω2(RQ)q, (1) where kloss is the loss factor, ω is the resonant angular frequency of the mode, q is the bunch charge, and RQ is the normalized shunt impedance, which is a key factor for evaluating the beam loading effect. The monopole mode has the highest sensitivity of the resonant modes. Regarding the monopole mode of a pillbox cavity, RQ is calculated as follows: (RQ)010=2J02(j01xd)LT2ωϵ0πJ12(j01)d2. (2) Here, x is the beam offset, d is the cavity radius, L is the length of the pillbox cavity, T is the transit time factor, Jn is the nth-order Bessel function, and j01 is the 1st root. Notably, the RQ of the monopole mode of the pillbox is insensitive to the beam offset near the cavity center.

In principle, the loss factors of bunches passing through a cavity at a nonrelativistic velocity are lower than those at an ultrarelativistic velocity. The energy loss in the cavity should vanish when the velocity approaches 0. The influence of the bunch length must also be considered if the bunch length is not negligible with respect to the frequency of the TM010 mode. Assuming a Gaussian bunch with an rms length σs and velocity v=β c passing through a cylindrical pillbox, the line density is λ(s)=exp(s2/2σs2)/(2πσs). By integrating over all the particles in the bunch, the dependence of the beam loss factor on β and σs can be expressed as follow:[26, 27]. k010(β,σs)k010(1,σs)=exp(ω2σs2γ2β2c2)(βsinj01L2βdsinj01L2d)2, (3) where c is the speed of light, and β and γ are the relativistic factors. Describing the beam length using the longitudinal time scale σt=σs/v, the output voltage extracted from a coupling structure with an impedance Z can be expressed as follows: Vout=ωq2ZQext(RQ)exp(ω2σt22)(βsinj01L2βdsinj01L2d) (4) where Qext is the external quality factor, which mainly depends on the coupling part. The RQ in Eq.(4) is solely determined by the cavity shape and is unrelated to β. As shown in Eq.(4), the output voltage is proportional to the bunch charge q, and related to the bunch length σt and velocity β c. As the bunch length increases or the velocity decreases, the voltage decreases. The TM010 mode can be used for the measurement of these parameters, but the coupling effect between the parameters should be taken seriously.

Regarding the measurement of the bunch charge, cavities with a low resonant frequency and short cavity length are preferred to reduce the influence of the bunch length and energy jitter, especially under conditions of a long bunch length or low energy.

Vout in Eq.(4) is proportional to exp(ω2σt22), indicating that two cavities with different frequencies can be used for measuring the bunch length, which can be determined by dividing the two TM010 mode signals, simultaneously normalizing the bunch charge. Since Vout is also related to β, it may impact the bunch-length measurement. Regarding the ultrarelativistic case, Vout is independent of β. On the contrary, the energy jitter of the low energy beam will cause large errors in the accuracy of the beam length. The Vout can be rewritten as follows: Vout=qκexp(ω2σt22)RQ(β), (5) where κ=ω2ZQext, which is only determined by the cavity structure. Based on Eq.(5), by using two cavities with the same RQ(β), the effect of the bunch energy can be eliminated. The RQ depends only on the cavity shape and has no dependence on the material or surface condition of the cavity. Therefore, for the bunch length measurement, when the dimensions of the two cavities are proportional, the RQ values are the same but the frequencies differ. Additionally, the beam offset and longitudinal distribution can also cause errors, which will be simulated and analyzed in Sect. 3.

For measuring the bunch energy, the charge must be normalized; furthermore, the exponential term in Eq. (4), exp(ω2σt22), is also not negligible for long bunches. Therefore, two cavities with the same frequency but different L/d are required to eliminate the influence of the bunch length and charge when detecting the bunch energy. In this case, the dependence of the voltage amplitude on β is as follows: Voutβsinj01L2βdsinj01L2d. (6)

The variation in the amplitude of the TM010 mode with β and the ratio of the cavity length L and radius d is shown in Fig. 1(a). Note, a nearly linear relationship between β and the voltage amplitude can be observed Fig. 1 when β>0.86 and L/d<2.

