1 Introduction
We are constructing a small-angle neutron scattering spectrometer (SANS) for material research at the China Spallation Neutron Source (CSNS) [1]. The SANS requires three auxiliary beam monitors in addition to a main detector. Two gas electron multiplier detectors are employed as the beam monitors in front of the scattering sample for their high counting rate (∼1 MHz) and excellent spatial resolution [2]. The third beam monitor, behind the sample, is used to suppress the experimental uncertainty due to fluctuation of the neutron beam by counting the neutrons passing through the sample. A 3He counter could be a good option for this function. Considering the international supply of 3He, however, we turn to alternatives to the 3He counter. In recent years, many advanced micropattern gas detectors have been developed for use in neutron beam monitors [3-6]. These detectors can be potential options for use as this beam monitor. However, as an intervening detector, it needs to be repeatedly pulled away from the beam line during neutron scattering measurement and plugged in for calibration of transmission neutron counting. Furthermore, limited room is reserved for the monitor in the vacuum of the scattering chamber. It will even be impossible to maintain a gas detector under these conditions. Hence, a design in which a piece of lithium glass equipped with a low-power photomultiplier tube (PMT) is chosen.
Cerium-activated lithium glass has been widely used in nuclear engineering and research as a conventional neutron detector [7-9]. However, we still need to address numerous problems with this particular beam monitoring application. As a neutron beam monitor, it should have a high neutron detection efficiency. However, the beam spot at the SANS is approximately 2.4 cm × 2.4 cm, with a maximum fluence rate of 107n·cm-2·s-1. The beam intensity is so high that we should balance the detection efficiency and the dead time to guarantee stable operation of the monitor. The uncertainty (statistics) of the measured intensity is expected to be better than 1%. Owing to the high beam intensity, this value will be easy to meet. At the same time, low gamma sensitivity is also required. The disadvantage of using lithium glass as a neutron beam detector is its excessive gamma sensitivity compared with the 3He counter. In current applications, the sensitivity is expected to be below 10-4 for gamma rays around 1 MeV. A type of lithium-glass scintillator provided by China Building Materials Academy (CBMA) was studied using the 252Cf neutron source and 60Co gamma source at the Institute of High Energy Physics (IHEP). The charge spectra and neutron light yield were obtained. A neutron beam monitor based on this type of glass scintillator was designed using the Monte Carlo toolkit Geant4 [10].
2 Approaches
The neutron beam monitor is sealed in the vacuum of the scattering chamber, which contains the neutron scattering spectrometer (Fig. 1). This monitor is expected to acquire the beam intensity with high accuracy and appropriate efficiency. It is conventionally designed as an intervening detector. There should be a mechanical module to insert it into and extract it from the beam line. Considering that the dimensions of the monitor must not exceed ϕ80 mm × 150 mm and the monitor should be suitable for moving up and down, we devoted considerable effort to selecting the assembly of the scintillator and PMT. Ultimately, a side-on PMT (e.g., Hamamatsu type R4332) and a low-power high-voltage divider were selected for photoelectric conversion. The lithium-glass scintillator is optically coupled to the cathode window of the PMT by a synthetic quartz light guide. They are placed in a gas-sealed aluminum tube, which is responsible to insulate the outside vacuum and provides a darkroom for the PMT.
-201809/1001-8042-29-09-006/alternativeImage/1001-8042-29-09-006-F001.jpg)
2.1 Experiment
To investigate the performance of the selected lithium-glass scintillator, an experimental study was carried out. The cylindrical lithium-glass sample has dimensions of ϕ 40 mm × 4 mm. It has a total lithium content of 7.7 wt.% with 6Li enrichment of 93%. In the experiment, the glass was wrapped in aluminum foil to improve the light collection efficiency. The unwrapped end was mounted on the front window of a PMT (Philips XP2020) by silicone grease. A bias voltage of -1800 V was applied to the PMT. An oscilloscope (Agilent DSO-X3024A) with an interface developed using LabVIEW was used to perform data acquisition.
The single-photoelectron response of the XP2020 PMT was calibrated in advance. A light-emitting diode with a light wavelength of 395–400 nm, which was chosen according to the maximum scintillation light wavelength of 395 nm, was driven by a pulse generator as the calibration light source. We obtained a single-photoelectron response of 6.2× 102 fC/photoelectron.
