Introduction
The neutron spectrum, which is the neutron fluence rate or neutron density versus neutron energy, is one of the most critical parameters for describing the neutron irradiation field of a radiation source [1]. Obtaining the on-site energy spectrum of a neutron field is a relatively complicated issue in neutron measurements [2,3], which has motivated studies to not only determine more exact measurement methods, unfolding algorithms, and simulation methods [4] but also the application of rational criteria to ensure the accurate unfolding of the neutron spectrum. Furthermore, although the neutron ambient dose equivalent rate
To obtain the neutron spectra and
Methods and measurement
As shown in Fig. 1, the neutron spectrum measurement includes the following four main parts: (1) calculation of the BSS response function, (2) on-site measurement, (3) unfolding of the spectrum based on the response function and measurement data, and (4) rationality evaluation of the neutron spectrum using various criteria.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F001.jpg)
BSS and spectral unfolding methods
The BSS, manufactured by Centronic Ltd., UK, consists of an SP9 3He thermal neutron detector and a set of high-density polyethylene spheres (0.938 ± 0.009 g/m3) with the following 12 different diameters: 3, 3.5, 4, 4.5, 5, 6, 7, 8, 9, 9.5, 10, and 12 inches (1 inch=2.54 cm). The internal diameter of SP9 was 3.2 cm, the stainless-steel shell was 1 mm thick, and the nominal pressures were 230 kPa 3He and 120 kPa Kr at 20 ℃. The incident neutrons moderated by polyethylene spheres were detected by SP9 at the center of each sphere and a coupled series of suitable electronic equipment [12,13].
The neutron spectrum unfolding method of the BSS is based on the theory of the development of few channel data, which is an underdetermined system of equations [2,14] that has infinite solutions. When the response function Ri(E) and each detector count rate Ci are known, the neutron fluence rate
The maximum entropy, iteration, and GA methods were used to unfold the neutron energy spectra and provide mutual verification. The maximum entropy and iteration methods achieved by FORTRAN integrated into UMG, called MAXED and GRAVEL, respectively, were employed to obtain the neutron energy spectrum. A priori default spectrum for the preliminary determination of the structure of the neutron energy spectrum was required when using the maximum entropy and iteration methods. The two methods aim to change the default spectrum into one that fits the data but remains “as close as possible” to the default spectrum. Therefore, the a priori default spectrum significantly affects the final results of the spectral unfolding process. Detailed descriptions and analyses of these two methods are available in previous studies [15,16]. For MAXED, the chi-squared per degree of freedom was 1.8, and the highest energy, temperature, and temp reductions were set to 20, 1, and 0.85, respectively. For GRAVEL, the chi-square per degree of freedom was 1.8, and the highest energy and maximum number of iterations were set to 20 and 2000, respectively. The selection of the default spectrum used in this study is described in Sect. 2.3.
The GA, which imitates the Darwinian evolution paradigm, also known as the “survival of the fittest” strategy, was applied to solve the problem of unfolding the neutron energy spectrum [17]. It is a new type of unfolding method that has recently emerged and can eliminate the limitation of requiring a default energy spectrum during calculations. In this study, the GA code was achieved using C++ within the ROOT framework. The main parameters used in the code were as follows: the upper and lower boundaries of the spectral unfolding zone were 20 MeV and 0, respectively, genetic algebra was 500, population size was 30000, crossover probability was 0.3, mutation probability was 0.001, and genetic probability was 0.2.
Model of response functions calculation
The response functions, which are the key parameters of a BSS, were calculated using Geant4, a Monte Carlo code library used for simulating particle transport, coupled with ENDF/B-VII [18]. Figure 2 shows the geometry of a 4.5 inches polyethylene sphere with SP9. The incident neutrons were assumed to be a monoenergetic parallel beam emitted from a circular plane with a diameter equal to that of the simulated sphere, and the circular plane was placed at a distance of 20 cm from the SP9 center. The structure, dimensions, material, and material density of the SP9 and polyethylene spheres were the same as the actual parameters described above. S(α, β) was applied to the neutron transport calculation in polyethylene to solve the problem of neutron thermalization [19]. Excluding the polyethylene sphere and SP9, all the spaces were set to a vacuum. A neutron was considered to escape, and the trace was stopped once it occurred in the vacuum zone during the simulation. The calculated response energy ranged from 10-9 eV to 20 MeV, considering 20 energy blocks per order of magnitude from 10-9 eV to 10 MeV, and five energy blocks from 10 MeV to 20 MeV. In addition, the responses of six experimental energies, 144 keV, 250 keV, 565 keV, 1.2 MeV, 2.5 MeV, and 14.8 MeV, which were determined from an accelerator-driven neutron source were also calculated.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F002.jpg)
To validate the simulated BSS energy response values, the six aforementioned energy responses were calibrated in a monoenergetic neutron field using the shadow cone method, and the thermal neutron response of SP9 was calibrated in a thermal neutron reference irradiation field formed by a neutron source and graphite moderator [20] at PTB. We should also consider that the air gap between the SP9 holder and sphere is a key factor for BSS [21].