Fig. 1
(Color online) (a) Voltage amplitude deviations relative to the beam velocity βc and L/d value of the pillbox. (b) Voltage amplitude dependence on the beam energy Ek with different L/d values. The V0 value on the Y-axis represents the voltage amplitude when the beam speed approaches c
pic

Compared to β, the beam energy Ek is a more commonly used parameter. The relationship between the beam energy and β is expressed by the following formula: Ek=(11β21)m0c2. (7) According to Eqs. (6) and (7), the relationship between the amplitude of the TM010 mode and the beam energy Ek is shown in Fig. 1(b). When the energy of the particle is above 5 MeV, the amplitude variations as the energy decreases can be ignored. The amplitude of the TM010 mode begins to decrease when the energy is below 5 MeV and further rapidly decreases with a lower energy. Therefore, the TM010 mode can be an effective tool to measure the low beam energy Ek, but it is not suitable for high-energy situations.

3

Simulation and test results of different parameters

3.1
Simulation and analysis of the cavity beam length monitor
3.1.1
Influence of the beam energy and the resolution analysis

Based on the discussion in Sect. 2, two cavities with proportional dimensions have the same RQ(β) values and the division of the two TM010 mode signals can be employed to measure the bunch length, effectively mitigating the influence of the bunch charge and energy jitter. To verify the aforementioned analysis, two cavities were established in the CST software. The dimensions of the two models are shown in Fig. 2(a). The two cavities have the same shape, with a size ratio of 1:2. The frequencies of the two cavities were 7.912 GHz and 3.956 GHz; the simulated results of RQ with different β values are shown in Fig. 2(b). The RQ(β) values of the two cavities are entirely identical.

Fig. 2
(Color online) (a) Two models built in CST; the dimensional ratio of the two cavities is 1:2 (b) RQ deviations of the two cavities with different β values
pic

By cancelling the RQ(β) in Eq. (5), the bunch length measured by two TM010 modes at different frequencies can be expressed as follows: σt=2ω22ω12ln(κ2V1κ1V2). (8) The resolution Δσt of the detecting system can be derived from Eq. (8) as: Δσt=10SNR/20σt(ω12ω22), (9) where SNR is the signal-to-noise ratio of the TM010 mode voltage measurement system. As shown in Eq. (9), the resolution of this method depends on the difference between the two frequencies and is inversely proportional to the beam length. However, when the bunch length was excessively long, the voltage amplitude rapidly decreased, resulting in a poor signal-to-noise ratio of the system, thus worsening the resolution. When the two frequencies were 5 and 10 GHz, a 106-fs resolution was achieved for a 1-ps bunch measurement with the SNR assumed to be 70 dB.

3.1.2
Influence of the transverse offset

The output voltage is proportional to the RQ of the resonant mode. For a cylindrical cavity without the tube and coupling structure, Eq. (2) provides its RQ value of the TM010 mode. Based on Eqs. (2) and (4), the voltage in a cylindrical cavity is proportional to J0(kc01x) and is maximized on the cavity axis. To study the influence of the beam transverse offset x, a closed pillbox with a radius of 20 mm was established in the CST software [30], and the simulation was evaluated using the eigenmode and wakefield solver. The theory and simulation results are demonstrated in Fig. 3, which are in good agreement.

Fig. 3
Normalized R/Q of the TM010 mode in the pillbox cavity along the cavity diameter
pic

For cavities with tubes, the electric field intensity of the TM010 mode along the cavity diameter is inconsistent with the J0 function, especially for the pipe area. According to [31, 32], the RQ of a pillbox with a tube can be expressed as follows: RQ=(V0T(x))2ωU=(V0T(0)I0(Kx))2ωU. (10) where K=2πγβλ, I0 is the modified Bessel function, and T is the transit time factor. The effect of the transverse offset is included in the transit time factor, and the changes caused by the offset are negligible according to the properties of the I0(Kx) function. A model of the pillbox with a tube was also simulated with the eigenmode solver in CST, as shown in Fig. 4(a). The RQ values for different offsets on the z-axis were calculated considering the transit time factor, and the normalized results are presented in Fig. 4(b). The amplitude of the TM010 mode of the pillbox with a tube can be considered invariant with respect to the beam offset compared with that of the closed cavity.

