Nuclear Science and TechniquesVol.32, No.5Article number 45Published in print 01 May 2021Available online 08 May 2021
38800

The nitrogen doping/infusion of 650 MHz cavities for the circular electron positron collider (CEPC) is investigated in this study. Two 650 MHz 1-cell cavities are first treated via buffered chemical polishing (BCP), followed by nitrogen doping. A "2/6" condition is adopted, similar to that for 1.3 GHz cavities of Linear Coherent Light Source II. The quality factor of both cavities improved to 7 × 1010 in low fields, i.e., higher than that obtained from the baseline test. One 650 MHz two-cell cavity is nitrogen infused at 165 °C for 48 h with a BCP surface base. The intrinsic quality factor (Q0) reached 6 × 1010 at 22 MV/m in the vertical test, and the maximum gradient is 25 MV/m, which exceeds the specification of the CEPC (4 × 1010 at 22 MV/m).

Video Abstract

Superconducting radio frequency cavityNitrogen dopingNitrogen infusionQuality factorAccelerating gradient

1 Introduction

Currently, superconducting radio frequency (SRF) cavities with a high quality factor (Q) are adopted in large accelerators worldwide [1, 2]. The Linear Coherent Light Source II (LCLS-II) in America uses a specific condition for nitrogen doping to increase Q0 of 1.3 GHz 9-cell cavities. The average Q0 of 280 nitrogen-doped 9-cell cavities reached 3.1 × 1010 at 16 MV/m, which exceeded the specification of 2.7 × 1010. Owing to the excellent results of nitrogen doping, the LCLS-II High Energy program has continued to adopt nitrogen-doped cavities [3], which can be operated at 20.8 MV/m with Q0 of 2.7 × 1010. In China, the construction of the Shanghai high repetition rate XFEL and extreme light facility (SHINE) began in April 2018, for which six hundred 1.3-GHz 9-cell cavities were required to obtain an 8 GeV electron beam. Moreover, the nitrogen doping of 1.3 GHz cavities was conducted in several institutes including the Peking University in China [4-6], and preliminary improvement was observed.

The circular electron positron collider (CEPC) is a high-energy collider at the Higgs factory that can operate in the W and Z modes [7]. The SRF system of the CEPC is extremely challenging because of the associated wide beam energy and current [8]. Two hundred and forty 650 MHz two-cell cavities exist in the collider, and ninety-six 1.3 GHz nine-cell cavities exist in the booster of the CEPC tunnel, which is located 100 m underground. The 650 MHz two-cell cavity will be operated in continuous wave mode with high Q, rendering the cryogenics system economical. The Q0 was 4 × 1010 at a accelerating gradient (Eacc) of 22 MV/m for the vertical acceptance test, and 1.5 × 1010 at 19.7 MV/m for the operation [9]. This specification is critical for the SRF cavity of a circular collider, which has a more constrained environment than linacs (such as LCLS-II and SHINE).

Compared with the nitrogen doping/infusion of 1.3 GHz cavities, that of 650 MHz cavities are investigated less. Nitrogen doping of 650 MHz cavities was conducted at Fermi National Accelerator Laboratory (FNAL) [10]; it reached ~7 × 1010 at 22 MV/m and recorded the highest Q worldwide. Nitrogen doping/infusion of 650 MHz cavities for the CEPC was conducted in collaboration with the Institute of High Energy Physics and Peking University. Nitrogen doping/infusion studies worldwide are primarily based on electro-polishing (EP), which can achieve a higher Eacc than BCP. However, EP facilities are complicated and expensive, and EP conditions for 650 MHz cavities does not exist China. By contrast, BCP is considerably simpler and more convenient process than EP. Therefore, nitrogen doping/infusion based on BCP has been adopted and proven to be effective. Two 650 MHz one-cell cavities received nitrogen doping, and one 650 MHz two-cell cavity received nitrogen infusion. The nitrogen doping/infusion treatment and vertical test of 650 MHz superconducting cavities are presented herein.

2 Nitrogen doping of 650 MHz 1-cell cavities

In this study, nitrogen doping was conducted on two 650 MHz 1-cell cavities (650S1, 650S2) made of fine-grain Nb. The primary parameters of the Nb used are listed in Table 1.

