1 Introduction
The main goal of electromagnetic calorimeters is to measure precisely the energy of detected particles such as photons and electrons. These detectors are important elements in nuclear physics instrumentation and are widely used especially in hadronic and particle physics experiments [1]. Usually, the knowledge of the impact position of detected particles is required leading to a lateral segmentation of these calorimeters. Lead fluoride (PbF2) is commonly employed as a constituent material of such calorimeters [2-5]. Its high density offers a short radiation length (X0=0.93 cm) and a small Molière radius (rM=2.12 cm) leading to compact detector geometries [6, 7]. In practice, the segmentation size is close to rM while the longitudinal calorimeter length is equal to several X0 to ensure a full development of electromagnetic showers in the detector blocks. Each calorimeter block is generally connected to a photomultiplier tube (PMT) to collect the light induced by a shower and an electronic base to shape and amplify the PMT signal. The amplitude of the final signal is then related to the energy released by the particle in the considered block via a calibration coefficient.
A great deal of effort is still underway to optimize the performance of electromagnetic calorimeters and in particular, their energy resolution [8-10]. In addition to the design optimization, many energy calibration techniques are still investigated to improve the reconstructed energy of detected particles [11-14]. Indeed, the energy calibration is sensitive to many factors, such as the hadronic and electromagnetic background during the experiment and the gain drift of blocks due to an increasing loss of their transparency when exposed to high radiation rates and accumulated doses [15, 16]. The choice of the energy calibration method is then crucial to ensure a frequent and proper monitoring of the calibration coefficients of each calorimeter block keeping thus an acceptable energy resolution. In many experiments, dedicated calibration runs have to be taken in order to send particles of known energy in the calorimeter. Other methods consist to calibrate the detector with cosmic rays or with a cluster of LEDs of known light intensity placed in front of the calorimeter blocks [16]. All these standard methods have the disadvantage to change momentarily the experimental setting or trigger and to decrease the amount of time dedicated to physics runs. In addition, many parameters, such as the nature of particles used in the calibration procedure, their energies and incident angles, and the experimental background, could be different relative to the physics run conditions. The obtained calibration coefficients are then not necessarily optimized for the particles detected during physics runs. Finally, the calibration provided by these methods at a given time may not still be valid afterwards if the calibration coefficients are time dependent.
This paper presents an energy calibration method based on the detection of π0 mesons in the calorimeter. This method is performed to monitor daily the calibration of the electromagnetic calorimeter of the Jefferson Lab Hall A DVCS experiments [2, 3, 17, 18] and can be applied in similar experiments where two-photon decays of π0 can be detected in a laterally segmented calorimeter. It does not require specific runs since the calibration data are taken simultaneously with the primary experimental data at the same conditions. This method will be exposed and tested on the basis of a calorimeter simulation using the GEANT4 toolkit [19]. The relative accuracy on the obtained calibration coefficients will finally be discussed in terms of the number of detected π0.
2 Simulation of an experimental case
The main goal of the Jefferson Lab Hall A DVCS experiments is to study the Deeply Virtual Compton Scattering (DVCS) process on the nucleon eN → eNγ using a few GeV electron beams impinging on a liquid hydrogen (LH2) or deuterium (LD2) target [2, 3, 17, 18]. Many similar experiments in the world are concerned with this increasingly timely topic of the characterization of the nucleon structure with the DVCS process [20-23]. As shown in Fig. 1, the scattered electron is detected in a high resolution spectrometer (HRS) which determines accurately its momentum and angles as well as the reaction vertex coordinates in the target [24]. The emitted photon is detected in an electromagnetic calorimeter consisting of a 16 × 13 matrix of 3 × 3 × 18.6 cm3 PbF2 blocks placed at 1.1 m from the target. This short distance combined with the high luminosity of the experiment (∼1037 cm-2s-1) leads to an important background rate in the calorimeter as well as radiation damage near the front face of blocks. The energy calibration of the calorimeter is of direct relevance since it affects directly the eN → eNγ event identification based on a study of the missing mass squared:
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F001.jpg)
where k, k′, p, and
are respectively the 4-vectors of the incident electron, the scattered electron, the initial nucleon at rest and the emitted photon. Experimentally, k, k′, and p are known accurately and the resolution of
is dominated by the calorimeter energy resolution [16]. The photon 4-vector is determined for each event, j, from the energies deposited in the calorimeter blocks:
where Ai(j) is the output signal amplitude of block i, and
The correction factors, ϵi, and thus the new calibration coefficients, Ci, must then be known accurately in order to get the real deposited energies, Ei(j):
The goal of the calibration method, discussed hereafter, is to determine the ϵi for each block, i. This method is based on the study of π0 electroproduction events eN → eNπ0 → eNγγ where the two photons coming from the π0 decay are detected in the calorimeter. This reaction is very common in lepton-hadron scattering and is usually present in the data of DVCS experiments. The π0 energy is almost equal to the DVCS photon energy because of the kinematic similarity between these two reactions, and it is then well adapted for the calorimeter calibration.
