logo

An online monitor ionization chamber used in particle therapy

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

An online monitor ionization chamber used in particle therapy

TANG Bin
HU Zhengguo
MAO Ruishi
XU Zhiguo
WANG Jiansong
YUE Ke
TU Xiaolin
WU Dapeng
CHEN Jinda
ZHANG Jie
WANG Meng
SUN Zhiyu
ZHANG Xueheng
LI Qiang
XU Hushan
XIOA Guoqing
Nuclear Science and TechniquesVol.20, No.3pp.177-180Published in print 20 Jun 2009
35500

The clinical trials of tumor therapy using heavy ions beam 12C are now in progress at Institute of Modern Physics in Lanzhou. In order to achieve the precise radiotherapy with the high energy 12C beam in active pencil beam scanning mode, we have developed an ionization chamber(IC) as an online monitor for beam intensity and also a dosimeter after calibration. Through the choosing of working gas and voltage, optimizing of the electrics and the read-out system, calibrating the linearity, the detector system provide us one of the simple and highly reliable way to monitoring the beam during the active pencil beam scanning treatments. The measurement results of this detector system show that it could work well under the condition of high energy 12C beam in active pencil beam scanning mode.

Heavy ion beam therapyIntegral ionization chamberIrradiation doseAbsorb dose
References
[1] Tobias C A, Anger H O, Lawrence J H. Am J Roentgenol Ther Nucl Med, 1952, 67(1):1-27.
[2]

Particle therapy facilities in operation (March 2009)

, available at http://ptcog.web.psi.ch/ptcentres.html
Baidu ScholarGoogle Scholar
[3] Amaldi U. Phys Med, 1998, 14: 76-84.
[4] Eickhoff H, Böhne D, Debus J, et al.

The proposed accelerator facility for light ion cancer therapy in Heidelberg Proc

. 1999 Particle Accelerator Conf. New York, 1999: 2513-2515.
Baidu ScholarGoogle Scholar
[5] Carstro J R. Radiate Environ Biophys, 1995, 34: 45-48.
[6] TANG Bin, HU Zhengguo, XU Zhiguo, et al. Nucl Electron Detect Technol (in Chinese), 2008, 28(4): 699-701.
[7] MAO Ruishi, XIAO Guoqing, ZHAO Tiecheng, et al. High power laser and particle beams (in Chinese), 2008, 20(9): 1537-1540.
[8] Lyman J T, Awschalom M, Berardo P, et al.

Protocol for heavy charged-particle therapy beam dosimeter. AAPM Report No. 16

, 1986: 13-15.
Baidu ScholarGoogle Scholar
[9] Brusasco C, Voss B, Schardt D M, et al. Nucl Instrum Methods Phys Res B, 2000, 168: 578-592.
[10] IAEA,

Absorbed dose determination in external veam radiotherapy, IAEA Technical reports Serial No. 398

, International Atomic Energy Agency, Vienna, 2000: 57-63.
Baidu ScholarGoogle Scholar