Introduction
X-ray absorption fine structure[1-5] (XAFS) spectroscopy based on synchrotron radiation (SR) sources is a powerful tool for determining the electronic and atomic local structures, as well as the chemical-physical states of elements in condensed matter. The ion chamber (IC) is the most commonly used detector to collect XAFS signals, which exhibits a good linear response to the X-ray intensity for a wide range of energies[6]. However, a few drawbacks of IC include microphonic noise and the problematic electron-ion pair recombination[6, 7] enhanced by a high X-ray flux. Thus, finding a substitute for IC is desirable to accurately acquire a high-flux X-ray intensity. Silicon PIN photodiodes (SPPD) and diamond-based photodetectors (DD) were often used as substitutes for ion chambers and have been investigated for more than two decades. Sato[6] investigated the feasibility of SPPD as an X-ray intensity monitor and found that these detectors were adequately feasible for XAFS experiments because their spectra had nearly the same quality as that of IC, and it can be more effective than IC for a high X-ray flux. Dalba et al.[7] studied the properties of SPPD as X-ray detectors for Extended-XAFS (EXAFS) measurements and found that good-quality EXAFS measurements with SPPD were possible if diffraction effects were eliminated. Pace et al.[8] studied the application of DD to XAFS measurements between the energy range of 4–13 keV and found that all the XAFS spectra were significantly similar to the spectrum from IC and exhibited a good signal-to-noise ratio (SNR). De Sio et al.[9] performed a direct comparison between an IC and a nitrogen-doped single-crystal DD applied to Fe K-edge XAFS spectroscopy and found that the EXAFS data and parameter values obtained in the transmission mode by both the DD and a standard IC were in good agreement. Ravel et al.[10] evaluated the performance of DD as an incident beam monitor and verified that the effect of diffraction on XAFS measurements can be effectively eliminated if DD is combined with polycapillary lenses. Semiconductor-type detectors have the advantages[11] of a high response, no microphonic noise, good linearity, and a wide dynamic range. In particular, they make it easier to miniaturize detectors to a sufficiently small size. In addition to the limitations of the traditional IC used in high-flux X-ray detection, its large size limits its application[12-15] in certain combined detection techniques. Small-sized SPPD and DD may be more advantageous for high-flux XAFS measurements in certain combination techniques.
Compared to DD, SPPD is more economical and easier to obtain from commercial markets. Blanchet et al.[16] developed a small and robust SPPD built into a beamstop for detecting backscattered X-ray photons, demonstrating a good response to X-ray intensity between the energy range of 6–20 keV. Beale et al. [12] also successfully obtained XAFS signals of the Zn K-edge using a photodiode in a combined small-angle X-ray scattering (SAXS)/wide-angle X-ray scattering (WAXS)/XAFS setup. Over the past few decades, several measurement approaches have been developed to reduce the time required to collect an entire XAFS spectrum. The Quick-scanning EXAFS (QEXAFS) technique[17, 18] not only dramatically reduced the data collection time of an entire EXAFS spectrum to a few seconds or less, but also maintained a high SNR[19] of the obtained QEXAFS spectrum. Yu et al. [20] developed a QEXAFS method with a time resolution of 2 s based on the EPICS and LabVIEW software at the BL14W1 beamline of the Shanghai Synchrotron Radiation Facility (SSRF). Khalid et al.[21] refined the QEXAFS technique. The QEXAFS spectrum acquisition time was reduced to milliseconds by driving the channel-cut Si(111) monochromator to oscillate at a frequency of 5 Hz with a tangential arm. In addition, the use of high-speed counting cards (HSCC) is also an effective method used to accelerate XAFS data collection; here, it was named the high-frequency (HF) sampling scheme. Using the IK220 HSCC card and IC, a similar HF sampling mode[22] has been reported. In a compact combined technique, SPPD is apparently more advantageous than IC owing to its small size. However, the afterglow effect and detection efficiency of SPPD can be disadvantageous[6] for rapid measurements. Therefore, SPPD may be considered less suitable for QEXAFS measurements than routine IC. Although previous studies have demonstrated that SPPD can be sufficiently used in the conventional XAFS collection, the feasibility of SPPD applied to acquire HF sampling XAFS spectra has not yet been studied, especially when the SPPD is coupled into the SAXS beamstop of a combining setup. This study aims to determine whether the SPPD used as a rear detector is comparable to the DD used as the front detector in the HF sampling scheme. If the small-sized SPPD can be used for HF-sampling XAFS measurements, it will be significantly beneficial to develop these combined techniques including XAFS measurements.
