Capture-gated fast neutron spectroscopy with a composite scintillator

NUCLEAR ELECTRONICS AND INSTRUMENTATION

Capture-gated fast neutron spectroscopy with a composite scintillator

Hong-Zhao Zhou,
Tao Sun,
Hai-Xia Liu,
Yi-Na Liu,
Ye Chen,
Chong-Wei Li,
Xian Guan,
Dong-Xi Wang,
Wu-Yun Xiao
Nuclear Science and TechniquesVol.37, No.8Article number 141Published in print Aug 2026Available online 19 May 2026
20701

Recoil fast neutron spectroscopy with organic scintillators is hampered by the difficulty of its spectrum unfolding, which is caused by a flat response of the detectors to monoenergetic neutrons. Capture-gated (CG) fast neutron spectroscopy using composite scintillators is a promising method for addressing this problem as the CG events produce a characteristic peak in the detector response. Nevertheless, the realization of CG fast neutron spectroscopy has been hindered by the poor discrimination performance of currently available composite scintillators and difficulty in eliminating the effect of chance coincidence events. In this study, CG fast neutron spectroscopy was realized for the first time based on a composite scintillator proposed in our previous work. The results of this study show that this spectroscopy technique has a suitable energy range of approximately 0.4 to 20 MeV. For the 241Am-Be source, the number of iterations of the recoil and CG spectrum unfolding using a flat default spectrum were 10000 and 380, respectively. Using the simulated spectrum as a reference, the R2 values of the recoil and CG unfolded spectra were found to be 0.939 and 0.975, respectively, above 1 MeV. These results indicate that CG fast neutron spectroscopy significantly alleviates the difficulty of spectrum unfolding and exhibits better energy resolution and precision. The technique presented in this work can be applied in neutron metrology laboratories and other nuclear facilities.

Fast neutron spectroscopyComposite scintillatorCapture-gated eventResponse matrixSpectrum unfolding
1

Introduction

Neutron monitoring is necessary in nuclear science and industry as nuclear fission, fusion, and spallation reactions release a large number of neutrons. The neutron fluence energy spectrum reveals their energy distribution and can be used to calculate quantities such as the fluence rate, average energy, and radiation dose. Therefore, neutron spectroscopy is an important component of neutron monitoring. Nuclear fission, fusion, and spallation reactions mainly release fast neutrons, whose fluence-to-dose conversion coefficients are much higher than those of slow neutrons [1]. Consequently, fast neutrons usually dominate the neutron dose, and thus fast neutron spectroscopy becomes significant for monitoring the operation of neutron facilities such as nuclear power plants [2], fusion tokamaks [3], spallation neutron sources [4], space stations [5], and neutron metrology laboratories [6].

Other than the time-of-flight information, the fast neutron energy spectrum is typically measured using spectrometers or detectors via neutron moderation, reaction, and scattering [7]. The first category mainly refers to the Bonner sphere spectrometer (BSS) [8], which is characterized by a broad applicable energy range from 0.025 GeV to 10 GeV [9]. However, multiple moderation spheres make the BSS bulky, and its energy resolution is poor owing to few-channel unfolding. Similar to the BSS, the multifoil activation technique [10] and superheated drop detectors [11] also require few-channel unfolding. The multifoil activation technique is typically used to monitor neutrons in nuclear reactors [12], which can also be calculated theoretically [13]. Detectors of the second category include 3He proportional counters [14], CLYC scintillators [15], and diamond detectors [16]. These detectors have better energy resolutions than BSS as their response spectra to monoenergetic neutrons contain characteristic peaks. Nevertheless, the small cross section of fast neutrons hinder their usage. Detectors of the last category, such as hydrogen-filled proportional counters [17], proton-recoil telescopes [18], and organic scintillators [19], usually rely on hydrogen to scatter fast neutrons and release recoil protons. Compared to other neutron detectors, organic scintillators are characterized by a high sensitivity to fast neutrons, excellent n/γ discrimination performance, short decay time, and low cost. These advantages contribute to the wide use of organic scintillators in recoil fast neutron spectroscopy [20-24], which refers to fast neutron spectrum measurements that rely on recoil-charged secondary particles.

Since the quenching effect depends on the energy loss density of the charged particles resulting from the interaction of the incident neutrons with the scintillator material [25], organic scintillators have nonlinear light output for protons. In addition, isotropic scattering from hydrogen below 10 MeV [7] results in a flat response to monoenergetic neutrons, which makes spectrum unfolding difficult. The first issue can be resolved by calibrating the light output in a neutron field with a broad energy distribution using the time-of-flight imformation [26]. To address the second issue, 10B (or 6Li) is doped into liquid and plastic scintillators [27, 28]. As a result, the incident fast neutrons will first create a recoil pulse on scattering from hydrogen and then give rise to a capture pulse after the thermalized neutron is captured by 10B (or 6Li). The correlated recoil and capture pulses constitute a capture-gated event (CGE) [29], which can be recognized through time coincidence. As the kinetic energies of the captured neutrons are usually negligible, CGEs exhibit a unique signature that can be used to single out only those neutrons that transfer all their energy to the recoil protons. Therefore, a characteristic peak emerges in the capture-gated (CG) spectrum formed by the recoil pulses of CGEs of monoenergetic neutrons. CG fast neutron spectroscopy takes advantage of this feature and can alleviate the difficulty of spectrum unfolding.

Doping 10B (or 6Li) aggravates the quenching effect [30] and degrades the light output as well as the n/γ discrimination performance [27], thus making it difficult to identify the capture pulses and coincidence events. To avoid this problem, segmented spectrometers [31, 32] and composite scintillators [33, 34], which combine organic scintillators with 3He proportional tubes and 10B- (or 6Li)-loaded inorganic scintillators, respectively, have been developed. As the organic and inorganic scintillators use the same photomultiplier tube (PMT) and electronic circuit, composite scintillators are usually more compact than segmented spectrometers. Capture pulses generated in inorganic scintillators have a shape that is distinct from that of the recoil pulses generated in organic scintillators. Therefore, capture pulses and coincidence events can be easily recognized. Based on composite scintillators, coincidence events consisting of CGEs and chance coincidence events (CCEs) have been identified [35] and the characteristic peaks in the coincidence spectrum of monoenergetic neutrons have been verified in previous studies [35, 36]. However, the currently available composite scintillators have been mainly based on plastic scintillators, which make distinguishing between neutrons and γ rays difficult. In addition, the effect of CCEs on the coincidence spectrum has not been eliminated and the response matrix formed by CGEs has not been established yet [37]. These issues inhibit the realization of CG fast neutron spectroscopy.

