Nuclear Science and TechniquesVol.35, No.7Article number 121Published in print Jul 2024Available online 12 Jul 2024
70100

The hard X-ray nanoprobe beamline BL13U is a phase II beamline project at the Shanghai Synchrotron Radiation Facility. The beamline aims to enable comprehensive experiments at high spatial resolutions ranging from 50 to 10 nm. The X-ray energy range of the beamline, 5–25 keV, can detect most elements in the periodic table. Two operating modes were designed to accommodate the experimental requirements of high energy resolution or high photon flux, respectively. X-ray nanofluorescence, nanodiffraction, and coherent diffraction imaging are developed as the main experimental techniques for BL13U. This paper describes the beamline optics, end-station configurations, experimental methods under development, and preliminary test results. This comprehensive overview aims to provide a clear understanding of the beamline capabilities and potential applications.

Shanghai Synchrotron Radiation FacilityHard X-ray nanoprobeX-ray nanofocusing
1

Introduction:

Focused hard X-ray beams have emerged as powerful tools for studying the structures and functions of complex heterogeneous systems on a microscopic scale[1]. Recent advancements in synchrotron facilities and focusing-optic manufacturing have enabled the realization of nanofocused beams[2, 3]. Using reflective, refractive, and diffractive optics, beam sizes ranging from 50 to 7 nm have been successfully achieved in hard X-ray focusing[4-6]. Based on hard X-ray nanoprobes, a variety of techniques have found important applications in multiple fields, including biology, chemistry, physics, materials science, and nanotechnology[7-12]. Presently, synchrotron facilities are operating and developing hard X-ray nanoprobe beamlines worldwide[13-19].

The hard X-ray nanoprobe beamline (BL13U) is one of the phase-II project beamlines at the Shanghai Synchrotron Radiation Facility[20-23] (SSRF). It aims to satisfy comprehensive experimental research requirements with high spatial resolution and high sensitivity detection. The scientific case for BL13U centers on spectroscopy, fluorescence, and diffraction imaging across a broad range of areas, including life, materials, and environmental sciences. The techniques available at this beamline provide spatial mapping of the elemental composition by X-ray fluorescence (XRF), speciation by X-ray absorption near-edge spectroscopy (XANES), the structural phase from nano-X-ray diffraction (nano-XRD), and phase contrast or electron density from coherent diffraction imaging (CDI) and ptychography. To meet these demands, the design of the beamline has been optimized for flexibility in energy range and focusing mode, requiring different operating modes that allow for a wide range of available X-ray energies, beam sizes, and fluxes. Two monochromators were implemented in the beamline and can be switched to provide high energy resolution or high photon flux for spectroscopy and imaging techniques, respectively. At the end station, a multilayer Kirkpatrick–Baez (KB) system, combined with a liquid nitrogen-cooled double multilayer monochromator (DMM), was adopted to achieve the designed focal spot of 10 nm at 10 keV. A total reflection KB system was installed after the liquid nitrogen-cooled double crystal monochromator (DCM) to focus the beam to 50 nm in the 5–25 keV energy range.

This paper presents the design and initial commissioning results of BL13U in terms of the beamline, experimental end-station, experimental methods, based on the multilayer KB system.

2

Beamline

2.1
Light source

The SSRF comprises a 150 MeV electron linear accelerator, a booster that elevates the electron energy to 3.5 GeV, and a 3.5 GeV electron storage ring. To obtain continuously adjustable energy spectra, BL13U employs a U20 in-vacuum undulator as its X-ray source. This undulator has a total length of 4 m and can generate highly brilliant 1st–11th harmonics covering the entire designed energy range of 5–25 keV (minimum magnetic gap of 6 mm). The tuning curve for U20 was calculated using XOP software, as shown in Fig. 1. Calculations show that the brilliance is higher than 3.7×1019 phs/mrad2/mm2/s/ (0.1% bandwidth) at 10 keV. The specifications of the undulator are listed in Table 1.

Fig. 1
Tuning curve for U20 in-vacuum undulator
pic
Table 1
Main parameters of the storge ring and insertion device
Parameters Values
Storage ring energy (GeV) 3.5
Nominal design current (mA) 300
Current (operation 2023) (mA) 200
Electron beam emittance (x) (nm rad) 4.22
Electron source size (σx × σy) (μm2) 173.1 ×10
Electron source divergence (σx × σy) (μrad2) 34.07 × 4.08
Insertion device In-vacuum undulator
Minimum magnetic gap (mm) 6
Number of periods 200
Period length (mm) 20
Photon energy range (keV) 5-25
Magnetic field intensity (T) 0.84
Power(K=1.57) (kW/mrad2) 80
Show more
2.2
Beamline optics

The following guidelines were applied to the optical design of the hard X-ray nanoprobe beamline:

• The undulator source was focused onto the secondary source aperture by mirrors, and the beam from the second source aperture was focused onto the sample using nanofocusing KB optics.

