logo

Energy calibration of laterally segmented electromagnetic calorimeters based on neutral pion detection

NUCLEAR ENERGY SCIENCE AND ENGINEERING

Energy calibration of laterally segmented electromagnetic calorimeters based on neutral pion detection

Malek Mazouz
Nuclear Science and TechniquesVol.28, No.11Article number 155Published in print 01 Nov 2017Available online 25 Oct 2017
35200

This paper describes a method for energy calibration of laterally segmented electromagnetic calorimeters based on the detection of two-photon decays of π0 mesons. The calibration procedure performs a χ2 function minimization between the measured π0 energy in the calorimeter and its expected energy deduced from the π0 momentum direction. The performance of this technique is demonstrated with a Monte Carlo simulation of an experimental case where biased calibration coefficients are employed. The real calibration coefficients are restored with less than 1% relative accuracy when a sufficient number of π0 is detected. This technique is applied to monitor daily the calibration coefficients of the calorimeter used in the Jefferson Lab Hall A DVCS experiments.

Electromagetic calorimetersEnergy calibrationDetector modelling and simulationsData processing methods

1 Introduction

The main goal of electromagnetic calorimeters is to measure precisely the energy of detected particles such as photons and electrons. These detectors are important elements in nuclear physics instrumentation and are widely used especially in hadronic and particle physics experiments [1]. Usually, the knowledge of the impact position of detected particles is required leading to a lateral segmentation of these calorimeters. Lead fluoride (PbF2) is commonly employed as a constituent material of such calorimeters [2-5]. Its high density offers a short radiation length (X0=0.93 cm) and a small Molière radius (rM=2.12 cm) leading to compact detector geometries [6, 7]. In practice, the segmentation size is close to rM while the longitudinal calorimeter length is equal to several X0 to ensure a full development of electromagnetic showers in the detector blocks. Each calorimeter block is generally connected to a photomultiplier tube (PMT) to collect the light induced by a shower and an electronic base to shape and amplify the PMT signal. The amplitude of the final signal is then related to the energy released by the particle in the considered block via a calibration coefficient.

A great deal of effort is still underway to optimize the performance of electromagnetic calorimeters and in particular, their energy resolution [8-10]. In addition to the design optimization, many energy calibration techniques are still investigated to improve the reconstructed energy of detected particles [11-14]. Indeed, the energy calibration is sensitive to many factors, such as the hadronic and electromagnetic background during the experiment and the gain drift of blocks due to an increasing loss of their transparency when exposed to high radiation rates and accumulated doses [15, 16]. The choice of the energy calibration method is then crucial to ensure a frequent and proper monitoring of the calibration coefficients of each calorimeter block keeping thus an acceptable energy resolution. In many experiments, dedicated calibration runs have to be taken in order to send particles of known energy in the calorimeter. Other methods consist to calibrate the detector with cosmic rays or with a cluster of LEDs of known light intensity placed in front of the calorimeter blocks [16]. All these standard methods have the disadvantage to change momentarily the experimental setting or trigger and to decrease the amount of time dedicated to physics runs. In addition, many parameters, such as the nature of particles used in the calibration procedure, their energies and incident angles, and the experimental background, could be different relative to the physics run conditions. The obtained calibration coefficients are then not necessarily optimized for the particles detected during physics runs. Finally, the calibration provided by these methods at a given time may not still be valid afterwards if the calibration coefficients are time dependent.

This paper presents an energy calibration method based on the detection of π0 mesons in the calorimeter. This method is performed to monitor daily the calibration of the electromagnetic calorimeter of the Jefferson Lab Hall A DVCS experiments [2, 3, 17, 18] and can be applied in similar experiments where two-photon decays of π0 can be detected in a laterally segmented calorimeter. It does not require specific runs since the calibration data are taken simultaneously with the primary experimental data at the same conditions. This method will be exposed and tested on the basis of a calorimeter simulation using the GEANT4 toolkit [19]. The relative accuracy on the obtained calibration coefficients will finally be discussed in terms of the number of detected π0.