Fig. 4
(Color online) Pillbox with a tube in CST and normalized R/Q of the TM010 mode of the pillbox with a tube
pic

Coupling structures and cavity types, such as the reentrant type, can lead to noncylindrically symmetric structures of the cavity. In these cases, the beam offset effect no longer conforms to the aforementioned theoretical analysis. To study the influence, a simulation of the reentrant cavity monitor with different coupling ports was performed in the CST software. Figure 5 demonstrates the four cavities established for the simulation and the amplitude variation of the TM010 modes. For the cavity with a single port, the beam offset had a significant impact on the output signal from the port; the change was approximately 1% with a 3 mm offset. For the four-port cavity, the amplitude variation owing to the transverse offset was less than 1.

Fig. 5
(Color online) Models of the reentrant cavities with different ports and the normalized R/Q variation of the TM010 modes of the aforementioned models
pic
3.1.3
Influence of the beam longitudinal distribution

The measurement of the bunch length based on the TM010 mode was achieved by evaluating the amplitudes of two frequencies of the longitudinal spectrum of the beam and should be independent of the distribution. Because the shape of the bunch is not Gaussian in an actual scenario, the influence of different distributions should be considered. For a bunch with a longitudinally symmetric distribution, the beam spectrum F(ω) expanded in the Taylor’s series can be written in the following form [13]: F(ω)=+f(t)cos(ωt)dtt0+t0f(t)(1ω2t22+ω4t424+O(ω6t6))dtq(112ω2σt2+ρ+324ω4σt4), (11) where f(t) is the beam longitudinal density function. ρ is the kurtosis, describing the tailedness of a distribution relative to a Gaussian distribution, and is 0 if f(t) is Gaussian. For ρ>0, the beam distribution is thin-tailed, and referred to as leptokurtic. For ρ<0, the distribution tends to be platykurtic, which is more realistic for electron beams. The probability densities of certain distributions with different kurtosis values are shown in Fig. 6. The difference between the spectra of the Gaussian and symmetric distributions with different kurtosis values can be described as follows: |F(ω)FG(ω)FG(ω)|=|ρ24ω4σt4|112ω2σt2+324ω4σt4. (12)

Fig. 6
(Color online) Various symmetric distributions with different kurtosis values
pic

If the distribution of a longitudinally symmetric bunch is considered, in which the kurtosis significantly differs from that of the Gaussian distribution, such as the double exponential distribution (ρ=3), and if the error is limited to be within 1, the criterion of the beam length and frequency can be obtained as follows: 0<ωσt<0.296. (13) This is a rather strict criterion since the real shape of the bunch is usually between the Gaussian and rectangular distributions, 0<|ρ|>1.2.

In conclusion, the resolution in Eq.(9) is better with a higher frequency for short bunches, but the frequency is limited to ensure that the influence of different bunch shapes is negligible.

3.2
Simulation and analysis for low-energy measurement
3.2.1
Influence of the resonant frequency

As analyzed in Sect. 2, the TM010 mode can be used for low-energy measurements. However, for a low energy, the bunch length and distribution may significantly vary owing to the space charge effects and low magnetic stiffness, which will affect the accuracy. The Eq. (4) can be rewritten as follows: VoutqF(ω0)(βsinj01L2βdsinj01L2d), (14) F(ω0) represents the component of the longitudinal beam spectrum at the cavity frequency ω0. Therefore, when using two cavities of the same frequency, both the beam charge and beam length can be normalized by dividing the TM010 mode signals of the two cavities.