Neutron spectral analysis of reactor building
Neutron energy spectra have different structures for incident neutrons with various energies because the absorption cross-section of the moderator varies with the neutron energy. Neutrons above 20 MeV are produced by a proton-induced reaction on a lithium target in the laboratory; its characteristics can be described by a quantum molecular dynamics model [22]. However, in fission reactors or other common neutron fields, the neutron energy is lower than 20 MeV and the energy spectra can be expressed by the following three equations:
In the thermal region (<0.5 eV), the energy spectrum follows the Maxwellian distribution as follows (Eq. (3)) [23]:
In the intermediate energy region (0.5 eV-0.1 MeV), the classical 1/E distribution was abandoned because the function diverges for E→0 and decreases very slowly for an energy of ~1 keV; the distribution was replaced by Eq. (4) as follows [24]:
Moreover, for the fission energy spectrum of U-235 (fast neutrons, 1-20 MeV), the distribution is approximated by the Watt distribution (Eq. (5)) [3,25], which can be explained by the nuclear evaporation model as follows [26]:
Normally, fission neutrons emitted from U-235 are moderated by both the coolant and moderator, such as light water and heavy water, and then become thermal neutrons and a small number of intermediate energy neutrons, epithermal neutrons, and fission neutrons [27].
According to the physical process, a hypothesis indicating that the spectrum of a reactor building can be explained as a superposition of these energy spectra with different parameters and weight factors can be proposed as follows (Eq.(6)):
Neutrons in the thermal energy region account for a high proportion of the final spectra, and the most probable energy of a fast neutron may move to a lower energy (<1 MeV) owing to the surrounding moderators. In addition, the neutron-absorbing material also affects the final superposition energy spectrum around the measurement locations. In addition, moderators with a high resonance absorption cross-section distort the energy spectrum in the intermediate energy region, although this effect can be neglected owing to the low energy resolution of the BSS in the intermediate energy region.
According to the analysis of the neutron energy spectrum of reactor buildings, a previously published [28] neutron spectrum at the workplace of a nuclear power plant (Fig. 3), with a covering thermal, intermediate energy, and fast neutrons, was chosen as the default spectrum for the MAXED and GRAVEL methods.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F003.jpg)
On-site measurement
The neutron energy spectra were measured at the following six locations: F1-A (305), F1-B (access route), F2-C (1R326), F2-D (1R511), F3-E (position between pumps 1 and 3 in 501), and F3-F (501 ladder). F1-A and F1-B are located in factory 1, F2-C and F2-D in factory 2, and F3-E and F3-F in factory 3. Table 1 summarizes the measurement devices used. Extra polyethylene holders should fill the holes of the Bonner spheres as much as possible (Fig. 4(a)) because the air gap between the holder and sphere has a significant influence on the response function. The direction of SP9 was parallel to the floor during the measurement (Fig. 4(b)). The height of each sphere support varied to make the geometric center of SP9 with different moderators 35 cm above the ground. According to the uncertainty formula of the linear model (Eq. (7)), the time and total count of a single measurement should be greater than 200 s and 13000, respectively, to ensure that the relative uncertainty of the count rate is less than 1%.