2 Simulation of an experimental case

The main goal of the Jefferson Lab Hall A DVCS experiments is to study the Deeply Virtual Compton Scattering (DVCS) process on the nucleon eNeNγ using a few GeV electron beams impinging on a liquid hydrogen (LH2) or deuterium (LD2) target [2, 3, 17, 18]. Many similar experiments in the world are concerned with this increasingly timely topic of the characterization of the nucleon structure with the DVCS process [20-23]. As shown in Fig. 1, the scattered electron is detected in a high resolution spectrometer (HRS) which determines accurately its momentum and angles as well as the reaction vertex coordinates in the target [24]. The emitted photon is detected in an electromagnetic calorimeter consisting of a 16 × 13 matrix of 3 × 3 × 18.6 cm3 PbF2 blocks placed at 1.1 m from the target. This short distance combined with the high luminosity of the experiment (1037 cm-2s-1) leads to an important background rate in the calorimeter as well as radiation damage near the front face of blocks. The energy calibration of the calorimeter is of direct relevance since it affects directly the eNeNγ event identification based on a study of the missing mass squared:

Figure 1:
Experimental setup of the Hall A DVCS experiments and definition of particle 4-vectors. q'=qγ'q'=q2'+q1' is the photon (π0) 4-vector. The recoil nucleon, not detected, is identified with a cut on the missing mass squared defined by Eq. (1) (Eq. (12)) if a photon (π0) is detected in the calorimeter.
pic
Mx2=(k+pk'qγ')2, (1)

where k, k′, p, and qγ'

are respectively the 4-vectors of the incident electron, the scattered electron, the initial nucleon at rest and the emitted photon. Experimentally, k, k′, and p are known accurately and the resolution of Mx2

is dominated by the calorimeter energy resolution [16]. The photon 4-vector is determined for each event, j, from the energies deposited in the calorimeter blocks:

Ei'(j)=Ci'Ai(j), (2)

where Ai(j) is the output signal amplitude of block i, and Ci' is the corresponding calibration coefficient. Let’s assume that the Ci' coefficients are roughly known at a given time with any standard calibration method. As mentioned above, many factors, such as different experimental conditions and transparency losses of blocks, could modify the calibration by 20% in average [25] and by up to 40% for some particular blocks:

Ci=ϵiCi'. (3)

The correction factors, ϵi, and thus the new calibration coefficients, Ci, must then be known accurately in order to get the real deposited energies, Ei(j):

Ei(j)=ϵiCi'Ai(j)=ϵiEi'(j). (4)

The goal of the calibration method, discussed hereafter, is to determine the ϵi for each block, i. This method is based on the study of π0 electroproduction events eNeNπ0eNγγ where the two photons coming from the π0 decay are detected in the calorimeter. This reaction is very common in lepton-hadron scattering and is usually present in the data of DVCS experiments. The π0 energy is almost equal to the DVCS photon energy because of the kinematic similarity between these two reactions, and it is then well adapted for the calorimeter calibration.

To demonstrate the validity of the calibration method, a GEANT4 simulation of the experimental setup is performed and N=2.106 eNeNπ0 events are generated following the kinematics presented in Table 1. This particular kinematics corresponds to one setting of the Hall A DVCS experiments [26]. The electromagnetic showers created by the π0 photons in the calorimeter are fully simulated [27]. The generation and tracking of Čerenkov photons induced by the showers in the PbF2 blocks are not considered in this study because of unrealistic computing times and strong sensitivity to exact optical properties of crystals and their wrapping surfaces. However, one can attribute an average number of Nph=330 photo-electrons per GeV deposit collected by the PMT of each block to be coherent with the reported experimental energy resolution σ(E)/E=3.1% [26]. This number is deduced from a H(e,eCalo'pHRS) elastic calibration where Ee'=3.16 GeV scattered electrons are detected in the calorimeter:

Table 1:
Simulated kinematics. θkk′ (θkq′) is the polar angle between the scattered electron (π0) and the incident electron. ϕkq is the azimuthal angle of the π0 relative to the incident electron. <q0'> is the mean π0 energy within the calorimeter acceptance
k0(GeV) k0'(GeV) θkk′(deg) <q0'>(GeV) θkq′(deg) ϕkq′(deg)
4.455 [1.78, 1.96] [25, 28] 2.4 [9, 28] [-50, 50]
Show more
3.1%=σ(Ee')Ee'=1Nph Ee'(GeV). (5)

Finally, real radiative effects, where the incident or the scattered electron emits a real photon, are taken into account in the simulation following the procedure described in [16].