A simulation of two cavities was conducted in the CST software to validate the analysis. The two cavities were designed to have the same frequency but a different lengths, as shown in Fig. 7(a). Using the CST Wakefield solver, the voltage signals of the two cavities with different beam lengths and β values were simulated. The amplitude dependence on β are plotted in Fig. 7(b). The amplitude varies with β at beam lengths of 3 and 6 ps, but the division of the signals of the two cavities is irrelevant to the beam length, and only depends on the energy. Generally, cavities with long lengths demonstrate a high dependence on β, and two cavities with the same frequency are required to obtain the beam energy. According to Eq. (7), the deviations in the amplitude of two cavities at different energies can be obtained, as illustrated in Fig. 7(c). the amplitude apparently decreases more rapidly at lower energies, aligning with the discussion in Sect. 2.

Fig. 7
(Color online) (a) Two models established in the CST with the same frequencies and different lengths. (b) Simulated amplitude deviations simulated with different β values of the two cavities. (c) Simulated amplitude deviations at different beam energies obtained by the relationship between Ek and β. The V0 value on the Y-axis represents the voltage amplitude when the beam speed approaches c
pic
3.2.2
Resolution analysis for low-energy measurement

The relationship between β and the voltage can be approximated to be linear when β>0.86 and L/d<2, as indicated in Sect. 2. To normalize the beam charge and beam length, two cavities with the same frequency are required to measure the beam energy. The β parameter can be calculated from the two voltages of the TM010 modes, which is expressed as follows: β=a×Vout1Vout2+(1a), (15) where a is the linear fitting coefficient that represents the sensitivity of the voltage signal to β, Vout is the normalized voltage, and β reaches 1 when the particle velocity is c. By obtaining the derivative of both sides of Eq. (15), the resolution of β can be obtained. (Δβa)2=(ΔVout1Vout1)2+(ΔVout2Vout2)2 (16) Δβ=2a×SNR (17) By substituting the resolution of β into the relationship between Ek and β, as shown in Eq. (7), the expression for the resolution of Ek can be derived as follows: ΔE=β(Ek+m0c2)3Δβm02c4. (18) Because Δ E is proportional to β and (Ek+m0c2)3, the resolution rapidly deteriorates as the energy increases. Simulations based on our 4.76-GHz BAM pickup [33] were performed to apply the aforementioned analysis. The TM010 mode dependence on β can be obtained by changing the velocity of the particles. Figure 8 presents the simulation results and the linear fitting equation. The ΔE values were calculated by using Eq. (18) and are summarized in Table 1, assuming that SNR is 70 dB, and the other low-frequency cavity is nearly independent of β and only used for charge normalization. The resolution can reach 3.17 keV at 0.5 MeV, but only 151.99 keV at 3 MeV.

Fig. 8
The TM010 mode amplitude dependence on β of the 4.76 GHz pickup
pic
Table 1
Simulation results of the beam energy resolution with the 4.76GHz pickup
Ek(MeV) Resolution (keV) Precision ()
0.5 3.17 6.34
1 11.52 11.52
2 55.02 27.51
3 151.99 50.66
Show more
3.3
Evaluation test of the bunch length and energy

Evaluation experiments were performed based on a 4.76-GHz BAM pickup [33], which was installed after the RF gun in our test beamline of the Tsinghua Thomson scattering X-ray source facility [34], as seen in Fig. 9. The main beam parameters are summarized in Table 2.

Fig. 9
(Color online) Scheme of the test beamline
pic
Table 2
Test beamline beam parameters
Parameters Test beamline
Bunch charge (pC) 20-30
Laser pulse length (ps) 3
Electron energy (MeV) 0.5-3
Pulse repetition rate (Hz) 10
Show more

In the test, the beam energy ranged from 0.5 MeV to 3 MeV and the raw rf signals from the pickup were captured by a 20 GHz bandwidth oscilloscope. Because the variation in the bunch charge also affects the amplitude of the pickup, an adjacent Faraday cup was used to calibrate the charge information. The amplitude variation of the pickup with the energy is plotted in Fig. 10. The sensitivity is 130.54 mV/pC at an energy of 3 MeV, but when the energy is reduced to 0.5 MeV, the sensitivity is only 70.92 mV/pC, decreasing to 54%.