In our previous work [38], a stilbene-lithium glass composite scintillator was proposed. This detector exhibited a much superior triple-discrimination performance than composite scintillators based on plastic scintillators [34, 35, 39-41]. The coincident events were identified from the pulses generated by this detector. The CG spectrum was calculated by subtracting the chance coincidence (CC) spectrum formed by the CCEs from the coincidence spectrum [42]. The subtraction was necessary because CCEs usually dominated the coincidence events. Triple discrimination was employed to identify the CCEs caused by γ rays and capture pulses. In this study, CG fast neutron spectroscopy was fully realized to obtain the energy spectrum of the fast neutron fluence based on the proposed detector and acquisition method of the CG spectrum. Energy spectra obtained from recoil fast neutron spectroscopy, which utilized recoil pulses without coincidence, were also considered for comparison.

2

Principle of fast neutron spectroscopy

The proposed composite scintillator was based on a ϕ4 cm × 4 cm stilbene crystal, with each of its end surface covered with a piece of ϕ4 cm × 1 mm lithium glass. One piece of lithium glass was coupled with a ϕ5.08 cm ETL 9266 PMT. As shown in Fig. 1, the recoil, γ-ray, and capture pulses were generated by the neutrons and γ rays entering the end surface of the detector vertically were digitized. Recoil and γ-ray pulses were generated in stilbene, whereas capture pulses were generated in the lithium glass. Due to its small thickness, the γ-ray pulses generated in the lithium glass were below the threshold and thus were not recorded.

Fig. 1
(Color online) Schematic showing the generation and processing of signals generated in the composite scintillator used for fast neutron spectroscopy
pic

The measured spectrum was expressed aspic(1)where R is the convolution of the response matrix and Φ is the energy spectrum of fluence. R consists of a series of response spectra corresponding to monoenergetic neutrons of different energies. The goal of fast neutron spectroscopy is to obtain the energy spectrum of the fluence of the fast neutrons through spectrum unfolding, which involves deconvolution using the measured spectrum and response matrix. For the composite scintillator, there are two types of fast neutron spectroscopy techniques. Recoil fast neutron spectroscopy is the conventional method employing common organic scintillators. It involves recoil spectrum unfolding using the measured recoil spectrum and recoil response matrix. As a specific method for composite scintillators, CG fast neutron spectroscopy requires CG spectrum unfolding using the measured CG spectrum and CG response matrix.

From the digitized pulses, the measured recoil spectrum was formed by the recoil pulses that were identified by performing triple discrimination. Coincident events were recognized via time coincidence. Coincident recoil events consisting of CGEs and recoil CCEs survived triple discrimination. As the two pulses of each CCE were uncorrelated, the recoil CC spectrum had the same distribution as the recoil spectrum measured without a time coincidence. Its strength was calculated using the counts of the capture CCEs and capture events collected without a time coincidence. The measured CG spectrum was obtained by subtracting the recoil CC spectrum from the coincidence recoil spectrum, which was formed by coincident recoil events. The detector response matrix was typically established using Monte Carlo (MC) simulation.

3

Spectrum measurement

The recoil and CG spectra corresponding to monoenergetic neutrons and neutrons with broad energy distributions were measured. The measured monoenergetic neutron spectra were used for verifying the MC simulation and the feasibility of recoil and CG fast neutron spectroscopy. The measured spectra for neutrons with broad energy distributions were used for recoil and CG spectrum unfolding. Pulses were collected and processed offline to implement triple discrimination and spectrum recording.

3.1
Pulse collection

The detector was operated at a bias voltage of -810 V and was powered by an Iseg SHR4060 desktop power supply. Pulses were collected using a CAEN DT5730 digitizer controlled using the CoMPASS software. The digitizer had a resolution of 14 bits and a sampling rate of 500 MS/s. Each pulse consisted of 200 samples. The trigger threshold was set to 30 LSB, which is equivalent to approximately 50 keVee. Energy calibration was done using a 152Eu γ source via the Compton coincidence measurement [43] reported in our previous work [38]. The long gate used for calculating the pulse amplitude was set to 300 ns.

To obtain the measured monoenergetic neutron spectra, pulse collection measurements were performed in the reference monoenergetic neutron radiation fields at the China Institute of Atomic Energy. Neutron fields with energies of 1.2, 2.5, 5, and 14.84 MeV were generated using a tandem accelerator. A detector was placed 123.5 cm away from the target at the same level. Pulses generated by room-scattered neutrons were recorded by placing a shadow cone 10 cm away from the target along the detector-target axis. The shadow cone was composed of 20 cm iron and 30 cm boron-loaded polyethene.

For measuring the spectra of neutrons with broad energy distributions, neutron fields were generated separately by an 241Am-Be and 252Cf sources at the State Key Laboratory of NBC Protection for Civilians. The current neutron emission rates of the two sources were 4.19 × 107 s-1 and 9.86 × 108 s-1, respectively, after decay correction. As shown in Fig. 2, a 5.2-cm-thick lead brick was placed near the source tube for absorbing the γ rays and increase the probability of CGEs. The source–detector distance was 1.2 m when the 241Am-Be source was used. This distance was increased to 2 m when using the 252Cf source due to its higher neutron emission rate. Shadow cones consisting of 20 cm iron and 30 cm boron-loaded polyethene were used for collecting the pulses generated by room-scattered neutrons.

Fig. 2
(Color online) Experimental setup for measuring the spectra of the neutron sources (a) without and (b) with the shadow cone
pic
3.2
Spectrum recording

Each pulse was discriminated using the linear separation method [44]. Pulse-height spectra, formed by the recoil pulses without time coincidence, were recorded and further converted into measured recoil spectra by applying energy calibration to them. The effect of room scattering was eliminated by subtracting the spectrum corresponding to the room-scattered neutrons from the measured recoil spectrum.

To obtain the measured CG spectra, the timing gate of the time coincidence was set to 20 μs, according to our previous work [42]. For each capture pulse, its previous pulse was checked. If the time difference between two adjacent pulses was less than the timing gate, these two pulses were recognized as a coincident event, which could be a CGE or CCE. The previous pulse of a CCE was randomly generated by the recoil protons, γ rays, or neutron capture. Therefore, coincidence events were divided into γ-ray CCEs, capture CCEs, recoil CCEs, and CGEs. γ-ray CCEs and capture CCEs were easily identified by performing triple discrimination. However, recoil CCEs and CGEs were mixed together since their previous pulses were intrinsically identical. Consequently, CGEs could not be recognized individually. Since the recoil CC and CG spectra had different distributions, the effect of the recoil CCEs was eliminated by performing spectrum subtraction. The recoil CC spectrum was calculated as follows:pic(2)where Srec is the recoil spectrum obtained without time coincidence, and CCCE,cap and Ccap are the counts of capture CCEs and capture events without time coincidence, respectively. The requirement of coincidence and subtraction of the recoil CC spectrum significantly reduced the CGE efficiency.

4

Establishment of the response matrix

The recoil and CG response matrices were simulated using the Geant4 [45] MC toolkit. The amplitudes of the measured pulses are affected by fluctuations in the light output, amount of light transmission, light-electron conversion, and electronic noise. These factors can smear the spectrum and their effect is seen in the degraded energy resolution of the detector. To keep the simulated spectra consistent with the measured ones, energy calibration of the detector was done. In addition, a light output function was used for calculating the light output according to the energies of the secondary charged particles. The light output was expressed in MeVee, the unit of electron equivalent energy. The MC simulations were verified by comparing the measured recoil and CG spectra of the monoenergetic neutrons with the simulated ones. Details of the establishment of the recoil and CG response matrices are presented below.

