1 Introduction
The Am-Be and 252Cf neutron sources are widely used in neutron dosimetry laboratories. The important features of Am-Be source is its long half-life and emission of fast neutrons with rather wide energy spectra, while 252Cf yields a highly concentrated and reliable neutron spectrum from a small assembly. The last feature of both sources is helpful in covering the energy domain of interest for many applications in ambient and personal dosimetry. Nevertheless, the neutron energy spectrum depends on materials and dimension of the capsule and on amount and physio–chemical properties of active material, thus affecting relevant quantities such as spectrum-averaged fluence-to-dose equivalent conversion coefficient [1-6]. It is important to be know the neutron flux accurately so as to measure different quantities related to a neutron source [7].
Neutron detection has a key role in development of nuclear technology and its applications [8,9]. High sensitive neutron detectors are required. However, most of the neutron detectors are also sensitive to gamma rays. Therefore, an effective technique is required to eliminate the gamma background for accurate neutron measurements. Generally, methods used to recognize neutron interactions within a detector rely on second-order effects. Two neutron interactions used for a variety of thermal neutron detectors are the 10B(n,α)7Li [9, 10] and 6Li(n,α)3H reactions [11, 12]. The α particles can be easily detected.
Low atomic number materials, such as high-density polyethylene (HDPE), which has a high concentration of hydrogen, tend to have relatively high elastic scattering cross sections for fast neutrons, and often (n,p) reactions from fast neutrons interacting in hydrogen-filled materials are manipulated for fast neutron detection. For detection of thermal or low energy neutrons, polymers-based SSNTDs, i.e. Solid State Nuclear Track Detectors (such as CR-39), can be used only in combination with a neutron reactive film which converts neutrons into other detectable radiations, recoil nuclei [13, 14]. Use of a BF3 or 3He neutron detector, one can measure the optimum thickness of the moderator accurately, which is essential to avoid excessive thermal neutron absorption in the moderator. The neutron activation analysis can determine accurately the major and trace elements in different samples [10–12, 15]. In practice, aromatic hydrocarbons are considered as the most radiation-resistant hydrogenous substances and have properties to moderate slow neutrons more effectively [16].
A paraffin wax moderator assembly has been developed to find out optimum thickness of paraffin wax. It can be utilized for fast neutrons from various neutron-producing reactions on an accelerator or for measurement of neutron yield from a Mather type plasma focus device [17–20]. The assembly has been utilized to find out neutron flux of a neutron probe which contains Am-Be as a neutron source [21]. Neutron flux comparison and neutron activation are used to the measure neutron flux [24]. A 50 mCi Am-Be neutron probe is used as a neutron source for the measurement of moisture in soils [21].
The moderating materials should be of large scattering cross-section, small absorption cross-section and large energy loss per collision [23], and paraffin wax is just a desired moderator. Its optimum thickness was calculated for 3 MeV neutrons [24], and was experimentally determined at 7 cm and 2.5–4.5 cm for Am-Be and 252Cf neutron sources, respectively [25, 26]. The results were based on the maximum gamma ray yield due to neutron activation. The optimum thickness of HDPE moderator was measured at 4–6 cm for 2.8 MeV neutrons. For theoretical calculations, different codes such as MNCP, MNCPX, SRIM etc. are used [27, 28]. The present paper is designed to determine the optimum moderator thickness that is required to be wrapped around a BF3 detector used in monitoring fissile material. The experimental results are compared with MC simulations for Am-Be neutron source using MCNP5 code version 1.40 [29].
2 Material and Methods
2.1. Theoretical calculations
The work of Goshal is referred for theoretical calculations [24]. The slowing down process involves elastic scattering by light nuclei like hydrogen and carbon in paraffin wax (ρ= 0.93 g/cm3) [5]. The average logarithmic decrease in neutron energy per collision is calculated. The average number of collisions needed to thermalize fast neutrons is also calculated. The transport, thermal neutron mean free paths are calculated for paraffin wax and finally the optimum moderator thickness is calculated at 6 cm for 3 MeV neutrons.
2.2. Experimental Approach
The paraffin wax moderator assembly consists of two parts as shown in Fig.1. The first part consists of a removable paraffin wax discs, neutron detector and a neutron source (i.e. Am-Be/ 252Cf neutron sources). The second part is a mounting frame with two parallel spring-loaded Φ12 mm steel rods, each fixed at one end with the frame. The other two ends are fixed to a movable plastic ring around a 1.4 cm thick paraffin wax disc. The purpose of the plastic ring is to give strength to the paraffin wax discs. A number of paraffin wax discs are used. Sized at Φ27 cm × 0.6–1.4 cm, each disk is mounted in a 6 mm thick plastic ring. The paraffin wax discs can be assembled easily on the mounting frame, between two semi cylindrical fixed pipes with the assembly at the same axes, one for the neutron detector and the other for the neutron source. The two parallel rods act as a guide for the paraffin wax discs, and keep the discs in close contact with each other. The Φ27 cm discs keep the axis same with respect to the axes of the neutron detector and the neutron source. Paraffin wax discs can be added or removed for measurement of optimum thickness of paraffin wax, where the neutron detector counts is the maximum. Two BF3 (Φ1″and 2″) and one 3He (Φ0.5″) neutron detectors are used for measuring the optimum thickness with Am-Be or 252Cf neutron sources. The associated electronics with detector assembly consisted of pre-amplifier, amplifier, single channel analyzer and counter/timer manufactured by the ORTEC.