• A DMM in conjunction with multilayer KB mirrors was used to achieve high flux at 10 nm focusing spot.

• A DCM with a total-reflection KB mirror was used to achieve higher energy resolution at 50 nm focusing spot.

A schematic overview of the beamline with the major optical components is shown in Fig. 2.

Fig. 2
(Color online) Schematic layout of the Beamline
pic

The optical layout of the beamline comprises three pairs of slits, two horizontal mirrors, and two monochromators. The photon source profile in the undulator is elliptical; thus, the main mirrors and the monochromator deflect the beam in the horizontal direction. This approach also reduces the vibrations in the vertical direction.

A water-cooled four-knife slit S1, located at 19.1 m from the source, is used to define the acceptance angle (70 µrad ×30 µrad) of the beamline and reduce the heat load on the horizontal collimating mirror (HCM, 22.5 m). The HCM (parameters listed in Table 2) is placed vertically to collimate incident X-rays in the horizontal direction. As the first optical element of the beamline, the HCM can reject high-order harmonics and reduce the thermal loading of high-energy X-rays on the monochromator crystals. To decrease the heat load on the monochromator, a four-knife slit S2 is placed at 24.7 m to limit the acceptance angle of the monochromator to 60 µrad ×20 µrad.

Table 2
Parameters of the reflective mirror at beamline optics
  Horizontal collimating mirror Pre-focusing mirror
Position (m) 22.5 31.4
Mirror shape Bendable curvature circular cylinder, horizontally deflecting Toroidal, vertically and horizontally deflecting
Incidence angle (mrad) 2.5 2.5
Mirror size (L × W) (mm2) 800 ×10 650 × 5
Cross-section size of beam (L × W) (mm2) 1.3 ×0.7 1.3 × 0.96
Slope error (μrad) 0.5 0.5
Micro roughness (RMS) (nm) 0.3 0.3
Coating Rh/Pt, 50 nm Rh, 50 nm
Substrate Si Si
Cooling system Water-cooling system /
Show more

The beamline is outfitted with a pair of monochromators, each tailored for a distinct operating mode. The DMM, located 26.1 m downstream of the source, is used for the high-flux fluorescence mode. Conversely, the DCM, stationed 28.6 m from the source, is employed for the high energy resolution mode. Depending on the experimental requirements, the two monochromators (parameters listed in Table 3) can be switched.

Table 3
Parameters of the monochromators
  Double multilayer monochromator Double-crystal monochromator
Position (m) 26.1 28.6
Coating Ru/C, Ni/C /
Energy range (keV) 8.3-25 (Ru/C), 5-8.3 (Ni/C) 5-25
Energy resolution (ΔE/E) 2×10-2 at 10 keV 2×10-4 at 10 keV
Bragg angle range (°) 0 – 2.5 4–24
Size of crystals (L × W × H) (mm3) 230 ×35 ×40 50 ×50 ×20 (first crystal)
  230 ×35 ×40 100 ×30 ×20 (second crystal)
Effective size of crystals (L × W) (mm2) 230 ×10 (two stripes) 50 × 50
    100 × 50
Substrate Si Si
Beam offset in horizontal direction (mm) 13.5 13.5
Show more

The multilayer monochromator can be coated with suitable multilayer pairs with different periodic thicknesses as required, which can significantly improve the flux to meet the requirements of the high-flux fluorescence mode. Two multilayers were adopted for different energy ranges: a Ru/C system with a periodic thickness of 3.1 nm for the 8.3–25 keV energy range and a Ni/C system with a periodic thickness of 3.5 nm for the 5–8.3 keV energy range, respectively. Calculations show that the reflectivities at 10 and 20 keV are higher than 75% and 85%, respectively. The multilayer stripes can be switched according to experimental requirements using a lateral motor inside the vacuum chamber. The energy resolution of the DMM is 2×10−2E/E) at 10 keV.

To meet the requirements of experiments such as XAFS, the energy resolution of the high-energy-resolution mode was designed to be better than 2×10−4E/E) at 10 keV. To achieve this, a Si(111) crystal DCM is used. The energy range of the DCM is 5–25 keV, with a minimum energy step of 0.05 eV. Both the DMM and DCM are cooled by a liquid-nitrogen coolant circulation unit. The structure of the multilayer changes under thermal loads and in low-temperature environments; we conducted relevant in-situ experiments on this in the early stage[24-27]. The DMM has a much smaller grazing incidence angle (0.4°–2°) than that of the DCM (4°–24°). To ensure that both systems exhibit the same deviation of light in the horizontal direction, a constant offset of 13.5 mm is set.

After the monochromators, the toroidal prefocusing mirror (PFM, 31.4 m) focuses the beam both horizontally and vertically with respect to the secondary source aperture (S3, at 53 m). The PFM (parameters listed in Table 2) is installed vertically with a Rh surface coating and a grazing incidence angle of 2.5 mrad. The incident beam is monochromatic and the heat is very low; therefore, no special cooling is required.

