Introduction
In rare-event detection experiments targeting dark matter particles [1-3], neutrinoless double-beta decay [4, 5], and neutrino property studies [6], cosmic-ray muons constitute a significant background. The high penetration power of muons enables them to traverse multi-kilometer-thick rock overburdens and directly interact with detectors, contaminating rare event signals [7]. High-energy muons can also generate secondary particles through interactions with materials, such as muon-induced neutrons and electrons. Muon-induced neutrons present particularly severe challenges: in low-energy regimes, neutron signals become indistinguishable from Weakly Interacting Massive Particle (WIMP)-nucleon collisions [8], neutron inelastic scattering produces gamma backgrounds masking neutrinoless double-beta decay events, and neutron elastic scattering signals overlap with neutrino interactions [9]. Consequently, rare-event experiments are predominantly conducted in underground laboratories, where multi-kilometer-thick rock layers effectively shield cosmic-ray muons.
The world’s underground laboratories with depths exceeding 2000 m include the China Jinping Underground Laboratory (CJPL) (2400 m) [10, 11] and Canada’s Sudbury Neutrino Observatory (SNO) (2000 m) [12], where the cosmic-ray muon flux is as low as 10-10 cm-2 s-1. The CJPL, located in the Jinping Tunnel in Sichuan Province, China, is currently the deepest underground laboratory in the world, with over 1,500 m of rock covering the facility and a maximum vertical rock depth of up to 2,400 m [10]. The thick rock surrounding the laboratory creates an environment with exceptionally low background levels throughout the facility. Nevertheless, the backgrounds from cosmic-ray muons and muon-induced neutrons remain non-negligible, making it essential to measure their flux at CJPL.
Wu et al. used a plastic scintillator telescope to measure cosmic-ray muons in the vertical direction. The first measured cosmic ray flux, without applying angular corrections, was (2.0 ± 0.4) × 10-10 cm-2 s-1 [13]. Guo et al. carefully considered the geometric structure of the mountain and corrected for the acceptance angle of the detector, thereby obtaining a more accurate measurement of the cosmic-ray flux, which was (3.53 ± 0.22stat. ± 0.07syst.)×10-10 cm-2 s-1 [14].
The flux of muon-induced neutrons is typically one order of magnitude lower than that of cosmic-ray muons [15]. For neutron detection, a 0.5% gadolinium-doped liquid scintillator (Gd-LS) detector was developed. Previous experiments have demonstrated that the Gd-LS detector exhibits excellent n-γ discrimination and neutron detection capabilities [16-20]. This capability is achieved using the pulse shape discrimination (PSD) method. The underlying mechanism relies on distinct modes of energy deposition: Gamma-ray-induced electronic recoils rapidly deposit energy via ionization and excitation, leading to the instantaneous generation of numerous charge carriers. Consequently, the signal current increased sharply to its peak and decayed quickly, exhibiting a steep trailing edge. In contrast, neutron-induced nuclear recoils deposit energy via elastic scattering, resulting in a slower energy transfer process and prolonged charge carrier generation, which manifests as gradual signal decay. Therefore, an integral-dependent discrimination factor (Dis factor), defined as the ratio of the charge integral over the trailing edge of the waveform to the total charge integral, can be employed to discriminate between electronic and nuclear recoil signals. Neutron detection is accomplished by analyzing the fast signal (generated by the elastic scattering of neutrons with nuclei), slow signal (resulting from neutron capture on gadolinium), and time interval between them. Specifically, neutrons entering the Gd-LS detector produce fast signals via elastic scattering with nuclei. As neutrons slow down and lose energy, they may be captured by gadolinium nuclei, emitting gamma rays with energies of approximately 8 MeV, which generate slow signals. To reject background events caused by gamma radiation, a neutron event is validated only if the time interval between the fast and slow signals exceeds 2 μs and is less than 40 μs.
