Introduction
Mass is a fundamental property of atomic nuclei and plays a crucial role in studies of nuclear structure and nuclear astrophysics [1, 2]. Significant efforts has been devoted to developing mass-measurement technologies, particularly for short-lived nuclides [3, 4]. Isochronous mass spectrometry (IMS) [5, 6] on heavy-ion storage rings has proven to be a powerful tool, providing numerous new mass values for short-lived nuclides [7-12].
In conventional IMS equipped with a single time-of-flight (TOF) detector, nuclear masses are determined by measuring the revolution times (T) of ions stored in the ring. Given that a storage ring operates within a certain acceptance of magnetic rigidity Bρ (or momentum p), stored ions of the same species exhibit a velocity spread. Although ions with higher velocities are expected to circulate in the ring along larger orbital paths and vice versa, the revolution time of ions in different orbits will still differ if the velocity variation is not perfectly compensated by variations in orbital length. This velocity spread results in a broad distribution of T. The dependence of T on velocity can be effectively removed for ions that satisfy isochronous conditions [5]. However, for most ions, the velocity dependence of T leads to significant deterioration in mass precision and may introduce systematic errors [13] in the measured masses.
To reduce the effect of velocity spread on the T distribution and enhance the performance of IMS, several methods have been proposed. One such method, known as Bρ-tagging IMS [14], employs a mechanical slit at the dispersive focal plane of the beam line or in the ring to restrict the Bρ acceptance, thereby narrowing the T distribution [14, 15]. However, this approach reduces transmission efficiency and is unfavorable for mass measurements of exotic nuclei with extremely low production yields. Another method, referred to as Bρ-defined IMS or Bρ-IMS [16, 13], employs two TOF detectors to measure the velocity of ions directly and determine Bρ values [17]. With the measured Bρ and velocity, the mass-to-charge ratio m/q of the ions is calculated, resulting in improved mass-resolving power [16, 13]. However, compared to conventional IMS utilizing only one TOF detector, Bρ-IMS introduces double the energy loss for ions during each revolution. Since energy loss increases with proton number, heavier ions experience more pronounced losses. Repeated traversal through the carbon foils of the two detectors can cause the ion momentum to exceed the storage ring’s acceptance range, leading to ion loss and a significant reduction in storage time, complicating the precise determination of ion velocity and Bρ. Additionally, the installation of two TOF detectors in a straight section of the ring presents practical challenges, such as the precise determination of the distance between detectors and managing time delay differences [18]. Furthermore, a previous study revealed that using a single oscilloscope to record flight time signals is necessary to avoid jitter at the start times of two oscilloscopes [19]. However, using only one oscilloscope introduces other issues, such as signal attenuation due to longer cables and excessive writing times caused by the large volume of data being recorded.
In this paper, we introduce a new method called Tune-IMS to improve conventional IMS. The motion of a charged particle with nonzero emittance in a storage ring exhibits betatron oscillations around a closed orbit [20, 21]. The number of oscillations per revolution, known as the tune (Q) value, is directly related to the ions’ Bρ and their revolution time T [21]. A previous study [22] proposed a method for measuring Q using a single TOF detector. By using the correlation between T and Q for each ion species, the measured T can be transformed into a new revolution time, TQ, corresponding to a reference closed orbit with a fixed Q. As a result, for ions that do not satisfy the isochronous condition, the new TQ distribution is narrower, leading to improved mass-resolving power.
In this study, we briefly present the principle of this method and apply it to an IMS experiment [23, 24] conducted at the experimental cooling storage ring (CSRe) of the Heavy Ion Research Facility in Lanzhou (HIRFL-CSR). The re-determined mass values of 63Ge, 65As, 67Se and 71Kr with higher precision were compared with those from a recent Bρ-IMS experiment [25, 26]. The Coulomb displacement energies (CDEs) extracted from these new mass values were also compared with theoretical estimates.