The size of the secondary source aperture is defined using vertical and horizontal slit blades. The beam transverse coherence is proportional to the distance between the sample and secondary source and inversely proportional to the size of the secondary source. Thus, for XRF experiments, larger secondary source apertures are used to obtain a high photon flux, whereas for CDI experiments, coherence requires a small aperture to achieve partial or full coherence illumination.

The stability of the photon beam is the key to achieving nanofocus. The mirrors and monochromators may vibrate with changes in the thermal load over time, leading to misalignment of the photon beam on the secondary source aperture. To stabilize the beam cone before S3, a dual-frequency feedback system is implemented that reads the current signals of the beam position monitor (BPM, in front of S3) to control both the second crystals of the DMM and DCM in their pitch and roll angles and the pitch angles of the PFM using a piezoelectric actuator.

2.3
End-station setup

The experimental hutch was divided into three functional areas: the extension area, 50 nm focusing area, and 10 nm focusing area, as shown in Fig. 2. The extension area was designed for CDI and optic-testing experiments.

The 50 nm focusing area was designed to focus X-rays to 50 nm. As shown in Fig. 3(a), a total-reflection KB mirror with a fixed mirror shape was chosen as the nanofocusing optic and mounted on a high-precision manipulator inside a vacuum chamber. According to the beamline schematic, the nominal grazing incidence angle of the KB mirror is set to 2.1 mrad. The nominal focal distances from the centers of the horizontal and vertical mirrors are 45 and 160 mm, respectively, resulting in a working distance of 20 mm. This working distance provides sufficient space for in-situ environments. A laser interferometric feedback system was integrated with each mirror to maintain the distance between the mirror and sample within a change of less than 5 nm.

Fig. 3
(Color online) Schematic diagram of the focusing set-up at BL13U. (a) The 50-nm focusing set-up. (b) The 10-nm focusing set-up
pic

A high-precision piezoelectrically actuated sample stage (piezo stage) was mounted downstream of the total-reflection KB mirror system in the vacuum chamber. As shown in Fig. 4 (a), the sample stage is a combination of three linear translation stages and one hexapod stage, and is mounted upside down in the chamber. The base stage provides a maximum travel range of 5 cm in the X, Y, and Z directions. The upper level is a hexapod stage which can provide rotational motion with a resolution of 1 mrad, partial pendulum motion with a travel range of approximately 20°, and a travel range of 10 mm in the Z direction. The 10-mm stage provides coarse positioning in the X and Y directions. The top nanocube stage provides a 1 nm positioning resolution for nanoscanning experiments. Capacitive sensors are used to correct the movements of the piezo stage. This allows 2D mapping and tomographic measurements to be performed on a wide variety of samples. Different types of sample holders were designed to load films, capillaries, and well-designed in-situ setups. Images of the KB system and sample stage in the vacuum chamber are shown in Fig. 5.

Fig. 4
(Color online) CAD drawing of the sample stages. (a) Sample stage used in the 50-nm focusing area. (b) Sample stage used in the 10-nm focusing area
pic
Fig. 5
(Color online) Images of the vacuum-chamber setups. (a) Total reflection KB system and sample stage in the 50-nm vacuum chamber. (b) Multilayer KB system and sample stage in the 10-nm vacuum chamber. (c) The robotic arm, which is the same in both vacuum chambers
pic

A robotic arm system (as shown in Fig. 5(c)) inside the chamber and a sample transfer system outside the chamber jointly perform automatic sample changes in the vacuum chamber. The robot can carry a load of more than 0.1 kg and achieve a positioning accuracy of 0.1 mm.

Several types of detectors are available to meet the requirements of different experiments (as shown in Fig. 6). A fluorescence detector (Vortex-ME3, Hitachi, USA) with a high-performance pulse processor (Xspress3, Quantum Detectors, UK) mounted on the fluorescence detector stage is available for nano-XRF experiments. A pixel detector (Pilatus 3 S 2M, Dectris, Switzerland), CMOS detector (ORCA-Lightning, Hamamatsu, Japan), and ion chambers (Digital Ion Chamber, OKEN, Japan) are mounted on the detector platform downstream of the vacuum chamber and can be selected for use by the motor on the platform. Bragg geometry experiments are conducted using a photon-counting pixel detector (Merlin 4X, Quantum Detectors, UK). An articulated robotic diffractometer is used to position the detector at the Bragg peak of interest. The diffractometer has seven degrees of freedom, allowing the spherical movement to reach the desired Bragg peak. In combination with the laser tracker, a positioning accuracy better than 20 μm can be achieved.

Fig. 6
(Color online) Images of the (a) detector platform, (b) robotic diffractometer, and (c) fluorescence detector and stage
pic

In contrast to the setup for 50-nm focusing, a phase-compensating mirror (PCM, at 129.5 m) and a multilayer KB mirror nano-focusing system (at 130 m) are utilized to achieve a 10-nm focal spot as an ultimate goal.