In this study, a commercial SPPD was coupled with a beamstop used in conventional SAXS measurements to protect the subsequent two-dimensional detector from irradiation damage. This beamstop-coupled SPPD obtains the role of a rear detector in transmission XAFS measurements. By comparing the performance of this miniaturized SPPD with those of DD and IC in transmission XAFS measurements, its feasibility and validity for acquiring the conventional and HF sampling XAFS spectra are discussed.
Preparation of Silicon PIN photodiode
In a typical XAFS transmission mode, an ion chamber is used to monitor the X-ray intensity (I0) in front of the sample, referred to as the XAFS front detector (FD). Another ion chamber was used to monitor the X-ray intensity (I1) behind the sample, here named the XAFS rear detector (RD). For a sample with a thickness d, the absorption coefficient (μ) can be expressed as follows: μ(E)d=ln(I0/I1). However, owing to the large size of the ion chamber, it is often difficult to meet the requirements of XAFS measurements using a combined technique involving an ion chamber. A transmission-type detector should generally be used to record the incident X-ray intensity; evidently, the SPPD is not suitable for I0 detection in front of the sample. In this study, DD[23] or IC was used as the front detector for the XAFS measurement to detect the incident X-ray intensity (I0). A mini SPPD was used as the rear detector to record the X-ray transmission intensity (I1) behind the sample. To maintain a maximum distance of the rear detector used for XAFS measurements from other experimental signal paths, its size should be minimized. To optimize the coupling of a mini-SPPD into a beamstop, it is typically mounted in a vacuum tube for SAXS measurements and maintained at a distance from the sample to avoid blocking other experimental signals. This type of SPPD can not only be used for XAFS signal acquisition but can also prevent irradiation damage of the direct beam to the SAXS detector.
In this study, the SPPD was purchased from Beijing Lightsensing Technologies Ltd., as shown in Fig. 1(a). The model used was LSSPD-SPB3; the effective photosensitive area was 3×3 mm2. The maximum spectrum response at 940 nm was 0.5 A/W, which is larger than the previously reported 0.43 A/W[6] of S1227-1010BR SPPD at 720 nm. Other factory parameters for this SPPD are listed in Table 1. To achieve spatial compatibility for the simultaneous measurement of the SAXS, XRD, and XAFS signals, the SPPD was coupled into a beamstop with a volume of 12×6×6 mm3. The beamstop was supported by a hollow cylindrical rod with an outer diameter of 3 mm, as shown in Fig. 1(b). The beamstop had a rectangular opening, and the SPPD embedded in the beamstop faced the transmission beam behind the sample. Note, M. Kocsisa and A. Somogyi[24] developed a miniature IC for X-ray microprobe measurements at the European Synchrotron Radiation Facility (ESRF); however, this miniature IC was insufficiently small for SAXS/XRD/XAFS combined measurements. If the aforementioned miniature IC was to be used for recording the transmission X-ray intensity, it would need to be placed at the downstream end; otherwise, it would obscure the SAXS or XRD signal. In addition, its design, which lacks window material, contains dozens of micron entrance pinholes, and has a constant Ar pressure flow of approximately 1 atm, will severely restrict its effectiveness in the SAXS/XRD/XAFS combined technique. Therefore, it is necessary to develop a mini SPPD detector for XAFS measurements in the SAXS/XRD/XAFS combined technique.