To mimic the uncertainty on the experimental calibration coefficients, the energies, Ei, deposited in the calorimeter blocks are multiplied by random numbers, Ki, varying uniformly between 1-Kmax and 1, where Kmax=40%, and become equal to:

Ei'(j)=KiEi(j). (6)

The calibration method consists then to determine, from Ei'(j) and the known 4-vectors {k, k′(j), p}, the correction factors, ϵi, which have to be applied in order to get the correct energies. It is evident from Eqs. 4 and 6 that a successful calibration should give ϵi ≈ 1/Ki with δi=ϵiKi-1 being the relative deviation of ϵi from 1/Ki.

3 Calibration method

The calibration method is based on a χ2 minimization between the measured π0 energy, Eπ0rec, reconstructed from Ei', and the expected π0 energy, Eπ0cal, calculated from the π0 momentum direction. The following subsections detail the different steps of the calibration.

3.1 Particle 4-vector reconstruction

The reconstructed π0 energy in the simulation is the sum of the two shower energies created by the two photons of the π0 decay. Experimentally, one can have multiple showers in the calorimeter coming from other meson decays, accidentals or any additional final state particles. In this case, a clustering algorithm based on a cellular automata is used to determine the number of showers and to separate the different cluster contributions [28]. The energy of a shower is then the sum of the energies deposited in the Nclus blocks belonging to the corresponding cluster:

Esh'=iNclusEi'. (7)

Figure 2 shows an example of a π0 event in the calorimeter where the corresponding clusters are indicated in red. The transverse coordinates (xc,yc) of a shower centroid are determined with a center of gravity based method from the coordinates (xi,yi) of the blocks belonging to the corresponding cluster [29]:

Figure 2:
(Color online) Numbering of the calorimeter blocks. The blocks located at the calorimeter edge are shown in blue. An example of a two-cluster event created by the π0γγ decay, is shown with the red blocks.
pic
xc=iwixiiwi, (8)

with a similar equation for yc. The weight, wi, is given by:

wi=max{0;W0+ln(Ei'Esh')}, (9)

where W0 is a free dimensionless parameter optimized for the considered energies [30]. The showers whose centroid coordinates belong to a block of the calorimeter edge (see Fig. 2) are excluded from the calibration procedure. Indeed, the energy of a particle impinging these particular blocks is not well reconstructed because of the shower energy leakage near the calorimeter edge.

The knowledge of the shower centroid coordinates, with a 3 mm accuracy [26], and the interaction vertex coordinates allows us to reconstruct the momentum direction of the particle creating the shower and thus its 4-vector q'(Esh',q'), assuming a photon as a detected particle. The different steps of this reconstruction procedure are applied for each simulated event to be coherent with the experimental data analysis. It is worth noting that the obtained particle direction is not very sensitive to the deposited energies, Ei', and thus to the calibration coefficients, but depends mainly on the block coordinates where a maximum energy is released by the shower [30]. Consequently, the angle between the particle direction and the virtual photon momentum, q=kk', is well reconstructed even if inadequate calibration coefficients are employed.

3.2 eNeNπ0 reaction identification

In the simulated data, only eNeNπ0eNγγ events are generated. Experimentally, we have to identify this reaction by selecting only 2-cluster events and by computing the invariant mass of the two reconstructed particles in the calorimeter:

Minv=(q1'+q2')2, (10)

where q1' and q2' are the particle 4-vectors determined as described in the previous subsection. If these particles are photons coming from a π0 decay, then their invariant mass should be equal to the π0 mass Mπ00.135 GeV, within detector resolution. In an experimental or simulated Minv distribution, π0 events are located in a peak around Mπ0 and cannot be confused with other mesons or events. However, a miscalibration of the calorimeter blocks can shift the peak position (Mpeak) and increase its width (σMinv) as shown in the simulated Minv histogram of Fig. 3. A cut around the peak position, instead of Mπ0, is then applied to ensure the selection of π0 events:

Figure 3:
(Color online) Simulated two-gamma invariant mass distribution before (black) and after (red) the calibration. The vertical dashed lines represent the Minv cut of Eq. (11).
pic
|MinvMpeak| < 3 σMinv, (11)

where σMinv is the resolution of the reconstructed Minv variable.