Fig. 10
(Color online) Measured data and simulated results of the 4.76 GHz pickup dependence on beam energy and length
pic

Figure 10 also presents the simulation results obtained in the CST with a beam length of 3 ps, which is the same as the laser pulse length. However, the simulated sensitivity is only reduced to 95% at 0.5 MeV, which is inconsistent with the experimental result. This occurred because the beam length increases at lower energies owing to the space charge force effects, which are not considered in the CST Wakefield solver. A simulation using Astra was performed with the space charge force activated, and the beam length at different energies was obtained. Subsequently, we imported the bunch energy and bunch-length parameters into CST to recalculate the amplitude of the pickup. As shown in Fig. 10, the experimental data and the simulation results obtained with the CST and Astra [35] are in good agreement.

4

A design of three cavities for the four-parameter measurement

Two cavities can be used for the bunch length or low-energy measurement. For the TM010 mode in the cavity, a high frequency is sensitive to the beam length, and a long cavity length is sensitive to the beam energy, while the TM010 mode with a low frequency and short length was used as the BAM monitor and for the charge normalization. A set of three cavities was designed to measure the beam charge, arrival time, beam length, and energy. Figure 11 presents the three cavities established in CST. The frequency of Cavity1 and Cavity2 was 1.904 GHz for both, but the cavity lengths were significantly different. The two cavities can be used to measure the beam energy with the beam length normalized. The frequency of Cavity3 increased to 11.424 GHz, and the sizes of Cavity3 and Cavity2 were proportional. Therefore, Cavity2 and Cavity3 can be used for measuring the beam length. Note, the three cavities were designed with four ports to reduce the influence of the beam offset.

Fig. 11
(Color online) Three cavities designed for measuring the four parameters
pic

In the simulation, we found that increasing the length of the Cavity2 demonstrated a more apparent energy dependence, but the frequency of the TM011 mode gradually approached that of the TM010 mode. The long cavity length may also cause errors owing to the tilted incidence of the beam. The cavity length should be carefully chosen to avoid the effect of the TM011 mode and the beam tilted incident. The final optimized dimensions and simulated rf parameters are summarized in Table 3.

Table 3
Optimized dimensions and simulated rf parameters of the three cavities
Parameter Cavity1 Cavity2 Cavity3
Cavity radius (mm) 32.50 61.62 10.27
Cavity length (mm) 20.00 84.00 14.00
Cut radius (mm) 20.00
Cut length (mm) 15.00
Pipe radius (mm) 5.00 24.00 4.00
Frequency (GHz) 1.904 1.904 11.424
R/Q(Ω) 84.29 106.09 106.09
Show more

The amplitudes of the TM010 modes in Cavity1 and Cavity2 were monitored while scanning the β values. As shown in Fig. 12, the voltage amplitude of Cavity1 was barely reduced compared to that of Cavity2. The ratio of the two normalized signals was proportional to β with a coefficient of 1.951. The energy measurement resolutions calculated by Eq. (18) are summarized in Table 4, with a 70 dB SNR. The theoretical resolution reached 0.70 keV at 0.5 MeV and decreased to 37.59 keV at 3 MeV.

Fig. 12
(Color online) (a) Simulated amplitude dependence on β of Cavity1 and Cavity2. (b) Simulated amplitude deviations at different beam energies of Cavity1 and Cavity2. The V0 value on the Y-axis represents the voltage amplitude when the beam speed approaches c
pic
Table 4
Theoretical energy resolution of the dual-cavity design
Ek(MeV) Resolution (keV) Precision ()
0.5 0.78 1.56
1 2.84 2.84
2 13.61 6.80
3 37.59 12.53
Show more

The RQ values of the TM010 modes in Cavity2 and Cavity3 were also simulated with different β values. As shown in Fig. 13(a), the RQ(β) value of the two cavities are consistent. For the bunch-length measurement, a parameter scan of the bunch length was applied in the simulation. The simulated amplitude deviations of Cavity2 and Cavity3 are shown in Fig. 13(b), which are consistent with the theoretical analysis. The maximum bunch length was 4.2 ps with a limit of Eq. (13). A simulation of the effect of the beam offset was also conducted for the two cavities, both containing four coupling ports. The results indicate that the TM010 mode varies by less than 1, implying insensitivity to the beam offset. According to Eq. (9), the resolution can reach 63.13 fs at a bunch length of 1 ps bunch with an SNR of 70 dB.