Compared with the total-reflection KB mirror, a fixed-surface multilayer KB mirror can increase the grazing incidence angle, resulting in a significant increase in the numerical aperture and realizing a 10-nm diffraction-limited focal spot. The downstream short mirror (horizontal focusing) determines the working distance, and the upstream long mirror (vertical focusing) has a higher fabrication difficulty and requires higher mechanical stability. Considering these aspects, the lengths of the vertical and horizontal mirror are 40 and 10 mm, respectively. The grazing incidence angle is set as 10.5 mrad.

The surface shape of the long vertical mirror is difficult to control and introduces a wavefront error that deteriorates the focal spot. To solve this problem, a phase-compensation mirror (PCM) was implemented in front of this long mirror. The PCM consists of 18 piezoelectric actuators. The voltage applied to each piezoelectric material leads to piezoelectric deformation in the transverse direction and causes the optical substrate to stretch or compress, ultimately producing the required local curvature along the meridian direction. The application of a global optimization method based on measured wavefront errors allows the precise determination of the optimal voltage distribution on the PCM[28]. This ultimately compensates for the errors originating from the KB mirror and other upstream optics. The wavefront error can be calculated from the aberrations of stripes or speckles detected by a CMOS detector mounted on a detection platform[29, 30]. To minimize the flux loss at high energies, the nominal grazing incidence angle of the Rh-coated PCM is 2.8 mrad. The parameters of the total-reflection, KB mirror, the multilayer KB mirror and the phasecompensating mirror are shown in Table 4.

Table 4
KB and phase-compensating mirror parameters
  Total reflection KB mirror Multilayer KB mirror Phase compensating mirror
Position (m) 72 ~72 77 ~77.03 76.55
Mirror Vertical focusing Horizontal focusing Vertical focusing Horizontal focusing Vertical focusing
Mirror material Single crystal silicon Single crystal silicon Single crystal silicon Single crystal silicon Single crystal silicon
Substrate shape Elliptical cylinder Elliptical cylinder Elliptical cylinder Elliptical cylinder Flat, piezoelectric deformation
Incidence angle (mrad) 2.1 2.1 10.5 10.5 2.8
Numerical aperture 0.0014 0.0017 0.007 0.007
Substrate size (L × W × H) (mm3) 162 ×30 ×30 52 × 20 ×30 42 × 30 ×30 12 × 20 ×30 160 × 50 ×10
Active optical surface (L × W) (mm2) 160 ×5 50 × 5 40 × 5 10 × 5 150 ×8
Figure error (tangential) (nm) PV<2 PV<2 PV<1 PV<1 PV<2
Micro roughness (RMS) (nm) <0.3 <0.3 <0.3 <0.3 <0.3
Reflective coating Rh Rh Ru/C Ru/C Rh
Mirror centre to focal point (mm) 160 45 40 10 /
Show more

In the 10-nm focusing area, the high-precision piezo sample stage is a sophisticated assembly of three translation stages, one partial pendulum motion stage, and one rotation stage, as shown in Fig. 3(b). This configuration allows for a maximum travel range of 6.9 cm and a resolution of 2 nm. It’s worth noting that the detectors, detector platform, and robotic diffractometer in this area are identical to those in the 50-nm focusing area.

3

Experimental station

3.1
Experimental methods
3.1.1
Scanning nano-XRF and XRF tomography

Scanning X-ray fluorescence is a powerful technique for imaging the elemental distribution within a sample[31, 32]. Using the DMM monochromator and increasing the size of the second source aperture, we achieved a higher flux, thereby increasing the efficiency of the fluorescence experiments. The implementation of a 10–50nm focal spot can extend the spatial elemental resolution of XRF to the nanoscale, paving the way for advancements in numerous research fields. Furthermore, by performing 2D fluorescence scanning at different angles, we can achieve nanoscale imaging of 3D elemental distributions through tomographic reconstruction. However, it’s worth noting that the combination of XRF and tomography requires a substantial amount of time for data acquisition. To overcome this, a fly-scan mode has been adopted to increase the experimental efficiency[33, 34].

3.1.2
Diffraction method

X-ray diffraction, an important basic experimental method, is widely used to study the crystal structure of materials[35, 36]. By positioning the area detector (Pilatus detector) downstream of the sample in transmission mode, we can capture XRD patterns. Two-dimensional XRD can be realized by raster scanning the sample. Additionally, XRF and XRD can be performed simultaneously, providing a simultaneous investigation of both the crystalline structure and elemental distribution.

CDI is a lensless imaging technique capable of recovering the shape and strain of a sample[37-39]. This is achieved by analyzing the diffraction pattern of the detector using a phase-recovery algorithm[40]. Coherence is a key factor for achieving CDI. To fulfill the full coherence requirement, the second source aperture is set to 11 µm vertically and 46 µm horizontally. The balance between the signal strength and oversampling ratio of the CDI experiment is crucial and can be accomplished by adjusting the distance between the detector and the sample.

