logo

A method to estimate gas magnification of multi-gap resistive plate chamber

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

A method to estimate gas magnification of multi-gap resistive plate chamber

WU Yuelei
HU Huasi
QIN Juan
WANG Xiaolian
ZHENG Xiangyang
XU Zizong
Nuclear Science and TechniquesVol.18, No.5pp.298-301Published in print 20 Oct 2007
61400

The gas magnification of multi-gap resistive plate chamber (MRPC) is very difficult to be measured using the present instruments thoroughly in the world. The gas magnifications about ~107 were obtained in the various operation-high-voltages of multi-gap resistive plate chamber (MRPC) by means of Monte-Carlo simulation and the presented experiment method.

MRPCMagnificationMonte-Carlo simulation
References
[1] Zeballos E C, Crotty I, Williams M C S, et al. Nucl Instr Meth, 1996, A 374: 132.
[2]

Addendum to the Time of Flight ALICE Technical Design Report

. CERN/LHCC 4, 24 2004.
Baidu ScholarGoogle Scholar
[3] Akindinov A, Anselmo F, Basile M, et al. Nucl Instr Meth, 2000, A 456: 16.
[4] Shao M, Ruan L J, Chen H F, et al. Nucl Instr Meth, 2002, A 492: 344.
[5] Chen Hongfang, Li Cheng, Wang Xiaolian, et al. HEP & NP (in Chinese), 2002, 26: 201.
[6] Williams M C S. Nuclear Physics B (Proc. Suppl.) 1998, 61B: 250-257.
[7] Riegler W, Lippmann C, Veenhot R. Nucl Instr Meth, 2003, A 500: 144-162.