Table 1.
Primary parameters of fine-grain niobium
Parameter Value
RRR 350–400
ASTM size 6.0–5.5
HV10 ~47.8
Yield strength (MPa) ~64
Tensile ultimate strength (MPa) ~170
Show more

BCP was performed prior to nitrogen doping/infusion. The polishing was performed at the vertical BCP facility of Ningxia Orient Superconductor Technology Co., Ltd. (OSTEC), as shown in Fig. 1, which has been used for all types of SRF cavities. Both cavities received bulk BCP (150 μm), annealing at 750 °C for 3 h, and light BCP (30 μm). Subsequently, they were assembled in a clean room and subjected to low-temperature baking (120 °C for 48 h). The vertical test results (baseline) of the two cavities are shown in Fig. 2. The maximum gradients of 650S1 and 650S2 reached 19.3 and 24.7 MV/m, respectively. Field emission, which might have resulted from contamination during the process, was observed at low fields in both cavities.

Fig. 1
(Color online) Setup for BCP of 650 MHz 1-cell cavity
pic
Fig. 2
(Color online) Vertical test results of 650S1 and 650S2 (at 2.0 K)
pic

After performing the vertical test on the BCP baseline, both cavities received another light BCP (30 μm) to eliminate contamination on the inner surface. The two cavities were nitrogen doped based on a “2/6” recipe, similar to that for the 1.3 GHz cavity of LCLS-II. First, the furnace was pumped until the pressure was less than 2 × 10-4 Pa. Second, the furnace was heated by ramping its temperature to 800 °C for 3 h. Third, the furnace was maintained at 800 °C for 3 h for annealing. Finally, nitrogen gas was injected into the furnace for 2 min at ~3.5 Pa, followed by annealing at 800 °C for 6 min. Finally, the furnace was cooled to ~50 °C under vacuum.

Because it is extremely difficult to achieve uniform removal via BCP, neither cavity underwent etching after nitrogen doping. The vertical test results at 2.0 K are shown in Fig. 2. The Q0 of both cavities exceeded 7 × 1010 at low fields, which was higher than that obtained via the BCP baseline test. The Q-slope of 650S1 was evident above 10 MV/m; cavities 650S1 and 650S2 quenched at 16.6 and 8.8 MV/m, respectively. To remove the niobium nitride phase layer, the inner surface of the cavity after nitrogen doping must be removed lightly, and this is generally achieved via EP. Because no EP device exists, it is more suitable to improve the Q and Eacc of 650 MHz cavities via nitrogen infusion, which requires no EP of the inner surface after infusion [11].

3 Nitrogen infusion of 650 MHz 2-cell cavity

Nitrogen infusion was performed on a 650 MHz 2-cell cavity. One 650 MHz 2-cell cavity was designed, manufactured, post-processed, nitrogen-infused, and tested.

Design of 650 MHz 2-cell cavity

The radio frequency (RF) design of a 650 MHz 2-cell cavity (650D1) was optimized to minimize Bpeak/Eacc and Epeak/Eacc. Meanwhile, the R/Q and cell coupling (k) should be maximized. The final RF parameters are presented in Table 2[8].

Table 2.
Primary parameters of 650 MHz 2-cell cavity
Parameter Value
Beam aperture (mm) 156
Bpeak/Eacc (mT/(MV/m)) 4.2
Epeak/Eacc 2.4
R/Q (Ω) 213
G (Ω) 284
k 2.86%
Show more

The two important mechanical parameters of the 650 MHz 2-cell cavity were the Lorentz force detuning coefficient (KL) and frequency sensitivity to pressure (df/dp). The geometry and dimensions of the cavity are shown in Fig. 3. Two cavity wall thicknesses (3 and 4 mm) were compared, as summarized in Table 3. Finally, the 4 mm thickness was adopted because of the lower KL and df/dp.

Table 3.
Mechanical parameters of 650 MHz 2-cell cavity
Parameter Condition Units 3 mm 4 mm
Max stress (under 2 atm) Beam ports fixed MPa 24.7 20
df/dp Beam ports free Hz/mbar -256 -185
Beam ports fixed Hz/mbar -6.4 0.95
Tuning sensitivity / kHz/mm 558 313
Tuning stiffness / kN/mm 3.7 6.4
Tuning range (293 K) / kHz 705 353
KL Beam ports free Hz/(MV/m)2 -8.2 -7.4
Beam ports fixed Hz/(MV/m)2 -0.88 -0.6
Show more
Fig. 3
(Color online) Geometry and dimensions of 650 MHz 2-cell cavity
pic
4 BCP baseline test of 650 MHz 2-cell cavity