Device | Model |
---|---|
Detector | SP9 and 12 polyethylene spheres (Centronic, UK) |
Preamplifier | 142PC (Ortec, USA) |
Main amplifier | 570 (Ortec, USA) |
High voltage | 556 (Ortec, USA) |
Multichannel analyzer | USB - MCA4 CH (TechnoAP, Japan) |
Neutron ambient dose equivalent meter | LB6411 (Berthold, Germany) |
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F004.jpg)
In addition, a neutron dose equivalent ratemeter, LB6411, a 3He detector located at the center of a 25 cm diameter polyethylene moderator, was employed to directly measure the neutron dose equivalent rate
The on-site process of measurement relied on the following three assumptions: (1) The fluence rate of the neutron radiation field is homogeneous in the local area of measurement, and the minor positional changes when replacing different spheres are negligible. (2) The fluence rate of the neutron radiation field was stable and unchanged during the measurement. (3) The disturbance induced by BSS can be ignored in the measurement area of the neutron radiation field.
Results and discussion
Response functions and uncertainty
The response functions that were simulated using Geant4 are shown in Fig. 5. The thermal neutron response of SP9 was 3.2 cm2 based on the calibration results in PTB, and the real gas pressure of 3He in SP9 was 248.7 kPa, which is 7.5% higher than the nominal value. Therefore, in subsequent simulations, the actual value of the gas pressure was used. Figure 6 shows the deviation of the simulation compared to the experiment, which indicates the D-values between the ratio of the simulation to the experimental value and 1; excluding the 3 inches and 3.5 inches spheres at a neutron energy of 14.8 MeV, all the results were within 0.9~1.1 at each energy point, indicating that they sufficiently coincided with one another. The largest relative deviation was observed at a high neutron energy (14.8 MeV) for the smallest sphere (3 inches), and the maximum was 14%. Table 2 lists the combined uncertainty (k=1) and components, including the 3He pressure, density of the polyethylene sphere, air gap, isotropy of the sphere, statistic of uncertainty by simulation, maximum deviation, and experimental uncertainty. The uncertainty of the 3He pressure was fixed at 1% for all the spheres and was mainly due to the absorption of the thermal energy by stainless steel. The uncertainty of the density of polyethylene caused by the polyethylene nonuniformity and measurement error ranged from 5% to 13%, which increased as the diameter of the sphere increased. A constant air gap between the SP9 holder and the sphere of 0.4 mm was used in the response function simulations, and the estimations of the deduced uncertainties caused by this factor were based on a 0.4 mm variation. The largest uncertainty (approximately 12 %) occurred for the 3 inches sphere responses. The isotropy of the sphere uncertainty and statistical uncertainty of the simulation changed monotonically, decreasing from 3% to 1% and increasing from 1% to 4%, respectively, when the sphere diameter varied from 3 inches to 12 inches The largest experimental uncertainty of each sphere was relatively stable, which was approximately 4-5% during the calibration based on a monoenergetic neutron field. For each sphere, the maximum deviation between the experimental and simulation results was also selected among the six energy points as an uncertainty component of the response function, which are representative because they cover the thermal neutrons, intermediate energy neutrons, and fast neutrons. Finally, the maximum combined uncertainty, namely 20% (k = 1), was considered as the final response function uncertainty owing to the inability to obtain the experimental values of all the energy points.
