I. INTRODUCTION
Photon activation analysis (PAA), in most cases nowadays, is an accelerator-based radioanalytical technique. After the irradiation of the samples with high energy photons created by bremsstrahlung radiation of electrons, qualitative and quantitative information on the target elements can be obtained by detection of nuclear emissions following the irradiation with LINAC. The first article of PAA dated back to 1951 when Gaudin and Pannell of MIT tried to determine the amount of beryllium in low grade beryl ores by photodisintegration of Beryllium [1]. They commented on this method as "rapid, simple, and nondestructive". In 1954, Basile proposed to analyze some of the light elements using photonuclear reaction induced by bremsstrahlung radiation from a betatron [2]. His paper led to the wide use of the bremsstrahlung radiation as the high energy photon source of PAA. A chapter in Encyclopedia of Analytical Chemistry gave a brief but complete summary of PAA research [3]. In recent years, PAA has expanded its applications in radiotherapy, meteorology, geochemistry, archeology, industrial material, environmental studies, etc. [4-10].
Figure 1 shows the basic principle of PAA. Photon activation usually occurs in the giant dipole resonance (GDR) region of the target nucleus. A high energy photon above the particle emission threshold enters the nucleus and interacts with nucleons. Then the excited nucleus equilibrates and turns into a statistical "compound nucleus" state. This "compound nucleus" will de-excite through residual strong interaction by emitting prompt gamma rays, neutrons, protons, and even alpha particles. Following the "compound" nucleus stage, the product nucleus generally is still unstable because it is usually a proton-rich (or neutron-rich) nucleus. It continues to de-excite to the final nucleus through weak interactions, such as electron capture and beta decay. Delayed gamma rays are emitted after those weak interactions. In most cases of PAA study, researchers concentrated on delayed gamma ray emission with a relatively long half-life. Those gamma rays can be detected by HpGe spectrometers and generated spectra with the assistance from computers. Hence, the procedure of PAA is quite straightforward: from the spectra of delayed gamma rays, one acquires their energy lines and intensity, which corresponds to certain product nuclides. From the information on the products nuclides, one can trace back to quantity and species of the target nuclides according to the corresponding nuclear reaction channels.
-201405/1001-8042-25-05-003/alternativeImage/1001-8042-25-05-003-F001.jpg)
II. CALCULATIONS
According to Ref. [11-13], the net peak area (or counts) P of the characteristic gamma line in the spectrum is
where η is the detector efficiency, θ is the branching ratio of the reaction channel, ζ is the absolute intensity of the gamma line, A is the activity of radioactive nuclide, m is the mass of the total sample, cm is the concentration of the nuclides of interest, h is the natural abundance of the target nuclide, L is the Avogadro constant, λ is the decay constant of product nuclide, Ar is the atomic mass of the target nuclide, φ(E) is the energy differential photon flux, σ(E) is the cross section of the photonuclear reaction of interest, ti is the irradiation time, tc is the counting time, and td is the decay time from the end of irradiation to the start of spectrum collection.
Equation (1) can be rearranged as
On the left side of Eq. (2) is the concentration cm of a certain element in the sample, which is the ultimate value PAA seeks. On the right side mostly are the parameters we can measure (P, η, m, ti, tc, td), or parameters we can find the values in existing databases (θ, ζ, h, L, λ, Ar, σ(E)). The only parameter we normally cannot measure directly is the energy differential photon flux φ(E).
How to find the energy differential photon flux φ(E)? As mentioned before, we need assistance from computer simulations. From the simulated photon yield, it is feasible to derive photon flux and then calculate the concentration of the target nuclide directly based on the Eq. (2).
The output of the computer simulation is the photon yield Y(E). To obtain photon flux φ(E), one needs to multiply it with average beam current, Ibeam, of the LINAC as below.
where qe is the charge of an electron. Inserting Eq. (3) into Eq. (2), one gets
To simplify Eq. (4), we define the integral in the denominator as the reduced reaction rate IR, the Eq. (4) evolves into
In Eq. (5), the uncertainty from atomic mass Ar, decay constant λ, electron charge qe, detector efficiency η, branching ratio θ, absolute intensity ζ, natural abundance h, Avogadro’s number L, and time parameters (ti, tc, td) usually can be ignored. The primary source of uncertainty comes from uncertainty in the net peak area P, mass of sample m, beam current Ibeam, and reduced reaction rate IR. Hence, the uncertainty propagation of Eq. (5) follows
One should note that the uncertainty of the reduced reaction rate in Eq. (6) is the combination of the uncertainties of photon yield and cross sections.