The 650 MHz two-cell cavity (650D1) was manufactured with fine-grain Nb, same as the 650 MHz one-cell cavity. Before nitrogen infusion, the cavity received bulk BCP (200 μm), annealing at 800 °C for 3 h, and light BCP (30 μm); subsequently, it was assembled in a clean room. A vertical test was performed after these treatments, as shown in Fig. 4. The vertical test system comprised double magnetic shields, and the residual magnetic field around the cavity was less than 5 mGs. No compensation coils were used during the vertical test. Fast cooling was performed at a speed of ~5 K/min. The vertical test results of the BCP baseline at 2.0 K are shown in Fig. 5. The Q0 of 650D1 reached 3.3 × 1010 at 22 MV/m. Finally, the cavity quenched at 26.5 MV/m with a Q0 of 1.4 × 1010. Strong multipacting occurred when the gradient was 11–18 MV/m; however, it can be avoided by increasing the input power. The simulation shows that multipacting can occur between 8 and 20 MV/m, which is consistent with the vertical test [9]. The radiation resulting from field emission began to increase at 18.8 MV/m, indicating contamination on the inner surface of the cavity.

Fig. 4
(Color online) Vertical test of 650 MHz 2-cell cavity
pic
Fig. 5
(Color online) Vertical test results of BCP baseline (at 2.0 K)
pic

5 Nitrogen infusion of 650 MHz 2-cell cavity

The 650 MHz two-cell cavity (650D1) received another light BCP (30 μm) and high-pressure rinsing (HPR) to remove possible contamination and reset the inner surface after the BCP baseline test. Subsequently, the cavity was transferred to a furnace for nitrogen infusion [12]. All six flanges were covered with clean Nb foil to avoid contamination and dust during nitrogen infusion, as shown in Fig. 6.

Fig. 6
(Color online) Nitrogen infusion of 650 MHz 2-cell cavity
pic

The “165 °C 48 h” condition, which may achieve a higher Q than the “120 °C 48 h” condition, was adopted in nitrogen infusion[11]. The pressure and temperature during nitrogen infusion are shown in Fig. 7. First, the furnace was heated by ramping the temperature to 800 °C and maintaining it at that temperature for 3 h for annealing. Second, the furnace was cooled to 165 °C and maintained for half a minute. Third, nitrogen gas was injected into the furnace for 48 h at ~3.5 Pa. Finally, the furnace was cooled to ~50 °C under vacuum.

Fig. 7
(Color online)Temperature and pressure during nitrogen infusion
pic

The inner surface of the nitrogen-infused 650D1 was analyzed using a new inspection camera. A typical image of the equator is shown in Fig. 8 (a); it is considerably rougher compared with that observed after EP, as shown in Fig. 8 (b). No defects were observed on the entire inner surface of the 650D1.

Fig. 8
(Color online) Inspection of cavity equator. (a) 650D1 (nitrogen infused); (b) 1.3 GHz cavity (after EP)
pic

After nitrogen infusion, 650D1 underwent HPR and was assembled in a clean room for vertical testing. The vertical test results of 650D1 at 2.0 K are shown in Fig. 9. In the first vertical test, Q0 reached 7.2 × 1010 at 9 MV/m, which was 40% higher than the BCP baseline. However, Q0 decreased rapidly above 9 MV/m because of strong field emissions induced by contamination during clean room assembly and transfer. Finally, the gradient reached 13.6 MV/m with a Q0 of 5.2 × 109. To eliminate the field emission, 650D1 was subjected to HPR and re-assembled, followed by a second vertical test. At this time, the cavity achieved a higher Q0 than the BCP baseline over the entire range of gradient. The Q0 was 6 × 1010 at 22 MV/m, which exceeded the vertical test specification (4 × 1010 at 22 MV/m) for the CEPC. This result is similar to that recorded by FNAL (7 × 1010 at 22 MV/m), which was based on EP. Therefore, nitrogen infusion based on BCP is applicable in medium fields (~ 20 MV/m). No field emission was observed until the quench gradient reached 25 MV/m with a Q0 of 3.3 × 1010. Multipacting was avoided by increasing the input power. To eliminate the multipacting effect, a new 650 MHz 2-cell cavity has been designed and proven to be effective [13].

Fig. 9
(Color online) Vertical test results of 650 D1 (at 2.0 K)
pic

6 Conclusion

Nitrogen doping and infusion studies were conducted on 650 MHz superconducting cavities for CEPC. Owing to the lack of EP conditions for the 650 MHz cavity and to identify other possible methods, a BCP surface baseline was adopted, and no BCP or EP was conducted after nitrogen doping/infusion.