-201705/1001-8042-28-05-004/alternativeImage/1001-8042-28-05-004-F001.jpg)
The background counts per second were measured without the neutron sources, and they were subtracted from the actual neutron count rates emitted from neutron sources. For different paraffin wax thicknesses, counts emitted from either Am-Be or 252Cf neutron sources, were taken for 100 seconds by using the BF3 and one 3He neutron detectors. For both the Am-Be and 252Cf neutron sources, the paraffin wax thicknesses were increased to find the maximum counts, and on further increase in thickness, the counts started decreasing. Here, the maximum counts represent optimum paraffin wax thickness that is required to thermalize fast neutron (of course, the process is the tradeoff between absorption and moderation). To minimize scattering from the walls, the heavy shielding was placed about 4 m away from the source. The scattering contribution at the detector position was checked with and without the shielding, with the source at its position. The difference was negligible and hence an evidence of no scattering contribution from shielding walls was seen. Furthermore, BF3 or 3He detectors can detect only slow or thermal neutrons and the gases have very low interaction cross-section for fast or scattering neutrons.
The moderator assembly (Fig. 1) was also used for the measurement of neutron flux emitted from the Am-Be and 252Cf neutron sources to check the validity of the optimum thickness of the paraffin wax moderator, using the neutron flux comparison method and the neutron activation analysis method [22].
2.2.1. Neutron flux comparison method
The comparison of counting rates was carried out, between the standard Am-Be neutron source of known flux (F1) and that from the neutron probe of unknown flux (F2) under conditions insensitive to the energy of neutrons. Let C1 be the average counts per sec for F1 recorded by a BF3 (Φ1″) neutron detector in contact with the 6 cm thick paraffin wax moderator and C2 be the average neutron counts per sec, with their background subtracted, from the neutron probe under the same condition. The neutron flux F2 from the neutron probe is given by F2 = F1 C2/ C1.
2.2.2. Neutron activation analysis method
The paraffin wax moderator assembly was used to measure the neutron flux of the neutron probe through activation analysis technique by using silver foil, and the 6 cm optimum thickness of the paraffin wax moderator was validated.
Thermal neutron activation analysis is the absolute method to determine the neutron flux of a neutron probe. The silver isotopes of 108Ag (T1/2 =0.39 min) and 110Ag (T1/2 = 24.6 sec) are produced through nuclear reactions of 107Ag(n, γ)108Ag and 109Ag(n, γ)110Ag with thermal neutron. After a few half-lives of 110Ag most of the radioactivity in silver foil is due to 108Ag [30]. A G.M counter with Mylar end-window is used for the neutron activation analysis. The silver foil is placed at the Mylar window. Both the neutron probe and the end-window of G.M detector are placed at the same axis in opposite and in contact with the paraffin wax moderator (Fig.2).
-201705/1001-8042-28-05-004/alternativeImage/1001-8042-28-05-004-F002.jpg)
The optimum paraffin wax thickness of 6 cm was used in the moderator assembly. At the termination of activation, the flux φ of the neutron probe (in n/cm2/sec) is given by φ =[A0w(1−e−λt)/(σAavαm)]ε(4π/Ω) [24], where A0 is the activity of 108Ag at t =0, w is the atomic weight of silver, λ is the decay constant, t is the irradiation time, σ = 37.6 barns (taken from the ENDF library [31]) is the thermal neutron capture cross section of 107Ag, Aav is the Avogadro’s number, α is the natural isotopic abundance of 107Ag, m is the mass in grams of 107Ag, ε is the intrinsic efficiency of the detector, and Ω is the solid angle subtended by the G.M detector at the neutron source (Am-Be) of the probe.
The intrinsic efficiency of the G.M detector was determined from a 137Cs γ-ray source of known activity. The silver foil was activated for 5 min. Although both 108Ag and 110Ag are produced, most of the activity is due to 108Ag at the time of counting. The neutron probe was taken far away from the moderator assembly as quickly as possible, so the G.M detector background count was made almost negligible at the time of counting. The activated silver foil was counted immediately, and was repeated several times for better results. The background of the G.M detector was measured in advance.
