Sensitivity of a new-developed neutron detector

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

Sensitivity of a new-developed neutron detector

PENG Tai-Ping
ZHU Xue-Bin
YANG Hong-Qiong
YANG Jian-Lun
YANG Gao-Zhao
LI Lin-Bo
SONG Xian-Cai
Nuclear Science and TechniquesVol.16, No.1pp.40-42Published in print 01 Feb 2005
22601

We develop a kind of neutron detector, which consists of a polyethylene thin film and two PIN semiconductors connected face-to-face. The detector is insensitive to γ-rays. Its sensitivity to neutron has been calculated with MCNP program and calibrated by experiments, and the results indicate that the neutron sensitivity of the compensation detector will vary with polyethylene converter. The compensation PIN detector can be employed to measure pulse neutron in neutron and gamma mixture radiation field.