The second step of the eNeNπ0 reaction identification consists to compute for each π0 event the missing mass squared defined by:

Mx2=(k+pk'q1'q2')2, (12)

which must be equal to the nucleon mass squared MN20.88

GeV2 within detector resolution if the considered event corresponds to a eNeNπ0 reaction and if the calorimeter is well calibrated. Figure 4 shows the simulated Mx2 distribution where only eNeNπ0 events are generated. The peak position in Fig. 4 is not centered around MN2 because of the Ei smearing by the random factors, Ki, which is equivalent experimentally to a wrong calibration of the blocks. The non-Gaussian behavior of the right side of the peak is due to real radiative effects. Experimentally, deep inelastic scattering (DIS) events, where additional mesons are created in the final state, are also present in a Mx2 distribution and are located after the Mx2 peak position (see Fig.2 in Ref. [26] for more details). To avoid the contamination from DIS events, the selection of eNeNπ0 is performed, as shown in Fig. 4, by applying the following Mx2 cut:

Figure 4:
(Color online) Simulated missing mass squared distribution before (black) and after (red) the calibration. The vertical dashed line represents the Mx2 cut of Eq. (13).
pic
Mx2 < Mpeak2+σMx2, (13)

where Mpeak2 is the peak position in the Mx2 distribution and σMx2 its resolution. Both Eq. (11) and Eq. (13) cuts are applied in the simulation to be coherent with the experimental data analysis.

3.3 Expected π0 energy calculation

For each selected eNeNπ0 event, the reconstructed π0 energy writes:

Eπ0rec=i=1208=16×13Ei' di, (14)

where di=1 if the block, i, belongs to one of the two clusters created by the π0 and di=0 otherwise. This reconstructed energy leads to particular values of Mx2 and Minv which can be different from MN2 and Mπ0 because of energy resolution effects or a bad calibration of blocks. Actually, it is possible to find for each event a more realistic estimation of the pion energy, called Eπ0cal hereafter, giving exactly Mx2=MN2 and Minv=Mπ0. Equations (10) and (12) lead to:

MN2=(k+pk'q1'q2')2=(k+pk')2+Mπ022(k0k0'+MN)Eπ0cal+2q(Eπ0cal)2Mπ02 cosθ, (15)

where θ is the angle between the momenta of the virtual photon and the π0. As mentioned above, θ is well defined experimentally even if biased calibration coefficients are employed. The physical solution of Eq. (15) is given by:

Eπ0cal=b+b24ac2a, (16)

where:

a=4(k0k0'+MN)24q2cos2θ,b=4(k0k0'+MN)[MN2kk'+p2Mπ02],c=4Mπ02q2cos2θ+[MN2kk'+p2Mπ02]2. (17)

Figure 5 shows the histogram of the relative difference between Eπ0cal and the real π0 energy, Eπ0real, known for each event in the simulation. The comparison between the reconstructed π0 energy, Eπ0rec, and Eπ0real is also shown in Fig. 5. The mean value and the small RMS of the (Eπ0calEπ0real) distribution relative to the (Eπ0recEπ0real) distribution indicate that Eπ0cal is much closer to the real π0 energy than Eπ0rec. The calculated π0 energy can then be used to adjust the calorimeter calibration and determine the correction factors ϵi as detailed in the following subsection.

Figure 5:
(Color online) Relative difference between the real π0 energy, Eπ0real, and the reconstructed π0 energy, Eπ0rec, before (black) and after (red) the calibration. The relative difference between, Eπ0real, and the calculated π0 energy, Eπ0cal, (Eq. (16)) is shown with the blue histogram.
pic
3.4 χ2 minimization

For each eNeNπ0 event, j, Eπ0cal(j) gives a realistic estimation of the π0 energy and thus of what the energy deposited in the calorimeter should be. The correction factors, ϵi, are then those minimizing the χ2 defined by:

χ2=j=1Nπ0(Eπ0cal(j)Eπ0rec(j))2, (18)

where Nπ0 is the total number of events and Eπ0rec(j) is the reconstructed π0 energy determined with the correction factors ϵi:

Eπ0rec(j)=i=1208ϵiEi'(j)di(j). (19)

The minimization of Eq. (18) writes:

dχ2dϵk=2j=1Nπ0[Eπ0cal(j)i=1208ϵiEi'(j)di(j)]Ek'(j)dk(j)=0 k=1,2...208 (20)

and yields to the following linear set of equations:

i=1208[j=1Nπ0Ei'(j)di(j)Ek'(j)dk(j)]ϵi=j=1Nπ0Eπ0cal(j)Ek'(j)dk(j) k=1,2...208 (21)

The correction factors, ϵi, are then obtained by inverting the 208 × 208 matrix, αik=j=1Nπ0Ei'(j)di(j)Ek'(j)dk(j). Experimentally, this minimization procedure provides the adjusted calibration coefficients defined by Eq. (3).

