The role of AGFA high-energy CR in the calibration and quality control of multileaf collimators (MLC)

LOW ENERGY ACCELERATOR, RAY TECHNOLOGY AND APPLICATIONS

The role of AGFA high-energy CR in the calibration and quality control of multileaf collimators (MLC)

LI Zhaobin
XIONG Fei
HUANG Guofeng
CAO Zheng
JIANG Ruiyao
FU Shen
Nuclear Science and TechniquesVol.18, No.3pp.145-149Published in print 20 Jun 2007
57500

Linear accelerators equipped with multileaf collimators (MLC) are becoming more common and are widely used in the intensity-modulated radiotherapy (IMRT). There is an imperative need to ensure the commissioning specification of the linear accelerators for the sake of quality assurance (QA) and quality control (QC). This paper is aimed to investigate the role of AGFA high-energy CR (Computed Radiography) in calibrating dynamic multileaf collimators and evaluating the accuracy of the leaf position. The result shows that AGFA high-energy CR can easily and conveniently be used to calibrate MLC and verify its position. Hence, the application of AGFA high-energy CR is proved to be an accurate and time-saving method for routine MLC QC, especially when MLC calibration adjustments are required.

High-energy CRMLC calibrationQuality control
References
[1] Boyer A L, Ochran T G, Nyerick C E, et al. Med Phys, 1992, 19: 1255-1261.
[2] Galvin J M, Smith A R, Lally B. Int J Radiat Oncol Biol Phys, 1993, 25: 181-192.
[3] Huq S M, Yu Y, Chen Z P, et al. Med Phys, 1995, 22: 241-247.
[4] Jordan T J, Williams P C. Phys Med Biol, 1994, 39: 231-251.
[5] Alan R H, Thomas J J. Radiotherapy and Oncology, 1997, 45: 225-233.
[6] Sandra C V, René A B, Maarten L P D. Radiotherapy and Oncology, 2006, 80: 86-92.
[7] James H V, Atherton S, Budgell G J, et al. Phys Med Biol, 2000, 45(2): 495-509.
[8] ZHU Zhengfei, XU Zhiyong, FU Xiaolong, et al. China Oncology, 2006, 16(2): 154-157.
[9] Marion E, Mark D L, Maarten L P D, et al. Radiotherapy and Oncology, 2001, 60: 215-224.
[10] De Wagter C, Martens C, De Deene Y, et al. Cancer/Radiother, 1999, 3(Suppl 1): 171-182.
[11] Don S, Hildebolt C F, Sharp T L, et al. Radiology, 1999, 213(2): 455-460.
[12] Swee R G, Gray J E, Beabout J W, et al. American Journal of Roentgenology, 1997, 168(2): 539-542.
[13] Sanfridsson J, Holje G, Svahn G, et al. Acta Radiologica, 2000, 41(4): 310.