The 252Cf neutron source at the IHEP, which has a neutron yield of approximately 1× 106 n/s, was used to study the neutron response of the lithium glass. A diagram of the experimental layout is shown in Fig. 2. The source itself is surrounded by lead sphere shielding to suppress gamma rays. The neutron moderator layer consists of 35-cm-thick paraffin. A layer of borax with a thickness of 11 cm is placed as the outermost layer for dumping of the moderated neutrons. A neutron beam collimator tunnel with a diameter of 10 cm is prepared starting from the lead sphere and continuing through the neutron moderating and dumping materials. The lithium glass coupled with the PMT is positioned next to the collimator tunnel. Because most neutrons coming through the tunnel are fast neutrons, a cylindrical moderator with a thickness of 10 cm is placed in the tunnel to thermalize the neutrons in this experiment. To study the n-gamma discrimination capability of this glass, a 60Co gamma ray source experiment was also carried out using the same experimental setup.
-201809/1001-8042-29-09-006/alternativeImage/1001-8042-29-09-006-F002.jpg)
2.2 Monte Carlo simulation
As a traditional choice of neutron beam monitor, the lithium-glass detector has some advantages over the gas detector. i) It can be processed into a needed shape easily. ii) The volume is smaller for the same detection efficiency. iii) No gas-maintaining accessories are required. At the same time, a well-known shortcoming is its excessive sensitivity to gamma rays. Along the neutron beam line of the SANS, the intensity of the associated gamma rays is high. Some of them are direct side products of the neutrons on the spallation target; the others come from neutron reactions with the materials around the spectrometer. To significantly reduce the neutron misidentification due to gamma rays and to establish a proper design, a simulation was performed using the Geant4 Monte Carlo simulation toolkit (version 10.1). The lithium glass that we use is similar to the commercial product GS20; hence, it was modeled with a composition of 18% Li2O (6Li enriched to 93%), 4% MgO, 18% Al2O3, 57% SiO2, and 3% Ce2O3 [11]. All the physical processes related to detection were added within the framework provided by Geant4. The high-precision neutron nuclear reactions model driven by cross section data was implemented to handle the neutron transport. The PENELOPE code embedded in Geant4 was used to simulate the electromagnetic progresses. The scintillation and light collection were modeled using the optical photon package of Geant4 [12]. The attenuation length of the scintillation light is ∼50 cm [13], which is much larger than the glass thickness. Consequently, the attenuation effect is not significant. The reflectivity of the aluminum foil wrapping was set to 70%. The other scintillating and optical parameters come from reference [14].
3 Results and discussion
3.1 Charge spectra
Typical neutron and gamma-ray signals are displayed in Fig. 3(a). The corresponding charge spectra of the lithium glass for 60 Co and 252Cf are shown in Fig. 3(b). In the spectrum of 252Cf, the left "peak" is attributed to the background on the basis of the experimental result obtained without a source. The right peak is the neutron peak. The light yield of a neutron, Y, can be obtained by the following equations:
-201809/1001-8042-29-09-006/alternativeImage/1001-8042-29-09-006-F003.jpg)
where Qn is the average charge corresponding to a neutron, and Qs is the charge response of a single photoelectron. The charge response of a neutron, Qn, is 5.3×102 pC, which is derived from the peak position of the charge spectrum. The scintillation light collection efficiency, εc 80%, is acquired by a light transition simulation with the Geant4 toolkit, which has been verified in our previous work [14].