Fig. 13
(a) RQ(β) of Cavity2 and Cavity3. (b) Simulated amplitude variations of the dual-cavity with the Gaussian distribution bunches
pic

As shown in Fig. 12, the amplitude variation of Cavity1 was significantly small with β>0.86. The frequency of Cavity1 was 1.904 GHz, which is the same as that of Cavity2, which indicates that the amplitude of Cavity1 did not decrease when the bunch length was less than 6 ps. Therefore, Cavity1 can be used for monitoring the bunch charge and arrival time, as previously analyzed.

5

Summary

In this study, we analyzed beam diagnostics based on the TM010 mode; the impacts of multiple beam parameters on the TM010 mode were studied. The bunch length and energy measurements both require two TM010 modes. Regarding the measurement of the bunch length, the errors caused by the beam energy, beam offset, and longitudinal distribution were thoroughly considered. For low-energy measurements, the amplitude of the TM010 mode was nearly linear with β when β>0.86, and the bunch length was normalized with the same frequency of the two cavities. This method was used for low-energy measurements, achieving a high accuracy with a long cavity length. An evaluation test of the bunch length and energy was performed with a 4.76 GHz BAM pikcup. Finally, a set of three cavities with frequencies of 1.904 GHz and 11.424 GHz was proposed for measuring the four parameters, which reached a 63.13-fs resolution at 1 ps and a 1.56 precision at 0.5 MeV with a 70 dB SNR.

References
1 P. Strehl, Beam instrumentation and diagnostics. (Springer, Berlin, 2006), pp. 110
2 J.D. Fan, Y.J. Tong, Y.G. Nie et al.,

First commissioning results of the coherent scattering and imaging endstation at the Shanghai soft X-ray free-electron laser facility

. Nucl. Sci. Tech. 33, 114 (2022). https://doi.org/10.1007/s41365-022-01103-0
Baidu ScholarGoogle Scholar
3 C. Feng, H.X. Deng,

Review of fully coherent free-electron lasers

. Nucl. Sci. Tech. 29, 160 (2018). https://doi.org/10.1007/s41365-018-0490-1
Baidu ScholarGoogle Scholar
4 L. Wang, W.C. Fang, Z.T. Zhao,

Design and optimization of low-emittance C-band photocathode RF electron gun

. Nuclear Techniques 44, 060201 (2021) (in Chinese). https://doi.org/10.11889/j.0253-3219.2021.hjs.44.060201
Baidu ScholarGoogle Scholar
5 R.K. Li, C.X. Tang, Y.C. Du et al.,

Experimental demonstration of high quality MeV ultrafast electron diffraction

. Rev. Sci. Instrum. 80, 083303 (2009). https://doi.org/10.1063/1.3194047
Baidu ScholarGoogle Scholar
6 Z.T. Zhao, D. Wang, Q. Gu et al.,

SXFEL: A Soft X-ray Free Electron Laser in China

. Synchrotron Radiation News. 30, 29-33 (2017). https://doi.org/10.1080/08940886.2017.1386997
Baidu ScholarGoogle Scholar
7 Z. Hui, L. Yu, H. Zhou et al.,

X-ray crystallography experimental data screening based on convolutional neural network algorithms

. Nuclear Techniques 46, 030101 (2023) (in Chinese). https://doi.org/10.11889/j.0253-3219.2023.hjs.46.030101
Baidu ScholarGoogle Scholar
8 H. Chen, L.M. Zheng, B. Gao et al.,