4 Results and discussion

Figure 6 shows the relative difference δi=ϵiKi-1 between the obtained correction factors, ϵi, and the smearing factors, 1/Ki, as a function of the calorimeter block number. Figure 6 proves that the present calibration method succeeds to find the real calibration coefficients for all the calorimeter blocks, within 1% deviation, except for those located at the calorimeter edge. Since the clusters centered around the latter blocks are excluded from the calibration procedure, the α-matrix diagonal elements corresponding to these blocks are filled with low deposited energies and become sensitive to energy fluctuations of shower tails. The corresponding calibration coefficients are then slightly overestimated. The energy resolution improvement is shown in Fig. 5 where a clear decrease of the (Eπ0recEπ0real) distribution RMS is visible after this calibration. The calibration quality can also be seen in Fig. 3 and Fig. 4 where the peaks corresponding respectively to π0 events and eNeNπ0 events are now centered around Mπ0 and MN2 with a smaller width.

Figure 6:
(Color online) The relative difference, δi, between the correction factors, ϵi, and the smearing factors, 1/Ki, as a function of the calorimeter block number, i. The blocks located at the calorimeter edge (see Fig. 2) are represented with blue open circles.
pic

The average relative accuracy on the obtained correction factors or calibration coefficients can be defined as:

Δ=1154iδi2, (22)

where the sum runs over the block numbers not belonging to the calorimeter edge. This relative accuracy decreases when the number, Nπ0, of eNeNπ0 events, used in the calibration procedure, increases. For other segmented calorimeters composed of Nb blocks, Δ should rather depend of Nπ0/Nb assuming a uniform π0 distribution in the calorimeter acceptance. Figure 7 shows the evolution of Δ as a function of Nπ0/Nb and demonstrates that less than 1% average accuracy on the calibration coefficients can be obtained if Nπ0/Nb is larger than 100. A fit of Fig. 7 points leads to the following empirical power law:

Figure 7:
Average relative accuracy, Δ, on the calibration coefficients as a function of the ratio between the number of π0 used in the calibration procedure and the number of calorimeter blocks (Nb=208). The dashed line represents Eq. (23) fit to the points.
pic
Δ(%)=5.22 (Nπ0/Nb)0.386. (23)

The number Nπ0/Nb depends experimentally on the integrated luminosity and the differential cross section of the eNeNπ0 process in the studied kinematics. For example, the average π0 production rate for the kinematics of Tab. 1 is Nπ0/Nb90 per day. A compromise on the calibration frequency during the experiment should then be found taking into account the drift rate of the block gains. For the Jefferson Lab Hall A DVCS experiments, a daily monitoring of the calorimeter calibration coefficients is achieved at 1% accuracy level with this calibration method [26].

In some extreme cases when some initial calibration coefficients are different from the real ones by a considerable amount, iterations of the different steps discussed in Sect. 3 could be necessary. This is mainly due experimentally to an identification issue of the eNeNπ0 events because of a possible contamination of the selected yield (Eq. (13)) by DIS events. A more selective Mx2 cut can then be applied in the first iteration to minimize this contamination. The calibration coefficients obtained after the first iteration are used as initial calibration coefficients for the second iteration and so on. A convergence of the calibration coefficients is generally obtained after the second or the third iteration.

5 Conclusion

The present work discussed an energy calibration technique of laterally segmented electromagnetic calorimeters where the two photons of the π0 decay produced by the eNeNπ0 reaction are detected. The calibration procedure is based on a χ2 minimization between the reconstructed π0 energy, from the calorimeter block data, and its expected energy. This energy is calculated for each event from the π0 momentum direction, exploiting the good spatial resolution of the calorimeter. Contrary to other standard calibration methods, this technique allows us to have a set of calibration coefficients optimized for the studied particles under real experimental conditions. The average relative accuracy on these coefficients is less than 1% if the number of detected π0 events relative to the number of calorimeter blocks is larger than 100. This technique has been successfully applied to monitor daily the calorimeter calibration of the Jefferson Lab Hall A DVCS experiments at a 1% accuracy level and with a continuous loss of block transparencies. The calibration method can be applied similarly using any other exclusive meson electroproduction reaction if all-photon decays of these mesons are detected in the calorimeter.