where I(λ) is the normalized scintillation light spectrum, and εq(λ) is the quantum efficiency of the PMT for a wavelength λ. We did not measure the scintillation spectrum of this glass. The average quantum efficiency
A thermal neutron is absorbed in lithium glass by the following reaction:
The charged ions (the triton and alpha) are emitted with initial kinetic energies of 2.73 and 2.05 MeV, respectively. Their ranges in lithium glass are 40 and 7 μm, respectively, which are far smaller than the thickness of the glass sample. Therefore, the energy deposited by the neutron-produced ions does not generate additional smearing of the neutron spectral peak. The peak width indicates an energy resolution of 16%. In the spectrum of 60Co, the left "peak" comes from the background cut by the threshold. The Compton edges of the two energy lines at 1.173 and 1.332 MeV merge and form a peak-like structure (the right "peak" in the spectrum at ∼300 pC). Lithium glass is composed of light elements, and the tested sample is thin; therefore, the peak-to-Compton ratio is low. As a result, the photoelectric peaks are merged by the Compton plateau due to the energy resolution. The 60Co spectrum from the Monte Carlo simulation agrees well with that shown in Fig. 3(b), where the Compton-edge "peak" position is ∼0.9 MeV. The neutron peak at ∼530 pC is distinguished from the gamma spectrum. However, many events overlap each other owing to the gamma sensitivity of the lithium glass. The neutron peak corresponds to an electron-equivalent energy of 1.6 MeV [16]. The positions of the neutron peak and the Compton-edge "peak" together verify the linearity of the PMT gain.
3.2 Gamma sensitivity
Decreasing the thickness of the glass plate is an effective way to reduce the gamma sensitivity of the neutron monitor. The electron-equivalent energy of a thermal neutron Ene 1.6 MeVee [16]. An electron with an energy of 1.6 MeV has a range of ∼2.9 mm in lithium glass. It has a high probability of leaving a glass plate with a thickness smaller than its range before exhausting its kinetic energy. Therefore, the energy deposited by a gamma ray is limited by the thickness of the lithium glass regardless of the energy of a gamma photon. The gamma sensitivity ξ is introduced to evaluate the n-gamma discrimination ability. It represents the probability that a gamma signal with a specific energy is mistaken for a neutron signal. Its value can be derived using the formula [17]
where Cγ is the number of gamma rays mistaken for neutron signals, and Ct is the total number of detected gamma rays. According to the experimental result, the energy resolution of the thermal neutron peak given by this type of lithium glass is 16%. A threshold for eliminating gamma events with Ene- FWHM=1.344 MeV is implemented. Thus, gamma rays depositing energy above the threshold are expected to be mistaken for neutrons.
To validate the code, thermal neutrons and gamma rays with an energy of 1.332 MeV from 60Co are generated to strike a glass scintillator sample 4 mm in thickness in the simulation. The energy spectra are shown in Fig. 4(a). They are consistent with the experimental result. When the glass thickness decreases to 2.0 mm, there is no apparent change in the neutron spectrum. However, the neutron-gamma separation becomes larger. Under this condition, even higher energies (e.g., 5 and 10 MeV) lead to a further decrease of the overlap between the neutron spectrum and gamma spectrum. The systematic results for the gamma sensitivity are given in Fig. 4(b). When the thickness is 0.5 mm for a gamma energy of 1.5 MeV, the sensitivity of the lithium glass reaches a magnitude of 10-6. This value is close to that of a well-behaved 3He detector with an appropriate threshold [17].
-201809/1001-8042-29-09-006/alternativeImage/1001-8042-29-09-006-F004.jpg)
3.3 Lithium glass selection
The width of the pulse output from the lithium glass is 200–300 ns, as shown in Fig. 3(a). Accordingly, the counting capability of the monitor based on the lithium glass is approximately 3.3×106s-1 if we ignore the dead time of the electronics. Therefore, the detection efficiency of the monitor should be kept below 5.7% considering the beam intensity. The simulated neutron absorption efficiency of scintillators with different thicknesses is shown in Fig. 5. The neutron spectrum at the SANS is supposed to peak at the thermal energy (1.8 angstrom). Thus, we consider a detector design based mainly on thermal neutrons. The thermal neutron absorption efficiency is approximately 59% for 0.5-mm-thick lithium glass. To reduce the scintillator thickness is an effective and direct method to decrease the efficiency. The minimum thickness that the vendor of this type of lithium glass can supply is 0.1 mm. At the same time, thinner glass also has a lower gamma sensitivity. However, the monitor is supposed to work under a vacuum environment and may be moved frequently into and out of the beam line. When the thickness is less than 0.5 mm, the glass tends to be deformed or even break under the strain caused by the difference in pressure. If defects are created during operation, the measurement accuracy will be affected. Thus, the use of low-enriched or natural (6Li) lithium glass is an appropriate option that yields acceptable gamma sensitivity without mechanical problems. The natural enrichment of 6Li is 7.4%, and that of the glass used in the experiment is 93%. Accordingly, a 0.5-mm-thick lithium glass with natural lithium has a neutron detection efficiency of approximately 4.7%, which satisfies both the gamma sensitivity and efficiency requirements.