CDI can only reconstruct isolated samples, which limits its application. To overcome this, ptychography is performed. Ptychography can be regarded as the combination of CDI and scanning transmission X-ray microscopy[41]. The high-precision sample stage in BL13U fulfills the requirement for the overlapping scanning of the probe. The diffraction pattern can be captured using the detector (Merlin detector) on a robotic diffractometer. Phase retrieval algorithms[42] are used to reconstruct the electron densities of the samples.

The data acquisitions of fluorescence and ptychography have no conflict, which grants the opportunity to simultaneously obtain the elemental distribution and microstructure of the sample. Moreover, by combining this with tomography, 3D nanoscale localization of elements in a sample can provide more information for scientific research.

3.1.3
Absorption spectrum

The oxidation state of a given element can be investigated by analyzing the XANES. At BL13U, XANES measurements can be performed in either the transmission or fluorescence modes using an ion chamber and a fluorescence detector (Vortex-ME3), respectively. The distance between the fluorescence detector and sample can be adjusted using the fluorescence detector stage to optimize the intensity of the fluorescence signal. Furthermore, the spatial resolution of the chemical states of the elements within the sample can be determined through XANES mapping.

3.2
Control and data-acquisition system

The control system consists of software and hardware platforms. The Experimental Physics and Industrial Control System (EPICS) is used as the software platform. It provides a more efficient data communication protocol (PVAS) and more flexible data structure (pvData). Additionally, EPICS provides a rich set of C-language interface functions that can be seamlessly integrated into C- or C++-based image data processing software packages. The core of the hardware platform is a motion controller. The VME motion controller (TURBO-PMAC2-VME, DELTA TAU, US) offers a wide range of control methods, supports laser interferometer-based position encoders, and has standard PID feedback control algorithms that have been adopted as motion controllers in BL13U. The control system uses a server/client architecture. The human–computer interface platform uses computers running Linux or Windows operating systems, and the experimental and graphic processing software packages can be run on any computer on the network as required. The network system has a bandwidth of 1 GHz.

4

First commissioning results

The major phase of the beamline construction was successfully completed. We have performed a focusing test for the high-flux fluorescence mode at 10 keV using the multilayer KB mirror, and the energy resolution of the DMM is approximately 0.76×10−2 (shown in Fig. 7). Speckles have been proven to modulate wavefronts and provide sufficient wavefront information[43]. The wavefront of the focal beam was tested using X-ray speckle scanning metrology technique (SBT)[44-46]. To facilitate this, we mounted a 15000-mesh sandpaper 5-mm downstream of the focal point and positioned an Eiger detector 3591-mm downstream from the sandpaper. The test was performed by scanning the sandpaper with a step size of 200 nm over 50 steps with an exposure time of 1 s.

Fig. 7
The rocking curve of DMM
pic

The rows and columns were extracted from 50 speckle images to create a series of new speckle images. The deviations between speckle images were then calculated. The local wavefront curvatures were calculated using these deviations. Quadratic integration was performed to obtain the wavefronts at the detector position. Wavefront errors in the vertical direction can be corrected using a PCM based on a global iterative algorithm[28].

Subsequently, we employed the wavefront propagation method to calculate the focal-spot size in both horizontal and vertical directions. The final focal size was determined based on eight repeated scans. The full width at half maximum (FWHM) of the intensity profile was considered asthe focal spot size. Figure 8 illustrates the intensity profiles of the focus spot in both the horizontal and vertical directions. Based on the beamline schemes at BL13U, we achieved a focal spot of 27.53 nm in the horizontal direction and 23.31 nm in the vertical direction in the high-flux fluorescence mode at 10 keV.

Fig. 8
Horizontal and vertical intensity profiles in the focal plane obtained from the wavefront propagation
pic
5

Summary and conclusion

The hard X-ray nanoprobe beamline BL13U at the SSRF was designed for use in a wide range of scientific fields such as life, environmental, and materials sciences. The beamline has two different operating modes: a high-flux fluorescence mode and a high energy resolution mode, which are realized by switching the monochromator and KB mirrors. BL13U offers adjustable beam coherence owing to the modifiable size of its second source aperture. This flexibility allows it to meet the flux requirements of experiments such as XRF and XRD, as well as the coherence prerequisites of CDI and ptychography, and makes it possible for the combined use of different techniques. In addition, the high-energy-resolution mode can be used to perform XANES and in-situ experiments. We successfully constructed the beamline and tested the high-flux fluorescence mode at 10 keV with a focusing spot of approximately 30 nm. The performance results of the beamline are in good agreement with the design for a 10-nm focusing station, achieving nanoscale focusing along with a substantial flux of 5×109 phs/s and a 0.76×10-2E/E) energy resolution. Next, we will focus on optimizing the sample experiments and adjusting the 50-nm focusing station. The equipped detectors are already capable of conducting XRF and XRD experiments. Upon completion of the high-energy-resolution mode and CDI-method testing, BL13U will be well adapted to the needs of scientific research.