Owing to nitrogen doping, the Q0 of the 650 MHz 1-cell cavity exceeded 7 × 1010 at low fields, which was higher than the 5–6 × 1010 obtained via the BCP baseline test. The 650 MHz 2-cell cavity attained a Q0 of 6 × 1010 at 22 MV/m after nitrogen infusion based on the “165 °C 48 h” condition. It was evident that nitrogen infusion increased Q0. Both Q0 and Eacc exceeded the specifications of the CEPC.

References
[1] H.T. Hou, J.F. Chen, Z.Y. Ma et al.,

Prototypes fabrication of 1.3 GHz superconducting RF components for SHINE project

, Paper Presented at the Proceeding of SRF2019 (Dresden, Germany, 2019). https://doi.org/10.18429/JACoW-SRF2019-MOP049
Baidu ScholarGoogle Scholar
[2] D. Gonnella, S. Aderhold, A. Burrill et al.,

Industrialization of the nitrogen-doping preparation for SRF cavities for LCLS-II

. Nucl. Instrum. Methods Phys. Research Sect. A: Accel. Spectrom. Detect. Assoc. Equip. 883, 143-150 (2018). doi: 10.1016/j.nima.2017.11.047
Baidu ScholarGoogle Scholar
[3] D. Gonnella, S. Aderhold, D. Bafia et al.,

The LCLS-II HE high Q and gradient R&D program

, Paper Presented at the Proceeding of SRF2019 (Dresden, Germany, 2019). https://doi.org/10.18429/JACoW-SRF2019-MOP045
Baidu ScholarGoogle Scholar
[4] S. Chen, J.-K. Hao, L. Lin et al.,

Successful nitrogen doping of 1.3GHz single cell superconducting radio-frequency cavities

. Chin. Phys. Lett. 35, 037401 (2018). doi: 10.1088/0256-307X/35/3/037401
Baidu ScholarGoogle Scholar
[5] C. Dong, P. Sha, B.Q. Liu et al.,

Nitrogen doping experiment of 1.3 GHz superconducting cavity

. High Power Laser Particle Beams 32, 045105 (2020). doi: 10.11884/HPLPB202032.190141
Baidu ScholarGoogle Scholar
[6] B. Liu, P. Sha, C. Dong et al.,

Nitrogen doping with dual-vacuum furnace at IHEP

, Nucl. Instrum. Methods Phys. Research Sect. A: Accel. Spectrom. Detect. Assoc. Equip. 993, 165080 (2021). doi: 10.1016/j.nima.2021.165080
Baidu ScholarGoogle Scholar
[7] The CEPC-SPPC Study Group,

CEPC Conceptual Design Report. IHEP-CEPC-CDR-2018-09

. arXiv:1809.00285
Baidu ScholarGoogle Scholar
[8] J.Y. Zhai, D.J. Gong, H.J. Zheng et al.,

Design of CEPC superconducting RF system

. Int. J. Modern Phys. A 34, 1940006 (2019). doi: 10.1142/S0217751X19400062
Baidu ScholarGoogle Scholar
[9] H.J. Zheng, J. Gao, J.Y. Zhai et al.,

RF design of 650-MHz 2-cell cavity for CEPC

. Nucl. Sci. Tech. 30, 155 (2019). doi: 10.1007/s41365-019-0671-6
Baidu ScholarGoogle Scholar
[10] K.E. McGee,

New results on N-doping applied to 5-cell elliptical β= 0.65 644 MHz cavities

, Tesla Technology Collaboration Meeting 2021. https://indico.desy.de/event/27572/contributions/94301/attachments/63521/77637/McGee_NAL644MHz_TTC_19Jan2021_v2-1.pdf
Baidu ScholarGoogle Scholar
[11] A. Grassellino, A. Romanenko, Y. Trenikhina et al.,

Unprecedented quality factors at accelerating gradients up to 45MVm−1 in niobium superconducting resonators via low temperature nitrogen infusion

. Supercond. Sci. Technol. 30, 094004 (2017). doi: 10.1088/1361-6668/aa7afe
Baidu ScholarGoogle Scholar
[12] S. Chen,

Nitrogen doping study of improving intrinsic quality factor for superconducting radio-frequency cavities, Dissertation

(Peking University, 2020).
Baidu ScholarGoogle Scholar
[13] H. Zheng, P. Sha, J. Zhai et al.,

Development and vertical tests of 650 MHz 2-cell superconducting cavities with higher order mode couplers

. Nucl. Instrum. Methods Phys. Research Sect. A: Accel. Spectrom. Detect. Assoc. Equip. 995, 165093 (2021). doi: 10.1016/j.nima.2021.165093
Baidu ScholarGoogle Scholar