logo

Characteristics of a prototype CdZnTe detector

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

Characteristics of a prototype CdZnTe detector

XU Huichao
CHENG Cheng
ZHAO Cuilan
ZHANG Jinzhou
PU Shijie
Nuclear Science and TechniquesVol.18, No.6pp.358-361Published in print 20 Dec 2007
31501

Cadmium zinc telluride (CZT) is a preferred material for X-ray and gamma-ray detector. Thanks to the relatively high atomic number, high density and wide band-gap, CZT detector possesses sharp energy resolution and high detection efficiency without cryogenic cooling. We have developed a CdZnTe detector with an energy resolution of 3.45% (FWHM) at 59.54 keV at room temperature, and it is used for X-ray fluorescence analysis. In this paper, leakage current, energy resolution and long term stability of the CZT detector are discussed.

CdZnTe (CZT) detectorFabricationCharacteristicsStability
References
[1] Schlesinger T E, Toney J E, Yoon H, et al. Mater Sci Eng, 2001, 32: 103-189.
[2] Owens A, Peacock A. Nucl Instr Meth Phys Res, 2004, A531: 18-37.
[3] Limousin O. Nucl Instr Meth Phys Res, 2003, A504: 24-37.
[4] Li Longxia, Lu Fengying. The large-size single crystals of CdZnTe growth at Yinnel Tech, DOE-NN-20 Review March1st, 2001, SNL, Albuquerque
[5] Burger A, Cui Y, Groza M, et alJournal of Electronic Materials, 2003, 32(7): 756-760.
[6] CHENG Cheng, WEI Yongbo, XU Huichao. Nucl Sci Tech, 2006, 17: 106-112.
[7] XU Huichao, ZHANG Jinzhou, SHEN Haoyuan, et al. Nuclear Techniques (in Chinese), 2005, 28(10): 791-793.