References
1 M. Cotte, A. Genty-Vincent, K. Janssens et al.,

Applications of synchrotron X-ray nano-probes in the field of cultural heritage

. Comptes Rendus Phys. 19, 575-588 (2018). https://doi.org/10.1016/j.crhy.2018.07.002
Baidu ScholarGoogle Scholar
2 A. Björling, S. Kalbfleisch, M. Kahnt et al.,

Ptychographic characterization of a coherent nanofocused X-ray beam

. Opt. Express 28, 5069 (2020). https://doi.org/10.1364/OE.386068
Baidu ScholarGoogle Scholar
3 L. Chayanun, S. Hammarberg, H. Dierks et al.,

Combining nanofocused X-rays with electrical measurements at the NanoMAX beamline

. Crystals 9, 432 (2019). https://doi.org/10.3390/cryst9080432
Baidu ScholarGoogle Scholar
4 X.J. Huang, H.F. Yan, E. Nazaretski et al.,

11 nm hard X-ray focus from a large-aperture multilayer Laue lens

. Sci. Rep. 3, 3562 (2013). https://doi.org/10.1038/srep03562
Baidu ScholarGoogle Scholar
5 H. Mimura, S. Handa, T. Kimura et al.,

Erratum: breaking the 10 nm barrier in hard-X-ray focusing

. Nat. Phys. 6, 146 (2010). https://doi.org/10.1038/NPHYS1457
Baidu ScholarGoogle Scholar
6 S.P. Krüger, H. Neubauer, M. Bartels et al.,

Sub-10 nm beam confinement by X-ray waveguides: design, fabrication and characterization of optical properties

. J. Synchrotron Radiat. 19, 227-236 (2012). https://doi.org/10.1038/S0909049511051983
Baidu ScholarGoogle Scholar
7 G.Z. Zhou, Z. Hu, S.M. Yang et al.,

Preliminary exploration of hard X-ray coherent diffraction imaging method at SSRF

. Acta Phys. Sin. 69, 034102 (2020). https://doi.org/10.7498/APS.69.20191586
Baidu ScholarGoogle Scholar
8 S.L. Liu, Z.J. Xu, Z.J. Xing et al.,

Periodic artifacts generation and suppression in X-ray ptychography

. Photonics 10, 532 (2023). https://doi.org/10.3390/photonics10050532
Baidu ScholarGoogle Scholar
9 S.L. Liu, Z.J. Xu, X.Z. Zhang et al.,

Position-guided ptychography for vibration suppression with the aid of a laser interferometer

. Opt. Lasers Eng. 160, 107297 (2023). https://doi.org/10.1016/j.optlaseng.2022.107297
Baidu ScholarGoogle Scholar
10 C.H. Cui, Z.M. Zhou, L.F. Wei et al.,

Millisecond dynamics of colloidal suspension studied by X-ray photon correlation spectroscopy at the Shanghai Synchrotron Radiation Facility

. Nucl. Sci. Tech. 35, 1 (2023). https://doi.org/10.1007/s 41365-023-01358-1
Baidu ScholarGoogle Scholar
11 D. Dzhigaev, Z. Zhang, L.A.B. Marçal et al.,

Three-dimensional coherent X-ray diffraction imaging of ferroelastic domains in single CsPbBr3 perovskite nanoparticles

. New J. Phys. 23, 063035 (2021). https://doi.org/10.1088/1367-2630/ac02e0
Baidu ScholarGoogle Scholar
12 L.A.B. Marçal, E. Oksenberg, D. Dzhigaev et al.,

In situ imaging of ferroelastic domain dynamics in CsPbBr3 perovskite nanowires by nanofocused scanning X-ray diffraction

. ACS Nano 14, 15973-15982 (2020). https://doi.org/10.1021/acsnano.0c07426
Baidu ScholarGoogle Scholar
13 P.D. Quinn, L. Alianelli, M. Gomez-Gonzalez et al.,

The hard X-ray nanoprobe beamline at diamond light source

. J. Synchrotron Radiat. 28, 1006-1013 (2021). https://doi.org/10.1107/S1600577521002502
Baidu ScholarGoogle Scholar
14 U. Johansson, D. Carbone, S. Kalbfleisch et al.,

NanoMAX: the hard X-ray nanoprobe beamline at the MAX IV Laboratory

. J. Synchrotron Radiat. 28, 1935-1947 (2021). https://doi.org/10.1107/S 1600577521008213
Baidu ScholarGoogle Scholar
15 B.H. Lin, Y.H. Wu, X.Y. Li et al.,

Capabilities of time-resolved X-ray excited optical luminescence of the Taiwan Photon Source 23A X-ray nanoprobe beamline

. J. Synchrotron Radiat. 27, 217-221 (2020). https://doi.org/10.1107/S16000577519013675
Baidu ScholarGoogle Scholar
16 H.F. Yan, X.J. Huang, Y.S. Chu et al.,

