logo

A SPHERICAL GRAPHITE IONIZATION CHAMBER

A SPHERICAL GRAPHITE IONIZATION CHAMBER

Chen Lishu
Nuclear Science and TechniquesVol.3, No.3pp.167-170Published in print 01 Aug 1992
25100

This paper describes an exposure ratemeter made of graphite ion chamber. The sensitive volume is 14 cm3. The wall is 5 mm in thickness. The DC amplifier with a high input impedance consists of several stage differential amplifications, and is used for measurement of the ionization currents. Six ranges are able to cover a wide range of exposure rates. The full scale of the maximum range is 75 C/(kg • h). The lowest measurable limit is 5 mC/(kg • h). The zero drift within 8 h is less than 5% of each full scale. This instrument provides with a relatively convenient method for the measurement. The main advantages of the ionization chamber over other methods lie in rapidity and accuracy, and the change in radiation field with time can be indicated directly. It is particularly successful in acting as a routine dosimeter under high exposure rates.

High doseExposure ratemeterGamma doseSpherical ionization chamberCavity ionization chamber
REFERENCES
[1] Chen Lishu, Shi Dezhou, Wen Youqin et al.

Cavity ionization chamber

. In: A compilation of monitoring technology reports on radiation protection. Beijing: Atomic Energy Publication House, 1978: 145-160.
Baidu ScholarGoogle Scholar
[2] Chen Lishu, Wang Wenchao, Lei Zhaoqi.

An extrapolation ionization chamber

. In: The physics methods for measurement of ionization radiation (proceedings of symposium), Beijing: Atomic Energy Publication House,, 1975: 173-189.
Baidu ScholarGoogle Scholar
[3] Attix F H. Radiation dosimetry. Vol.1 and Vol.2. New York: Academic Press, 1968: 32, 229.
[4] International Commission on Radiation Units and Measurements.

Radiation quantities and units, ICRU Report 33

. Washington, D C: ICRU, 1980.
Baidu ScholarGoogle Scholar
[5] Mayburg S, Lawrence W L. J Appl Phys, 1952, 23: 1006.
[6] Coleman J H, Bohm David. J Sppl Phys, 1953, 24: 497.
[7] Fowler J F, Farmer F T. Brit J Radiol, 1956, NS 24: 118.
[8] Boag J W. Brit J Phys, 1952, 3: 222.
[9] Moriuchi Y, Katon A, Takata N.

Practical accurate methods for measuring the m-values of ionizing gases

. In: National and international standardization of radiation dosimetry (Proc Symp). Atlanta, 1977. Vienna: IAEA, 1978, 2: 45.
Baidu ScholarGoogle Scholar