Principle of Tune-IMS
Under ideal conditions in a stable magnetic field, the magnetic rigidity Bρ, revolution time T, and tune Q are uniquely determined for a specific ion circling a closed orbit. If the closed orbit of the same ion is slightly altered, the corresponding variation in T with Bρ (denoted as Δ T and Δ Bρ, respectively) is described by [27]:
In contrast, the tune number Q represents the betatron oscillations around the closed orbit. In a linear storage ring equipped solely with dipoles and quadrupoles, the ratio of variations in Q (denoted as Δ Q) to variations in Δ Bρ/Bρ (or equivalently, Δ p/p), is defined as the natural chromaticity ξ [28]:
Experimental results
Measurement of T
The new tune-IMS method was applied to an early IMS experiment conducted at the HIRFL-CSR [23, 24]. In the experiment, a 78Kr28+ beam was accelerated to 483 MeV/u and bombarded with a 15-mm thick beryllium target. The fragments, predominantly bare ions, were selected using the second Radioactive Ion Beam Line at Lanzhou (RIBLL2) and injected into the CSRe (see Fig. 1). The magnetic rigidity Bρ of the RIBLL2-CSRe system was set to 5.9464 Tm, and the transition point
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F001.jpg)
A TOF detector equipped with a 19-μg/cm2 thin carbon foil was installed in the CSRe [29]. Each ion passes through the detector, releasing secondary electrons from the foil, which are then transmitted isochronously through perpendicularly arranged electric and magnetic fields to a microchannel plate (MCP) counter. The electrical signal from the MCP anode was directed to a fast digital oscilloscope for full-waveform sampling without amplification. The detection efficiency of the signal at each revolution turn mainly depends on the atomic number of the ions. After each injection, a measurement time of 200 μs was set, corresponding to approximately 300 turns for the ions stored in the CSRe. For each ion, a time sequence t(N) as a function of turn number N was obtained.
To account for energy loss, a third-order polynomial was fitted to the t(N) sequence. In this case, T is the first derivative of t(N) at a specific N, such as the middle of the ion’s full storage period.
In reality, the magnetic field instability can induce drifts in T. This study used the method reported in Refs. [30] to correct T drifts caused by unstable magnetic fields. Compared to the previous approach [24], this method can handle injections with only a few stored ions. A portion of the obtained T-spectrum is shown in Fig. 2.
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F002.jpg)
Determination of Q and ξ
The method for measuring Q and ξ using a single TOF detector is described in detail in Refs. [22]. The remainder of this paper is organized as follows:
Due to betatron oscillations, the time sequence t(N) recorded by the TOF detector includes an oscillation term [17]. Using a 51Fe26+ ion measured in this experiment as an example, Fig. 3 shows the results obtained by applying a Digital Fourier Transform (DFT) [31, 22] to the residuals of t(N) fitted with a third-order polynomial. Three significant peaks, labeled A, B, and C, were identified, indicating three periodic components within the time sequence. This analysis was applied to all recorded time sequences, consistently revealing a three-peak pattern.
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F003.jpg)
A detailed study of the origin of these peaks can be found in Ref. [22]. To accurately model the three periodic components, a comprehensive fitting function [17] was employed to refit the passing time sequences as follows:
The accuracy of the measured Qi value depends on the amplitude Ai of the oscillation component, which is proportional to the square root of the emittance [22]. For ions with low emittance, the amplitude of the betatron oscillation is small, resulting in low peaks in the P(f)-spectrum. Moreover, for lighter ions with lower charge numbers, the detection efficiency of the time signals was significantly reduced, increasing the background noise in the P(f) spectrum. We defined the signal-to-noise ratio as the ratio of the peak amplitude to the mean amplitude of the background noise. Obviously, a low signal-to-noise ratio can lead to inaccurate determination of the Q value. Therefore, only the (T,Qi) datasets with a signal-to-noise ratio exceeding 7 were used for further analyses.
Using all 51Fe26+ ions as an example, Fig. 4 presents scatter plots of the selected Qi against T, revealing a clear correlation. As shown in the figure, QA exhibits the highest counts and Pearson correlation coefficients (PCC) [33]. Consequently, QA was selected for subsequent analyses.
