logo

DEVELOPMENT AND APPLICATIONS OF AN ON-LINE THERMAL NEUTRON PROMPT-GAMMA ELEMENT ANALYSIS SYSTEM

DEVELOPMENT AND APPLICATIONS OF AN ON-LINE THERMAL NEUTRON PROMPT-GAMMA ELEMENT ANALYSIS SYSTEM

Liu Yuren
Lu Yunxin
Xie Yali
Wang Yonghai
Du Youling
Tan Jin
Meng Bonian
R.S. Seymour
Nuclear Science and TechniquesVol.2, No.2pp.71-78Published in print 01 May 1991
21800

This paper introduces the principles, instrumentation, implementation, and industrial applications of an on- line thermal neutron prompt- gamma element analysis system (using a 252Cf neutron source, Am- Be neutron source, or neutron generator). The energy resolution of the system at the H prompt- gamma full- energy photopeak (2.22325 MeV) is 3.6 keV. The concentration measurement error of Al2O3, Fe2O3, CaO and SiO2 is ±0.3%, ±0.1%, ±0.4% and ±0.4%, respectively. The system has been tested on- site at both the Shandong and the Zhengzhou Aluminum Works. Our preliminary on- site measurements confirm that the stability, reliability, measurement range, and accuracy of the system can meet the requirements of the aluminum production process. Facilitation of this measurement at aluminum plants is expected to reduce plant costs by over 3 million dollars annually through reduced energy consumption, more rapid qualification of pulps being mixed during the production process, and in reduced labor costs.

Prompt- gammaNeutron activation analysisOn- lineAluminum Pulp
REFERENCES
[1] D.L. Anderson et al., J. Radioanal. Chem., 63 (1981), 97.
[2] M.P. Failey et al., Anal. Chem., 51 (1979), 2209.
[3] A.G. Hanna et al., Nucl. Instr. Meth., 188 (1981), 619.
[4] C. Chung and T. Tseng, Nucl. Instr. Meth., A267 (1988), 223.
[5] M.S. Germani et al., Anal. Chem., 52 (1980), 240.
[6] J.H. McQuaid et al., IEEE Trans. Nucl. Sci., NS 28 (1981), 1: 304.
[7] E.R. Karlashev, All- Union Research Institute of Radiation Engineering, Moscow, USSR, 1987.
[8] Liu Yuren, Nucl. Tech. (Shanghai), 11 (1988), 12: 9.
[9] Liu Yuren, Papers of China fifth nuclear electronic and detective technique and third national high energy experimental physics proceedings, 1988, p.180- 190.