References
[1] R-Y Zhu et al., Proceedings of the Tenth International Conference on Calorimetry in Particle Physics, Pasadena, California, USA, 25-29 March 2002.
[2] C. Munoz Camacho, A. Camsonne, M. Mazouz et al. (Jefferson Lab Hall A Collaboration),

Scaling tests of the cross-section for deeply virtual compton scattering

, Phys. Rev. Lett. 97, 262002 (2006). doi: 10.1103/PhysRevLett.97.262002
Baidu ScholarGoogle Scholar
[3] M. Mazouz, A. Camsonne, C. munoz Camacho et al. (Jefferson Lab Hall A Collaboration),

Deeply virtual compton scattering off the neutron

, Phys. Rev. Lett. 99, 242501 (2007). doi: 10.1103/PhysRevLett.99.242501
Baidu ScholarGoogle Scholar
[4] F.E. Maas et al. (A4 Collaboration),

Measurement of strange quark contributions to the nucleon’s form-factors at Q2=0.230 (GeV/c)2

, Phys. Rev. Lett. 93, 022002 (2004). doi: 10.1103/PhysRevLett.93.022002
Baidu ScholarGoogle Scholar
[5] J. Grange et al.,

Muon (g-2) Technical Design Report, FERMILAB-FN-0992-E

, (2015). doi: 10.2172/1251172
Baidu ScholarGoogle Scholar
[6] D.F. Anderson, M. Kobayashi, Y. Yoshimura and C.L. Woody,

Lead Fluoride: An Ultracompact Cherenkov Radiator for EM Calorimetry

, Nucl. Instrum. Meth. A 290, 385 (1990). doi: 10.1016/0168-9002(90)90553-I
Baidu ScholarGoogle Scholar
[7] K.A. Olive et al.,

Review of Particle Physics

, Chin. Phys. C 38, 090001 (2014). doi: 10.1088/1674-1137/38/9/090001
Baidu ScholarGoogle Scholar
[8] A.T. Fienberg et al.,

Studies of an array of PbF<sub>2</sub> Cherenkov crystals with large-area SiPM readout

, Nucl. Instrum. Meth. A 783, 12 (2015). doi: 10.1016/j.nima.2015.02.028
Baidu ScholarGoogle Scholar
[9] I. Balossino et al.,

The HPS electromagnetic calorimeter

, Nucl. Instrum. Meth. A 854, 89 (2017). doi: 10.1016/j.nima.2017.02.065
Baidu ScholarGoogle Scholar
[10] C. Zhang,

FoCal - A high granularity electromagnetic calorimeter for forward direct photon measurements

, Nucl. Instrum. Meth. A 845, 542 (2017). doi: 10.1016/j.nima.2016.06.075
Baidu ScholarGoogle Scholar
[11] S. Baunack et al.,

Real-time calibration of the A4 electromagnetic lead fluoride (PbF<sub>2</sub>) calorimeter

, Nucl. Instrum. Meth. A 640, 58 (2011). doi: 10.1016/j.nima.2011.02.099
Baidu ScholarGoogle Scholar
[12] Y.-F. Xu, J.-H. Chen et al.,

Physics performance of the STAR zero degree calorimeter at relativistic heavy ion collider

, Nucl. Sci. Tech. 27, 126 (2016). doi: 10.1007/s41365-016-0129-z
Baidu ScholarGoogle Scholar
[13] R.T. Jones, M. Kornicer et al.,

A bootstrap method for gain calibration and resolution determination of a lead-glass calorimeter

, Nucl. Instrum. Meth. A 566, 366 (2006). doi: 10.1016/j.nima.2006.07.061
Baidu ScholarGoogle Scholar
[14] M. Mazouz and E. Voutier, Monitoring of the calibration constants of the DVCS calorimeter, JLAB-TN-05-033 (2005).
[15] P. Kozma, R. Bajgarb and P. Kozma,

Radiation resistivity of PbF2 crystals

, Nucl. Instrum. Meth. A 484, 149 (2002). doi: 10.1016/S0168-9002(01)02011-3
Baidu ScholarGoogle Scholar
[16] M. Defurne et al. (Jefferson Lab Hall A Collaboration),