-202207/1001-8042-33-07-011/alternativeImage/1001-8042-33-07-011-F001.jpg)
Parameter | Value |
---|---|
Absolute maximum ratings | |
Operating temperature, Top (oC) | -25 ~ +85 |
Storage temperature, Tstg (oC) | -40 ~ +85 |
Forward current, If (mA) | 10 |
Reverse current, Ir (μA) | 1000 |
Soldering temperature (10s), Ts (oC) | 260 |
The SPPD is relatively sensitive to visible and infrared radiation. To avoid interference from signals other than X-rays, the SPPD must be kept at a distance from visible and infrared lights. Therefore, the SPPD was coupled into a beamstop and further wrapped with black 3M tape, as shown in Fig. 1(c). The dark current (Idark) of the as-prepared SPPD was measured to be approximately 17 nA, as shown in Fig. 1(d). Previous studies [25, 26] have demonstrated that the dark current Idark consists of a bulk leakage current (Ibulk) and surface leakage current (Isurf). The former is proportional to the mesa area of the diode, whereas the latter is perimeter-dependent on the mesa sidewall; both originate from defects. A decrease of approximately 0.5 nA was noted in the dark current at the beginning when the dark current of the as-prepared SPPD detector was measured. This may be attributed to the small number of dangling bonds on the surface of the SPPD, which initially captured a certain amount of charge. When an external electric field was applied to the SPPD, these captured charges were gradually released and depleted, causing the dark current to decrease at the beginning and reach a quasi-equilibrium state. When the incident X-ray energy value was near the Cu K-edge, approximately 9 keV, the X-ray intensity was approximately 1012 cps at beamline 1W2B of the Beijing Synchrotron Radiation Facility (BSRF), which corresponds to an irradiation power of approximately 1.44 mW. The response current induced by the transmission beam was usually in the range of several hundreds of μA to a few mA when measuring the K-edge X-ray absorption spectra of the transition metals using this SPPD. Compared to the dark current of 17 nA, the dynamic range of this SPPD at nearly 9 keV was approximately 104–105. Theoretically, this SPPD can be sufficiently used for the K-edge XAFS measurements of transition metals.
XAFS data acquisition
The feasibility of this SPPD used for XAFS measurements was verified at beamline 1W2B of the BSRF. The storage ring was run at 2.5 GeV with an electron-beam current of approximately 250 mA. The X-ray energy range of beamline 1W2B was between 5 to 18 keV, which can be monochromized by a Si(111) monochromator with an energy resolution (ΔΕ/Ε) of approximately 2×10-4. The X-ray spot size was approximately 1.0 (H) × 0.5 (V) mm2 at the sample position. The X-ray photon flux was approximately 1.0×1012 photons/s at the Cu K-edge (8979 eV). To assess the feasibility of the SPPD used for XAFS measurements, conventional K-edge XAFS spectra and HF sampling K-edge XAFS spectra of the standard Cu foil were collected. By replacing the detectors, three groups of detector combinations labeled FD-RD (including IC-IC, IC-SPPD, and DD-SPPD) were used in the conventional XAFS measurements, where FD represents the front detector (I0) in front of the sample and RD represents the rear detector (I1) behind the sample. When the DD was used as the front detector[23], the sensitive area was 10×10 mm2. The thickness of the diamond film was 0.5 mm and a bias voltage of 300 V was applied to both sides of the diamond film. Because the large size of the ion chamber is unavoidable, the ion chamber is not considered in the HF sampling mode that is mainly used for the combined technique. Therefore, in the HF sampling XAFS measurements, only one group of detector combinations, labeled as DD-SPPD, was used to collect the XAFS spectra. The main difference between the conventional XAFS mode and the HF sampling XAFS mode was that the 974 counter was used in the conventional XAFS mode, whereas a high-speed counter card was used in the HF sampling XAFS mode. A schematic of the two sampling modes is illustrated in Fig. 2. For comparison, the XAFS spectra collected with different detector combinations were directly labeled as FD-RD, where the front detector (FD) was used to record the X-ray intensity (I0) in front of the sample and the rear detector (RD) was used to record the X-ray intensity (I1) behind the sample. All collected XAFS data were processed and analyzed using a comprehensive Demeter [27] program.