Hard X-ray nanoprobe: a scanning hard X-ray microscopy beamline offering multi-modal imaging capabilities at 10 nm. SPIE Optical Engineering + Applications

. Proc SPIE 11112, X-Ray Nanoimaging: Instruments and Methods IV San Diego, California, USA. 11112, 1111202 (2019). https://doi.org/10.1117/12.2531196
Baidu ScholarGoogle Scholar
17 G. Martínez-Criado, J. Villanova, R. Tucoulou et al.,

ID16B: a hard X-ray nanoprobe beamline at the ESRF for nano-analysis

. J. Synchrotron Radiat. 23, 344-352 (2016). https://doi.org/10.1107/S1600577515019839
Baidu ScholarGoogle Scholar
18 C.G. Schroer, P. Boye, J.M. Feldkamp et al.,

Hard X-ray nanoprobe at beamline P06 at PETRA III

. Nucl. Instrum. Meth. Phys. Res. Sect. A Accel. Spectrometers Detect. Assoc. Equip. 616, 93-97 (2010). https://doi.org/10.1016/j.nima.2009.10.094
Baidu ScholarGoogle Scholar
19 A. Somogyi, K. Medjoubi, G. Baranton et al.,

Optical design and multi-length-scale scanning spectro-microscopy possibilities at the Nanoscopium beamline of Synchrotron Soleil

. J. Synchrotron Radiat. 22, 1118-1129 (2015). https://doi.org/10.1107/S1600577515009364
Baidu ScholarGoogle Scholar
20 H. Xu,

Current status and progresses of SSRF project

. Nucl. Sci. Tech. 19, 1-6 (2008). https://doi.org/10.1016/S 1001-8042(08)60013-5
Baidu ScholarGoogle Scholar
21 L.X. Yin, R.Z. Tai, D. Wang et al.,

Progress and future of Shanghai synchrotron radiation facility

. J. Vac. Soc. Jpn. 59, 198-204 (2016). https://doi.org/10.3131/JVSJ2.59.198
Baidu ScholarGoogle Scholar
22 J.H. He, Z.T. Zhao,

Shanghai synchrotron radiation facility

. Natl. Sci. Rev. 1, 171-172 (2014). https://doi.org/10.1093/nsr/nwt039
Baidu ScholarGoogle Scholar
23 W.Z. Zhang, J.C. Tang, S.S. Wang et al.,

The protein complex crystallography beamline (BL19U1) at the Shanghai Synchrotron Radiation Facility

. Nucl. Sci. Tech. 30, 170 (2019). https://doi.org/10.1007/s 41365-019-0683-2
Baidu ScholarGoogle Scholar
24 H. Jiang, Y. He, Y.M. He et al.,

Structural characterization and low-temperature properties of Ru/C multilayer monochromators with different periodic thicknesses

. J. Synchrotron Radiat. 22, 1379-1385 (2015). https://doi.org/10.1107/S1600577515017828
Baidu ScholarGoogle Scholar
25 Y. Liu, Q. S. Huang, H. Jiang et al.,

Thermal and temporal stability of nitridated Ru/B4C multilayer for high flux monochromator application

. in Optical Interference Coatings Conference (OIC) 2019 (2019), paper FB.2 FB.2 (Optica Publishing Group, 2019). https://doi.org/10.1364/OIC.2019.FB.2
Baidu ScholarGoogle Scholar
26 H. Jiang, W.Q. Hua, N.X. Tian et al.,

In situ GISAXS study on the temperature-dependent performance of multilayer monochromators from the liquid nitrogen cooling temperature to 600℃

. Appl. Surf. Sci. 508, 144838 (2020). https://doi.org/10.1016/j.apsusc.2019,144838
Baidu ScholarGoogle Scholar
27 S. Yan, H. Jiang, H. Wang et al.,

Temperature-dependent thermal properties of Ru/C multilayers

. J. Synchrotron Radiat. 24, 975-980 (2017). https://doi.org/10.1039/d0nr02260h
Baidu ScholarGoogle Scholar
28 H. Jiang, N.X. Tian, D.X. Liang et al.,

A piezoelectric deformable X-ray mirror for phase compensation based on global optimization

. J. Synchrotron Radiat. 26, 729-736 (2019). https://doi.org/10.1107/S1600577519003047
Baidu ScholarGoogle Scholar
29 N.X. Tian, H. Jiang, A.G. Li et al.,

Influence of diffuser grain size on the speckle tracking technique

. J. Synchrotron Radiat. 27, 146-157 (2020). https://doi.org/10.1107/s1600577519015200
Baidu ScholarGoogle Scholar
30 N.X. Tian, H. Jiang, L. Xue et al.,

Influence of photon beam and motor vibrations on at-wavelength X-ray speckle scanning metrology

. Front. Phys. 10, 864985 (2022). https://doi.org/10.3389/fphy.2022.864985
Baidu ScholarGoogle Scholar
31 J. Dik, K. Janssens, G. Van Der Snickt et al.,