4.1
Energy resolution

For the total light output, L, the energy resolution was calculated aspic(3)where ΔL is the full width at half maximum (FWHM) of L, and a1, a2, and a3 are the fitting parameters to be determined.

Four γ sources, namely, 137Cs, 60Co, 152Eu, and 22Na, were used for the energy calibration of the detector because the response of organic scintillators to electrons is linear for particle energies above approximately 125 keV [7]. For other detectors, such as NaI(Tl), the energy calibration is generally done using full energy peaks. However, a composite scintillator has a low atomic number, and full-energy peaks cannot be recognized in its γ spectra. Thus, Compton coincidence measurement was implemented to generate the characteristic peaks at the Compton edges. The energy of the Compton edge was calculated aspic(4)where Eγ is the energy of the incident γ rays and mec2 is the remaining energy of the electron. A ϕ5.08 × 5.08 cm BaF2 detector was placed 32 cm against the composite scintillator to detect γ rays scattered from stilbene. The measurements were carried out using the abovementioned four γ sources fixed successively at 2 cm in front of the composite scintillator. These two distances were optimized to balance the count rate and solid angle of the composite scintillator.

Similar to the measurement of CGEs, the Compton coincidence measurement is also time-coincident. Therefore, the coincidence events generated by the composite scintillator included true coincidence events (TCEs) and CCEs. For γ rays with a certain energy, the TCEs shared approximately the same amplitude and constituted a characteristic peak at the Compton edge in the true coincidence (TC) spectrum. The CCEs followed the same distribution as the spectrum obtained without coincidence. Consequently, the continuum in the coincidence spectrum was caused by CCEs. Using Eq. (2), the CC spectrum was calculated using the spectrum measured without coincidence and the counts over the continuum region with and without coincidence. This CC spectrum was further subtracted from the coincidence spectrum to obtain the TC spectrum.

22Na has a 90.20% probability of undergoing β+ decay, and the resultant positron emission and its subsequent annihilation will release two 511 keV γ rays that are 180° apart. If one 511 keV γ ray enters the composite scintillator, the other will probably enter the BaF2 detector. This triggers an annihilation coincidence (AC) event. The AC spectrum has the same distribution as the spectrum corresponding to the 511 keV γ rays without coincidence. Thus, after subtracting the CC spectrum for 22Na, the AC spectrum should also be removed.

The upper row of Fig. 3 shows that after subtracting the CC and AC spectra from the Compton coincidence spectra (labeled as “Total”), the TC spectra exhibited clear characteristic peaks at the Compton edges for all four γ sources. The proportion of the CC spectrum depends on the number and emission probabilities of the γ rays. 137Cs requires at least two decays to form a CCE since it releases a 661.66 keV photon in each decay. Therefore, the proportion of the CC spectrum for 137Cs was significantly lower than that for the other three sources.

Fig. 3
(Color online) Decomposition of the Compton coincidence spectra (upper row) and comparison of the measured and simulated γ spectra without coincidence (lower row)
pic

Due to the nonzero dimensions of the two detectors, Compton scattering at an angle slightly below 180° can trigger a coincidence event. Therefore, the deposited energy spectrum of the TCEs (SDES) was actually a peak rather than a line. The true coincidence peaks in the upper spectra of Fig. 3 were affected by SDES and energy resolution. SDES was simulated using Geant4 in order to obtain its width. The FWHM corresponding to the energy resolution was computed as follows:pic(5)where FWHMDES is the FWHM of SDES and FWHMTCP is the FWHM of the true coincidence peak. To determine a1, a2, and a3, Eq. (3) was used for fitting the FWHMs that were calculated based on the true coincidence peaks. Then, SDES without coincidence was broadened by the energy resolution, and a simulated spectrum was obtained. The lower row of Fig. 3 shows that all the simulated spectra agreed well with the measured ones. This indicates that the energy resolution parameters were correct.

4.2
Light output function for protons

As the scintillation light in organic scintillators is mainly induced by recoil protons, the light output function for protons, Lp(E), where E represents the energy of the protons, is crucial for obtaining the correct simulated spectra. Since the Lp(E) reported in the literature [46] resulted in a discrepancy between the simulated and measured spectra below 5 MeV, a new Lp(E) was established using the measured recoil spectra of monoenergetic neutrons.

As is typical of many organic scintillators, Lp(E) is proportional to E3/2 for energies below approximately 5 MeV and becomes approximately linear for higher energies [7]. For the measured recoil spectrum of neutrons with energy En, Lp(En) and M(En), where M(En) refers to the light output corresponding to the local maximum in the multi-scattering region, can be obtained by taking the first derivative of the spectrum [47]. Simulations showed that incident neutrons tended to transfer all their energy to the recoil protons via multiple scattering. Therefore, M(En) represents the minimum total light output of the recoil protons, whose total kinetic energy is En. As Lp(E) is a convex function, M(En) was calculated aspic(6)Using Lp(En) and M(En), the new Lp(E) was established as follows:pic(7)

4.3
Monte Carlo simulation

The composite scintillation detector was modeled using the Geant4-10.7.2 toolkit. Each end surface of the ϕ4 cm×4 cm stilbene crystal was connected to a piece of ϕ4 cm×1 mm lithium glass. The front and surrounding surfaces of the composite scintillator were covered by 0.4 mm aluminum. The PMT was neglected since it had little effect on the neutron transportation in stilbene and lithium glass. Monoenergetic neutrons entered the detector perpendicularly. Their positions were randomly sampled from a ϕ4.08 cm disk, which was kept 5 cm away from the end surface. The recommended PhysicsList FTFP_BERT_HP was used for neutron transportation below 20 MeV. The CENDL-3.1 [48] library was used to provide the neutron cross sections since the built-in G4NDL library did not have the cross sections for the 12C(n,n′)3α reaction [26]. For each run, in order to keep the statistical error of the total count below 3%, the number of primary neutrons was taken as , where is the Ceil of the integer value En (in MeV).

In the SteppingAction class, the kinetic energies of each proton track before and after a step created in stilbene were read as Epre and Epost. Using Eq. (7), the light output generated during step Lstep was represented by . The light outputs for the other charged particles were calculated in a similar manner and their light output functions were identical to those in our previous work [38]. The initial time, t1, corresponding to the light output generated in stilbene, was labeled as the first gate of the CGE. In the StackingAction class, the α track created in the two pieces of lithium glass represented the 6Li(n,α) 3H reaction. The corresponding time, t2, was labeled as the second gate of the CGE. Lstep, t1, and t2 were written into the EventAction class after their values were determined. At the end of each event, the standard deviation, σ, was computed using Eq. (3), with the total light output as Ltotal. The light output broadened by the energy resolution, Lreso, was sampled from a Gaussian distribution with a mean value of Ltotal and a standard deviation of σ. Lreso was written into the histogram H1 to record the simulated recoil spectrum. If t1 and t2 were both set, Lreso was also written into the histogram H2 to record the simulated CG spectrum. At the end of each run, H1 and H2 were saved into files.