III. EXPERIMENTAL SETUP, SIMULATIONS, CROSS SECITONS, AND PHOTON YIELD
A. Experimental setup
The experiment to realize quasi-absolute method was done by the 44 MeV L-band pulsed LINAC at the Idaho Accelerator Center. In the experiment, the peak current is about 300 mA, pulse width is 2.2 μs, repetition rate is 120 Hz, and the peak energy is about 30 MeV. The side view of the experimental setup is shown in Fig. 2.
-201405/1001-8042-25-05-003/alternativeImage/1001-8042-25-05-003-F002.jpg)
Electrons are first produced by the hot cathode and then accelerated by a series of alternating RF electric fields in the acceleration cells. They form a focused electron beam from magnetic fields. The radius of the electron beam is around 3 millimeters. The energy distribution of the electrons (Fig. 3) was measured with a Faraday cup, a magnetic spectrometer and beam-split to allow quasi-monochromatic electrons to pass after magnetic fields bend the track of electrons.
-201405/1001-8042-25-05-003/alternativeImage/1001-8042-25-05-003-F003.jpg)
After the electron beam passes through a water cooling system and a path of air, it is absorbed by a 3 mm-thick tantalum electron-photon converter (or radiator). Bremsstrahlung radiation is the primarily physical process in the radiator. After the radiator, the electron beam almost completely stops and the newly generated photon beam strikes an aluminum hardener of about 3 inches thickness. The hardener absorbs the residual electrons and ensures that the beam out of the radiator is entirely made of high energy photons.
The urban particulate matter—standard reference material 1648a [14]—from NIST was chosen as the sample for verifying quasi-absolute method. It was selected because it is a well-known certified reference material and widely used in instrumental analytical methods. The sample was wrapped into a 1 cm by 1 cm square area with Aluminum foil and put behind the hardener along the beam axis. The thickness of the sample is about 1 mm.
B. Photon beam simulations
Photon flux simulation was exactly based on the geometry and material of the experimental setup in Fig. 1. The simulated photon flux is on the surface of the sample. Sample matrix effects (e.g., in sample Compton scattering, γ induced further nuclear reactions, etc.) are not considered since the sample is small and thin. Geant4 is used as our primarily simulation software. MCNPX applied as an accessorial validation. The Geant4 toolkit we applied to the simulation is version 4.9.6 [15]. The class library of the high energy physics (CLHEP) used in this version is 2.1.3.1 [16]. The Geant4 toolkit is installed on a computer server running 64 bits GNU/Linux. The physics list of simulations includes all the electromagnetic processes, such as ionization, multiple scattering, positron annihilation, photoelectric effects, Compton scattering, pair production, and bremsstrahlung radiation. In the simulations, the total number of electrons we applied was 10 million. The statistical uncertainty from the simulation was small since we processed such a large number of electrons.
Figure 4 depicts the Geant4 simulation of a photon beam for the quasi-absolute method in PAA. In the figure, the yellow dot is the tracking point. The relationship of the track color and its corresponding particle is: photon–green, electron–red, positron–blue, neutron–yellow. The yellow square disk at the end of beam line is the sample.
-201405/1001-8042-25-05-003/alternativeImage/1001-8042-25-05-003-F004.jpg)
Figure 5 shows the difference between simulated photon yields from Geant4 and those from MCNPX with the same geometry and materials shown in Fig. 2. The two simulated results are consistent but with some slight variance. The percentage of differences is approximately 2% for the whole energy range (0∼30 MeV) and about 7.6% for the energy range of interest (10∼30 MeV) of PAA. The source of this discrepancy may originate from the different cross section libraries and the different algorithms used to implement calculation of electromagnetic processes in the two simulation codes.
-201405/1001-8042-25-05-003/alternativeImage/1001-8042-25-05-003-F005.jpg)
C. Selection of cross sections
Ten photonuclear reactions were selected for realizing the quasi-absolute method (see Table 2). They were chosen based on the following criteria: (1) concentrations of their target nuclides are certified; (2) the target nuclides in this sample have a wide atomic number distribution: from light elements to heavy ones; and (3) their product nuclides have clear interference-free energy lines (pile-up can be ignored at the low counting rates in these experiments); (4) most of them are (γ, n) reactions. For the same isotope, the (γ, n) reaction channel has the largest cross sections in GDR region.