Additionally, neutrons originating from environmental (, n) reactions and spontaneous fission induced by U/Th decay must be distinguished from muon-induced neutrons [21-29]. Environmental neutrons predominantly arise from the rocks surrounding the laboratory and detector materials, with energies typically below 10 MeV [30-32]. To shield against rock-origin neutrons, a 1-meter-thick polyethylene (PE) room was constructed. Utilizing the Gd-LS detector and PSD method, Du et al. obtained neutron energy spectrum measurements in the 1–10 MeV range within the CJPL Phase I laboratory hall and the PE room [16]. These measurements were conducted using the SAND-II algorithm to unfold the neutron spectra. The results indicated that the neutron flux in the CJPL Phase I laboratory hall was (1.51 ± 0.03stat. ± 0.1syst.)×10-7 cm-2 s-1, while in the PE room, it was (4.9 ± 0.9stat. ± 0.5syst.)×10-9 cm-2 s-1. These measurements provide valuable assessments of neutrons in the laboratory, facilitating the application of the Gd-LS detector for ultra-low flux measurements (10-11 cm-2 s-1). To obtain the source of the neutron background in the PE room, Zhong et al. employed a genetic algorithm to optimize the energy spectral continuum, thereby enabling the identification of the U-Th content in the material and determining the source and yield of the primary neutrons [19, 33]. The results indicated that 92.5% of the neutrons in the PE room originated from the aluminum protection plates near the detector, which were found to have a high thorium contamination of (1421.6±171.1) μg/kg.
Muons, muon-induced electrons, and muon-induced neutrons exhibit higher energies than the environmental background [34-36]. This energy difference allows an energy threshold (above 10 MeV) to effectively distinguish muon and muon-induced events from the environmental background. To differentiate between particle types, this study employed the PSD method. However, a critical limitation arises from the dynamic range of the Data Acquisition (DAQ) system: high-energy interactions with the scintillator produce saturated output signals that exceed the measurement range of the DAQ. Signal saturation introduces nonlinearities and waveform information loss, necessitating saturated signal reconstruction to enable accurate measurements of cosmic-ray muon flux, muon-induced neutron flux, and muon-induced electron flux in the high-energy regime.
In this study, a compact Gd-LS detector (28 L liquid scintillator) and a function-fitting method were used to conduct full-directional measurements of the cosmic muon flux, muon-induced electron flux, and muon-induced neutron flux in the PE room of the CJPL. This study demonstrates the feasibility of employing miniaturized detectors for the precise quantification of ultra-low-flux experimental backgrounds (10-10 cm-2 s-1). Section 2 details the experimental configuration of the detector and the data acquisition system. Section 3 describes the saturation signal reconstruction algorithm based on function fitting techniques. Section 4 outlines the Geant4 simulation framework [37] for detector efficiency calibration and the methodology for determining muon flux, muon-induced electron flux, and muon-induced neutron flux. Finally, Sect. 5 discusses the results and prospects for future applications in underground rare-event experiments.
28-liter gadolinium-doped liquid scintillator detector
The Gd-LS detector, located in the CJPL PE room, is a cylindrical detector with a diameter of 0.3 m and length of 0.4 m. It is filled with 28-liter of EJ-335 liquid scintillator contained in a quartz glass vessel with a thickness of only 4 mm. The scintillator was doped with 0.5% of the total mass of Gd. For EJ-335, the density was 0.89 g/cm3, and its scintillation light output was as reported in [38]. The scintillation light attenuation length was 4.5 m (applicable to large-volume detectors). The quartz glass vessel was coated with a polytetrafluoroethylene (PTFE) reflective layer on its sides to enhance the photon collection efficiency. Hamamatsu R5912-02 photomultiplier tubes (PMTs) are attached to both ends of the vessel, providing double-ended readout for signal collection. Additionally, lead plates, aluminum plates, and PE block shields surrounded the detector, and the entire experimental setup was positioned in one corner of the PE room.
A schematic of the DAQ system is shown in Fig. 1. Signals from the two PMTs and a high-precision pulse generator were split into two channels using three Fan-In/Fan-Out (FIFO) units. One channel was directly input into a Flash Analog-to-Digital Converter (FADC) module operating at a sampling rate of 500 MHz with an 8-bit resolution. The other channel is directed to three discriminators: signals from the two PMTs pass through the discriminators and are processed by a logical AND unit to generate a coincidence signal, whereas the pulse generator signal is processed by a discriminator and combined with the PMT coincidence signal via a logical OR unit. The output of the logical OR unit serves as an FADC trigger signal. When both PMTs simultaneously detected signals exceeding the threshold or the pulse generator emitted a pulse, the FADC initiated waveform sampling. The digitized waveforms were transmitted via optical fibers and stored on a computer. Fixed-frequency pulse signals from the generator were used to calculate the dead time and data selection efficiency.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F001.jpg)
Environmental neutrons and gamma rays typically exhibit energies below 10 MeV [39], and their full signals can be measured using the Gd-LS detector. Cosmic-ray muons, with an average energy of 340 GeV, generate extremely large signals when detected alongside their secondary particles. Owing to the parameter constraints of the DAQ system, the maximum amplitude of the recorded waveforms (excluding the baseline) was limited to 244.5 ADC units. The signals that exceeded this threshold were truncated. Previous studies have confirmed that events below 10 MeV are recorded successfully [16], whereas events above 10 MeV predominantly exhibit saturation. A function-fitting method was applied to reconstruct the saturated events. An artificial neural network (ANN) reconstruction was also implemented, and the results demonstrated satisfactory accuracy.