4.4
Comparison between the measured and simulated spectra

The upper row of Fig. 4 shows the measured and simulated recoil spectra of neutrons of four different energies. Each spectrum was normalized by the counts at M(En). The good agreement between the simulated and measured spectra validates the MC simulations and energy-resolution parameters. All recoil spectra exhibited a continuum owing to the energy variation of the recoil protons at different scattering angles. Because Lp(E) is proportional to E3/2 below 5 MeV, the recoil spectrum was approximately proportional to . Thus, the counts increased as L decreased.

Fig. 4
(Color online) Comparison between the measured and simulated spectra of monoenergetic neutrons of different energies. The upper row presents the recoil spectra and the lower row presents the CG spectra
pic

As can be seen from the figure, each recoil spectrum has two edges. The right edge represents Lp(En) and is smeared by the energy resolution. The left edge represents M(En), which is caused by multiple scattering from hydrogen. If the recoil protons have a fixed total kinetic energy, Ek, the total light output of multi-scattering Lmul is affected by the distribution of kinetic energy. Lmul reaches its maximum, Lmax, when Ek is transferred to one proton. This means that multiple scattering becomes single scattering. Lmul reaches its minimum Lmin when all the recoil protons have equal kinetic energy. Therefore, Lmul is a distribution ranging from Lmin to Lmax. The recoil spectrum caused by multiple scattering is the convolution of the energy distribution and Lmul. Above M(En), the energy range contributing to the recoil spectrum gradually decreases. Consequently, the recoil spectrum exhibits an evident decline after M(En) and a local maximum appears at M(En).

The lower row of Fig. 4 shows the measured and simulated CG spectra of 1.2, 2.5 and 5 MeV neutrons. The spectra for 14.84 MeV neutrons were discarded since the measured CG spectrum had low counts and failed to show an explicit distribution. Each spectrum was normalized by the counts at Lp(En).

Compared to the recoil spectrum, the most significant distinction in the CG spectrum is that a characteristic peak appears at Lp(En). This verifies the selection effect of the CGEs on the amplitude of the recoil pulses. The good agreement between the simulated and measured spectra indicate that CG spectra were correctly simulated. The local maximum at M(En), caused by multiple scattering, in CG spectrum is more apparent than that in the recoil spectrum. As the number of CCEs was remarkably higher than that of CGEs, subtraction of the CC spectrum resulted in fluctuations on the left side of the characteristic peak. This phenomenon became more evident for the latter two energies since the probability of CGEs decreased as the neutron energy increased. In addition, a few γ rays could be mislabeled as neutrons because the n/γ discrimination became ambiguous when L was small.

4.5
Response matrices

The recoil and CG spectra were simulated using Geant4. The light outputs of these spectra ranged from 50 keVee to 11.95 MeVee, with an interval of 50 keVee. The lower limit of the light output was set to discard noise and ambiguous pulses. The energy of the incident neutrons increased from 400 keV to 20 MeV in intervals of 100 keV. The lower limit of the neutron energy corresponded approximately to the lower limit of the light output. All the spectra were normalized by the incident neutron fluence. The response matrix was constructed with the simulated spectra ranked in the ascending order of neutron energy. Each column in the matrix represented a spectrum. Figure 5 illustrates the two response matrices, in which the neutron energy ranges from 1 to 20 MeV with an interval of 1 MeV.

Fig. 5
(Color online) Recoil and CG response matrices
pic

In the recoil-response matrix, each recoil spectrum could be separated into three stages as the light output increased. In the first stage, the recoil spectrum was mainly composed of recoil carbons and low-energy protons. This stage was characterized by high counts and a fast descent. In the second stage, the recoil spectrum was dominated by recoil protons of various total energies transferred by the incident neutrons. The recoil spectrum was flat at this stage due to the isotropic cross section for neutron elastic scattering from hydrogen. As the incident neutron energy increased, the maximum light output increased, and the counts in each bin decreased. The last stage began with Lp(En), in which the incident neutron transferred almost all of its energy to the recoil protons. Subsequently, the thermalized neutron was captured by a hydrogen nucleus and a deuterium nucleus was created. A 2.22 MeV γ ray was released during the deexcitation of the deuterium nucleus. Scintillation light induced by recoil protons and 2.22 MeV γ rays generated a mixed pulse. Recoil protons stimulated a fixed light intensity. However, the scintillation light stimulated by the 2.22 MeV γ rays depended on the Compton scattering angle. This resulted in the amplitude variation of the mixed pulses. The probability of a mixed pulse was very low as the counts in the last stage were much lower than those in the second stage. Therefore, the last stage had little effect on spectrum unfolding.

In the CG response matrix, each CG spectrum contained characteristic peaks at Lp(En). A ridge line was converged by the characteristic peaks of the different CG spectra. This ridge line was a clear distinction between the two matrices and was helpful for alleviating the difficulty of spectrum unfolding. The probability of CGEs and the characteristic peak decreased as the incident neutron energy increased. CGEs discarded the events in which an incident neutron escaped without losing all of its energy. Therefore, the counts in the first stage were relatively low, whereas those in the second stage were concentrated in the peak area. The 2.22 MeV γ ray would not occur because CGEs require thermalized neutrons to be captured by 6Li. Thus, the counts in the last stage were zero. Owing to the low probability of CGEs, the counts in the CG response matrix were smaller than those in the recoil response matrix by approximately three orders of magnitude.

5

Spectrum unfolding

Using the measured spectrum and response matrix, spectrum unfolding was realized. Spectrum unfolding of the measured recoil and CG spectra of monoenergetic neutrons was performed to verify the feasibility of spectrum unfolding. As neutrons in practical workplaces usually have various energies, the recoil and CG spectrum unfolding was performed using the measured recoil and CG spectra of neutrons with a broad energy distribution.

5.1
Implementation of spectrum unfolding

In this study, spectrum unfolding was performed using the MAXED [49] code from the UMG3.1 toolkit, which was developed by Physikalisch-Technische Bundesanstalt. This code applies the principle of maximum entropy. It uses a default spectrum as the initial estimate and enters an iterative loop to optimize the solution based on the measured spectrum and response matrix. The code was terminated when the chi-square or iteration number reached their respective preset limits.

The iteration limit was set to 10000, which was sufficiently large for convergence. The light output for the measured spectrum and columns of the response matrix ranged from 0.05 to 11.95 MeVee. The neutron energy for the default spectrum and rows of the response matrix ranged from 0.4 to 20 MeV.