Reaction | h% | E (keV) | T1/2 | Branch% | η | I (s-1) | P | C (μg/g) | Cq (μg/g) | Accuracy |
---|---|---|---|---|---|---|---|---|---|---|
48Ca(γ,n)47Ca | 0.187 | 1297.09 | 4d12h51m40s | 71.00 | 0.01950 | 5.79E-14 | 26660±290 | 58400±1900 | 47700±8300 | -18.3% |
55Mn(γ,n)54Mn | 100 | 834.85 | 312d6h26m40s | 99.98 | 0.03002 | 5.86E-14 | 30210±187 | 790±44 | 786±98 | -0.5% |
59Co(γ,n)58Co | 100 | 810.775 | 70d20h33m20s4 | 99.45 | 0.03089 | 6.51E-14 | 2280±310 | 17.93±0.68 | 13.7±2.5 | -23.5% |
58Ni(γ,n)57Ni | 68.08 | 1377.63 | 1d11h36m40s | 81.7 | 0.01839 | 2.20E-14 | 763±39 | 81.1±6.8 | 88±12 | 8.3% |
66Zn(γ,n)65Zn | 27.90 | 1115.55 | 244d5h6m40s | 50.6 | 0.02260 | 7.31E-14 | 19760±290 | 4800±270 | 3660±420 | -23.7% |
68Zn(γ,p) 67Cu | 18.75 | 184.577 | 2d13h50m | 48.7 | 0.12344 | 5.24E-15 | 30300±1100 | 4800±270 | 3610±380 | -24.8% |
75As(γ,n)74As | 100 | 595.85 | 17d18h23m20s | 59.40 | 0.04174 | 9.95E-14 | 35209±206 | 115.5±3.9 | 73.7±7.9 | -36.2% |
85Rb(γ,n)84Rb | 72.17 | 881.61 | 32d18h23m20s | 68.98 | 0.02846 | 1.44E-13 | 6560±230 | 51.0±1.5 | 28.9±3.2 | -43.4% |
123Sb(γ,n)122Sb | 42.79 | 564.119 | 2d17h21m40s | 70.67 | 0.04403 | 2.88E-13 | 8240±350 | 45.4±1.4 | 24.9±2.8 | -45.1% |
140Ce(γ,n)139Ce | 88.45 | 165.864 | 137d14h46m40s | 79.9 | 0.13253 | 3.16E-13 | 18080±270 | 54.6±2.2 | 32.2±3.4 | -41.0% |
Original data of cross sections is in exchange format (EXFOR), which contains an extensive compilation of experimental nuclear reaction data [17-37]. These records were retrieved and validated via a consortium of organizations, including the International Atomic Energy Agency (IAEA), the National Nuclear Data Center (NNDC), and the Center for Photonuclear Experiments (CDFE). Our criteria for choosing cross sections are: (1) find experimental reaction data and target isotopes as accurately as possible: (γ, n) data is preferred over (γ, total) and single target isotopes are preferred to targets with more than one isotope; (2) the cross section energy range should expand the energy range of interest (10∼30 MeV); (3) the data is as recent as available; (4) data from highly relevant sources in nuclear physics was preferred.
D. Photon yeild
With proper cross section data, we can apply quasi-absolute method by Eq. (5) if photon yield is available. Photon yield was obtained from photon beam simulations by the ratio between the number of photons in a certain energy bin and total incident electrons. The number of photons in different energy bins with different sizes was counted by the function "hist" in statistical software R [38]. We tabulated the cross section data and photon yield and summed the products of them to obtain the reduced reaction rate IR and its corresponding uncertainty. Table 1 gives an example of this process. In Table 1, E is the energy, σ is cross section, Δσ is uncertainty of cross section, "Energy Range" is the energy range centered on E, Np(E) is the number of photons in the energy range, Y(E) is photon yield in the energy range, Y(E)σ(E) is the reduced reaction rate in the energy range, ΔY(E)σ(E) is the uncertainty of reduced reaction rate in the energy range, and IR is reduced reaction rate. After having obtained the reduced reaction rate IR and its uncertainty ΔIR, we inserted them into Eq. (6) to calculate the estimated uncertainty for the quasi-absolute method.
