SYNCHROTRON SCIENCE AND TECHNOLOGY
Development of a PSD-based photon beam position measurement system
XIAO Yunyun,
GU Liming,
LU Ping,
WANG Jigang,
TANG Leilei,
YANG Yongliang,
ZHOU Zeran,
XU Hongliang
Nuclear Science and TechniquesVol.23, No.2pp.70-74Published in print 20 Apr 2012
As an important part of the beam diagnostic system, the synchrotron light beam position measurement has a very high value in the high quality and high stability light source applied research. A new photon beam position monitor based on position-sensitive detector (PSD) has been developed to measure the photon beam position in vertical and horizontal directions at the same time at HLS (Hefei Light Source). The new PBPM based on the PSD has fast response speed, high sensitivity and wide dynamic range. This PBPM system also includes the C4674 signal processing circuit, NI USB-9215 data acquisition device and the LABVIEW data acquisition program. This PBPM system has been calibrated vertically and horizontally on-line, and then has been applied in the beam line B3EA of HLS to measure the position of the synchrotron light. As the results shown, the resolution of the system is better than 3 μm.
Photon beam position monitorPosition-sensitive detectorCalibration