A high-resolution small animal SPECT system developed at Tsinghua

RADIOCHEMISTRY, RADIOPHARMACEUTICALS AND NUCLEAR MEDICINE

A high-resolution small animal SPECT system developed at Tsinghua

DAI Tiantian
MA Tianyu
LIU Hui
CUI Junjian
WEI Qingyang
LIU Yaqiang
WANG Shi
Nuclear Science and TechniquesVol.22, No.6pp.344-348Published in print 20 Dec 2011
37902

An SPECT system dedicated to small animal imaging shall be of a millimeter spatial resolution or even better. This study was aimed at achieving 0.5-mm spatial resolution for a small animal SPECT system at low cost. It was developed from a single-head clinical SPECT scanner, with a seven-pinhole collimator and a four-degree-of- freedom motion control stage. Several key techniques were developed, including high-resolution image reconstruction algorithm, high accuracy geometrical calibration method, and optimized system matrix derivation scheme. The system matrix was derived from Monte-Carlo simulation and de-noised by fitting each point spread function to a two dimensional Gaussian function. Experiments of point source and ultra micro hot rod phantom were conducted. With a spatial resolution of 0.5–0.6 mm, this system provides a practical way for low-cost high-resolution animal imaging on a clinic SPECT system.

SPECTMulti-pinhole collimatorHigh resolutionSystem matrixImage reconstruction
References
[1] Chatziioannou A F. Proc Am Thorac Soc, 2005, 2: 533-536.
[2] Meikle S R, Kench P, Kassiou M, et al. Phys Med Biol, 2005, 50: R45-R61.
[3] Beekman F J, Vastenhouw B. Phys Med Biol, 2004, 49: 4579-4592.
[4] Kastis G A, Barber H B, Barrett H H, et al. IEEE Trans Nucl Sci, 2000, 47: 1923-1927.
[5] Beekman F J, van der Have F, Vastenhouw B, et al. J Nucl Med, 2005, 46: 1194-1200.
[6] Schramm N U, Ebel G, Engeland U, et al. IEEE Trans Nucl Sci, 2003, 50: 315-320.
[7] Shao Y, Ma T, Yao R. IEEE Nucl Sci Symp Conf Rec, 2007, 4285-4288.
[8] DiFilippo F P. Phys Med Biol, 2008, 53: 4185-4201.
[9] Dai Q, Qi Y. At Energ Sci Technol, 2009, 43: 1034-1038.
[10] Rowe R K, Aarsvold J N, Barrett H H, et al. J Nucl Med, 1993, 34: 474-480.
[11] Metzler S D, Bowsher J E, Greer K L, et al. IEEE Trans Med Imag, 2002, 21: 878-886.
[12] Bal G, Acton P D. Phys Med Biol, 2006, 51: 4923-4950.
[13] Refecas M, Mosler B, Dietz M, et al. IEEE Trans Nucl Sci, 2004, 51: 2597-2605.
[14] Yao R, Ma T, Shao Y. IEEE Trans Nucl Sci, 2009, 56: 2651-2658.
[15] Dai T, Liu Y, Ma T, et al. At Energ Sci Technol, 2011, 45: 369-373.
[16] Liu H, Ma T, Dai T, et al.

M18-224: High accuracy geometrical calibration for half-mm animal SPECT imaging

. IEEE NSS MIC RTSD, Knoxville, Tennessee, Oct, 2010.
Baidu ScholarGoogle Scholar
[17] Ma T, Yao R, Shao Y, et al. IEEE Trans Med Imag, 2009, 28: 1929-1939.
[18] Shepp L A, Vardi Y. IEEE Trans Med Imag, 1982, MI-1: 113-122.
[19] Hudson H M, Larkin R S. IEEE Trans Med Imag, 1994, 12: 601-609.
[20] Jan S, Santin G, Strul D, et al. Phys Med Biol, 2004, 49: 4543-4561.
[21] Brun R, Rademakers F. Nucl Inst Meth Phys Res A, 1997, 389: 81-86. See also http://root.cern.ch/.