E (MeV) | σ (mb) | Δσ (mb) | Energy range (MeV) | Np(E) | Y(E) | Y(E)σ(E) | ΔY(E)σ(E) |
---|---|---|---|---|---|---|---|
10.82 | 3.5 | 0.1 | 10.82∼10.92 | 642 | 0.0000642 | 2.247E-31 | 6.42E-33 |
11.02 | 4.8 | 0.1 | 10.92∼11.12 | 1271 | 0.0001271 | 6.1008E-31 | 1.271E-32 |
11.22 | 7.3 | 0.1 | 11.12∼11.32 | 1194 | 0.0001194 | 8.7162E-31 | 1.194E-32 |
11.42 | 8.7 | 0.2 | 11.32∼11.52 | 1188 | 0.0001188 | 1.03356E-30 | 2.376E-32 |
⋯ | ⋯ | ⋯ | ⋯ | ⋯ | ⋯ | ⋯ | ⋯ |
23.82 | 16.1 | 5.1 | 23.72∼23.92 | 151 | 0.0000151 | 2.4311E-31 | 7.701E-32 |
24.02 | 19.8 | 5 | 23.92∼24.12 | 147 | 0.0000147 | 2.9106E-31 | 7.35E-32 |
24.22 | 13.8 | 5.1 | 24.12∼24.22 | 68 | 0.0000068 | 9.384E-32 | 3.468E-32 |
∑IR = 1.47582E-28 ± 8.08109E-30 |
IV. RESULTS AND DISCUSSION
Table 2 shows the experimental results from the quasi-absolute method. In Table 2, h is natural abundance, E is the energy line we chose for calculation, T1/2 is the half-life of the product nuclides, Branch% is the absolute intensity of the energy lines, η is the efficiency of the detector of that energy line, I is the reaction rate which was calculated from the simulation and the cross section database, P is the net peak area. c is the NIST certified value of concentration, cq is the concentration calculated from the quasi-absolute method, and "accuracy" is the discrepancy of the two values. The uncertainty of the NIST certified value is expanded uncertainty, with a coverage factor of k=2 (approximately 95% confidence) [14]. The uncertainty from quasi-absolute method is calculated by the standard uncertainty propagation formulas.
From Table 2, one can see that the calculated value is close to the certified value for Ca, Mn, Co, Ni, and Zn; but apart for As, Rb, Sb, and Ce. The calculated concentrations are within a factor of two from the known concentrations. For some elements (e.g. As, Rb, Sb, and Ce), the results from the quasi-absolute method are considerably lower than the known concentration, close to a factor of two. This may stem from the fact that the cross section data we applied to those isotopes is an IAEA-tabulated total photonuclear cross section. For several elements (e.g. Ca, Mn, Co, Ni, and Zn), the calculated concentrations are reasonably close to the known concentrations (around 20%). For Zinc, two different reactions—(γ, n) and (γ, p)—reach very similar results. In addition, our results from the quasi-absolute method are systematically lower, which could mean that the simulated photon flux is higher than the real one. Many experiments have shown that simulated flux of standard Monte Carlo codes is higher than the real experimental flux by around 20%, according to different experimental setups and simulation programs [39-41]. This difference may come from a combination of several factors, such as beam emittance, uncertainty of beam current, beam loading, beam wandering, cross sections, energy dissipation in experimental setup, etc.
Because of the limitations in measurement of experimental parameters, computer simulation of the Bremsstrahlung photon beam, and cross section data, the uncertainty of the quasi-absolute method may be still too high for practical use in precise radiochemistry and nuclear chemistry. In addition, the systematically higher photon flux of simulation is a source of systematic error of the quasi-absolute method.
V. CONCLUSION
We integrated the calculation of absolute measurement with the simulation of the bremsstrahlung photon yield and raised a question whether the quasi-absolute method could be developed and applied to practical radioanalytical chemistry. The experimental results show that the quasi-absolute method can be effective for some elements if the cross section data is accurate enough and simulated photon flux is close to a real situation.
Quasi-absolute method has its merit in practice and lots of room for further improvements: after systematic adjustment of photon flux and more accurate measurements on experimental parameters (especially photo-nuclear cross sections), it is possible to use the quasi-absolute method in actual radioanalytical practice for certain elements. On the other hand, if we have a priori knowledge of elemental concentrations and a tagged photon beam, it is possible to invert the calculations to obtain valuable cross section information. Theoretically, the quasi-absolute method will work well if the measurements and the corresponding simulation reach the required precision and accuracy. This should be a valuable direction in nuclear activation analysis.
CLHEP user guide
, http://proj–clhep.web.cern.ch/proj–clhep/manual/. Accessedthe R Core Team. An introduction to R, Version 3.0.1.
, 2013. http://cran.r–project.org/doc/manuals/r-release/R-intro.html