This study analyzed data collected over 412 days of stable detector operation. The dataset includes all signals that are unambiguously distinguishable from noise, with stable trigger rates, baselines, and baseline fluctuations.
Saturation signal events reconstruction
Function-fitting method
Saturation-induced signal truncation caused by the dynamic range limitation of the DAQ system leads to an irreversible loss of complete waveform morphology and charge integral data. Because the charge integral is a critical parameter directly proportional to the energy deposition and the Dis factor of detection events, the precise reconstruction of saturated signals is essential for comprehensive waveform recovery.
The hypothesis is that signal waveforms with different amplitudes exhibit analogous shape characteristics. To test this hypothesis, an artificial neural network (ANN) methodology was employed [40]. This study utilized a generalized radial basis function (GRBF) neural network, a topology validated for robust signal reconstruction performance [40, 41]. Four amplitude ranges were designated as training cohorts: 50–100, 100–150, 150–200, and 200–saturation (244.5 ADC units), with training sets containing 10,000, 10,000, 10,000, and 88 events, respectively. The scarcity of signals in the 200–244.5 range arises from the limited number of high-energy gamma-ray events emitted after neutron capture by Gd nuclei. All training samples were randomly selected from the detection datasets to ensure statistical representativeness, with a reduced population in the final cohort reflecting the natural amplitude distribution sparsity. The top panel of Fig. 2 displays the ANN outputs as characteristic scatter distributions: black inverted triangles (50–100), pink squares (100–150), purple circles (150–200), and blue pentagrams (200–saturation). These clustered distributions validate the waveform similarity hypothesis across the amplitude ranges.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F002.jpg)
Parallel reconstruction efforts apply a function-fitting method to the saturated signals. Red curves in Fig. 2 (top) represents the fits to all ANN output points using Eq. (1), where p1, p2, and p3 denote the peak amplitude, peak time, and half-height width, respectively. Table 1 compares the ANN and function-fitting reconstruction results, showing an integration average deviation within a few thousandths across all amplitude ranges. The bottom panel of Fig. 2 demonstrates the reconstruction of a saturated signal: black scatters represent raw data, the red curve is the fitted function, and 244.5 ADC units mark the DAQ dynamic range limit. Both panels in Fig. 2 confirm the ability of the function to approximate the waveform morphology. However, the function aligns closely with low-amplitude waveforms but deviates slightly at higher amplitudes than the others._2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M001.png)
| Amplitude range | ANN training events | Fitting range | ANN integration | Fitting integration | Average deviation |
|---|---|---|---|---|---|
| 50–100 | 10,000 | 0–60 ns | 1660 | 1669 | 0.0052 |
| 100–150 | 10,000 | 0–60 ns | 2847 | 2684 | 0.006 |
| 150–200 | 10,000 | 0–60 ns | 4107 | 4127 | 0.0048 |
| 200–244.5 | 88 | 0–60 ns | 5983 | 5996 | 0.0022 |
Analysis of the bias value of the function fitting
To quantify the systematic deviations introduced by the fitting function during the waveform reconstruction, we defined two distinct bias categories for systematic error characterization. This analysis specifically focuses on the reconstructed charge integral bias of physical events, which is calculated by statistical averaging over multiple events processed using the fitting procedure. Parameter estimation errors from the fitting process were not analyzed independently, as these errors were inherently embedded within the statistical ensemble treatment and thus contributed intrinsically to the observed charge integral bias.