-202207/1001-8042-33-07-011/alternativeImage/1001-8042-33-07-011-F002.jpg)
Results and discussion
Feasibility of SPPD for XAFS measurements
The Cu K-edge X-ray absorption spectrum of the Cu foil was first collected using the DD-SPPD detector combination in the conventional mode, as shown in Fig. 3(a). As a reference, the Cu K-edge X-ray absorption spectrum of the Cu foil was then collected with the IC-IC detector combination in the conventional sampling mode, as shown in Fig. 3(a). By comparing the two X-ray absorption spectra, it was found that the fine structures and oscillation amplitudes of IC-IC and DD-SPPD were nearly the same. Based on Figs. 3(b) and 3(c), these two XAFS spectra are highly consistent both in the k-space and R-space, further confirming the consistency of the spectral characteristics collected by the two detector combinations of IC-IC and DD-SPPD. Furthermore, the repeatability of the XAFS data was verified by comparing three independent XAFS scans of the Cu foil Cu K-edge, as shown in Fig. 3(d). From the three XAFS spectra repeatedly acquired using the DD-SPPD detector combination, the three Cu K-edge X-ray absorption spectra were found to be nearly the same, presenting excellent repeatability. This result indicates that in addition to the SPPD not being damaged by the continuous X-ray radiation, the XAFS data quality collected by the DD-SPPD detector combination was approximately the same as that collected by the widely used IC-IC detector combination. As indicated, the incident X-ray intensity was approximately 1012 cps in the Cu K-edge XAFS spectra measurements. As a conservative estimate, a total of 20 min were required for acquiring a complete XAFS spectrum in the conventional XAFS sampling mode or step scanning mode. The SPPD was continuously exposed to the transmission X-ray beam for tens of minutes in a single XAFS scan; whereas in the HF sampling mode, the continuous exposure time of the SPPD was only for a dozen seconds. The reproducibility of the XAFS spectral characteristics in the conventional XAFS measurements clearly indicates that irradiation damage did not occur in the SPPD, as it was used to record the XAFS signals. The lifetime of SPPD depends on the exposure time in addition to the irradiation dosage. Namely, this type of photodiode can be stably used for the collection of transmission XAFS data if the incident X-ray flux is not higher than 1012 cps. Based on these results, visible and infrared photodiodes can be sufficiently used for the acquisition of the transmission X-ray absorption spectra, verifying the feasibility of SPPD used for XAFS measurements.
-202207/1001-8042-33-07-011/alternativeImage/1001-8042-33-07-011-F003.jpg)
Comparison of detector combinations
The aforementioned results confirm the feasibility of the DD-SPPD detector combination for conventional transmission XAFS measurements. The IC-SPPD detector combination was also used to collect the Cu K-edge X-ray absorption spectrum of the Cu foil, and the obtained XAFS spectrum was labeled IC-SPPD. For comparison, the conventional Cu K-edge XAFS spectrum of the Cu foil collected with the IC-IC detector combination was used as a reference. The three Cu K-edge X-ray absorption spectra of the Cu foil are shown in Figs. 4(a) and 4(b). The three detector combinations can obtain nearly the same X-ray absorption spectra as that in the E-space or the XAFS curves in the k-space. Apparently, both DD and IC can be used as the front detectors when SPPD is used as the rear detector; the two-detector combinations of DD-SPPD and IC-SPPD are feasible for XAFS measurements.
-202207/1001-8042-33-07-011/alternativeImage/1001-8042-33-07-011-F004.jpg)
To quantitatively evaluate the difference between the two XAFS curves, a deviation factor (Q) was defined as follows:
Here, Y0, Y, and
The calculated Q values demonstrate that the DD-SPPD and IC-SPPD detector combinations can acquire XAFS data closer to the reference data collected by the IC-IC detector combination; however, the IC-SPPD detector combination presents slight glitches on the XAFS spectrum compared to that collected by the IC-IC detector combination. This is because the IC and SPPD exhibited slightly different responses to the X-ray energy, as shown in Fig. 4(c). Consequently, certain unnecessary and harmful glitches were not sufficiently removed and were superimposed on the XAFS spectrum. By comparing the responses of the IC, SPPD, and DD to the X-rays, nearly all the details can be found to have a one-to-one correspondence; however, the response ratio in pairs among the three detectors was only approximately linear, as shown in Fig. 4(d). The slightly different responses of the detectors to the X-ray leads to minor deviations between the X-ray spectra or the XAFS curves recorded by the different detectors. Regardless,, the IC-SPPD and DD-SPPD detector combinations for XAFS data acquisition remain feasible.
HF sampling XAFS data acquisition
In addition to the conventional XAFS measurement scheme, a new sampling scheme, HF sampling XAFS transmission, was tested when the beamstop-coupled SPPD was used as the rear detector. In the HF sampling scheme, the XAFS signals were collected in the transmission mode, and a high-speed counting module (NI 9223) was used for data collection. A high-speed counting module was used for the Cu K-edge XAFS measurements of the Cu foil at three sampling frequencies (1, 10, and 100 kHz). Here, the sampling frequency represents the number of times that an experimental signal (X-ray intensity) was repeatedly read in one second. Thus, the X-ray intensities in front and behind the sample were repeatedly read 103, 104, and 105 times per second at the three respective sampling frequencies. Note, the sampling rate of this high-speed counting card was higher than that (5000 data points/s) of IK220[21]. A higher sampling rate can significantly improve data quality by raising the statistics.
