Development of X-ray dark-field imaging towards clinical application

SYNCHROTRON TECHNOLOGY AND APPLICATIONS

Development of X-ray dark-field imaging towards clinical application

ANDO Masami
HASHIMOTO Eiko
HASHIZUME Hiroyuki
HYODO Kazuyuki
INOUE Hajime
ISHIKAWA Tetsuya
KUNISADA Toshiyuki
MAKSIMENKO Anton
PATTANASIRIWISAWA Wanwisa
RUBENSTEIN Edward
ROBERSON Joseph
SHIMAO Daisuke
SUGIYAMA Hiroshi
TAKEDA Ken
UENO Ei
WADA Hiroshi
Nuclear Science and TechniquesVol.15, No.3pp.129-139Published in print 01 Jun 2004
23100

Review of X-ray dark-field imaging under development is presented. Its goal is its application to clinical diagnosis of organs that have been invisible by the ordinary techniques. In order to clinically visualize tissues in detail one needs high contrast and high spatial resolution say ~50 μmm. This X-ray optics comprises a Bragg asymmetric monochro-collimator and a Bragg case or a Laue case filter with capability of analyzing angle in a parallel position. Their diffraction index is 4,4,0 and the X-ray energy 35 keV (λ= 0.0354 nm). The filter has 0.6 mm thickness in the Bragg case or 1.075 mm or 2.15 mm thickness in the Laue case. Under this condition only the refracted X-rays from object can transmit through the filter while the beam that may receive absorption and/or phase change will not. Soft tissues at human joints thus taken show high contrast images so that the DFI is promising for clinical diagnosis. Preliminary X-ray absorption images of another clinical candidates of ear bones are also shown.

X-ray dark-field imagingX-ray bright-field imagingBragg asymmetric monochro-collimatorBragg case filterLaue case filterClinical diagnosis
References
[1] Ando M, Hosoya S. Proc of the 6th Internationa1 Conference on X-Ray Optics and Microanalysis 63-68, Shinoda G, Kohra K, Ichinokawa T ed, Tokyo: University of Tokyo Press, 1972
[2] Momose A, Takeda T, Itai Y et al. Nat Med, 1995, 2: 473-475
[3] Bonse U, Hart M. Appl Phys Lett, 1965, 7: 99-101
[4] Podurets KM, Somenkov VA, Shil’shtein SSh. Physica B, 1989, 156 & 157: 691-693
[5] Ingal VN, Beliaevskaya EA. J Phys D: Appl Phys, 1995, 28: 2314-2317
[6] Davis TJ, Gao D, Gureyev TE et al. Nature, 1995, 373: 595-598
[7] Chapman D, Thomlinson W, Johnston RE et al. Phys Med Biol, 1997, 42: 2015-2025
[8] Ando M, Sugiyama H, Maksimenko A et al. Jpn J Appl Phys, 2002, 41: L1016-L1018
[9] Kohra K. J Phys Soc Jpn, 1962, 17: 589-590
[10] Chikawa J, Asada Y, Fujimoto I. J Appl Phys, 1970, 41: 1922-1925
[11] Dilmanian FA, Zhong Z, Rent B et al. Phys Med Biol, 2000, 45: 933-946
[12] Snigirev A, Snigireva I, Kohn V et al. Rev Sci Instrum 1995, 66(12): 5486-5492
[13] Cloetens P, Barrett R, Baruchel J et al. J Phys D: Appl Phys, 1996, 29: 133-146
[14] Margaritondo G, Tromba G. J Appl Phys, 1999, 85(7): 3406-3408
[15] Yagi N, Suzuki Y, Umetani K et al. Med Phys, 1999, 26: 2190-2193
[16] Ando M, Sugiyama H, Xiaowei Z et al. Jpn J Appl Phys, 2001, 40(3B): L298-L301
[17] Maksimenko A, Sugiyama H, Pattanasiriwisawa W et al. Jpn J Appl Phys 2003, 42(9AB): L1096-L1099
[18] Ishikawa T, Kohra K. Perfect crystal methods, Handbook on synchrotron radiation, Brown GS, Moncton DE, ed. Amsterdam: North-Holland, 1991, 63-104
[19] Rubenstein E, Hofstadter R, Zeman HD et al. Proc Natl Acad Sci USA, 1987, 83: 9724
[20] Thomlinson W. Proceedings of the International School of Physics “Enrico Fermi,” Course CXXVIII, Biomedical applications of synchrotron radiation, Burattini E, Balerna A, eds. IOS Press, 1996, 127-153
[21] Dix W-R, Graeff W, Heuer J et al. Rev Sci Instrum, 1989, 60: 2260
[22] Dementiev EN, Dolbnya IP, Zagorodnikov EI et al. Rev Sci Instrum, 1989, 60: 2264
[23] Elleaume H, Fiedler S, Estève F et al. Phys Med Biol, 2000, 45: L39-L43
[24] Dix W-R, Kupper W, Dill T et al. J Synch Rad, 2003, 10: 219-227
[25] Akisada M, Ando M, Hyodo K et al. Nucl Instr Meth, 1986, A264(1): 713-718
[26] Hyodo K, Nishimura K, Ando M. Handbook on synchrotron radiation vol 4, Ebashi S, Koch M, Rubenstein E, ed. Amsterdam: Elsevier Science Publishers, 1991, 55
[27] Ohtsuka S, Sugishita Y, Takeda T et al. Jpn Circ J, 1997, 61(5): 432
[28] Burattini E, Cossu E, Di Maggio C et al. Radiology, 1995, 159: 239-244
[29] Johnston RE, Washburn D, Posano E et al. Radiology, 1996, 200: 659-663
[30] Kim HJ, Hong JO, Lee KH et al. IEEE Trans on Nucl Sci, 2001, 48(3): 837-842
[31] Shimao D, Mori, K, Sugiyama H et al. Jpn J Appl Phys, 2003, 42: 5874-5875
[32] Li J, Zhong Z, Litdke R et al.

Radiography of soft tissue of the foot and ankle with diffraction enhanced imaging

, J Am Podiatric Med Assoc, in press
Baidu ScholarGoogle Scholar
[33] Gureyev TE, Stevenson AW, Paganin D et al. J of Digital Imaging, 2000, 13(2): 121-126
[34] Shimao D, Mori K, Hyodo K et al.

Application of X-ray refraction-contrast to medical joint imaging

, Proceedings of the 8th International Conference on Synchrotron Radiation Instrumentation at San Francisco, 2003, in press
Baidu ScholarGoogle Scholar
[35] Shimao D et al. a draft in preparation
[36] Sugiyama H, Kunisada T, Shimao D et al. Attempt at field size expansion in X-ray dark-field imaging with 35 keV for clinical application - Visualization of human excised articular cartilage of femoral head, submitted to JSR
[37] Wada H, Koike T, Kobayashi T et al. Ear and Hearing, 1998, 19(3): 240-249
[38] Kunisada T et al. a draft in preparation