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F004.jpg)
According to Eq.(3), we obtain
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F005.jpg)
The T transformation and the mass determination
Using Eq.(4), the measured (T,QA) values for each ion were transformed into (TQ,QA0) at a fixed QA0=0.298, which is the mean of all QA values.
Figure 6 illustrates the standard deviation (σ) of T and TQ for each ion species, corresponding to the revolution times before and after the transformation, respectively. As shown in Fig. 6, the farther an ion species is from the one with the minimum σ(T), the larger the ratio of σ(T)/σ(TQ). For ion species with
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F006.jpg)
Using both T and TQ, the mass of each nuclide was determined following the procedure outlined in Ref. [35]. Figure 7 shows the differences between the redetermined MEs and the literature values [34] for all reference nuclides. To assess the agreement, the χn value is defined as
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F007.jpg)
For the re-determined MEs using T, the calculated χn=1.35 exceeds the 1σ confidence level of
CSRe2011 | Using T | Using TQ | CSRe2023 | Final | |||||
---|---|---|---|---|---|---|---|---|---|
Atom | ME (keV) | Counts | ME (keV) | Counts | ME (keV) | Counts | ME (keV) | Counts | ME (keV) |
63Ge | -46921(37) | 212 | -46887(26) | 391 | -46927(18) | 125 | -46978(15) | 279 | -46957(25) |
65As | -46937(85) | 37 | -46798(61) | 56 | -46803(53) | 15 | -46806(42) | 33 | -46805(33) |
67Se | -46580(67) | 67 | -46482(43) | 138 | -46534(32) | 46 | -46549(20) | 174 | -46545(17) |
71Kr | -46320(141) | 17 | -46150(92) | 30 | -46038(72) | 11 | -46056(24) | 148 | -46054(23) |
For comparison, ME values are shown in Fig. 8, with recent CSRe2023 results serving as a benchmark. Compared to the results from CSRe2011 and those using T, the results using TQ not only align more closely with CSRe2023 but also exhibit higher precision. The improvements in accuracy and precision are mainly due to the significantly reduced spread of TQ, as shown in Fig. 6, despite the lower statistics reported in Table 1. Notably, the mass precision using TQ is comparable to that achieved with the advanced Bρ-IMS method used in CSRe2023.
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F008.jpg)
Notably, for 63Ge, the ME value obtained using T exhibits a significant deviation compared to CSRe2023. Although the discrepancy is reduced with TQ, it remains as large as 2.2σ. Therefore, as shown in the last column of Table 1, the final ME value for 63Ge was adjusted, and the weighted average uncertainty was increased to 25 keV, based on a birth ratio [36] of 2.17.
Discussion
The newly obtained ME values facilitate the calculation of CDE [37], which is the binding-energy difference between mirror nuclei. The CDE arises from differing Coulomb interactions in mirror nuclei due to their varying proton numbers and provides valuable insight into nuclear forces and symmetries[38].
For mirror pairs with
Given that the CDE can be expressed as the CED between mirror nuclei [39, 40], a semi-classical approach to the Coulomb energy [39] can be used to estimate the CDE.
In Fig. 9, the CDEs obtained using experimental ME data from AME’20 [41], CSRe2011 [23] and the present study (using TQ) are represented by different symbols. The CDEs calculated based on the semi-classical approach are depicted as dashed lines. The CDEs from this work align more closely with the semi-classical approach compared to those from CSRe2011 and indicate a more regular change behavior.
To clearly illustrate the variation in CDEs, the quantity
In Fig. 10, experimental ΔCDEs are displayed alongside typical theoretical results. The zig-zag pattern in the theoretical Δ CDE is evident. Experimental ΔCDEs align well with various theoretical values for
-202411/1001-8042-35-11-014/alternativeImage/1001-8042-35-11-014-F010.jpg)
While the ME values of 63Ge from this work and CSRe2023 do not align well, the ΔCDEs at A=63 (and 65) are consistent with each other and agree with theoretical calculations (Fig. 10). Therefore, future mass measurements of 63Ge are required to resolve the mass discrepancy.