For a typical XAFS spectrum covering an energy range of 1000 eV, the total measurement time of the XAFS spectrum in the HF sampling scheme was controlled within a few seconds (8 s) in the transmission mode at BSRF. Namely, each XAFS spectrum contains 8000, 80000, and 800000 data points under the three sampling frequencies. Figure 5(a) presents the raw Cu K-edge XAFS spectra of the Cu foil collected using the DD-SPPD detector combination at sampling frequencies of 1, 10, and 100 kHz. Note, the HF sampling XAFS data are clearly different than that of the conventional XAFS data in appearance. It is difficult to determine the data qualities and features from the raw HF sampling XAFS data. To save time, the XAFS data are usually recorded in chronological order (time dimension) in these HF sampling XAFS measurements. The XAFS data is required to be converted from the time domain into the energy space for a subsequent data analysis. Another important issue is that the HF sampling scheme does not increase the energy resolution of the XAFS spectra, however drastically increases the number of data points. Therefore, in addition to the abscissa transformation, an appropriate statistical average for the HF sampling XAFS data is necessary. Typically, for an XAFS spectrum covering an energy range of approximately 1000 eV, it is sufficient to retain a few hundred data points. Under 103, 104, or 105 sampling frequencies, the entire X-ray absorption spectrum recorded by the HF sampling scheme can be roughly divided into 800 time segments in chronological order. Each time segment had approximately 10, 100, or 1000 raw data points. These raw data points in each time segment must be summed and averaged to form one reduced data point. After data pre-processing, each XAFS spectrum was statistically averaged into 800 data points. Conservatively, this data preprocessing did not reduce the energy resolution. The 800 data points reserved for each XAFS spectrum were generally more than those for the traditional XAFS spectrum. The statistical average not only reduces the number of data points, but also significantly improves the SNR of the XAFS spectrum, as shown in Fig. 5(b). Note, all three HF sampling XAFS spectra had 8 s data acquisition times; however, their SNRs after the statistical average preprocessing were different. For an energy region with a span smaller than the energy resolution, the XAFS signal can be regarded as approximately constant if the sampling frequency is sufficiently high. Therefore, the HF sampling scheme is equivalent to the repeated measurements of an approximately constant quantity and different sampling frequencies correspond to different repetitive measurement times. A higher sampling frequency leads to a greater statistic and lower noise. When the sampling frequency was as low as 1 kHz, the XAFS spectrum after the statistical average remained too noisy, resulting in the failure of the XAFS spectrum, as shown in the curve at the bottom of Fig. 5(b).
-202207/1001-8042-33-07-011/alternativeImage/1001-8042-33-07-011-F005.jpg)
To convert the abscissa of the HF sampling XAFS spectra from chronological order (time domain) to the energy space, the absorption-edge position was first determined and marked as E0, which corresponds to the first-derivative maximum of the XAFS spectrum. For a typical double-crystal monochromator, its rotation in the Bragg angle (θ) space is usually uniform. This indicates that the Bragg angle of the Si(111) monochromator linearly changed following the formula θ-θ0 = ω (t-t0). where θ0 indicates the Bragg angle corresponding to the absorption edge of the element to be measured, t0 is the time required for the monochromator to rotate to the absorption edge of the element, and ω is the rotation angular velocity of the monochromator. Based on the linear relationship between the Bragg angle (θ) and rotation time (t), the data point distribution of the HF sampling XAFS spectrum was also uniform in the Bragg angle space. The abscissa of the HF sampling XAFS spectra can be converted from the Bragg angle (θ) to the incident X-ray energy (E) using the following formula:
