logo

Reconstruction of 12 MV bremsstrahlung spectra from measured transmission data

ADVANCED NUCLEAR INSTRUMENTATION AND DETECTION

Reconstruction of 12 MV bremsstrahlung spectra from measured transmission data

LEI Jia-Rong
RAN Shu-Yang
YUAN Yong-Gang
FU Yu-Chuan
Nuclear Science and TechniquesVol.14, No.2pp.138-143Published in print 01 May 2003
23201

Transmission data for 12 MV bremsstrahlung beams produced by the process for continuous-pulsed 12 MV electrons coming from a linear accelerator hitting a 1.2mm thick plane tantalum target have been acquired with a combination of iron and Telfon attenuators. Two solid state dosimeters with LiF-TLD material surrounded Telfon were used as detectors. It has been checked that the experimental system achieves reasonably narrow-beam geometry by Monte Carlo simulations. From these transmission data, the original energy photon spectrum has been reconstructed using the iterative least-squares technique and compared with the spectrum calculated with Monte Carlo code system EGSnrc. The comparison shows that the numerical technique for analysis of transmission data can represent 12 MV bremsstrahlung spectrum acceptably well. The purpose of our work is to provide an effective way to reconstruct an unknown photon spectrum with high energy component and prove the correctness of this way.

Electron bremsstrahlung spectraAttenuation analysisLeast-squares methodMonte Carlo simulationEGSnrc system
References
[1] Huang P H, Kase K R, Bjärgard B E. Med Phys, 1981, 8(3): 368-374
[2] Gayton F M, Longworth J P. A passive gamma spectrometer for use in high fluxes in the energy range 0.01 to 10MV, RD/B/N2024
[3] Huang P H, Kase K R, Bjarngard B E. Med Phys, 1982, 9(5): 695-702
[4] Huang P H, Kase K R, Bjargard B E. Med Phys, 1983, 10(8): 778-785
[5] Xiao K Y, Cui G X, Liu H C. Atom Energy Sci Technol (in Chinese), 1998, 32(6): 499-502
[6] Nelson W R, Hirayama H, Rogers D W O.

The EGS4 code system, SLAC-Report-265

, Stanford Linear Accelerator Center, 1985
Baidu ScholarGoogle Scholar
[7] Bielajew A F, Rogers D W O. Nucl Instrum Methods Phys Res, 1987, B18: 165-81
[8] Kawrakow I, Rogers D W O.

The EGSnrc Code System: Monte Carlo simulation of electron and photon transport, NRCC Report PIRS-701

, 2000
Baidu ScholarGoogle Scholar
[9] Attix F H. Introduction to radiological physics and radiation dosimetry, New York: Wiley, 1986, 210-218
[10]

Technical Reports Series No.127, Determination of absorbed dose in reactors

, IAEA, VIENNA, 1971, 170-171
Baidu ScholarGoogle Scholar
[11] Kreyszig E. Advanced engineering mathematics, 2nd ed, New York: Wiley, 1962