Visualization of a lost painting by vincent van gogh using synchrotron radiation based X-ray fluorescence elemental mapping

. Anal. Chem. 80, 6436-6442 (2008). https://doi.org/10.1021/ac800965g
Baidu ScholarGoogle Scholar
32 P. Gueriau, S. Réguer, N. Leclercq et al.,

Visualizing mineralization processes and fossil anatomy using synchronous synchrotron X-ray fluorescence and X-ray diffraction mapping

. J. Roy. Soc. Interface. 17, 20200216 (2020). https://doi.org/10.1098/rsif.2020.0216
Baidu ScholarGoogle Scholar
33 Y. Jiang, J.J. Deng, J. Klug et al.,

High-speed three-dimensional imaging at the nanoscale via fly-scan ptycho-tomography

. Microsc. Microanal. 26, 1006-1008 (2020). https://doi.org/10.1017/S1431927620016645
Baidu ScholarGoogle Scholar
34 P.M. Pelz, M. Guizar-Sicairos, P. Thibault et al.,

On-the-fly scans for X-ray ptychography

. Appl. Phys. Lett. 105, 251101 (2014). https://doi.org/10.1063/1.4904943
Baidu ScholarGoogle Scholar
35 L. Romano Brandt, E. Salvati, C. Papadaki et al.,

Probing the deformation and fracture properties of Cu/W nano-multilayers by in situ SEM and synchrotron XRD strain microscopy

. Surf. Coat. Technol. 320, 158-167 (2017). https://doi.org/10.1039/d0nr02260h
Baidu ScholarGoogle Scholar
36 T.W. Cornelius, O. Thomas,

Progress of in situ synchrotron X-ray diffraction studies on the mechanical behavior of materials at small scales

. Prog. Mater. Sci. 94, 384-434 (2018). https://doi.org/10.1016/J.SURFCOAT.2017.01.065
Baidu ScholarGoogle Scholar
37 J.W. Miao, R.L. Sandberg, C.Y. Song,

Coherent X-ray diffraction imaging

. IEEE J. Sel. Top. Quantum Electron. 18, 399-410 (2012). https://doi.org/10.1109/jstqe.2011.2157306
Baidu ScholarGoogle Scholar
38 I. Robinson, R. Harder,

Coherent X-ray diffraction imaging of strain at the nanoscale

. Nat. Mater. 8, 291-298 (2009). https://doi.org/10.1038/nmat2400
Baidu ScholarGoogle Scholar
39 W.G. Yang, X.J. Huang, R. Harder et al.,

Coherent diffraction imaging of nanoscale strain evolution in a single crystal under high pressure

. Nat. Commun. 4, 1680 (2013). https://doi.org/10.1038/ncomms2661
Baidu ScholarGoogle Scholar
40 J.R. Fienup,

Reconstruction of a complex-valued object from the modulus of its Fourier transform using a support constraint

. J. Opt. Soc. Am. A 4, 118-123 (1987). https://doi.org/10.1364/JOSAA.4.00018
Baidu ScholarGoogle Scholar
41 F. Pfeiffer,

X-ray ptychography

. Nat. Photonics 12, 9-17 (2018). https://doi.org/10.1038/S41566-017-0072-5
Baidu ScholarGoogle Scholar
42 O. Mandula, M. Elzo Aizarna, J. Eymery et al.,

PyNX.Ptycho: a computing library for X-ray coherent diffraction imaging of nanostructures

. J. Appl. Crystallogr. 49, 1842-1848 (2016). https://doi.org/10.1107/S1600576716012279
Baidu ScholarGoogle Scholar
43 J.N. Xie, H. Jiang, A.G. Li et al.,

Deep learning for estimation of Kirkpatrick–Baez mirror alignment errors

. Nucl. Sci. Tech. 34, 122 (2023). https://doi.org/10.1007/s41365-01282-4
Baidu ScholarGoogle Scholar
44 H.C. Wang, J. Sutter, K. Sawhney,

Advanced in situ metrology for X-ray beam shaping with super precision

. Opt. Express 23, 1605-1614 (2015). https://doi.org/10.1364/OE.23.001605
Baidu ScholarGoogle Scholar
45 T. Zhou, H.C. Wang, O. Fox et al.,

Auto-alignment of X-ray focusing mirrors with speckle-based at-wavelength metrology

. Opt. Express 26, 26961 (2018). https://doi.org/10.1364/OE.26.026961
Baidu ScholarGoogle Scholar
46 H. Jiang, S. Yan, N.X. Tian et al.,

Extraction of medium-spatial-frequency interfacial waviness and inner structure from X-ray multilayers using the speckle scanning technique

. Opt. Mater. Express 9, 2878 (2019). https://doi.org/10.1364/OME.9.002878
Baidu ScholarGoogle Scholar
Footnote

The authors declare that they have no competing interests