Comparison of HF sampling XAFS spectra
To verify the data quality of the HF sampling XAFS spectra acquired by the SPPD, the Cu K-edge XAFS spectrum collected with the DD-SPPD detector combination and the 100 kHz sampling frequency was compared to the reference spectrum that was collected by the IC-IC detector combination in the conventional sampling mode. The E-space X-ray absorption spectra and k-space XAFS spectra of the Cu foil are shown in Fig. 6 (a) and (b), respectively, which demonstrates that regardless of the E-space or k-space, the HF sampling XAFS spectrum collected within a few seconds was very similar to the conventional XAFS spectrum collected within a few minutes, except for the fact that the SNR of the HF sampling XAFS spectrum was slightly worse. Considering the data with the IC-IC detector combination as a reference, the calculated Q factor was 0.017 for the X-ray absorption spectrum with the DD-SPPD detector combination shown in Fig. 6(a), or 0.213 for the XAFS spectrum with the DD-SPPD detector combination shown in Fig. 6(b). This result demonstrates that in addition to the SPPD detector being used for the collection of conventional XAFS spectra, it can also be used for the data acquisition of HF sampling XAFS. Furthermore, the Fourier transform spectra in the R-space and the isolated single-shell XAFS oscillations in the k-space were extracted, as shown in Fig. 6 (c) and (d), respectively. Similarly, the calculated Q factors were 0.121 and 0.068 for the Fourier transform spectrum and single-shell XAFS curve, respectively. Although the Q factor was defined in Eq. (1) to describe the similarity between the two curves, it was found that the calculated Q values were slightly different for the same data in the E-space, k-space, and R-space. Evidently, the Q-value calculated for the data in the k-space was the most sensitive to the curve deviation. A quantitative comparison between the HF-sampling and conventional XAFS spectra was also performed by fitting the nearest-neighbour single-shell XAFS oscillations of the Cu foil. The fitting curves are shown in Fig. 6(c) and (d), and the fitting parameters are listed in Table 2.
-202207/1001-8042-33-07-011/alternativeImage/1001-8042-33-07-011-F006.jpg)
Sample | Bond | R (Å) | N | σ2 (×10-3 Å2) | ∆E (eV) | R-factor (×10-3) |
---|---|---|---|---|---|---|
Cu[28] | Cu-Cu | 2.56 | 12 | - | - | - |
Cu foil(IC-IC) | Cu-Cu | 2.54±0.03 | 12.0±0.6 | 8.8±0.4 | 4.7±0.6 | 1.7 |
Cu foil(DD-SPPD) | Cu-Cu | 2.54±0.05 | 12.0±0.9 | 8.5±0.6 | 4.5±0.9 | 3.8 |
The XAFS fitting results indicate that the fitting parameters of the nearest-neighbor structure are nearly the same for the 100 kHz HF sampling XAFS and conventional XAFS spectra. This study demonstrates that the SPPD is not only capable of being sufficiently applied for conventional XAFS measurements, but can also be sufficiently used for HF sampling XAFS measurements. The small size of the SPPD and ease of assembly with a beamstop makes it capable for use in a combined technique, including the XAFS measurement. The good performance of this SPPD in the HF sampling XAFS measurements proves that it can be sufficiently applied to time-resolved experimental measurements within seconds.
Conclusion
In this study, a commercial LSSPD-SPB3 silicon PIN photodiode was coupled in a beamstop to form a small detector. This detector was tested for conventional and HF-sampling XAFS data collection. By comparing the XAFS spectra collected with IC-IC and DD-SPPD detector combinations, the following conclusions can be drawn: 1) The SPPD detector sufficiently matches with a diamond-film detector, and the DD-SPPD detector combination can be used for XAFS data collection. 2) The SPPD as the rear detector can be sufficiently used to collect data from conventional XAFS and HF sampling XAFS, verifying the feasibility of SPPD as the rear detector for XAFS acquisition. 3) The miniaturized SPPD detector is simple and convenient to use in a combined technique, including the XAFS measurement.