Conclusion and outlook
This study introduces Tune-IMS, a novel technique designed to enhance the performance of IMS. By effectively utilizing measured Q values, Tune-IMS was successfully implemented in IMS experiment at HIRFL-CSR. This approach led to a narrower revolution time spread, particularly for nuclides not satisfying the isochronous condition. Consequently, more precise mass values for fp shell nuclei 63Ge, 65As, 67Se, and 71Kr were re-determined, with validation supported by recent Bρ-IMS experiments. These redefined masses facilitated the calculation of CDEs, clarifying anomalies observed with previous mass values.
In Tune-IMS, the ion statistics are reduced due to the selection of Q values. To improve efficiency, a new type of TOF detector with positional sensitivity is currently under development [45]. This detector will provide additional position information, improving Q measurement performance and data utilization efficiency. Overall, Tune-IMS serves as a valuable complement to newly developed Bρ-defined IMS techniques and is particularly beneficial for conventional IMS systems equipped with a single TOF detector.
High-accuracy mass spectrometry with stored ions
. Phys. Rep. 425, 1-78 (2006). https://doi.org/10.1016/j.physrep.2005.10.011Precise mass measurements of radioactive nuclides for astrophysics
. Eur. Phys. J. A 59, 204 (2023). https://doi.org/10.1140/epja/s10050-023-01037-0Masses of exotic nuclei
. Prog. Part. Nucl. Phys. 120,Precision mass measurements of short-lived nuclides at HIRFL-CSR in lanzhou
. Front. Phys. 13,First isochronous mass spectrometry at the experimental storage ring ESR
. Nucl. Instrum. Methods Phys. Res., Sect. A 446, 569-580 (2000). https://doi.org/10.1016/S0168-9002(99)01192-4First isochronous mass measurements at CSRe
. Int. J. Mod. Phys. E 18, 352-358 (2009). https://doi.org/10.1142/S0218301309012380Nuclear structure studies of short-lived neutron-rich nuclei with the novel large-scale isochronous mass spectrometry at the FRS-ESR facility
. Nucl. Phys. A 812, 1-12 (2008). https://doi.org/10.1016/j.nuclphysa.2008.08.013Mass measurements of the neutron-deficient 41Ti, 45Cr, 49Fe, and 53Ni nuclides: first test of the isobaric multiplet mass equation in fp-shell nuclei
. Phys. Rev. Lett. 109,New results from isochronous mass measurements of neutron-rich uranium fission fragments with the FRS-ESR-facility at GSI
. Eur. Phys. J. A 52, 138 (2016). https://doi.org/10.1140/epja/i2016-16138-6Charge resolution in the isochronous mass spectrometry and the mass of 51Co
. Nucl. Sci. Tech. 32, 37 (2021). https://doi.org/10.1007/s41365-021-00876-0Isochronous mass measurements of neutron-deficient nuclei from 112Sn projectile fragmentation
. Phys. Rev. C 107,Ground-state mass of 22Al and test of state-of-the-art ab initio calculations
. Chin. Phys. C 48,Bρ-defined isochronous mass spectrometry and mass measurements of 58Ni fragments
. Eur. Phys. J. A 59, 27 (2023). https://doi.org/10.1140/epja/s10050-023-00928-6A new experimental approach for isochronous mass measurements of short-lived exotic nuclei with the FRS-ESR facility
. Hyperfine Interact. 173, 49-54 (2006). https://doi.org/10.1007/s10751-007-9541-4Improving the resolving power of isochronous mass spectrometry by employing an in-ring mechanical slit
. Nucl. Instrum. Methods Phys. Res., Sect. B 463, 138-142 (2020). https://doi.org/10.1016/j.nimb.2019.06.007Bρ-defined isochronous mass spectrometry: An approach for high-precision mass measurements of short-lived nuclei
. Phys. Rev. C 106,In-ring velocity measurement for isochronous mass spectrometry