(1) Default spectrum

As the initial solution, the default spectrum contains prior information. The enrichment of prior information contributes to the probability of spectrum unfolding converging to the global optimal solution. Simulated [49] and uniform spectrum [50] are the two commonly used default spectra. They represent the maximum and minimum prior information about the distribution of neutron fluence spectrum, respectively.

For fast neutron spectroscopy performed in a laboratory, the neutron fluence spectrum at the test position can be simulated with the entire laboratory modeled using the MC code. An unfolded spectrum close to the simulated one can be expected if the simulated default spectrum is used. However, if there is a difference between the unfolded and simulated spectra, it is difficult to determine whether the deviation originates from the distortion of the simulated spectrum or the inaccuracy of spectrum unfolding. In addition, it is difficult to simulate the fluence spectrum for in-situ fast neutron spectroscopy. On the other hand, a uniform spectrum has the same constant value for all counts. It only exerts nonnegative constraints on the solution. Therefore, a uniform default spectrum has better applicability in spectrum unfolding and improves the reliability of the unfolded spectrum.

Both simulated and uniform spectra were used as the default spectra in the spectrum unfolding for neutrons with broad energy distributions. Their influence on the recoil and CG spectrum unfolding was analyzed. Only a uniform default spectrum was used in the spectrum unfolding for monoenergetic neutrons since it was difficult to build a precise MC model of the accelerator laboratory.

(2) Chi-square

In it important to set an appropriate limit for chi-square. If this limit is too large, the number of iterations will be insufficient and spectrum unfolding will not reach the optimal resolution. This means that the unfolded spectrum will be “underfitted” and show deviation. If this limit is too small, the number of iterations will be excessive, and the spectrum unfolding will lose its physical signature. This means that the unfolded spectrum will be “overfitted” and oscillations will appear [51]. In this study, the chi-square limits were determined based on the initial chi-square of the default spectra. For the recoil and CG spectrum unfolding, the chi-square limits were set to 1 and 0.001, respectively.

5.2
Unfolded spectra

The recoil and CG unfolded spectra of the monoenergetic neutrons are illustrated in the upper row of Fig. 6. Since the fluence of the incident neutron beam was absent, each of these spectra was normalized to its total count to examine its distribution. For 14.84 MeV neutrons, there was no CG unfolded spectrum owing to the absence of the measured CG spectrum. The lower row of Fig. 6 shows the recoil and CG unfolded spectra for neutrons with broad energy distributions generated by 241Am-Be and 252Cf, respectively. The simulated spectra at the test points are shown for comparison. Details of the spectrum simulation performed using the SuperMC software [52] can be found in our previous work [53]. The two subfigures on the left in the lower row were obtained using simulated guess spectra, and the two subfigures on the right were obtained using flat guess spectra.

Fig. 6
(Color online) Recoil and CG unfolded spectra for monoenergetic neutrons (upper row) and neutrons with broad energy distributions (lower row)
pic

(1) Monoenergetic neutrons

A distinct peak emerges at En in each of the recoil and CG unfolded spectra, indicating that both types of spectrum unfolding are practicable. The narrower peaks in the CG unfolded spectra indicate that CG spectrum unfolding provides better energy resolution than recoil spectrum unfolding. As the neutron energy increases, the information entropies of the three CG unfolded spectra are 2.13, 2.47, and 2.97, respectively. Nonzero counts discretely appear on the left side of the peaks in the CG unfolded spectra of 2.5 MeV and 5 MeV neutrons. These counts are induced by multiple factors, such as deviations between the simulated and real response spectra, misclassification of neutrons and γ rays, as well as background and electronic noise.

Since the measured recoil spectra do not have characteristic peaks, it is difficult to determine the energy and fluence of the incident neutrons based on the measured recoil spectrum via unfolding. Therefore, the unfolded recoil spectrum tends to be dispersed. This trend becomes more obvious as the neutron energy increases. Consequently, as the neutron energy increases, the information entropies of the four recoil unfolded spectra are 2.62, 3.11, 4.73, and 4.82, whereas the peak at En gradually decreases. Each of the first three information entropies is higher than that of CG unfolded spectrum, indicating that the recoil unfolded spectra are more dispersed. Recoil spectrum unfolding is also sensitive to the errors in the measured spectrum and response matrix. This leads to abnormally high counts in the low-energy region of the four recoil unfolded spectra.

(2) Neutrons with broad energy distributions

Table 1 shows the iterations of the recoil and CG spectrum unfolding for the 241Am-Be and 252Cf sources using the simulated and flat default spectra, respectively. Benefiting from prior information, the simulated default spectra result in much fewer iterations than the flat default spectra. CG spectrum unfolding requires fewer iterations than recoil spectrum unfolding. This verifies that CGEs are helpful in reducing the complexity of the spectrum-unfolding process.

Table 1
Iterations of spectrum unfolding for the two neutron sources
Spectrum unfolding Simulated default spectra Flat default spectra
241Am-Be 252Cf 241Am-Be 252Cf
Recoil 1263 311 10,000 10,000
CG 5 13 380 78
Show more

The recoil and CG unfolded spectra in the lower row of Fig. 6 are consistent with the simulated spectra. This proves that for neutrons with broad energy distributions, the two types of spectrum unfolding are feasible with simulated or flat default spectra. According to Eq. (7), the lower limit of 0.05 MeVee corresponds to a threshold value of approximately 0.4 MeV. Therefore, the unfolded spectrum obtained using the flat-guess spectrum has approximately zero fluence rates below this threshold. The unfolded spectrum obtained with the simulated default spectrum could have nonzero fluence rates below this threshold because the default spectrum is close to the global optimal solution, and its energy distribution in this region probably undergoes only a slight change after spectrum unfolding iterations.

Owing to sufficient prior information of the simulated default spectra, the recoil and CG unfolded spectra for the 241Am-Be source revive the high fluence rates in the vicinities of 3, 5, 7.8, and 10 MeV, which are produced by the energy structure of 13C. Consequently, the recoil and CG unfolded spectra agree well with the simulated spectrum, and above 0.4 MeV, their R2 values were 0.951 and 0.988, respectively. For the 252Cf source, above 0.4 MeV, the R2 values of the recoil and CG unfolded spectra were 0.969 and 0.879, respectively. The CG unfolded spectrum exhibits a smaller R2 due to the lower fluence rate in the low-energy region. When the energy limit was increased to 1 MeV, R2 values of the two unfolded spectra were 0.953 and 0.986, respectively. Multiple parameters can be calculated using the unfolded spectra. For instance, the total fluence rates of the recoil and CG unfolded spectra for the 241Am-Be source were 149.01 cm-2 s-1 and 191.76 cm-2 s-1, respectively. The deviations from the total fluence rate of the simulated spectrum were approximately -17.48% and 6.20%, respectively.