The first type of bias arises from the imperfect alignment between Eq. (1) and the actual waveform, leading to systematic deviations when fitting the high-amplitude waveforms. The analysis framework employs unsaturated signals as baseline references across four amplitude ranges: 50–100, 100–150, 150–200, and 200–saturation (244.5 ADC units). The quantitative evaluation of the charge integral bias magnitudes within these ranges enables the systematic estimation of the uncertainty for saturated signal reconstruction. Equation (2) is further defined to characterize the first bias type:_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M002.png)
Figure 3(a) displays the amplitude-dependent bias distributions via color-coded histograms for the four amplitude ranges. The lower inset shows the expectation values and standard deviations derived from the Gaussian fits. As shown in Fig. 3(a), waveform integral biases using Eq. (1) remained below 0.006 (0.6%), with decreasing trends as the amplitudes increased. For the highest amplitude range (200–saturation), the bias measures 0.0023 and follows a decreasing pattern. This confirms the first-type bias for the saturated signal reconstruction using Eq. (1) is bounded at 0.0023. Systematic analysis validates that first-type biases remain within acceptable thresholds for practical applications, justifying their exclusion from subsequent analyses.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F003.jpg)
The second type of bias stems from reconstruction errors in the saturated signals caused by waveform truncation owing to information loss. Larger saturated signal amplitudes result in more severe truncation, exacerbating information loss, and thereby increasing the reconstruction bias.
Prior to analyzing this bias, the Cut parameter was defined to quantify the truncation extent using Eq. (3)._2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M003.png)
To quantify the integral bias between the actual and reconstructed waveforms, an Artificial Cut (AC) was applied to waveforms within the 100–150 amplitude range. Note that the Cut defined in Eq. (3) represents the percentage of the actual saturated signal truncation level relative to the full waveform amplitude, whereas AC denotes the same percentage-based Cut applied to unsaturated waveforms. The AC mimics truncation-induced information loss, which is analogous to the natural Cut. This amplitude range provides sufficient event statistics (10,000 events per Cut value from 10% to 80%) to reliably characterize the potential truncation levels in the saturated signals. Integral bias is calculated similarly to the first bias type (Eq. (2)), with the results shown in Fig. 3(b). Figure 3(b) displays the integral bias distributions for varying Cut values via color-coded histograms, with red dashed lines denoting Gaussian fits. The lower inset presents the Gaussian-fitted mean values. The results demonstrate that higher Cut values yield larger biases and broader distributions, validating the hypothesis that reconstructing heavily truncated waveforms introduces greater deviations. For 80% Cut, the bias reaches 10.7%, whereas 10% Cut results in 0.2% bias. This confirms that the reconstruction function significantly affects the impacts waveform fidelity, necessitating bias correction in both energy and Dis factor calculations for saturated signals.
Results of reconstruction
Following the evaluation of the reconstruction biases, saturated signal events were reconstructed. The correction factor (CF) is defined as CF=1-Bias, where Bias quantifies the systematic deviations described in the previous section. The reconstructed charge integral was calculated using Eq. (4), which incorporates CF to correct for the bias effects._2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M004.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M005.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M006.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F004.jpg)
Event selection
Based on the detector geometry, cosmic-ray muons deposit approximately 50 MeV of energy over a mean trajectory length of 30 cm within the detector. To suppress background contamination from environmental gamma events and peripheral muon interactions, muon candidates were selected with energy thresholds exceeding 10 MeV. Under this selection criterion, PMT flasher events and electronic noise are the dominant background sources in the high-energy region. PMT flasher events originate from spontaneous light emission caused by base discharge within PMTs, whereas electronic noise stems from baseline fluctuations in the readout electronics, which are induced by external interference. Both background types exhibited anomalously high waveform amplitudes localized to specific PMT channels. To discriminate these backgrounds from genuine muon signals, a parameter r related to the waveform amplitude is defined, which equals the ratio of the difference between the right and left waveform amplitudes to their sum (
We compared the reconstruction events with a large sample of pre-identified low-energy gamma events. These gamma events originate from neutron capture by Gd nuclei, which emit characteristic gamma rays. Characterized by low energy (typically below 5 MeV) and selected using time correlation criteria (described in Sect. 1), these events are uncontaminated by PMT flasher artifacts or electronic noise. Figure 5 compares the amplitude asymmetry parameter r between the low-energy gamma events and the reconstructed events. The left panel shows the distribution of r, revealing a higher density of reconstruction events than gamma events in the 0.2–0.4 range, suggesting the presence of background events. To reject background events, the right panel presents a two-dimensional scatter plot of the Q-value versus r, where the Q-values of the reconstruction events are scaled by a factor of 0.1 for enhanced visual discrimination, with the gamma event distribution defining the candidate region. Within the 0.2–0.4 range, five low-energy events were located outside the candidate region, leading to their classification as background. Furthermore, all reconstruction events exhibited large waveform amplitudes (with the smallest being approximately 244.5 ADC units), and the probability of simultaneous flasher event occurrence in the dual-PMT system was extremely low. Therefore, we attributed these background events to electronic noise. After stringent selection, 21 events remained as validated muon candidates.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F005.jpg)
Geant4 simulations of cosmic-ray Muons and their derivatives in Gd-LS detector
To calculate the detector’s active area, investigate potential neutron background contributions, determine neutron detection efficiency, calibrate the high-energy response of the detector system, and estimate the fluxes of muon, muon-induced eletron and muon-induced neutron, we performed Monte Carlo simulations using Geant4 software(version 4.10.06).