X-ray absorption fine structure spectroscopy in nanomaterials, Sci
. China Mater. 58, 313-341 (2015). doi: 10.1007/s40843-015-0043-4Fundamentals of XAFS
. Rev. Mineral. Geochem. 78, 33-74 (2014). doi: 10.2138/rmg.2014.78.2Local structural evolutions of CuO/ZnO/Al2O3 catalyst for methanol synthesis under operando conditions studied by in situ quick X-ray absorption spectroscopy
. Nucl. Sci. Teth. 28, 21 (2017). doi: 10.1007/s41365-016-0170-y.XAFS and SRGI-XRD studies of the local structure of tellurium corrosion of Ni-18% Cr alloy
. Nucl. Sci. Teth. 30, 153 (2019). doi: 10.1007/s41365-019-0673-4.Synchrotron radiation-based materials characterization techniques shed light on molten salt reactor alloys
. Nucl. Sci. Teth. 31, 6 (2020). doi: 10.1007/s41365-019-0719-7.Feasibility study of silicon PN photodiodes as X-ray intensity monitors for high flux X-ray beam with synchrotron radiation
. Nucl. Instrum. Meth. A 436, 285-290 (1999). doi: 10.1016/S0168-9002(99)00635-XPIN silicon diodes as EXAFS signal detectors
. J. Synchrotron Radiat. 3, 213-219 (1996). doi: 10.1107/S0909049596006073Photodiodes as detectors with high dynamical range for X-ray reflectivity measurements
. Nucl. Instrum. Meth. A 306, 544-548 (1991). doi: 10.1016/0168-9002(91)90050-ZDiamond photodetectors for X-Ray absorption spectroscopy applications
. AIP Conference Proceedings 1234, 97-100 (2010). doi: 10.1063/1.3463377Diamond solid state ionization chambers for X-ray absorption spectroscopy applications
. Appl. Phys. Lett. 93, 083503 (2008). doi: 10.1063/1.2976125Diamond sensors and polycapillary lenses for X-ray absorption spectroscopy
. Rev. Sci. Instrum. 84, 103106 (2013). doi: 10.1063/1.4824350A combined SAXS/WAXS/XAFS setup capable of observing concurrent changes across the nano-to-micrometer size range in inorganic solid crystallization processes
. J. Am. Chem. Soc. 128, 12386-12387 (2006). doi: 10.1021/ja062580rCombined x-ray absorption spectroscopy and X-ray powder diffraction
. AIP Conference Proceedings 258, 631-641 (1992). doi: 10.1117/12.49471Combined XRD and XAS
. Nucl. Instrum. Meth. A 649, 213-215 (2011). doi: 10.1016/j.nima.2010.11.07bA combined multi-technique in situ approach used to probe the stability of iron molybdate catalysts during redox cycling
. Top Catal. 52, 1400-1409 (2009). doi: 10.1007/s11244-009-9324-zA small and robust active beamstop for scattering experiments on high-brilliance undulator beamlines
. J. Synchrotron Radiat. 22, 461-464 (2015). doi: 10.1107/S160057751402829XQuick scanning EXAFS: first experiments
. Nucl. Instrum. Meth. A 270, 578-581 (1988). doi: 10.1016/0168-9002(88)90732-2Improvement in XAFS beamline BL01B1 at SPring-8
. J. Phys. Conference Series 190, 012041 (2009). doi: 10.1088/1742-6596/190/1/012041Quick-EXAFS implementation on the general purpose EXAFS beamline at ESRF
. J. Synchrotron Radiat. 18, 176-182 (2011). doi: 10.1107/S0909049510046546The XAFS beamline of SSRF
. Nucl. Sci. Teth. 26, 050102 (2015), doi: 10.13538/j.1001-8042/nst.26.050102Quick extended X-ray absorption fine structure instrument with millisecond time scale, optimized for in situ applications
. Rev. Sci. Instrum. 81, 015105 (2010). doi: 10.1063/1.3276679Data acquisition system modified by IK220 card for the SSRF XAFS station
. Nucl. Tech. 35, 561-567 (2012). doi: 10.1007/s11783-011-0280-zA polycrystalline diamond micro-detector for X-ray absorption fine-structure measurements
. J. Synchrotron Radiat. 29, 424-430 (2022). doi: 10.1107/S1600577521013011Miniature ionization chamber detector developed for X-ray microprobe measurements
. J. Synchrotron Radiat. 10, 187-190 (2003). doi: 10.1107/S0909049502019854Suppression of dark current in germanium-tin on silicon pin photodiode by a silicon surface passivation technique
. Opt. Express 23, 18611-18619 (2015). doi: 10.1364/OE.23.018611GeSn p-i-n photodetector for all telecommunication bands detection
. Opt. Express 19, 6400-6405 (2011). doi: 10.1364/OE.19.006400ATHENA, ARTEMIS, HEPHAESTUS: data analysis for X-ray absorption spectroscopy using IFEFFIT
. J. Synchrotron Radiat. 12, 537-541 (2005). doi: 10.1107/S0909049505012719High-temperature thermal expansion of six metallic elements measured by dilatation method and X-ray diffraction
. J. Mater. Sci. 23, 757-760 (1988). doi: 10.1007/BF01174717