The discrepancy between the unfolded and simulated spectra could be due to multiple factors such as insufficient precision of the MC model, fluctuation of the measured spectrum, and distortion of the response matrix. Fluence rates of the recoil-unfolded spectrum are lower than those of the simulated spectrum for the 241Am-Be source. This relationship is reversed for the 252Cf source. However, CG unfolded spectrum does not encounter this reversal for either of the two sources. This indicates that compared to CG spectrum unfolding, recoil spectrum unfolding is accompanied by a larger uncertainty owing to the higher sensitivity to errors in the measured spectrum and response matrix.

The recoil and CG unfolded spectra undergo remarkable changes after the flat default spectra are used. The R2 values for the 241Am-Be source above 0.4 MeV decreased to 0.775 and 0.831, respectively because the unfolded spectra had nearly zero fluence rates below this threshold. If the energy limit was increased to 1 MeV, R2 became 0.939 and 0.975. The recoil unfolded spectrum was smooth around the vicinity of 3, 5, 7.8, and 10 MeV. Further, the CG unfolded spectrum successfully restored the high fluence rates in these regions. This indicates that CG spectrum unfolding has a better energy resolution and less reliance on prior information. For the 252Cf source, R2 values of the recoil and CG unfolded spectra were 0.963 and 0.986, respectively, above 1 MeV. After exchanging the default spectrum, the CG unfolded spectrum remained unchanged. However, the fluence rates of the recoil unfolded spectrum between 0.4 and 1 MeV increased dramatically as the fluence rates below 0.4 MeV fell to zero.

6

Summary

Organic scintillators have been widely used to realize recoil fast neutron spectroscopy. However, unfolding the spectra obtained by organic scintillators is difficult because they have a flat response to monoenergetic neutrons. To address this issue, composite scintillators consisting of organic as well as inorganic scintillators have been proposed. These detectors can generate CGEs that form characteristic peaks in the response. Taking advantage of this feature, CG fast neutron spectroscopy was proposed to alleviate the difficulty of spectrum unfolding. Currently available composite scintillators are based on plastic scintillators and exhibit poor triple-discrimination performance. In addition, the effect of CCEs must be eliminated. Limited by these two factors, CG fast neutron spectroscopy has not been fully realized to date.

In this study, CG fast-neutron spectroscopy was fully realized for the first time based on a composite scintillator proposed in our previous work. This detector exhibited excellent triple-discrimination performance. The measured CG spectrum was obtained by subtracting the CC spectrum from the coincidence spectrum. The total count of the measured CG spectrum was smaller than that of the measured recoil spectrum by approximately three orders of magnitude. Recoil and CG response matrices were established through MC simulations. The good agreement between the measured and simulated spectra for monoenergetic neutrons verified these simulations. In the CG response matrix, the characteristic peaks generated by neutrons with different energies converged into a ridge line. Using the MAXED code, recoil and CG spectrum unfolding were performed for monoenergetic neutrons and neutrons with broad energy distributions. The results thus obtained showed that recoil and CG spectrum unfolding are both practicable for composite scintillators. Parameters such as the total fluence rate could be calculated based on the unfolded spectrum. The CG unfolded spectra corresponding to monoenergetic neutrons possessed narrower peaks and lower information entropies. Compared to recoil fast neutron spectroscopy, CG fast neutron spectroscopy was able to reduce the number of iterations by at least one order of magnitude. The CG unfolded spectra were more consistent with the simulated ones than the recoil unfolded spectra. When a flat default spectrum was used for the 241Am-Be source, R2 values of the recoil and CG unfolded spectra were 0.939 and 0.975, respectively. In addition, the CG unfolded spectrum successfully restored the high fluence rates in the vicinity of 3, 5, 7.8, and 10 MeV, whereas the recoil unfolded spectrum was smooth in these regions. This indicates that CG fast neutron spectroscopy can significantly alleviate the difficulty of spectrum unfolding and has better energy resolution and precision compared to recoil fast neutron spectroscopy. In a future work, the CGE efficiency will be improved by optimizing the composite scintillator and pulse collection.

References
1. ICRP Publication 116, Conversion coefficients for radiological protection quantities for external radiation exposures. (Elsevier, Amsterdam, 2010)
2. H.H. Xiong, T.S. Li, S.Z. Chen et al.,

Investigation of an online reactor neutron spectrum measurement method with ionization chambers

. Nucl. Technol. 202, 94–100 (2018). https://doi.org/10.1080/00295450.2017.1419780
Baidu ScholarGoogle Scholar
3. Y.M. Zhang, L.J. Ge, Z.M. Hu et al.,

The first experimental results of time-of-flight neutron spectrometer at EAST

. J. Fusion Energ. 40, 14 (2021). https://doi.org/10.1007/s10894-021-00304-6
Baidu ScholarGoogle Scholar
4. Y.H. Chen, G.Y. Luan, J. Bao et al.,

Neutron energy spectrum measurement of the Back-n white neutron source at CSNS

. Eur. Phys. J. A 55, 115 (2019). https://doi.org/10.1140/epja/i2019-12808-1
Baidu ScholarGoogle Scholar
5. M.L. Litvak, I. G. Mitrofanov, D. V. Golovin et al.,

Long-period variations of the neutron component of the radiation background in the area of the international space station according to the data of the BTN-neutron space experiment

. Cosmic Res. 60, 174–184 (2022). https://doi.org/10.1134/S0010952522030054
Baidu ScholarGoogle Scholar
6. Y.N. Liu, Z.Q. Wang, C.J. Li et al.,

Establishment of simulated neutron reference radiation field for workplaces of pressurized water reactor

. At. Energy Sci. Technol. 54, 2476–2480 (2020). https://doi.org/10.7538/yzk.2020.youxian.0465
Baidu ScholarGoogle Scholar
7. G.F. Knoll. Radiation detection and measurement, 4th edn. (John Wiley & Sons, New York, 2010)
8. C.W. Li, H.Z. Zhou, H.X. Liu et al.,

Neutron spectrometry of D2O-moderated 252Cf with Bonner sphere spectrometer

. Appl. Radiat. Isot. 197, 110824 (2023). https://doi.org/10.1016/j.apradiso.2023.110824
Baidu ScholarGoogle Scholar
9. Z.M. Hu, L.J. Ge, J.Q. Sun et al.,

Measurements of cosmic ray induced background neutrons near the ground using a Bonner sphere spectrometer

. Nucl. Instrum. Methods Phys. Res., Sect. A 940, 78–82 (2019). https://doi.org/10.1016/j.nima.2019.06.004
Baidu ScholarGoogle Scholar
10. B. Liu, H.R. Yang, H.W. Lv et al.,

Neutron spectrum unfolding of the multiple activation foils based on sparse representation

. Ann. Nucl. Energy 135, 106947 (2020). https://doi.org/10.1016/j.anucene.2019.106947
Baidu ScholarGoogle Scholar
11. S. Badiei, M.R. Kardan, G. Raisali et al.,

Unfolding of fast neutron spectra by superheated drop detectors using Adaptive Network-Based Fuzzy Inference System (ANFIS)