Simulation setting
In the particle source configuration, the GeneralParticleSource in Geant4 was used to generate cosmic-ray muons, with the cosmic-ray muon energy and incidence angle inputs derived from measurements conducted at CJPL by Guo et al. [14]. The total muon flux was 3.53 × 10-10 cm-2 s-1 with an average energy of 340 GeV. The distributions of muon energy, azimuthal and zenithal angles are illustrated schematically in Fig. 6(a), (b), and (c). In addition, the impact of the mountainous structure above the laboratory on muons was considered in recent measurements. Based on this consideration, more muon events originated from the southwest and southeast directions. The location of the particle source is shown in Fig. 6(d). Muons were generated in a circular plane 5 m directly above the detector, with a radius of 5 m.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F006.jpg)
The detector model in the detector geometry configuration is illustrated in Fig. 6(d). A PE room with dimensions 8 m (length) × 5.6 m (width) × 4.1 m (height) was simulated using 1 m thick PE sheets (green squares). The detector was positioned in one corner of the PE room and was fully enclosed by PE sheets. A 0.6 m × 0.6 m × 0.05 m lead plate (pink square) was installed at the front of the detector, whereas 0.005 m thick aluminum plates (yellow squares) shielded the rear and sides. The detector body comprised a cylindrical structure with a diameter of 0.3 m and a length of 0.4 m (brown cylinder) surrounded by a PTFE reflective layer (gray cylinder). In the experimental setup, glass light guides and PMTs were mounted at both ends of the detector body.
The active area of detector
The detector’s active area is estimated by_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M007.png)
An exposure unit is defined as 412 days of continuous detector operation, which is equivalent to 9,685 muons traversing a circular region with a 5-meter radius above the detector. Through 3,000 simulated exposure units, the event count distribution shown in Fig. 7. The simulations yielded a mean of 19.2 events per exposure unit. From this result, the mean event count was calculated as Nμ = 19.2, and the calculated active area S = 1528 cm2.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F007.jpg)
The energy deposition spectrum of the detector and the source of muon-induced neutrons
To estimate the composition of the reconstructed energy spectrum, correct the detector energy response, and investigate potential neutron background contributions, simulations were conducted with muon events equivalent to 1,000 times the experimental exposure. The simulation results are presented in Fig. 8. The upper panel of Fig. 8 displays the electron-equivalent energy (MeVee) spectrum below 100 MeVee in the detector, normalized to events per day per MeVee. This spectrum demonstrates that muon events dominate the visible events above 10 MeVee. In the 10–20 MeVee range, a small fraction of muon-induced electron events contributed to the spectrum. Additionally, recoil protons produced by the elastic scattering between muon-induced neutrons and hydrogen nuclei generate visible energy after the quenching effects. As shown by the pink band in the upper panel, muon-induced neutron events are approximately two orders of magnitude less frequent than muon events, with most exhibiting energies below 5 MeVee. The lower panel of Fig. 8 illustrates the spatial distribution of muon-induced neutron production within the PE room under these simulation conditions. The results revealed that while most neutrons originated from interactions at the PE room walls, detectable neutron-induced events primarily occured inside the detector. This spatial discrepancy arises because the 1-meter-thick polyethylene shielding effectively attenuates external neutrons, whereas internal neutron events result from muons directly penetrating the detector.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F008.jpg)
Muon-induced neutrons detection efficiency
The neutron detection efficiency is defined as the ratio of neutrons undergoing reactions in the LS to the total number of neutrons entering the detector, expressed as:_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M008.png)
Additionally, the geometric efficiency ϵg is defined as the ratio of neutrons entering from outside the sensitive volume of the detector to the total number of neutrons. This definition accounts for the fact that only external neutrons can be effectively detected, as the internal neutrons produced by muons traversing the detector are obscured by higher-amplitude muon signals and cannot be distinguished. The lower inset of Fig. 8 illustrates the simulated distribution of muon-induced neutrons, showing that external neutrons constitute 20.0% of the visible neutron population. Consequently, ϵg is determined to be 20.0%, where the 2.1% uncertainty stems from the MC statistic.
