COMPARISON OF ABSORBED DOSE MEASUREMENTS WITH AN EXTRAPOLATION CHAMBER AND A FARMER CHAMBER IN HIGH ENERGY ELECTRON BEAM

COMPARISON OF ABSORBED DOSE MEASUREMENTS WITH AN EXTRAPOLATION CHAMBER AND A FARMER CHAMBER IN HIGH ENERGY ELECTRON BEAM

S.C. Klevenhagen
Jin Tao
Wang Mi
Wu Jinzheng
Sun Qiangsheng
Nuclear Science and TechniquesVol.1, No.4pp.249-253Published in print 01 Nov 1990
38200

In this paper, the comparison of absorbed dose measurement performed with an extrapolation chamber and a Farmer chamber in high energy electron beam is described. The experiments performed in the course of this work show that the Xinjiang extrapolation chamber is suitable for calibration of high energy electron beams produced from medical linear accelerators.

Electron beamCalibrationExtrapolation chamberFarmer chamber
REFERENCES
[1] Jin Tao, Journal of Radiation Research and Radiation Processing, 5 (1987), 1: 54.
[2] G. Failla, Radiology, 29 (1937), 202.
[3] R. Loevinger, The Review of Scientific Instrument, 24 (1953), 907.
[4] J. Böhm and U. Schneider, Radiation Protection Dosimetry, 14 (1986), 193.
[5] J. Manson et al., Radiology, 115 (1975), 473.
[6] S.C. Klevenhagen. The Physics of electron beam therapy, Adam Higher, Bristol, ed. 1th, 1985.
[7] HPA (Hospital Physicist' Association, in London), Physics in Medical and Biology. 30 (1985), 1169.
[8] AAPM (American Association of Physicist in Medicine), Medical Physics, 10 (1983). 741.
[9] NACP (Nordic Association of Clinical Physicist), Acta Radiologica Oncology, 19(1981). 55.
[10] IAEA,

Absorbed dose determination in photon and electron beams, Technical reports series No. 277

, 1987.
Baidu ScholarGoogle Scholar
[11] L.O. Mattson, Radiotherapy Oncology, 4 (1985), 313.