. Nucl. Instrum. Methods Phys. Res., Sect. A 944, 162517 (2019). https://doi.org/10.1016/j.nima.2019.162517
Baidu ScholarGoogle Scholar
12. F. Molina, P. Aguilera, J. Romero-Barrientos et al.,

Energy distribution of the neutron flux measurements at the Chilean Reactor RECH-1 using multi-foil neutron activation and the Expectation Maximization unfolding algorithm

. Appl. Radiat. Isot. 129, 28–34 (2017). https://doi.org/10.1016/j.apradiso.2017.08.001
Baidu ScholarGoogle Scholar
13. Y.L Xiao, Z.X. Gu, Q.X. Zhang et al.,

New semi-analytical algorithm for solving PKEs based on Euler-Maclaurin approximation

. Ann. Nucl. Energy 141, 107308 (2020). https://doi.org/10.1016/j.anucene.2020.107308
Baidu ScholarGoogle Scholar
14. R. Batchelor, R. Aves, T. H. R. Skyrme,

Helium-3 filled proportional counter for neutron spectroscopy

. Rev. Sci. Instrum. 26, 1037–1047 (1955). https://doi.org/10.1063/1.1715182
Baidu ScholarGoogle Scholar
15. D.H. Hou, S.Y. Zhang, Y.G. Yang, et al.,

Neutron measurement and inversion based on CLYC scintillator

. J. Beijing Univ. Aeronaut. Astronaut. 47, 106–114 (2021). https://doi.org/10.13700/j.bh.1001-5965.2019.0643
Baidu ScholarGoogle Scholar
16. J. Liu, H.Y. Jiang, Z.Q. Cui, et al.,

Simultaneous measurement of energy spectrum and fluence of neutrons using a diamond detector

. Sci. Rep. 12, 12022 (2022). https://doi.org/10.1038/s41598-022-16235-x
Baidu ScholarGoogle Scholar
17. G. Pichenot, S. Guldbakke, B. Asselineau et al.,

Characterisation of spherical recoil proton proportional counters used for neutron spectrometry

. Nucl. Instrum. Methods Phys. Res., Sect. A 476, 165–169 (2002). https://doi.org/10.1016/S0168-9002(01)01424-3
Baidu ScholarGoogle Scholar
18. Y.H. Chen, Y.J. Qiu, Q. Li et al.,

Measurement of the neutron flux of CSNS Back-n ES#1 under small collimators from 0.5 eV to 300 MeV

. Eur. Phys. J. A 60, 63 (2024). https://doi.org/10.1140/epja/s10050-024-01272-z
Baidu ScholarGoogle Scholar
19. X. Yuan, X. Zhang, X. Xie et al.,

Neutron energy spectrum measurements with a compact liquid scintillation detector on EAST

. J. Instrum. 8, P07016 (2013). https://doi.org/10.1088/1748-0221/8/07/P07016
Baidu ScholarGoogle Scholar
20. R. He, X.Y. Niu, H.W. Liang et al.,

Advances in nuclear detection and readout techniques

. Nucl. Sci. Tech. 34, 205 (2023). https://doi.org/10.1007/s41365-023-01359-0
Baidu ScholarGoogle Scholar
21. T. He, P. Zheng, J. Xiao,

Measurement of the prompt neutron spectrum from thermal-neutron-induced fission in U-235 using the recoil proton method

. Nucl. Sci. Tech. 30, 112 (2019). https://doi.org/10.1007/s41365-019-0633-z
Baidu ScholarGoogle Scholar
22. S.Y. Zhang, X. Yang, Y.X. Wang et al.,

Measurement of neutron source characterization of the compact D–D neutron generator with unfolding algorithm

. Eur. Phys. J. A 59, 101 (2024). https://doi.org/10.1140/epja/s10050-023-01007-6
Baidu ScholarGoogle Scholar
23. D. Marocco, B. Esposito, F. Moro,

Combined unfolding and spatial inversion of neutron camera measurements for ion temperature profile determination in ITER

. Nucl. Fusion 51, 053011 (2011). https://doi.org/10.1088/0029-5515/51/5/053011
Baidu ScholarGoogle Scholar
24. N. Cruz, B. Santos, A. Fernandes et al.,

The design and performance of the real-time software architecture for the ITER radial neutron camera

. IEEE Trans. Nucl. Sci. 66, 1310–1317 (2019). https://doi.org/10.1109/TNS.2019.2907056
Baidu ScholarGoogle Scholar
25. C.C. Lawrence, M. Febbraro, M. Flaska et al.,

Warhead verification as inverse problem: Applications of neutron spectrum unfolding from organic-scintillator measurements

. J. Appl. Phys. 120, 064501 (2016). https://doi.org/10.1063/1.4960131
Baidu ScholarGoogle Scholar
26. H.Y. Bai, Z.M. Wang, L.Y. Zhang et al.,

Calibration of an EJ309 liquid scintillator using an AmBe neutron source

. Nucl. Instrum. Methods Phys. Res., Sect. A 863, 47–54 (2017). https://doi.org/10.1016/j.nima.2017.04.028
Baidu ScholarGoogle Scholar
27. C. Liao, H.R. Yang,

n/γ Pulse shape discrimination comparison of EJ301 and EJ339A liquid scintillation detectors

. Ann. Nucl. Energy 69, 57–61 (2014). https://doi.org/10.1016/j.anucene.2014.01.039
Baidu ScholarGoogle Scholar
28. Y.K. Sun, H. Zhang, X.K. Zhao et al.,

Identifying thermal neutrons, fast neutrons, and gamma rays by using a scintillator-based time-of-flight method

. Nucl. Instrum. Methods Phys. Res., Sect. A 940, 129–134 (2019). https://doi.org/10.1016/j.nima.2019.06.027
Baidu ScholarGoogle Scholar
29. J.B. Czirr, G.L. Jensen,

A neutron coincidence spectrometer

. Nucl. Instrum. Methods Phys. Res., Sect. A 284, 365–369 (1989). https://doi.org/10.1016/0168-9002(89)90303-3
Baidu ScholarGoogle Scholar
30. S. Hunt, C. Iliadis, R. Longland,

Characterization of a 10B-doped liquid scintillator as a capture-gated neutron spectrometer

. Nucl. Instrum. Methods Phys. Res., Sect. A 811, 108–114 (2016). https://doi.org/10.1016/j.nima.2015.12.001
Baidu ScholarGoogle Scholar
31. T.J. Langford, E.J. Beise, H. Breuer et al.,

Development and characterization of a high sensitivity segmented Fast Neutron Spectrometer (FaNS-2)

. J. Instrum. 11, P01006 (2016). https://doi.org/10.1088/1748-0221/11/01/P01006
Baidu ScholarGoogle Scholar
32. H.X. Huang, B. Gao, Y.T. Li et al.,

Low event rate neutron detector array using the coincidence between plastic scintillator and Helium-3 proportional counters

. Nucl. Instrum. Methods Phys. Res., Sect. A 1003, 165323 (2021). https://doi.org/10.1016/j.nima.2021.165323
Baidu ScholarGoogle Scholar
33. W. Li, M.R. Tao, Y.Y. Li et al.,

