Evaluation of dosimetry effect of lack of side-scatter volume on large field measurements with a MapCheck detector array

RADIOCHEMISTRY, RADIOPHARMACEUTICALS AND NUCLEAR MEDICINE

Evaluation of dosimetry effect of lack of side-scatter volume on large field measurements with a MapCheck detector array

MA Pan
DAI Jianrong
CUI Weijie
Nuclear Science and TechniquesVol.20, No.5pp.292-296Published in print 20 Oct 2009
32800

In this paper, we describe a quantitative evaluation of the dosimetry effect of lack of side-scatter volume (LSSV). MapCheck, an integration diode array, was used to measure dose maps for three large non-IMRT fields (30cm ×30 cm and 40 cm×40 cm open fields, and 30 cm×30 cm wedge field with wedge angle of 60°) and 19 large IMRT fields. For each field, measurement was performed twice, under the conditions of (1) full scatter volume and (2) LSSV. Condition 1 was satisfied by adding PMMA slabs against the side of the MapCheck, and Condition 2, without PMMA slabs. The measured dose maps were compared with pass rate and their difference was scored when the acceptance criterion was set to 0.5%, 1%, 2%, 3%, etc. For very large open fields, the effect of LSSV may be clinically significant, while for large wedge fields and IMRT fields, the effect is negligible.

Large fieldLack of side-scatter volumeDiode array
References
[1] LUO Wei, LI Jinsheng, Price Robert A, et al. Med Phys, 2006, 33: 2557-2564.
[2] YAN Yulong, Papanikolaou Nikos, WENG Xuejun, et al. Med Phys, 2005, 32: 1566-1570.
[3] Chen X, Yue N J, Chen W, et al. Phys Med Biol, 2005, 50: 2641-2645.
[4] Winkler P, Zurl B, Guss H, et al. Phys Med Biol, 2005, 50: 643-654.
[5] XING Lei, Li J G. Med Phys, 2000, 27(9): 1566-1570.
[6] Higgins P D, Alaei P, Gerbi B J, et al. Med Phys, 2003, 30: 3118-3123.
[7] Charland M P, Chetty I J, Yokoyama S, et al. J Appl Clin Med Phys, 2003, 4: 25-39.
[8] Arthur J. Olch Med Phys, 2002, 29: 2159-2168.
[9] YANG Y, XING L. Phys Med Biol, 2004, 49: 1521-1533.
[10] Buonamici F B, Compagnucci A, Marrazzo L, et al. Med Phys, 2007, 34: 1372-1379.
[11] Letourneau D, Gulam M, YAN D, et al. Radiother Oncol, 2004, 70: 199-206.
[12] Jursinic P A, Nelms B E. Med Phys, 2003, 30: 870-879.
[13] IAEA.

Technical Reports Series No. 277

, Vienna, 1987.
Baidu ScholarGoogle Scholar
[14] Sun Nuclear Corporation. MapCheck user guide Melbourne (FL), 2004.
[15] Nelms B E, Simon J A. J Appl Clin Med Phys, 2007, 8: 2448.