Correction of detector energy response
When the Hamamatsu R5912-02 PMT receives a large number of photons, the anode output current increases sharply. In the region of the last dynode, electron accumulation forms a space charge, generating a reverse electric field that weakens the accelerating field strength and reduces the secondary electron emission efficiency. Consequently, the detector exhibited nonlinear energy responses above 10 MeV. Correcting this nonlinearity is essential for accurately reconstructing the true energy of events.
The energy response below 10 MeV was calibrated using gamma sources 60Co and 137Cs. For energies above 10 MeV, the calibration was achieved by comparing the Q-values of the reconstructed saturated events with the simulated muon energy deposition peaks. These three calibration points are shown in the upper inset of Fig. 9. The small inset within the upper panel provides a magnified view of the low-energy region, where the red and green lines represent cubic polynomial and linear fits, respectively. The error of the calibration point was determined by the error of the mean value obtained from a Gaussian fit to the simulated peak and the uncertainty of the peak position of the integral spectrum. The minimal discrepancy between the two fits below 10 MeV deviates by only a few percent near 10 MeV, justifying the approximation of the response as linear in this range. The lower inset of Fig. 9 displays the reconstructed energy of muon candidate events using the corrected energy response curve.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F009.jpg)
Results
Systematic uncertainties
Table 2 summarizes the total systematic uncertainties. The dominant systematic uncertainties arise from the reconstructed energy, energy scale, radius of the quartz glass vessel, wall thickness of the PE room, PE shielding layer, and Pb shielding layer.
| Source | Parameter uncertainty | Flux Uncertainty | ||
|---|---|---|---|---|
| Muon (%) | Muon-induced Electrons (%) | Muon-induced Neutrons (%) | ||
| Energy scale | - | ±4.6 | ±4.6 | ±4.6 |
| Reconstructed energy | - | ±5.1 | ±5.1 | ±5.1 |
| Quartz glass vessel radius | ±0.2 cm | ±1.2 | ±1.3 | ±1.6 |
| PE room wall thickness | ±5.0 cm | ±0.3 | ±0.3 | ±0.7 |
| PE shielding thickness | ±1.0 cm | ±0.7 | ±1.4 | ±1.1 |
| Pb shielding thickness | ±0.1 cm | ±0.2 | ±0.4 | ±0.5 |
| Total systematic | - | ±7.0 | ±7.1 | ±7.2 |
The uncertainty in the energy scale originates from the uncertainty in the calibration point. The uncertainty of the calibration points arises from two aspects: the mean value error obtained through Gaussian fitting of the simulated energy spectrum peak and the uncertainty of the peak position of the integral spectrum. The uncertainty in the peak position in the integral spectrum is given by the difference between the distributions of the upper and lower errors in the reconstructed integral. Consequently, the uncertainty in the energy scale was determined by performing calibration curve fits using both the upper and lower error bounds of the calibration points and comparing the resulting flux differences between these two cases. The corresponding flux uncertainty is estimated to be 4.6%.
The uncertainty in the reconstructed saturated event energy affects the count in different energy bins. Its impact on the flux uncertainty is realized by affecting the best-fit value (k in the next section). To estimate this conservatively, we constructed two energy spectra: the first based on the reconstructed energy plus the upper error, and the second based on the reconstructed energy minus the lower error. By comparing the flux differences between these two distributions, we estimated the uncertainty in the reconstructed energy, and the corresponding flux uncertainty was estimated to be 5.1%.
The systematic uncertainties from the material dimensions were quantified through simulations. For example, a 0.2 cm uncertainty in the quartz glass vessel radius contributes 1.2% to the muon flux uncertainty, calculated as the relative change in reconstructed muon events when the radius varies by ± 0.2 cm. Finally, the total uncertainties of muon flux, muon-induced electron flux, and muon-induced neutron flux, derived through error propagation, were calculated as 7.0%, 7.1% and 7.2%, respectively.