-201809/1001-8042-29-09-006/alternativeImage/1001-8042-29-09-006-F005.jpg)
4 Summary
The third neutron beam monitor of the SANS at the CSNS must have small dimensions, convenient mobility, and easy maintenance. A monitor design based on a type of lithium glass supplied by CBMA and equipped with a side-window PMT was proposed. The performance of the selected lithium glass scintillator was studied using neutron and gamma sources. A subtle simulation of this monitor based on the Geant4 toolkit was performed. The gamma sensitivity can be suppressed close to the level of a 3He-based detector by decreasing the scintillator thickness. The detection efficiency is sufficiently high even for a piece of 0.5-mm-thick lithium glass. Considering the counting rate capability, we find that a glass scintillator of natural lithium 0.5 mm in thickness is an appropriate selection.
China Spallation Neutron Source: Design, R&D, and outlook
. Nucl. Instrum. Method A. 600, 10-13 (2009). doi: 10.1016/j.nima.2008.11.017Neutron beam monitor based on a boron-coated GEM
. Chin. Phys. C. 35, 668-674(2011). doi: 10.1088/1674-1137/35/7/012GEM-based thermal neutron beam monitors for spallation sources
. Nucl. Instrum. Method A. 732, 217-220(2013). doi: 10.1016/j.nima.2013.05.111Triple GEM gas detectors as real time fast neutron beam monitors for spallation neutron sources
. J. Instrum. 7, P07021 (2012). doi: 10.1088/1748-0221/7/07/P07021A single-crystal diamond-based thermal neutron beam monitor for instruments at pulsed neutron sources
. Nucl. Instrum. Method A. 610, 677-681(2009). doi: 10.1016/j.nima.2009.09.013Miller, Monte Carlo studies of a Micromegas neutron beam monitor
. Nucl. Instrum. Method A. 521, 441-453(2004). doi: 10.1016/j.nima.2003.09.064Pr 3+ -doped fluoro-oxide lithium glass as scintillator for nuclear fusion diagnostics
. Rev. Sci. Instrum. 80, 113504(2009). doi: 10.1063/1.3254444Fast-Response and Low-Afterglow Cerium-Doped Lithium 6 Fluoro-Oxide Glass Scintillator for Laser Fusion-Originated Down-Scattered Neutron Detection
. IEEE Trans. Nucl. Sci. 59, 2256-2259(2012). doi: 10.1109/TNS.2012.2212458Low-energy neutron flux measurement using a resonance absorption filter surrounding a lithium glass scintillator
. Radiat. Phys. Chem. 76, 917-920(2007). doi: 10.1016/j.radphyschem.2007.01.003Recent developments in Geant4
. Nucl. Instrum. Method A. 835, 186-225(2016). doi: 10.1016/j.nima.2016.06.125Phosphors and scintillators in radiation imaging detectors
. Nucl. Instrum. Method A. 546, 180-187(2005). doi: 10.1016/j.nima.2005.03.103Recent developments in Geant4
. Nucl. Instrum. Method A. 835, 186-225(2016). doi: 10.1016/j.nima.2016.06.125A photon transport model code for use in scintillation detectors
. Nucl. Instrum. Method A. 516, 116-121(2004). doi: 10.1016/j.nima.2003.07.058Monte Carlo simulation of a very high resolution thermal neutron detector composed of glass scintillator microfibers
. Appl. Radiat. Isotopes. 108, 100-107(2016). doi: 10.1016/j.apradiso.2015.12.035Inorganic thermal-neutron scintillators
. Nucl. Instrum. Method A. 529, 260-267(2004). doi: 10.1016/j.nima.2004.04.163An improved Li6-loaded glass scintillator for neutron detection
. Nucl. Instrum. Method. 13, 313-316(1961). doi: 10.1016/0029-554X(61)90221-XNeutron detection gamma ray sensitivity criteria
. Nucl. Instrum. Method A. 654, 412?416(2011). doi: 10.1016/j.nima.2011.07.030