A compact gamma-ray detector using wavelength-shifting fibers coupled to YAP scintillation crystal

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

A compact gamma-ray detector using wavelength-shifting fibers coupled to YAP scintillation crystal

ZHU Jie
MA Hongguang
MA Wenyan
ZENG Hui
WANG Zhaomin
XU Zizhong
Nuclear Science and TechniquesVol.19, No.1pp.45-49Published in print 20 Feb 2008
41300

The performance of a compact position sensitive gamma-ray detector based on wavelength-shifting fibers coupled to YAlO3:Ce scintillation crystal was evaluated using a Monte-Carlo simulation method. The simulation model has been setup using the GEANT4 codes. Compared with the gamma-ray detector based on the YAlO3:Ce scintillation crystal coupled to Hamamastu R2486 position sensitive photomultiplier tube, the results indicate that the gamma-ray detector based on wavelength-shifting fibers readout has good position linearity, good spatial resolution and larger effective field of view. The image and point spread function of measured point were presented. The spatial resolution response as a function of position was obtained. The factors influencing spatial resolution and position linearity were discussed.

Monte-Carlo simulationWSFSpatial resolution
References
[1] Bird A J, Ramsden D. Nucl Instrum Methods Phys Res A, 1990, A299: 480-483.
[2] Majewski S, Farzanpay F, Goode A, et al. Nucl Instrum Methods Phys Res A, 1998, A409: 520-523.
[3] Garibaldi F, Cisbani E, Cusanno F, et al. Nucl Instrum Methods Phys Res A, 2001, A471: 222-228.
[4] Position sensitive photomultiplier tube with crossed wire anodes R2486 series, Specifications manual (in Japanese). Hamamatsu: K.K. Electron Tube Center, 1993, 1-53.
[5] Knoll F. Proceedings of the IEEE, March 1983, 71: 320-323.
[6] Zeng H N, Xu S B, Liu S T, et al. Acta Photonica Sinica, 2001, 30: 1321-1324.
[7] Baccaro S, Blazek K, de Notaristefani F, et al. Nucl Instrum Methods Phys Res A, 1995, A361: 209-215.
[8] Kuraray’s Scintillation Materials, Specifications manual. Japan: Kuraray Company Limited, 1995: 1-36.
[9] Belcari N, Damiani C, Guerra A D, et al. Nucl Instrum Methods Phys Res A, 2001, A461: 413-415.
[10] Zhu J, Cheng L, Wang Z M. Acta Photonica Sin, 2006, 35: 1497-1500.
[11] Zhu Jie, Ma Hongguang, Ma Wenyan, et al. Nucl Sci Tech, 2007, 18: 302-306.