Beam dynamics optimization of very-high-frequency gun photoinjector

. Nucl. Sci. Tech. 33, 116 (2022). https://doi.org/10.1007/ s41365-022-01105-y
Baidu ScholarGoogle Scholar
9 H. Maesaka, H. Ego, C. Kondo et al.,

Electron beam diagnostic system for the Japanese XFEL

, in Proceedings of the 1st International Beam Instrumentation Conference. (Tsukuba, Japan, 2012), pp. 38-46
Baidu ScholarGoogle Scholar
10 Y. Otake,

Advanced diagnosis of the temporal characteristics of ultra-short electron beams

. Nucl. Instrum. Methods A 637, S7-S11 (2011). https://doi.org/10.1016/j.nima.2010.02.010
Baidu ScholarGoogle Scholar
11 S.S. Cao, Y.B. Leng, R.X. Yuan et al.,

Optimization of beam arrival and flight time measurement system based on cavity monitors at the SXFEL

. IEEE Trans. Nucl. Sci. 68, 2-8 (2020). https://doi.org/10.1109/TNS.2020.3034337
Baidu ScholarGoogle Scholar
12 Q. Luo, B. G Sun, Z.R. Zhou et al.,

Design of cavity beam monitor at HLS

. in Proceedings of IPAC2011. (San Sebastián, Spain, 2011), pp. 1278-1280
Baidu ScholarGoogle Scholar
13 Z.C. Chen, W.M. Zhou, Y.B. Leng et al.,

Subpicosecond beam length measurement study based on the TM010 mode

. Phys. Rev. ST Accel. Beams 16, 072801 (2013). https://doi.org/10.1103/PhysRevSTAB.16.072801
Baidu ScholarGoogle Scholar
14 J. Chen, Y.B. Leng, L.Y. Yu et al.,

Beam test results of high Q CBPM prototype for SXFEL

. Nucl. Sci. Tech. 28, 51 (2017). https://doi.org/10.1007/s41365-017-0195-x
Baidu ScholarGoogle Scholar
15 J. Chen, Y.B. Leng, L.Y. Yu et al.,

Optimization of the cavity beam-position monitor system for the Shanghai Soft X-ray Free-Electron Laser user facility

. Nucl. Sci. Tech. 33, 124 (2022). https://doi.org/10.1007/s41365-022-01117-8
Baidu ScholarGoogle Scholar
16 A. Brachmann, C. Bostedt, J. Bozek et al.,

Femtosecond operation of the LCLS for user experiments

. in Proceedings of the IPAC2010. (Kyoto, Japan, 2010), pp. 2287-2289
Baidu ScholarGoogle Scholar
17 S.S. Cao, R.X Yuan, J. Chen et al.,

Dual-cavity beam arrival time monitor design for the Shanghai soft X-ray FEL facility

. Nucl. Sci. Tech. 30, 72 (2019). https://doi.org/10.1007/s41365-019-0593-3
Baidu ScholarGoogle Scholar
18 Y. Otake, T. Ohshima, N. Hosoda et al.,

LLRF and timing system for the SCSS test accelerator at SPring-8

. Nucl. Instrum. Methods A 696, 151-163 (2012). https://doi.org/10.1016/j.nima.2012.06.067
Baidu ScholarGoogle Scholar
19 D.X. Wang, G.A. Krafft, C.K. Sinclair,

Measurement of femtosecond electron bunches using a rf zero-phasing method

. Phys. Rev. E 57, 2283 (1998). https://doi.org/10.1103/PhysRevE.57.2283
Baidu ScholarGoogle Scholar
20 J.H. Tan, W.C. Fang, D.C. Tong et al.,

Design, RF measurement, tuning, and high-power test of an X-band deflector for Soft X-ray Free Electron Lasers (SXFEL) at SINAP

. Nucl. Instrum. Methods A 930, 210-219 (2019). https://doi.org/10.1016/j.nima.2019.03.093
Baidu ScholarGoogle Scholar
21 Y. Bian, W.Y. Zhang, B. Liu et al.,

Sub-picosecond electron bunch length measurement using coherent transition radiation at SXFEL