Sphere diameter (inch) | 3He pressureuncertainty | Density uncertainty of PE sphere | Air gapuncertainty | Isotropy uncertainty of sphere | Statistical uncertainty of simulation | Experimentuncertainty*1 | Maximum deviation*2 | Combined uncertainty |
---|---|---|---|---|---|---|---|---|
3 | 1% | 5% | 12% | 3% | 1% | 5% | 14% | 20% |
3.5 | 1% | 5% | 10% | 3% | 1% | 5% | 13% | 18% |
4 | 1% | 5% | 8% | 2% | 1% | 4% | 9% | 14% |
4.5 | 1% | 6% | 6% | 2% | 1% | 5% | 7% | 12% |
5 | 1% | 6% | 5% | 2% | 1% | 5% | 7% | 12% |
6 | 1% | 7% | 5% | 1% | 1% | 5% | 5% | 11% |
7 | 1% | 8% | 5% | 1% | 1% | 4% | 5% | 12% |
8 | 1% | 9% | 4% | 1% | 1% | 5% | 5% | 12% |
9 | 1% | 10% | 4% | 1% | 2% | 5% | 4% | 13% |
9.5 | 1% | 11% | 4% | 1% | 2% | 5% | 4% | 14% |
10 | 1% | 12% | 4% | 1% | 2% | 5% | 4% | 14% |
12 | 1% | 13% | 4% | 1% | 4% | 4% | 4% | 15% |
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F005.jpg)
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F006.jpg)
Normalization
Figure 7 displays the normalized measurement results of the BSS relative to the average count rate with polyethylene spheres and a bare detector at the six locations. The normalization constants at the six locations were 3616, 729, 979, 343, 1354, and 1871 cps, respectively, which were the average count rates of each location. The peak package of F1-A was approximately 3 inches, which was lower than the other locations, indicating that F1-A would have the lowest neutron energy among these positions according to the response function shown in Fig. 5. The peak locations of F1-B and F3-E corresponded to a larger sphere, approximately 4-5 in, thus the neutron energy could be higher than that of the other locations. The peak locations were essentially the same and were located at 3.5 inches or 4 inches for F2-C, F2-D, and F3-F. Therefore, a preliminary conclusion indicates that the neutron energy at each location would be in the following order: F1-B and F3-E > F2-C, F2-D, and F3-F > F1-A.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F007.jpg)
Unfolding of energy spectra
Figure 8 shows the neutron energy spectra at six positions obtained by MAXED and GRAVEL; the two methods had nearly the same distribution. The thermal and fast neutron peak positions occurred at approximately 0.01 eV and 100 keV, respectively. Although the peak height of the GA unfolded spectrum was different from that of the other methods at certain locations, their corresponding peak positions were essentially in unity in the entire energy region, and the most probable energies of the fast neutrons were also near 10-100 keV. We also observed that the spectrum given by GA at F1-B and F3-E did not have a small thermal neutron peak compared to MAXED and GRAVEL at the same position. This may be caused by the constraint of the default spectrum required when using MAXED and GRAVEL, which presents an apparent peak for thermal neutrons. In addition, Fig. 9 shows the proportion of neutrons with different energy regions over the total neutrons at the six locations. Compared to MAXED and GRAVEL, there were larger proportions of intermediate-energy neutrons and smaller proportions of thermal neutrons and fast neutrons given by GA. This may be due to the influence of the default spectrum when using MAXED and GRAVEL.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F008.jpg)
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F009.jpg)
According to the shapes and structures of the unfolded energy spectra (Figs. 8 and 9), three groups were identified; this grouping was consistent with the pre-analysis by the count rate. For group 1, including the spectra obtained at F2-C, F2-D, and F3-F, which was in good agreement with the aforementioned analysis in Section 2.3, most of the neutrons were concentrated in the thermal region, and there was an apparent spike in the fast neutron energy region. For F2-D, the energy of the thermal neutron peak given by GA (1 eV) was higher than that of MAXED and GRAVEL (0.1 eV). Thermal, intermediate energy, and fast neutrons accounted for approximately 40%, 50%, and 10% of the total, respectively. For group 2, including the spectra obtained at F1-A, nearly all the neutrons were located in the thermal (approximately 50%) and intermediate energy (approximately 50%) regions, and the proportion of fast neutrons was near 0%, as shown in Fig. 9. The mean energy given by the three methods was 0.3 eV, indicating that a thicker material with a higher neutron moderation efficiency may be present in the environment. For group 3, including the neutron spectra obtained at F1-B and F3-E, the intermediate energy and fast neutrons were the vast majority (> 80 %), whereas the thermal neutron proportions were only 10-20%. Their corresponding average energies were 49.8 keV and 91.6 keV, respectively, which was significantly higher than those of the other locations in the same buildings, 0.3 keV of F1-A (Fig. 8(a)) and 42.9 keV of F3-F (Fig. 8(f)). However, considering the high mean fluence rates at F1-A (3167 cm-2·s-1) and F3-F (1474 cm-2·s-1), the mean fluence rates at F1-B and F3-E were relatively low at 513 and 967 cm-2·s-1, respectively, as shown in Fig. 10. A higher neutron energy and lower fluence rate indicates that there should be an extra absorption material of thermal neutrons nearby; intermediate and fast neutrons that were not totally moderated were detected by the BSS.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F010.jpg)
Neutron ambient dose equivalent rate
It is easy to derive the neutron ambient dose equivalent rate
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F011.jpg)
The
Evaluation of neutron spectra
There is no determined and unique spectrum; owing to the few-channel data of the BSS, certain criteria and conditions are required to verify the results of the unfolding spectra and ensure that they approximate the true values. However, all criteria are necessary and insufficient for the true spectrum. Three criteria, including the count rates of the inverse solutions, neutron spectrum structures, and key parameter comparison among the three methods, were used to evaluate the quality of the spectrum unfolding results.