The flux measurements of muons, muon-induced electrons, and muon-induced neutrons
It is assumed that the energy spectrum of cosmic-ray muons in the detector comprises contributions from three distinct components: muons, muon-induced electrons, and muon-induced neutron recoils. By employing the simulated energy distributions of the detector, we can fit the energy distribution of the experimental data points to derive both the spectral composition and flux in the measured spectrum. This fitting procedure is mathematically expressed as:_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M009.png)
The left panel of Fig. 10 shows the measured energy spectrum of the detector. The χ2/n.d.f. of this fit is calculated to be 8.25/11 for the binning shown in the figure. k is determined to be 1.03 ± 0.19, where the uncertainty is dominated by the statistics of data. The right panel of Fig. 10 displays the neutron and alpha recoil events. Data in the energy region below 10 MeVee were derived from prior studies, and spontaneous fission of 238U was considered [16, 19]. Data above 10 MeVee originate from neutron recoil events selected via n-γ discrimination applied to the reconstructed saturated events. The uncertainties were calculated using Poisson statistics at the 90% confidence level. In addition, the pink area in the figure represents the energy spectrum of the recoil protons after calculating the quenching effect in the simulation, which originates from neutrons entering from outside the detector.
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-F010.jpg)
Furthermore, because muon-induced electrons and neutrons originate externally to the detector, the fluxes of muon, muon-induced electrons, and muon-induced neutrons can be expressed as follows:_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M010.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M011.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M012.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M013.png)
_2026_07/1001-8042-2026-07-119/alternativeImage/1001-8042-2026-07-119-M014.png)
Summary
In this study, we measured the muon flux, muon-induced neutron flux, and muon-induced electron flux inside the PE room of the CJPL Underground Laboratory using a 28 L Gd-LS detector. For saturated signals exceeding the dynamic range of the DAQ system, we reconstructed the charge integration Q-values via a function-fitting method, with the maximum deviation in reconstruction calculated to be below 10%. Geant4 simulation software was employed for the detector efficiency estimation and energy calibration curve correction. The measured muon flux was determined as (3.64 ± 0.69stat. ± 0.25syst.) × 10-10 cm-2 s-1, the upper limit of the muon-induced neutron flux at the 90% confidence level is 3.52 × 10-9 cm-2 s-1, and the muon-induced electron flux is (5.59 ± 1.06stat. ± 0.40syst.) × 10-10 cm-2 s-1. The measurement results indicated that the cosmic-ray flux at CJPL was exceptionally low, with the flux of muon-induced neutrons being an order of magnitude lower than that of muons. Therefore, the contribution of muon-induced neutrons in the entire laboratory can be neglected.
Progress of underground nuclear astrophysics experiment JUNA in China
. Nuclear Physics News 63, 43 (2022). https://doi.org/10.1007/s00601-022-01735-3Limits on light WIMPs from the CDEX-1 experiment with a p-type point-contact germanium detector at the China Jingping Underground Laboratory
. Phys. Rev. D 90,Dark matter is darker
. Nucl. Sci. Tech. 34, 81 (2023). https://doi.org/10.1007/S41365-023-01249-5GERDA results and the future perspectives for the neutrinoless double beta decay search using 76Ge
. Int. J. Mod. Phys. A 33,Neutrinoless double beta decay: 2015 review
. Adv. High Energy Phys. 2016,Prospects for light dark matter searches at large-volume neutrino detectors
. Phys. Rep. 133,Study on cosmogenic radioactive production in germanium as a background for future rare event search experiments
. Nucl. Sci. Tech. 31. 55 (2020). https://doi.org/10.1007/s41365-020-00762-1The moment of truth for WIMP Dark Matter
. Nature 468, 389-393 (2010). https://doi.org/10.1038/nature09509Search for neutrinoless double-beta decay in 136Xe with EXO-200
. Phys. Rev. Lett. 109,The China Jin-ping underground laboratory and its early science
. Ann. Rev. Nucl. Part. Sci. 67, 231 (2017). https://doi.org/10.1146/annurev-nucl-102115-044842Status and prospects of a deep underground laboratory in China