. Nucl. Sci. Tech. 29, 74 (2018). https://doi.org/10.1007/s41365-018-0399-8
Baidu ScholarGoogle Scholar
22 T. Watanabe, J. Sugahara, T. Yoshimatsu et al.,

Overall comparison of subpicosecond electron beam diagnostics by the polychromator, the interferometer and the femtosecond streak camera

. Nucl. Instrum. Methods A 480, 315 (2002). https://doi.org/10.1016/S0168-9002(01)00952-4
Baidu ScholarGoogle Scholar
23 G. Berden, S.P. Jamison, A.M. MacLeod et al.,

Electro-Optic Technique with Improved Time Resolution for Real-Time, Nondestructive, Single-Shot Measurements of Femtosecond Electron Bunch Profiles

. Phys. Rev. Lett. 93, 114802 (2004). https://doi.org/10.1103/PhysRevLett.93.114802
Baidu ScholarGoogle Scholar
24 Q. Wang, Q. Luo, B.G. Sun,

Design and performance study of an improved cavity bunch length monitor based on an optimized offline test scheme

. Nucl. Instrum. Methods A 968, 163975 (2020). https://doi.org/10.1016/j.nima.2020.163975
Baidu ScholarGoogle Scholar
25 Q. Wang, Q. Luo, B.G. Sun,

Study on the influence of beam transverse position on the cavity bunch length measurement

. J. Phys. Conf. Ser. 1350, 012153 (2019). https://doi.org/10.1088/1742-6596/1350/1/012153
Baidu ScholarGoogle Scholar
26 S.S. Kurennoy,

Dependence of bunch energy loss in cavities on beam velocity

. Phys. Rev. ST Accel. Beams 2, 032001 (1999). https://doi.org/10.1103/PhysRevSTAB.2.032001
Baidu ScholarGoogle Scholar
27 D.O. Jeon,

Measurement of beam loading at the SNS superconducting linac for a beam with β<1

. Nucl. Instrum. Methods A 597, 132 (2008). https://doi.org/10.1016/j.nima.2008.08.145
Baidu ScholarGoogle Scholar
28 E. Kongmon, P. Wongkummoon, S. Rimjaem,

Simulations and measurements of the dipole magnet using in the 4-MeV electron spectrometer

. J. Phys. Conf. Ser. 1144, 012108 (2018). https://doi.org/10.1088/1742-6596/1144/1/012108
Baidu ScholarGoogle Scholar
29 S.A. Aseyev, E.A. Ryabov, B.N. Mironov et al.,

The development of ultrafast electron microscopy

. Crystals 10, 452 (2020). https://doi.org/10.3390/cryst10060452
Baidu ScholarGoogle Scholar
30

CST Microwave Studio, Germany

, www.cst.com, 2018.
Baidu ScholarGoogle Scholar
31 T.P. Wangler, RF Linear Accelerators, 2nd edn. (Wiley, New York,2008), pp. 3247
32 P.M. Lapostolle,

Proton linear accelerators: A theoretical and historical introduction

, Los Alamos National Laboratory Report No. LA-11601-MS, 1989.
Baidu ScholarGoogle Scholar
33 C.Y. Song, Y.Q. Jia, Z.Y. Liu et al.,

High-resolution cavity bunch arrival-time monitor for charges less than 1 pC

. Phys. Rev. Accel. Beams 26, 012803 (2023). https://doi.org/10.1103/PhysRevAccelBeams.26.012803
Baidu ScholarGoogle Scholar
34 C.X. Tang, W.H. Huang, R.K. Li et al.,

Tsinghua Thomson scattering X-ray source

. Nucl. Instrum. Methods A 608, S70-S74 (2009). https://doi.org/10.1016/j.nima.2009.05.088
Baidu ScholarGoogle Scholar
35 K. Floettmann,

Astra: A space charge tracking algorithm

. Version 3.2 March 2017. https://www.desy.de/~mpyflo/Astra_manual/Astra-Manual_V3.2.pdf
Baidu ScholarGoogle Scholar
Footnote

The authors declare that they have no competing interests