Count rates of the inverse solutions and experiment
The count rate of the inverse solution of the neutron spectrum is one of the most important constraints. Although the spectra may not indicate the correct results despite the inverse solution values sufficiently matching the experimental results, poor results of the unfolding spectra can be excluded if they are inconsistent with one another. Figure 12 shows the absolute experimental count rate, inverse solution count rate, and their ratios for the three unfolding methods applied in this study. The inverse count rate of the GA approximated the measurement results more closely than MAXED and GRAVEL at most locations, and the maximum relative deviation was 18% for the 9 inches sphere at F1-A. It is difficult to quantify the uncertainty of the neutron spectrum involving a response function, measurement, experiment (temperature, scattering, etc.), unfolding spectrum, and other unknown factors, especially in an on-site complex environment. Therefore, a ±20% relative deviation between the measurement and inverse solution values may be considered satisfactory owing to the 20% response function uncertainty and 1% measurement uncertainty.
-202212/1001-8042-33-12-014/alternativeImage/1001-8042-33-12-014-F012.jpg)
Neutron spectrum structures
Generally, an approximation of the spectrum can be obtained for the measurement spot. First, according to the peak package of the measurement count rate and the response function of the BSS, we qualitatively evaluated the neutron energy. In particular, this can provide a comparison of the neutron energy for the measurement of multiple locations. In this study, the grouping based on neutron spectrum structures was consistent with the grouping based on the count rate analysis. The order of the spectra average energy in the three groups was as follows: group 3 > group 1 > group 2. Second, according to the hypothesis of the energy spectrum superposition of reactor buildings, the spectra of group 1 were in good agreement with the hypothesis. The neutron energy spectra in groups 2 and 3 can also be expressed by Eq. (7) with different weight factors. In group 3, C3 of the spectra had the greatest value compared to C1 and C2, whereas in group 2, C3 was close to 0. Therefore, the six spectra met the theoretical hypothesis considering that they had different parameters and weight factors, indicating that the unfolding spectrum results were credible.
Key parameter comparisons of three methods
The key parameters of a neutron spectrum include the total fluence rate, average energy, and
Total fluence rate (cm-2·s-1) | Average energy (keV) | ||||||||
---|---|---|---|---|---|---|---|---|---|
MXD | GRV | GA | MXD | GRV | GA | MXD | GRV | GA | |
F1-A | 1.6% | 1.0% | 2.5% | 33% | 33% | 0 | 2.6% | 1.7% | 4.3% |
F1-B | 0.1% | 0.3% | 0.4% | 0 | 10% | 10% | 2.2% | 5.0% | 7.2% |
F2-C | 0.2% | 1.2% | 1.0% | 3.7% | 13% | 9.3% | 1.5% | 4.1% | 2.6% |
F2-D | 0.7% | 1.0% | 1.7% | 3.7% | 1.2% | 4.8% | 0.8% | 1.1% | 1.9% |
F3-E | 0.3% | 0.1% | 0.4% | 2.9% | 9.4% | 12% | 1.4% | 0.8% | 2.1% |
F3-F | 0 | 0.5% | 0.5% | 0.7% | 6.1% | 5.4% | 2.5% | 0.2% | 2.6% |
Conclusion
Neutron spectra measurements were conducted by NIM using a BSS at six on-site locations in the complex environment of the Qinshan Nuclear Power Plant, and the maximum entropy, iteration, and GA unfolding methods were applied to obtain more precise spectra and parameters. The following three conclusions were drawn from the analysis.
(1) The neutron spectrum shape, total fluence rate
(2) For the measurement of the neutron ambient dose equivalent rate
(3) The results of the unfolding energy spectra obtained by the three methods were relatively reliable based on the following three criteria: count rates of the inverse solutions, neutron spectrum structures, and comparison of the key parameters derived from the spectra. More criteria need to be identified for a better evaluation of the neutron spectra and unfolding methods in the future.