NaI:Tl,6Li-PVT composite scintillator toward neutron and gamma discrimination

. Radiat. Meas. 159, 106882 (2022). https://doi.org/10.1016/j.radmeas.2022.106882
Baidu ScholarGoogle Scholar
34. M. Mayer, J. Nattress, V. Kukharev et al.,

Development and characterization of a neutron detector based on a lithium glass–polymer composite

. Nucl. Instrum. Methods Phys. Res., Sect. A 785, 117–122 (2015). https://doi.org/10.1016/j.nima.2015.03.014
Baidu ScholarGoogle Scholar
35. K. Wilhelm, J. Nattress, I. Jovanovic,

Development and operation of a 6LiF:ZnS(Ag)-scintillating plastic capture-gated detector

. Nucl. Instrum. Methods Phys. Res., Sect. A 842, 54–61 (2017). https://doi.org/10.1016/j.nima.2016.10.042
Baidu ScholarGoogle Scholar
36. J. Nattress, M. Mayer, A. Foster et al.,

Capture-gated spectroscopic measurements of monoenergetic neutrons with a composite scintillation detector

. IEEE Trans. Nucl. Sci. 63, 1227–1235 (2016). https://doi.org/10.1109/TNS.2016.2537761
Baidu ScholarGoogle Scholar
37. S.D. Jastaniah, P.J. Sellin,

Digital techniques for n/γ pulse shape discrimination and capture-gated neutron spectroscopy using liquid scintillators

. Nucl. Instrum. Methods Phys. Res., Sect. A 517, 202–210 (2004). https://doi.org/10.1016/j.nima.2003.08.178
Baidu ScholarGoogle Scholar
38. H.Z. Zhou, W.Y. Xiao, T. Sun et al.,

Simultaneous detection of fast and thermal neutrons with a stilbene-6Li glass composite scintillator

. Nucl. Instrum. Methods Phys. Res., Sect. A 1054, 168396 (2023). https://doi.org/10.1016/j.nima.2023.168396
Baidu ScholarGoogle Scholar
39. M. Sharma, J. Nattress, K. Wilhelm et al.,

Triple pulse shape discrimination and capture-gated spectroscopy in a composite heterogeneous scintillator

. Nucl. Instrum. Methods Phys. Res., Sect. A 857, 75–81 (2017). https://doi.org/10.1016/j.nima.2017.03.019
Baidu ScholarGoogle Scholar
40. A. Foster, A. Meddeb, M. Wonders et al.,

On the fabrication and characterization of heterogeneous composite neutron detectors with triple-pulse-shape-discrimination capability

. Nucl. Instrum. Methods Phys. Res., Sect. A 954, 161681 (2020). https://doi.org/10.1016/j.nima.2018.11.140
Baidu ScholarGoogle Scholar
41. F. Pino, J.C. Delgado, S.M. Carturan et al.,

Novel flexible and conformable composite neutron scintillator based on fully enriched lithium tetraborate

. Sci. Rep. 13, 4799 (2023). https://doi.org/10.1038/s41598-023-31675-9
Baidu ScholarGoogle Scholar
42. H.Z. Zhou, T. Sun, H.X. Liu et al.,

Chance coincidence analysis for capture-gated neutron spectrometry with a composite scintillator

. Instrum. Methods Phys. Res., Sect. A 1056, 168636 (2023). https://doi.org/10.1016/j.nima.2023.168636
Baidu ScholarGoogle Scholar
43. L. Swiderski, M. Moszyński, W. Czarnacki et al.,

Measurement of Compton edge position in low-Z scintillators

. Radiat. Meas. 45, 605–607 (2010). https://doi.org/10.1016/j.radmeas.2009.10.015
Baidu ScholarGoogle Scholar
44. H.Z. Zhou, W.Y. Xiao, H.X. Liu et al.,

A linear separation method for neutrongamma discrimination with organic scintillators

. Instrum. Methods Phys. Res., Sect. A 1048, 167879 (2023). https://doi.org/10.1016/j.nima.2022.167879
Baidu ScholarGoogle Scholar
45. S. Agostinelli, J. Allison, K. Amako et al.,

Geant4—a simulation toolkit

. Instrum. Methods Phys. Res., Sect. A 506, 250–303 (2003). https://doi.org/10.1016/S0168-9002(03)01368-8
Baidu ScholarGoogle Scholar
46. R.A. Weldon, J.M. Mueller, C. Awe et al.,

Characterization of stilbene’s scintillation anisotropy for recoil protons between 0.56 and 10 MeV

. Instrum. Methods Phys. Res., Sect. A 977, 164178 (2020). https://doi.org/10.1016/j.nima.2020.164178
Baidu ScholarGoogle Scholar
47. A. Enqvist, C.C. Lawrence, B.M. Wieger et al.,

Neutron light output response and resolution functions in EJ-309 liquid scintillation detectors

. Instrum. Methods Phys. Res., Sect. A 715, 79–86 (2013). https://doi.org/10.1016/j.nima.2013.03.032
Baidu ScholarGoogle Scholar
48. Y.X. Zhuang, T.J. Liu, J.S. Zhang et al.,

CENDL-3 – Chinese evaluated nuclear data library, version 3

. J. Nucl. Sci. Technol. 39, 37–39 (2002). https://doi.org/10.1080/00223131.2002.10875033
Baidu ScholarGoogle Scholar
49. M. Reginatto, B. Wiegel, A. Zimbal et al.,

UMG 3.3, Analysis of data measured with spectrometers using unfolding techniques

. NEA (2004).
Baidu ScholarGoogle Scholar
50. M. Reginatto, A. Zimbal,

Bayesian and maximum entropy methods for fusion diagnostic measurements with compact neutron spectrometers

. Rev. Sci. Instrum. 79, 023505 (2008). https://doi.org/10.1063/1.2841695
Baidu ScholarGoogle Scholar
51. L. Giacomelli, M. Reginatto et al.,

Optimization of MAXED input parameters with applications to the unfolding of neutron diagnostics data from the Joint European Torus

. Rev. Sci. Instrum. 90, 093505 (2019). https://doi.org/10.1063/1.5097784
Baidu ScholarGoogle Scholar
52. Y.C. Wu, S.P. Yu, M.Y. Cheng et al.,

Design and implementation of multi-physics coupling analysis modeling program SuperMC/MCAM5.2

. At. Energy Sci. Technol. 49, 1–6 (2015). https://doi.org/10.7538/yzk.2015.49.S0.0001
Baidu ScholarGoogle Scholar
53. H.X. Liu, T. Sun, C.W. Li et al.,

Design and testing of an irradiation room with low room-scattering for neutron calibration

. Appl. Radiat. Isot. 211, 111402 (2024). https://doi.org/10.1016/j.apradiso.2024.111402
Baidu ScholarGoogle Scholar
Footnote

The authors declare that they have no competing interests.