E00-110 experiment at Jefferson Lab Hall A: Deeply virtual Compton scattering off the proton at 6 GeV

, Phys. Rev. C 92, 055202 (2015). doi: 10.1103/PhysRevC.92.055202
Baidu ScholarGoogle Scholar
[17] C. Munoz Camacho, P. Bertin, C. E. Hyde, J. Roche, et al. (Hall A DVCS Collaboration), Separation of Deeply Virtual Photon and π0 Electroproduction Observables of Unpolarized Protons, JLab Experiment E07-007 (2007).
[18] M. Mazouz, A. Camsonne, C. E. Hyde et al. (Hall A DVCS Collaboration), Measurement of the Deeply Virtual Compton Scattering off the Neutron, JLab Experiment E08-025 (2008).
[19] S. Agostinelli et al. (GEANT4 Collaboration),

GEANT4: A Simulation toolkit

, Nucl. Instrum. Meth. A 506, 250 (2003). doi: 10.1016/S0168-9002(03)01368-8
Baidu ScholarGoogle Scholar
[20] F.-X. Girod et al. (CLAS Collaboration),

Measurement of Deeply Virtual Compton Scattering Beam-Spin Asymmetries

, Phys. Rev. Lett. 100, 162002 (2008). doi: 10.1103/PhysRevLett.100.162002
Baidu ScholarGoogle Scholar
[21] S. Chekanov et al. (ZEUS Collaboration),

Measurement of deeply virtual Compton scattering at HERA

, Phys. Lett. B 573, 46 (2003). doi: 10.1016/j.physletb.2003.08.048
Baidu ScholarGoogle Scholar
[22] F. D. Aaron et al. (H1 Collaboration),

Measurement of deeply virtual Compton scattering and its t-dependence at HERA

, Phys. Lett. B 659, 796 (2008). doi: 10.1016/j.physletb.2007.11.093
Baidu ScholarGoogle Scholar
[23] A. Airapetian et al. (HERMES Collaboration),

Beam-helicity asymmetry arising from deeply virtual Compton scattering measured with kinematically complete event reconstruction

, J. High Energ. Phys. 10, 042 (2012). doi: 10.1007/JHEP10(2012)042
Baidu ScholarGoogle Scholar
[24] J. Alcorn et al.,

Basic instrumentation for Hall A at Jefferson Lab

, Nucl. Instrum. Meth. A 522, 294 (2004). doi: 10.1016/j.nima.2003.11.415
Baidu ScholarGoogle Scholar
[25] E. Fuchey, A. Camsonne, C. Munoz, M. Mazouz et al. (Jefferson Lab Hall A Collaboration),

Exclusive Neutral Pion Electroproduction in the Deeply Virtual Regime

, Phys. Rev. C 83, 025201 (2011). doi: 10.1103/PhysRevC.83.025201
Baidu ScholarGoogle Scholar
[26] M. Defurne, M. Mazouz et al. (Jefferson Lab Hall A Collaboration),

Rosenbluth Separation of the π0 Electroproduction Cross Section

, Phys. Rev. Lett. 117, 262001 (2016). doi: 10.1103/PhysRevLett.117.262001
Baidu ScholarGoogle Scholar
[27] G. Grindhammer, M. Rudowicz and S. Peters,

The Fast Simulation of Electromagnetic and Hadronic Showers

, Nucl. Instrum. Meth. A 290, 469 (1990). doi: 10.1016/0168-9002(90)90566-O
Baidu ScholarGoogle Scholar
[28] V. Breton et al.,

Application of neural networks and cellular automata to interpretation of calorimeter data

, Nucl. Inst. and Meth. A 362, 478 (1995). doi: 10.1016/0168-9002(95)00217-0
Baidu ScholarGoogle Scholar
[29] T. C. Awes, F. E. Obenshain, F. Plasil, S. Saini, S. P. Sorensen, et al.,

A simple method of shower localization and identification in laterally segmented calorimeters

, Nucl. Instrum. Meth. A 311, 130 (1992). doi: 10.1016/0168-9002(92)90858-2
Baidu ScholarGoogle Scholar
[30] M. Mazouz, L. Ghedira and E. Voutier,

Determination of shower central position in laterally segmented lead-fluoride electromagnetic calorimeters

, Journal of Instrumentation 11, P07001 (2016). doi: 10.1088/1748-0221/11/07/P07001
Baidu ScholarGoogle Scholar