Neutronics analysis of a stacked structure for a subcritical system with LEU solution driven by a D-T neutron source for 99Mo production
. Nucl. Sci. Tech. 32(11), 123 (2021). doi: 10.1007/s41365-021-00968-x.Investigation of an online reactor neutron spectrum measurement method with ionization chambers
. Nucl. Technol. 202, 94-100 (2018). doi: 10.1080/00295450.2017.1419780What is neutron metrology and why is it needed
, Metrologia 48, S225-S238 (2011). doi: 10.1088/0026-1394/48/6/S01Gamma ray multiplicity of a 240Pu solid sphere simulated by JMCT
. Nucl. Sci. Tech. 33(5), 53 (2022) doi: 10.1007/s41365-022-01043-9Prediction and measurement of neutron energy spectrum in a material test research reactor
. Ann. Nucl. Energy 27, 311-327 (2000). doi: 10.1016/S0306-4549(99)00057-2Conversion coefficients for use in radiological protection against external radiation
. ICRP Publication 74. Ann ICRP 26(3-4), 1996.Unfolding neutron spectra from water-pumping-injection multilayered concentric sphere neutron spectrometer using self-adaptive differential evolution algorithm
. Nucl. Sci. Tech. 32(3), 26 (2021). doi: 10.1007/s41365-021-00864-4Measurement with Bonner spheres spectrometer in pulsed neutron fields
. Radiat. Measurements 45, 1245-1249 (2010). doi: 10.1016/j.radmeas.2010.06.011Bonner sphere spectrometers—a critical review
. Nucl. Instrum. Meth. Phys. Res. Sect. A. 476, 12-20 (2002). doi: 10.1016/S0168-9002(01)01379-1Improved neutron spectrometer based on Bonner spheres
. Radiat. Protect. Dosim. 70, 285-289 (1997). doi: 10.1093/oxfordjournals.rpd.a031961Response matrix of an extended Bonner sphere system
. Nucl. Instrum. Meth. Phys. Res. Sect. A 476, 26-30 (2002). doi: 10.1016/S0168-9002(01)01383-3Calculations of the response functions of Bonner spheres with a spherical 3He proportional counter using a realistic detector model
. Health Phys. Radiat. Effects 26, 93,BONDI-97: A novel neutron energy spectrum unfolding tool using a genetic algorithm
. Nucl. Instrum. Meth. Phys. Res. A 432, 305-312 (1999). doi: 10.1016/S0168-9002(99)00535-5Geant4—a simulation toolkit
. Nucl. Instrum. Meth. Phys. Res. Sect. A 506, 250-303 (2003). doi: 10.1016/S0168-9002(03)01368-8Reference Manual for ENDF Thermal Neutron Scattering Data, General Atomics Report GA-8744, Revised (ENDF-269)
, 1978. doi: 10.2172/4075168New PTB thermal neutron calibration facility: first results
. Radiat. Protect. Dosim. 161, 352-356 (2014). doi: 10.1093/rpd/nct354Experimental evaluation of the Geant4-calculated response functions of a Bonner sphere spectrometer on monoenergetic neutron sources
. Nucl. Instrum. Meth. Phys. Res. Sect. A 965,163836 (2020). doi: 10.1016/j.nima.2020.163836Production of high energy neutron beam from deuteron breakup
. Nucl. Sci. Tech. 33(7), 92 (2022) doi: 10.1007/s41365-022-01075-1Reference thermal neutron field at KRISS for calibration of neutron detectors
. Radiat. Measurements 107, 73-79 (2017). doi: 10.1016/j.radmeas.2017.10.001Neutron spectrometry around a high-energy electron-positron collider using a multi-sphere system with passive detectors
. Radiat. Protect. Dosimetry 126, 541-545 (2007). doi: 10.1093/rpd/ncm109Measurement of spectrum and dose rate of natural neutron using bonner spheres
. Nucl. Sci. Eng. 33, 51-56 (2013). (in Chinese).Measurement of neutron energy spectra and ambient dose equivalent rates at workplaces of a nuclear power plant
. J. Astron. Metrol. Measure. 34, 6 (2014).doi: 10.12060/j.issn.1000-7202.2014.03.10Neutron dosemeter responses in workplace fields and the implications of using realistic neutron calibration fields. NPL Report. CIRM 27
, 1999.