. Phys. Conf. Ser. 203,Measurement of the cosmic ray and neutrino-induced muon flux at the Sudbury neutrino observatory
. Phys. Rev. D 80,Measurement of cosmic ray flux in the China JinPing underground laboratory
. Chinese Physics C 37,Muon flux measurement at China Jinping Underground Laboratory
, Chinese Physics C 45,Neutron- and muon-induced background in underground physics experiments, Eur
. Phys. J. 36: 171 (2008). https://doi.org/10.1140/epja/i2007-10539-6Measurement of the fast neutron background at the China Jinping Underground Laboratory
. Instrum. Meth. A 889, 105–112 (2018). https://doi.org/10.1016/j.nima.2018.01.098Stability of gadolinium-doped liquid organic scintillators
. Technical Physics Letters, 44, 251–254 (2018). https://doi.org/10.1134/S1063785018030185Response of gadolinium doped liquid scintillator to charged particles: measurement based on intrinsic U/Th contamination
. J. Instrum. 13,Identification of neutron sources and background levels in the polyethylene room of the China Jinping Underground Laboratory
. J. Instrum. 16:Performance of real-time neutron/gamma discrimination methods
. Nucl. Sci. Tech. 34, 8 (2023). https://doi.org/10.1007/s41365-022-01160-5Discrimination of neutron and gamma ray using the ladder gradient method and analysis of filter adaptability
. Nucl. Sci. Tech. 33, 159 (2022). https://doi.org/10.1007/S41365-022-01136-5Pulse-shaping method for real-time neutron/gamma discrimination at low sampling rates, Nucl
. Sci. Tech. 34, 165 (2023). https://doi.org/10.1007/s41365-023-01306-zThermal neutron background measurement in CJPL
. Nucl. Instrum. Meth. A 804, 108 (2015). https://doi.org/10.1016/j.nima.2015.09.043Simulation method for measurement of the cross-section of the 14N(n, α)11B reaction using a gridded ionization chamber
. Nucl. Sci. Tech. 32, 78 (2021). https://doi.org/10.1007/s41365-021-00921-yDevelopment of a low-background neutron detector array
. Nucl. Sci. Tech. 33, 41 (2022). https://doi.org/10.1007/s41365-022-01030-0α Decay in extreme laser fields within a deformed Gamow-like model
. Nucl. Sci. Tech. 35, 27 (2024). https://doi.org/10.1007/s41365-024-01371-yApplication of Geant4 simulation in measurement of cosmic-ray muon flux and studies of muon-induced background
. Eur. Phys. J. Plus 138, 1006 (2023). https://doi.org/10.1140/epjp/s13360-023-04639-1Resolution analysis of thermal neutron radiography based on accelerator-driven compact neutron source
. Nucl. Sci. Tech. 34, 76 (2023). https://doi.org/10.1007/s41365-023-01227-xNeutron spectrum at the underground laboratory for the ultra low background experiment
. Appl. Radiat. Isot. 81, 302–306 (2013). https://doi.org/10.1016/j.apradiso.2013.03.068Measurement of the neutron-induced total cross sections of natPb from 0.3 eV to 20 MeV on the Back-n at CSNS
. Nucl. Sci. Tech. 35, 18 (2024). https://doi.org/10.1007/s41365-024-01370-zSource-less density measurement using an adaptive neutron-induced gamma correction method
. Nucl. Sci. Tech. 34, 125 (2023). https://doi.org/10.1007/s41365-023-01274-4A method for neutron-induced gamma spectra decomposition analysis based on Geant4 simulation
. Nucl. Sci. Tech. 33, 154 (2022). https://doi.org/10.1007/s41365-022-01144-5A novel neutron energy spectrum unfolding tool using a genetic algorithm
. Nucl. Instrum. Meth. A 432, 305 (1999). https://doi.org/10.1016/S0168-9002(99)00535-5Gamma-, neutron-, and muon-induced environmental background simulations for 100Mo-based bolometric double-beta decay experiment at Jinping Underground Laboratory
. Nucl. Sci. Tech. 34, 135 (2023). https://doi.org/10.1007/s41365-023-01299-9Development of a scintillating-fiber-based beam monitor for the coherent muon-to-electron transition experiment
. Nucl. Sci. Tech. 35, 79 (2024). https://doi.org/10.1007/s41365-024-01442-0Underground laboratories
. Nucl. Instrum. Meth. Phys. Res. A 626-627, S64–S68 (2011). https://doi.org/10.1016/j.nima.2010.05.017GEANT4–a simulation toolkit
. Nucl. Instrum. Meth. A 506, 250–303 (2003). https://doi.org/10.1016/S0168-9002(03)01368-8Manufacture of gadolinium-doped liquid scintillator detector
. High Power Laser and Particle Beams 27,Neutron background measurements at China Jinping underground laboratory with a Bonner multi-sphere spectrometer
. Nucl. Instrum. Meth. A 859, 37–40 (2017). https://doi.org/10.1016/j.nima.2017.03.048Recovery of saturated signal waveform acquired from high-energy particles with artificial neural networks