To demonstrate the validity of the calibration method, a GEANT4 simulation of the experimental setup is performed and N=2.106 eN → eNπ0 events are generated following the kinematics presented in Table 1. This particular kinematics corresponds to one setting of the Hall A DVCS experiments [26]. The electromagnetic showers created by the π0 photons in the calorimeter are fully simulated [27]. The generation and tracking of Čerenkov photons induced by the showers in the PbF2 blocks are not considered in this study because of unrealistic computing times and strong sensitivity to exact optical properties of crystals and their wrapping surfaces. However, one can attribute an average number of Nph=330 photo-electrons per GeV deposit collected by the PMT of each block to be coherent with the reported experimental energy resolution σ(E)/E=3.1% [26]. This number is deduced from a
k0(GeV) | θkk′(deg) | θkq′(deg) | ϕkq′(deg) | ||
---|---|---|---|---|---|
4.455 | [1.78, 1.96] | [25, 28] | 2.4 | [9, 28] | [-50, 50] |
Finally, real radiative effects, where the incident or the scattered electron emits a real photon, are taken into account in the simulation following the procedure described in [16].
To mimic the uncertainty on the experimental calibration coefficients, the energies, Ei, deposited in the calorimeter blocks are multiplied by random numbers, Ki, varying uniformly between 1-Kmax and 1, where Kmax=40%, and become equal to:
The calibration method consists then to determine, from
3 Calibration method
The calibration method is based on a χ2 minimization between the measured π0 energy,
3.1 Particle 4-vector reconstruction
The reconstructed π0 energy in the simulation is the sum of the two shower energies created by the two photons of the π0 decay. Experimentally, one can have multiple showers in the calorimeter coming from other meson decays, accidentals or any additional final state particles. In this case, a clustering algorithm based on a cellular automata is used to determine the number of showers and to separate the different cluster contributions [28]. The energy of a shower is then the sum of the energies deposited in the Nclus blocks belonging to the corresponding cluster:
Figure 2 shows an example of a π0 event in the calorimeter where the corresponding clusters are indicated in red. The transverse coordinates (xc,yc) of a shower centroid are determined with a center of gravity based method from the coordinates (xi,yi) of the blocks belonging to the corresponding cluster [29]:
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F002.jpg)
with a similar equation for yc. The weight, wi, is given by:
where W0 is a free dimensionless parameter optimized for the considered energies [30]. The showers whose centroid coordinates belong to a block of the calorimeter edge (see Fig. 2) are excluded from the calibration procedure. Indeed, the energy of a particle impinging these particular blocks is not well reconstructed because of the shower energy leakage near the calorimeter edge.
The knowledge of the shower centroid coordinates, with a 3 mm accuracy [26], and the interaction vertex coordinates allows us to reconstruct the momentum direction of the particle creating the shower and thus its 4-vector
3.2 eN → eNπ0 reaction identification
In the simulated data, only eN → eNπ0 → eNγγ events are generated. Experimentally, we have to identify this reaction by selecting only 2-cluster events and by computing the invariant mass of the two reconstructed particles in the calorimeter:
where
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F003.jpg)
where σMinv is the resolution of the reconstructed Minv variable.