2.3. Monte Carlo simulation
MCNP is an advanced Monte Carlo code for simulation of the neutron transport [29, 32]. It has the ability to model complex geometry of designed experiment and contains all necessary cross-section data for neutron, photon and electron transport calculations [32]. For a neutron detection process, the history of each starting neutron is followed between collisions, its energy deposition is recorded, throughout its life to its death, until its energy is low enough to be neglected. In present study, the experimental set up was simulated in the School of Physics, University Science Malaysia, Penang, Malaysia, using the MCNP5 code version 1.40 [29, 32]. The experimental setup as described in Fig. 1 was modeled. According to this model, the neutron source was located at one side of the paraffin wax moderator while the neutron detector was simulated facing the source at the other side of the moderator. The experimental setup was simulated on wooden table top. To calculate the detector neutron flux, F2 tally (calculates the flux averaged over a surface [29] was used. Surface segmentation was utilized to calculate the flux at a surface segment facing the neutron source.
3 Results and Discussion
The measured counts per 100 seconds for different wax thicknesses by using the three neutron detectors for the Am-Be and 252Cf neutron sources plotted against wax thickness are shown in Fig.3. The maximum counts for each neutron detector and source represent the optimum paraffin wax thickness that is required to thermalize fast neutrons. From Fig.3, the optimum paraffin wax thicknesses were 6 cm and 5.5 cm for Am-Be and 252Cf neutron sources, respectively. The yield of thermal neutrons increases with the thickness of paraffin wax from 1.6 to 6 cm, due to fact that the fast neutrons lose their energy through elastic collision with hydrogen atoms of paraffin wax. The yield of thermal neutrons at these thicknesses is found to be maximum, as a large number of fast neutrons lose their maximum energy through elastic collisions. However, on further increase in thickness of the paraffin wax from 6.5 cm to 12 cm, the yields of thermal neutron decreases due to elastic and inelastic scattering reactions. The neutrons lose their energy through the interactions until they are captured by paraffin wax (shielding material) [23]. Normally, the neutron capture cross section is larger only for thermal neutron energies, that is why the neutrons slowing down by scattering are important before capture [23].
-201705/1001-8042-28-05-004/alternativeImage/1001-8042-28-05-004-F003.jpg)
To validate the optimum thickness of paraffin wax, the neutron flux was calculated. By the neutron flux comparison method, the average neutron counts per second, C1, were 141.3 ± 0.66 with the known neutron flux of F1 = (8.3±0.66)×104 n/s. With C2 =199.1 ± 0.79 of the neutron probe, the F2 was (1.17 ± 0.09)×105 n/s. On the other hand, the neutron flux deduced by the activation method was (1.19 ± 0.1)×105 n/sec. The two results are virtually the same, considering the measurement errors. The neutrons emerging out of paraffin wax are of both thermal and epithermal in energy. The fluence of epithermal neutrons was estimated from Ref. [30]. The ratio of epithermal to the total neutrons at a distance of 18.9 cm in air from the neutron source is 5.56% where the total fluence is maximum as calculated from Ref. [33]. Therefore, this ratio is the same for the paraffin moderator at 6 cm in thickness.
By using MCNP5, the energy spectra of Am-Be neutron source at 3, 5, and 8 MeV were produced with the relative probabilities of 1.000, 0.986 and 0.551, respectively. The simulation results for the Am-Be source are shown in Fig.4, and the counts are maximized at 6 cm thickness of the paraffin wax moderator, agreeing well with the experimental results.
-201705/1001-8042-28-05-004/alternativeImage/1001-8042-28-05-004-F004.jpg)
4 Conclusion
The paraffin wax moderator assembly can be used easily to thermalize fast neutrons of various energies produced from neutron-producing reactions on an accelerator. Neutron flux comparison and neutron activation techniques are used for accurate determination of neutron flux of the neutron probe to validate the optimum thickness of the paraffin wax moderator in the assembly. The neutron flux of (1.17 ± 0.09) ×105 n/s by the neutron flux comparison method agree well with the flux of (1.19 ± 0.1) ×105 n/s by neutron activation analysis method. The techniques can be utilized for the measurement of neutron yield from Mather type plasma focus device. The optimum thickness of the paraffin wax moderator is 6 cm for Am-Be and 5.5 cm for 252Cf neutron sources.