. Phys. Rev. Accel. Beams 24,Characterization of a double time-of-flight detector system for accurate velocity measurement in a storage ring using laser beams
. Nucl. Instrum. Methods Phys. Res., Sect. A 931, 52-59 (2019). https://doi.org/10.1016/j.nima.2019.03.058First isochronous mass measurements with two time-of-flight detectors at CSRe
. Phys. Scr. 166,Theory of the alternating-gradient synchrotron
. Ann. Phys. 281, 360-408 (2000). https://doi.org/10.1006/aphy.2000.6012Incoherent tune measurement of an ion storage ring using single ions
. Phys. Rev. Accel. Beams 27,Direct mass measurements of short-lived A=2Z-1 nuclides 63Ge, 65As, 67Se, and 71Kr and their impact on nucleosynthesis in the rp process
. Phys. Rev. Lett. 106,Precision isochronous mass measurements at the storage ring CSRe in Lanzhou
. Nucl. Instrum. Methods Phys. Res., Sect. A 654, 213-218 (2011). https://doi.org/10.1016/j.nima.2011.07.018Mass measurements show slowdown of rapid proton capture process at waiting-point nucleus 64Ge
. Nat. Phys. 19, 1091-1097 (2023). https://doi.org/10.1038/s41567-023-02034-2Mass measurement of upper fp-shell N=Z-2 and N=Z-1 nuclei and the importance of three-nucleon force along the N=Z line
. Phys. Rev. Lett. 130,Storage ring mass spectrometry for nuclear structure and astrophysics research
. Phys. Scr. 91,A high performance time-of-flight detector applied to isochronous mass measurement at CSRe
. Nucl. Instrum. Methods Phys. Res., Sect. A 624, 109-113 (2010). https://doi.org/10.1016/j.nima.2010.09.001Particle identification and revolution time corrections for the isochronous mass spectrometry in storage rings
. Nucl. Instrum. Methods Phys. Res., Sect. A 941,Studies in astronomical time series analysis. ii. statistical aspects of spectral analysis of unevenly spaced data
. Astrophys. J. 263, 835-853 (1982). https://doi.org/10.1086/160554Certain topics in telegraph transmission theory
. Trans. Am. Inst. Electr. Eng. 47, 617-644 (1928). https://doi.org/10.1109/T-AIEE.1928.5055024VII. note on regression and inheritance in the case of two parents
. Proc. R. Soc. London 58, 240 (1895). https://doi.org/10.1098/rspl.1895.0041The AME 2020 atomic mass evaluation (I). evaluation of input data, and adjustment procedures
. Chin. Phys. C 45,Isochronous mass measurements of Tz=-1 fp-shell nuclei from projectile fragmentation of 58Ni
. Phys. Rev. C 98,The calculation of errors by the method of least squares
. Phys. Rev. 40, 207-227 (1932). https://doi.org/10.1103/PhysRev.40.207Coulomb energy differences between high-spin states in isobaric multiplets
. Prog. Part. Nucl. Phys. 59, 497-561 (2007). https://doi.org/10.1016/j.ppnp.2006.10.001Variation in displacement energies due to isospin-nonconserving rorces
. Phys. Rev. Lett. 110,Coulomb energies
. Annu. Rev. Nucl. Part. Sci. 19, 471-526 (1969). https://doi.org/10.1146/annurev.ns.19.120169.002351Nuclear coulomb energies
. Rep. Prog. Phys. 41, 957-1026 (1978). https://doi.org/10.1088/0034-4885/41/7/001The AME 2020 atomic mass evaluation (II). Tables, graphs and references
. Chin. Phys. C 45,Proton drip-line calculations and the rp process
. Phys. Rev. C 65,Mass relations of mirror nuclei in terms of Coulomb energies based on relativistic continuum Hartree-Bogoliubov calculations
. Phys. Rev. C 103,A survey of Coulomb displacement energies and questions on the anomalous behavior in the upper fp-shell
. J. Phys. G: Nucl. Part. Phys. 41,Electrostatic-lenses position-sensitive TOF MCP detector for beam diagnostics and new scheme for mass measurements at HIAF
. Nucl. Sci. Tech. 30, 152 (2019). https://doi.org/10.1007/s41365-019-0676-1The authors declare that they have no competing interest