The second step of the eN → eNπ0 reaction identification consists to compute for each π0 event the missing mass squared defined by:
which must be equal to the nucleon mass squared
GeV2 within detector resolution if the considered event corresponds to a eN → eNπ0 reaction and if the calorimeter is well calibrated. Figure 4 shows the simulated
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F004.jpg)
where
3.3 Expected π0 energy calculation
For each selected eN → eNπ0 event, the reconstructed π0 energy writes:
where di=1 if the block, i, belongs to one of the two clusters created by the π0 and di=0 otherwise. This reconstructed energy leads to particular values of
where θ is the angle between the momenta of the virtual photon and the π0. As mentioned above, θ is well defined experimentally even if biased calibration coefficients are employed. The physical solution of Eq. (15) is given by:
where:
Figure 5 shows the histogram of the relative difference between
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F005.jpg)
3.4 χ2 minimization
For each eN → eNπ0 event, j,
where
The minimization of Eq. (18) writes:
and yields to the following linear set of equations:
The correction factors, ϵi, are then obtained by inverting the 208 × 208 matrix,
4 Results and discussion
Figure 6 shows the relative difference δi=ϵiKi-1 between the obtained correction factors, ϵi, and the smearing factors, 1/Ki, as a function of the calorimeter block number. Figure 6 proves that the present calibration method succeeds to find the real calibration coefficients for all the calorimeter blocks, within 1% deviation, except for those located at the calorimeter edge. Since the clusters centered around the latter blocks are excluded from the calibration procedure, the α-matrix diagonal elements corresponding to these blocks are filled with low deposited energies and become sensitive to energy fluctuations of shower tails. The corresponding calibration coefficients are then slightly overestimated. The energy resolution improvement is shown in Fig. 5 where a clear decrease of the
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F006.jpg)
The average relative accuracy on the obtained correction factors or calibration coefficients can be defined as:
where the sum runs over the block numbers not belonging to the calorimeter edge. This relative accuracy decreases when the number,
-201711/1001-8042-28-11-002/alternativeImage/1001-8042-28-11-002-F007.jpg)
The number Nπ0/Nb depends experimentally on the integrated luminosity and the differential cross section of the eN → eNπ0 process in the studied kinematics. For example, the average π0 production rate for the kinematics of Tab. 1 is
In some extreme cases when some initial calibration coefficients are different from the real ones by a considerable amount, iterations of the different steps discussed in Sect. 3 could be necessary. This is mainly due experimentally to an identification issue of the eN → eNπ0 events because of a possible contamination of the selected yield (Eq. (13)) by DIS events. A more selective
5 Conclusion
The present work discussed an energy calibration technique of laterally segmented electromagnetic calorimeters where the two photons of the π0 decay produced by the eN → eNπ0 reaction are detected. The calibration procedure is based on a χ2 minimization between the reconstructed π0 energy, from the calorimeter block data, and its expected energy. This energy is calculated for each event from the π0 momentum direction, exploiting the good spatial resolution of the calorimeter. Contrary to other standard calibration methods, this technique allows us to have a set of calibration coefficients optimized for the studied particles under real experimental conditions. The average relative accuracy on these coefficients is less than 1% if the number of detected π0 events relative to the number of calorimeter blocks is larger than 100. This technique has been successfully applied to monitor daily the calorimeter calibration of the Jefferson Lab Hall A DVCS experiments at a 1% accuracy level and with a continuous loss of block transparencies. The calibration method can be applied similarly using any other exclusive meson electroproduction reaction if all-photon decays of these mesons are detected in the calorimeter.