Investigation of the neutron spectrum of americium-beryllium sources by Bonner sphere spectrometry
. Nucl Instru and Meths in Phys Res Section A, 763: 547-552. (2014) doi: 10.1016/j.nima.2014.06.040252Cf fission-neutron spectrum using a simplified time-of-flight setup: An advanced teaching laboratory experiment
. American Journal of Physics, 81: 112-119. (2013) doi: 10.1119/1.4769032Group representation of the prompt fission neutron spectrum of 252Cf
. Esarda Bulletin: 112-114. (2011) https://esarda.jrc.ec.europa.eu/images//Bulletin/Files/B_2011_046.pdfHigh resolution measurements of neutron energy spectra from Am-Be and Am-B neutron sources
. Nucl Instru Meths in Phys Res Section A, 366: 340-348. (1995) doi: 10.1016/0168-9002(95)00613-3Energy spectrum measurement and dose rate estimation of natural neutrons in Tibet region
. Nucl. Sci. Tech., 26: 060202. (2015) doi: 10.13538/j.1001-8042/nst.26.060202Development of a spherical tissue equivalent proportional counter for neutron monitoring
. Nucl. Sci. Tech., 26: 060403. (2015) doi: 10.13538/j.1001-8042/nst.26.060403Special absorber neutron detector moderator assembly: A new detector system for flux measurements of collimated 2.5 MeV neutrons
. Review of Scientific Instruments, 70: 1194-1196. (1999) doi: 10.1063/1.1149322Research and applications of nuclear tracks: Developments in Pakistan and global comparison
. Natural Science, 2012. (2012) doi: 10.4236/ns.2012.431124Measurement of thermal neutron cross section and resonance integral for the 165Ho(n,γ) 166Ho reaction
. Radiochimica Acta International Journal for Chemical Aspects of Nuclear Science and Technology, 97: 63-69. (2009) doi: 10.1524/ract.2009.1575Determination of boron contents in water samples collected from the Neelum valley, Azad Kashmir, Pakistan
. Biological Trace Element Research, 139: 287-295. (2011) doi: 10.1007/s12011-010-8665-6Neutron and β/γ radiolysis of water up to supercritical conditions. 1. β/γ yields for H2, H· atom, and hydrated electron
. The Journal of Physical Chemistry A, 114: 5034-5034. (2010) doi: 10.1021/jp101790pThermal neutron detection using a silicon pad detector and 6LiF removable converters
. Review of Scientific Instruments, 84: 033503. (2013) doi: 10.1063/1.4794768New source-moderator geometry to improve performance of 252 Cf and 241 Am-Be source-based PGNAA setups
. Nucl Instru Meths in Phys Res Section A: 562: 358-364. (2006) doi: 10.1016/j.nima.2006.02.004Enhanced and reproducible neutron emission from a plasma focus with pre-ionization induced by depleted uranium (U238)
. Plasma Physics and Controlled Fusion, 48: 745. (2006) doi: 10.1088/0741-3335/48/6/003The world's first pelletized cold neutron moderator at a neutron scattering facility
. Nucl Instru Meths in Phys Res Section B, 320: 70-74. (2014) http://www1.jinr.ru/Preprints/2012/113(P13-2012-113).pdfStudy of neutron emission in a low-energy plasma focus with β-source-assisted breakdown
. Plasma Sources Science and Technology, 12: 443. (2003) doi: 10.1088/0963-0252/12/3/320Correlation of plasma electron temperature with neutron emission in a low-energy plasma focus
. Plasma Science, IEEE Transactions on, 29: 62-68. (2001) doi: 10.1109/27.912943Comparative study of ion, X-ray and neutron emission in a low energy plasma focus
. Plasma Sources Science and Technology, 7: 206. (1998) doi: 10.1088/0963-0252/7/2/015Neutron Activation Analysis (Slow Neutrons)
, in United States Patent. 1982.Neutron capture cross section measurements of paraffin wax
. International Journal of Application or Innovation in Engineering of Management, 3: 112-114. (2014) http://ijaiem.org/volume3issue4/IJAIEM-2014-04-20-034.pdfStudies on optimization of moderator thickness for BF3 detectors used for monitoring of fissile material
. Indian Journal of Pure and Applied Physics, 50: 811-813. (2012) http://nopr.niscair.res.in/handle/123456789/14910Verification of design calculations of a PGNAA setup using nuclear track detectors
. Radiation Measurements, 38: 37-41. (2004) doi: 10.1016/S1350-4487(03)00252-XOptimizing moderator dimensions for neutron scattering at the spallation neutron source
. Review of Scientific Instruments, 84: 125104. (2013) doi: 10.1063/1.4841875SRIM-The stopping and range of ions in matter (2010)
. Nucl Instru Meths in Phys Res Section B, 268: 1818-1823. (2010) DOI: 10.1016/j.nimb.2010.02.091ENDF/B-VII. 1 nuclear data for science and technology: cross sections, covariances, fission product yields and decay data
. Nuclear Data Sheets, 112: 2887-2996. (2011) doi: 10.1016/j.nds.2011.11.002A dual-hemisphere (spherical) irradiation facility for Am-Be neutron field optimization
. (2009) http://www.iaea.org/inis/collection/NCLCollectionStore/Public/42/009/42009903.pdf.