Scaling tests of the cross-section for deeply virtual compton scattering
, Phys. Rev. Lett. 97, 262002 (2006). doi: 10.1103/PhysRevLett.97.262002Deeply virtual compton scattering off the neutron
, Phys. Rev. Lett. 99, 242501 (2007). doi: 10.1103/PhysRevLett.99.242501Measurement of strange quark contributions to the nucleon’s form-factors at Q2=0.230 (GeV/c)2
, Phys. Rev. Lett. 93, 022002 (2004). doi: 10.1103/PhysRevLett.93.022002Muon (g-2) Technical Design Report, FERMILAB-FN-0992-E
, (2015). doi: 10.2172/1251172Lead Fluoride: An Ultracompact Cherenkov Radiator for EM Calorimetry
, Nucl. Instrum. Meth. A 290, 385 (1990). doi: 10.1016/0168-9002(90)90553-IReview of Particle Physics
, Chin. Phys. C 38, 090001 (2014). doi: 10.1088/1674-1137/38/9/090001Studies of an array of PbF<sub>2</sub> Cherenkov crystals with large-area SiPM readout
, Nucl. Instrum. Meth. A 783, 12 (2015). doi: 10.1016/j.nima.2015.02.028The HPS electromagnetic calorimeter
, Nucl. Instrum. Meth. A 854, 89 (2017). doi: 10.1016/j.nima.2017.02.065FoCal - A high granularity electromagnetic calorimeter for forward direct photon measurements
, Nucl. Instrum. Meth. A 845, 542 (2017). doi: 10.1016/j.nima.2016.06.075Real-time calibration of the A4 electromagnetic lead fluoride (PbF<sub>2</sub>) calorimeter
, Nucl. Instrum. Meth. A 640, 58 (2011). doi: 10.1016/j.nima.2011.02.099Physics performance of the STAR zero degree calorimeter at relativistic heavy ion collider
, Nucl. Sci. Tech. 27, 126 (2016). doi: 10.1007/s41365-016-0129-zA bootstrap method for gain calibration and resolution determination of a lead-glass calorimeter
, Nucl. Instrum. Meth. A 566, 366 (2006). doi: 10.1016/j.nima.2006.07.061Radiation resistivity of PbF2 crystals
, Nucl. Instrum. Meth. A 484, 149 (2002). doi: 10.1016/S0168-9002(01)02011-3E00-110 experiment at Jefferson Lab Hall A: Deeply virtual Compton scattering off the proton at 6 GeV
, Phys. Rev. C 92, 055202 (2015). doi: 10.1103/PhysRevC.92.055202GEANT4: A Simulation toolkit
, Nucl. Instrum. Meth. A 506, 250 (2003). doi: 10.1016/S0168-9002(03)01368-8Measurement of Deeply Virtual Compton Scattering Beam-Spin Asymmetries
, Phys. Rev. Lett. 100, 162002 (2008). doi: 10.1103/PhysRevLett.100.162002Measurement of deeply virtual Compton scattering at HERA
, Phys. Lett. B 573, 46 (2003). doi: 10.1016/j.physletb.2003.08.048Measurement of deeply virtual Compton scattering and its t-dependence at HERA
, Phys. Lett. B 659, 796 (2008). doi: 10.1016/j.physletb.2007.11.093Beam-helicity asymmetry arising from deeply virtual Compton scattering measured with kinematically complete event reconstruction
, J. High Energ. Phys. 10, 042 (2012). doi: 10.1007/JHEP10(2012)042Basic instrumentation for Hall A at Jefferson Lab
, Nucl. Instrum. Meth. A 522, 294 (2004). doi: 10.1016/j.nima.2003.11.415Exclusive Neutral Pion Electroproduction in the Deeply Virtual Regime
, Phys. Rev. C 83, 025201 (2011). doi: 10.1103/PhysRevC.83.025201Rosenbluth Separation of the π0 Electroproduction Cross Section
, Phys. Rev. Lett. 117, 262001 (2016). doi: 10.1103/PhysRevLett.117.262001The Fast Simulation of Electromagnetic and Hadronic Showers
, Nucl. Instrum. Meth. A 290, 469 (1990). doi: 10.1016/0168-9002(90)90566-OApplication of neural networks and cellular automata to interpretation of calorimeter data
, Nucl. Inst. and Meth. A 362, 478 (1995). doi: 10.1016/0168-9002(95)00217-0A simple method of shower localization and identification in laterally segmented calorimeters
, Nucl. Instrum. Meth. A 311, 130 (1992). doi: 10.1016/0168-9002(92)90858-2Determination of shower central position in laterally segmented lead-fluoride electromagnetic calorimeters
, Journal of Instrumentation 11, P07001 (2016). doi: 10.1088/1748-0221/11/07/P07001